K180975 · Inova Diagnostics, Inc. · LLL · Jun 25, 2018 · Immunology
Device Facts
Record ID
K180975
Device Name
QUANTA Flash HMGCR Reagents
Applicant
Inova Diagnostics, Inc.
Product Code
LLL · Immunology
Decision Date
Jun 25, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.5100
Device Class
Class 2
Indications for Use
QUANTA Flash HMGCR is a chemiluminescent immunoassay for the semi-quantitative determination of IgG autoantibodies against HMGCR (3-hydroxy-3-methylglutaryl-coenzyme A reductase) antigen in human serum. The presence of anti-HMGCR antibodies, in conjunction with clinical findings and other laboratory tests, is an aid in the diagnosis of idiopathic inflammatory myopathy (IIM).
Device Story
QUANTA Flash HMGCR is a semi-quantitative chemiluminescent immunoassay (CIA) for detecting IgG autoantibodies against HMGCR in human serum. The device uses paramagnetic beads coated with recombinant HMGCR antigen. Patient serum is diluted and incubated with the beads; anti-HMGCR antibodies bind to the antigen. After washing, isoluminol-labeled anti-human IgG is added, binding to the captured antibodies. Trigger reagents induce a luminescent reaction, measured as Relative Light Units (RLU) by the BIO-FLASH analyzer. The system uses a lot-specific Master Curve and two calibrators to calculate results in Chemiluminescent Units (CU). The device is intended for use in clinical laboratories by trained personnel. Results assist clinicians in diagnosing idiopathic inflammatory myopathy (IIM) when combined with other clinical and laboratory findings.
Clinical Evidence
Clinical validation performed on 723 characterized samples (257 IIM patients, 466 controls). Results: 25.3% sensitivity (95% CI: 20.4%-30.9%), 99.8% specificity (95% CI: 98.8%-100.0%), 98.5% PPV, and 70.8% NPV. Study confirms diagnostic utility for IIM sub-classification.
Technological Characteristics
Solid-phase chemiluminescent immunoassay using paramagnetic microparticles coated with recombinant HMGCR. Energy source: BIO-FLASH analyzer (luminometer). Dimensions/Form factor: Reagent cartridge format. Connectivity: Automated closed system. Software: Rule-based calculation of CU from RLU using lot-specific Master Curve and calibrator-derived Working Curve. Standards: EP05-A3, EP06-A, EP07-A2, EP17-A2, EP28-A3c.
Indications for Use
Indicated for the semi-quantitative determination of IgG autoantibodies against HMGCR in human serum to aid in the diagnosis of idiopathic inflammatory myopathy (IIM). For prescription use only.
Regulatory Classification
Identification
An antinuclear antibody immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoimmune antibodies in serum, other body fluids, and tissues that react with cellular nuclear constituents (molecules present in the nucleus of a cell, such as ribonucleic acid, deoxyribonucleic acid, or nuclear proteins). The measurements aid in the diagnosis of systemic lupus erythematosus (a multisystem autoimmune disease in which antibodies attack the victim's own tissues), hepatitis (a liver disease), rheumatoid arthritis, Sjögren's syndrome (arthritis with inflammation of the eye, eyelid, and salivary glands), and systemic sclerosis (chronic hardening and shrinking of many body tissues).
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
A. 510(k) Number:
K180975
B. Purpose for Submission:
New device
C. Measurand:
Human anti-HMGCR (3-hydroxy-3-methylglutaryl-coenzyme A reductase) IgG autoantibodies
D. Type of Test:
Semi-quantitative chemiluminescent immunoassay (CIA)
E. Applicant:
INOVA Diagnostics, Inc.
F. Proprietary and Established Names:
QUANTA Flash HMGCR Reagents
G. Regulatory Information:
1. Regulation sections:
21 CFR §866.5100, Antinuclear Antibodies Immunological Test System
2. Classification:
Class II
3. Product codes:
LLL – Extractable antinuclear antibody, antigen and control
4. Panel:
Immunology (82)
{1}
H. Intended Use:
1. Intended use:
QUANTA Flash HMGCR is a chemiluminescent immunoassay for the semi-quantitative determination of IgG autoantibodies against HMGCR (3-hydroxy-3-methylglutaryl coenzyme A reductase) antigens in human serum. The presence of anti-HMGCR antibodies, in conjunction with clinical findings and other laboratory tests, is an aid in the diagnosis of idiopathic inflammatory myopathy (IIM).
2. Indications for use:
Same as Intended Use
3. Special conditions for use statement:
For Prescription Use only
4. Special instrument requirements:
BIO-FLASH® chemiluminescent analyzer (K083518)
I. Device Description:
The QUANTA Flash HMGCR Reagents kit contains the following materials:
a. One QUANTA Flash HMGCR Reagent Cartridge
b. One tube of Resuspension Buffer
c. One transfer pipette
The QUANTA Flash HMGCR reagent cartridge contains the following reagents for 50 determinations:
a. HMGCR coated paramagnetic beads, lyophilized
b. Assay buffer – colored pink, containing protein stabilizers and preservatives
c. Tracer IgG – Isoluminol labeled anti-human IgG antibody, containing buffer, protein stabilizers and preservative
QUANTA Flash HMGCR Calibrators and QUANTA Flash HMGCR Controls are sold separately.
The QUANTA Flash HMGCR Calibrators kit contains two vials each of Calibrator 1 and Calibrator 2:
a. QUANTA Flash HMGCR Calibrator 1: Two barcode labeled tubes containing 0.3 mL pre-diluted, ready to use reagent. Calibrators contain human antibodies to HMGCR in
{2}
stabilizers and preservatives.
b. QUANTA Flash HMGCR Calibrator 2: Two barcode labeled tubes containing 0.3 mL pre-diluted, ready to use reagent. Calibrators contain human antibodies to HMGCR in stabilizers and preservatives.
The QUANTA Flash HMGCR Controls kit contains two vials of Negative Control and two vials of Positive Control:
a. QUANTA Flash HMGCR Negative Control: Two barcode labeled tubes containing 0.5 mL, ready to use reagent. Controls contain human antibodies to HMGCR in stabilizers and preservatives.
b. QUANTA Flash HMGCR Positive Control: Two barcode labeled tubes containing 0.5 mL, ready to use reagent. Controls contain human antibodies to HMGCR in stabilizers, and preservatives.
# J. Substantial Equivalence Information:
1. Predicate device name:
QUANTA Flash Jo-1
2. Predicate 510(k) number:
K151429
2. Comparison with predicate:
QUANTA Flash HMGCR Reagents Kit:
| Similarities | | |
| --- | --- | --- |
| Item | Device QUANTA Flash HMGCR Reagents | Predicate QUANTA Flash Jo-1 Reagents |
| Assay methodology | Solid phase (heterogenous) immunoassay | Same |
| Solid phase | Paramagnetic microparticles (beads) | Same |
| Detection/Operating principle | Chemiluminescent immunoassay | Same |
| Conjugate | Isoluminol conjugated anti-human IgG | Same |
| Shelf life | One year | Same |
| Sample type | Serum | Same |
| Calibration | Lot specific Master Curve with two calibrators (sold separately) | Same |
| Units | Chemiluminescent units (CU) | Same |
{3}
| Differences | | |
| --- | --- | --- |
| Item | Device
QUANTA Flash HMGCR Reagents | Predicate
QUANTA Flash Jo-1 Reagents |
| Intended Use | QUANTA Flash HMGCR is a chemiluminescent immunoassay for the semi quantitative determination of IgG autoantibodies against HMGCR (3- hydroxy-3 methylglutaryl-coenzyme A reductase) antigen in human serum. The presence of anti HMGCR antibodies, in conjunction with clinical findings and other laboratory tests, is an aid in the diagnosis of idiopathic inflammatory myopathy (IIM). | QUANTA Flash Jo-1 is a chemiluminescent immunoassay for the semi quantitative determination of IgG anti-Jo-1 antibodies in human serum. The presence of anti-Jo-1 antibodies, in conjunction with clinical findings and other laboratory tests, is an aid in the diagnosis of idiopathic inflammatory myopathy. |
| Analytical Measuring Range | 1.5 CU – 550.0 CU | 2.2 – 1147.2 CU |
| Antigen | Recombinant HMGCR | Recombinant Jo-1 |
K. Standard/Guidance Document Referenced:
EP05-A3, Defining, Establishing, and Verifying Reference Intervals in the Clinical Laboratory; Approved Guideline, Third Edition
EP06-A, Evaluation of Linearity of Quantitative Measurement, Approved Guideline, Second Edition
EP07-A2, Interference Testing in Clinical Chemistry, Approved Guideline, Second Edition
EP17-A2, Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline, Second Edition
EP28-A3c, Defining, Establishing, and Verifying Reference Intervals in the Clinical Laboratory; Approved Guideline, Third Addition
L. Test Principle:
The principle of the assay is chemiluminescent microparticle immunoassay, a variation of solid phase immunoassay. The QUANTA Flash HMGCR assay is designed to run on the BIO-FLASH instrument. This platform is a fully automated closed system with continuous load and random access capabilities that automatically processes the samples, runs the assay and reports the results. It includes liquid handling hardware, luminometer and computer with software-user interface. The QUANTA Flash HMGCR assay utilizes a reagent cartridge format, which is compatible with the BIO-FLASH instrument. HMGCR antigen is coated on
{4}
to paramagnetic beads, which are stored in the reagent cartridge lyophilized. When the assay cartridge is ready to be used for the first time, a buffer solution is added to the tube containing the beads, and the beads are re-suspended with the buffer. The reagent cartridge is then loaded onto the BIO-FLASH instrument. A patient serum sample is diluted 1:17 by the instrument in a disposable plastic cuvette. An aliquot of the diluted patient serum, HMGCR-coupled beads, and assay buffer are combined into a second cuvette, and mixed. This cuvette is incubated at $37^{\circ}\mathrm{C}$ . The beads are then magnetized and washed several times. Isoluminol conjugated anti-human IgG antibody is then added to the cuvette, and incubated at $37^{\circ}\mathrm{C}$ . Again, the beads are magnetized and washed repeatedly. The isoluminol conjugate produces a luminescent reaction when "Trigger" reagents are added to the cuvette. The light produced from this reaction is measured as Relative Light Units (RLU) by the BIO-FLASH optical system. RLU values are proportional to the amount of bound isoluminol conjugate, which in turn is proportional to the amount of anti-HMGCR antibodies bound to the antigen on the beads. The QUANTA Flash HMGCR assay utilizes a predefined lot specific Master Curve that is uploaded into the instrument through the reagent cartridge barcode. Based on the results obtained by running two calibrators, an instrument specific Working Curve is created, which is used by the software to calculate chemiluminescent units (CU) from the RLU value obtained for each sample.
# M. Performance Characteristics:
# 1. Analytical performance:
All the results below met the manufacturer's pre-specified acceptance criteria.
# a. Precision/Reproducibility:
Precision: The precision of the QUANTA Flash HMGCR assay was evaluated by testing seven serum samples prepared to contain various concentrations of anti-HMGCR antibody. Each sample was run in duplicate, twice a day, for 20 days with one reagent lot (total of 80 replicates per sample). The results are summarized in the table below.
| Sample | Mean (CU) | Within-Run | | Between-Run | | Between-Day | | Total | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | SD | CV (%) | SD | CV (%) | SD | CV (%) | SD | CV (%) |
| 1 | 10.9 | 0.5 | 4.6 | 0.2 | 2.2 | 0.5 | 4.3 | 0.7 | 6.7 |
| 2 | 48.3 | 1.1 | 2.4 | 1.5 | 3.1 | 2.6 | 5.3 | 3.2 | 6.6 |
| 3 | 16.3 | 0.3 | 2.0 | 0.5 | 3.4 | 0.8 | 5.0 | 1.0 | 6.3 |
| 4 | 23.4 | 0.4 | 1.9 | 0.9 | 3.8 | 0.9 | 4.1 | 1.4 | 5.9 |
| 5 | 76.2 | 2.1 | 2.8 | 1.7 | 2.2 | 4.0 | 5.2 | 4.8 | 6.3 |
| 6 | 175.5 | 3.8 | 2.1 | 5.4 | 3.1 | 11.4 | 6.5 | 13.1 | 7.5 |
| 7 | 400.5 | 15.5 | 3.9 | 20.4 | 5.1 | 22.6 | 5.6 | 34.1 | 8.5 |
{5}
Reproducibility: A total of eight samples were tested according to CLSI EP05-A03 at three different sites with one reagent lot to evaluate the site-to-site reproducibility. Each sample was run in replicates of five, once a day for five days, to generate 25 data points per sample at each site (N=75 per sample for all sites combined). Data were analyzed for within-run, between-day, between-site, and total imprecision. The results are summarized in the tables below.
| Sample | Mean (CU) | Within-Run | | Between-Day | | Between-Site | | Total | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | SD | CV (%) | SD | CV (%) | SD | CV (%) | SD | CV (%) |
| 1 | 10.6 | 0.6 | 3.4 | 0.7 | 6.5 | 0.5 | 5.1 | 0.9 | 8.8 |
| 2 | 16.1 | 1.0 | 3.1 | 0.7 | 4.4 | 0.8 | 4.9 | 1.2 | 7.5 |
| 3 | 22.8 | 0.8 | 3.4 | 1.1 | 4.6 | 0.0 | 0.0 | 1.3 | 5.7 |
| 4 | 23.9 | 0.5 | 2.2 | 1.0 | 4.2 | 0.5 | 1.9 | 1.2 | 5.1 |
| 5 | 55.6 | 1.8 | 3.2 | 3.0 | 5.5 | 0.0 | 0.0 | 3.5 | 6.3 |
| 6 | 123.9 | 3.6 | 2.9 | 6.2 | 5.0 | 4.7 | 3.8 | 8.6 | 6.9 |
| 7 | 134.2 | 4.1 | 3.0 | 7.7 | 5.7 | 0.4 | 0.3 | 8.7 | 6.5 |
| 8 | 344.2 | 22.0 | 6.4 | 18.3 | 5.3 | 14.8 | 4.3 | 32.2 | 9.4 |
To evaluate lot-to-lot variability, eight samples with anti-HMGCR antibody concentration at various levels across the measuring were tested. Each sample was tested in replicates of five, one run per day for five days using three difference reagent lots. The results are summarized below.
| Sample | Mean (CU) | Repeatability | | Between-Day | | Within-Lot | | Between-Lot | | Within Laboratory | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | SD | CV (%) | SD | CV (%) | SD | CV (%) | SD | CV (%) | SD | CV (%) |
| 1 | 10.9 | 0.4 | 3.7% | 0.8 | 7.7% | 0.9 | 8.5% | 0.2 | 1.8% | 1.0 | 8.7% |
| 2 | 15.1 | 0.3 | 2.2% | 0.9 | 5.6% | 0.9 | 6.0% | 1.1 | 7.2% | 1.4 | 9.4% |
| 3 | 22.8 | 0.7 | 3.0% | 1.3 | 5.8% | 1.5 | 6.5% | 0.8 | 3.6% | 1.7 | 7.4% |
| 4 | 22.3 | 0.7 | 3.0% | 1.4 | 6.1% | 1.5 | 6.8% | 0.2 | 0.8% | 1.5 | 6.8% |
| 5 | 53.8 | 2.4 | 4.4% | 2.9 | 5.3% | 3.7 | 6.9% | 1.7 | 3.1% | 4.1 | 7.6% |
| 6 | 122.6 | 3.3 | 2.7% | 7.0 | 5.7% | 7.8 | 6.3% | 6.0 | 4.9% | 9.8 | 8.0% |
| 7 | 130.3 | 2.9 | 2.2% | 9.3 | 7.1% | 9.7 | 7.5% | 2.6 | 2.0% | 10.1 | 7.7% |
| 8 | 349.8 | 13.5 | 3.8% | 19.8 | 5.7% | 24.0 | 6.8% | 6.9 | 2.0% | 24.9 | 7.1% |
{6}
b. Linearity/assay reportable range:
Linearity: The analytical measuring range (AMR) of the assay is defined by the lowest and highest points on the master curve, e.g., 1.5–550 CU. The linearity across this range was evaluated by a study according to CLSI EP6-A. The linearity was evaluated using four human serum samples with various anti-HMGCR antibody concentrations which were combined with another human serum sample containing low levels of anti-HMGCR antibodies in 10% increments (from 0% to 90% of low sample) to obtain values that cover the entire AMR. Each dilution was tested in duplicate. Percentage recovery of obtained mean results was calculated compared to the expected results. Results were analyzed by performing linear regression analysis and identifying the best fitting polynomial. For all samples, it was found that the best fitting polynomial is linear, except for sample 3, where the best fitting polynomial was a second order polynomial. The deviation from linearity between the linear regression and the second order polynomial met the acceptance criteria. The linear regression analysis was performed using the samples falling within the master curve and the results of samples within AMR are summarized as follows:
| Sample | Test Range (CU) | Slope (95% CI) | Y-intercept (95% CI) | R² | % Recovery |
| --- | --- | --- | --- | --- | --- |
| 1 | 61.0–610.2 | 0.97 (0.92–1.01) | 3.9 (-3.5–11.3) | 1.00 | 92.0–105.2% |
| 2 | 13.8–138.0 | 0.98 (0.95–1.01) | -0.1 (-1.3–1.05) | 1.00 | 90.8–100.0% |
| 3 | 2.2–21.5 | 0.94 (0.91–0.98) | 0.0 (-0.2–0.2) | 1.00 | 90.2–100.0% |
| 4 | 1.0–10.4 | 1.03 (0.99–1.06) | 0.0 (-0.1–0.1) | 1.00 | 98.6–106.7% |
| Combined | 1.0–610.2 | 0.98 (0.96–0.99) | 0.1 (0.0–0.2) | 1.00 | 90.2–106.7% |
Auto-rerun: To validate the auto-rerun function with 1:20 dilutions, three high positive specimens with anti-HMGCR antibody concentrations above assay measuring range were run with the auto-rerun function enabled on the BIO-FLASH. The same set of samples prepared manually with a 1:20 fold dilution were used as reference values and tested with the concentration of 702.0, 1056.0 and 1272.0 CU, respectively. The percent (%) recovery values for results obtained with the auto-rerun results compared to the results obtained with the manual dilution were between 100% and 105.7%.
Hook effect: Two high positive samples with anti-HMGCR antibody concentrations above the assay measuring range (5193.6 CU and 744.0 CU) were tested to assess for a potential hook effect. No hook effect was observed up to 5193.6 CU.
{7}
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Traceability: There is no international reference material for anti-HMGCR antibodies that allows for the standardization of anti-HMGCR antibodies detection assays. Calibrator and Control values are directly traceable to in-house Standards that are used to create the Master Curves for the QUANTA Flash HMGCR assay.
Value assignment: The QUANTA Flash HMGCR Calibrators and Controls are manufactured by diluting human serum that contains a high concentration of anti-HMGCR antibodies with a stabilizer and preservative solution. The human serum is obtained from commercial sources and it is tested for markers of infectious substances. Upon completion of the manufacturing process, the Calibrators and Controls are tested on at least two instruments, with at least two lots of reagent cartridge, and in replicates of five to obtain a minimum of 10 data points to determine final value assignment.
| | Target Value (CU) | Target Range (CU) |
| --- | --- | --- |
| QUANTA Flash HMGCR Calibrators | | |
| Calibrator 1 | 10 | 8–12 |
| Calibrator 2 | 200 | 180–220 |
| QUANTA Flash HMGCR Controls | | |
| Negative control | 10 | 8–12 |
| Positive control | 50 | 40–60 |
Stability:
Kit stability (unopened): Kit stability studies were performed. Accelerated stability study was performed for three weeks at 37°C using three lots and six samples run in duplicate at the following concentrations: 10.6 CU, 19.4 CU, 49.3 CU, 62.7 CU, 139.0 CU, and 317.2 CU. The accelerated stability study supports a claim for a 12-month shelf life. Real-time stability study has been performed at six-month intervals on the QUANTA Flash HMGCR reagents, to verify the 12-month expiration date assigned based on the accelerated stability study. A negative sample (Negative Control), a low positive sample (Positive Control) and a high positive patient sample were tested in replicates of six (replicates of nine at time zero). At the time of the submission, results were available up to 12 months for the Reagent Cartridge. The results of the real-time stability study performed to date support a claim of 12 months stability for unopened reagent cartridge.
On-board (In-use) stability: On-board stability study was performed for the reagent cartridge:
Two lots of cartridges were tested with four serum specimens (with different reactivity levels) along with the Negative and Positive Controls. The specimens were tested periodically up to 61 days. Percent recoveries were calculated compared to the
{8}
day zero average values, and linear regression analysis was performed by plotting % recovery against the number of days.
The results support on-board stability claim of 60 days.
**Sample stability**: The study was performed with four samples (negative, around the cut-off, moderate positive and high positive samples) for 14 days while stored at 2–8°C, and up to 48 hours while stored at room temperature (RT). In addition, the samples were tested for stability for three repeated freeze/thaw cycles. The results support sample stability for 48 hours of storage at RT, for 14 days of storage at 2–8°C, and for three freeze/thaw cycles when samples are stored at or below -20°C.
d. Detection limit:
**Limit of Blank (LoB)** was determined by assaying four blank samples in five replicates per sample over three days with two reagent lots. Sixty data points were generated for each lot. LoB was calculated as the 95th percentile using the parametric method for one lot, and non-parametric method for the other lot, as the dataset showed non-normal distribution. The LoB was determined to be 0.0 CU for each lot.
**The Limit of Detection (LoD)** was determined by assaying four low level samples prepared by mixing human serum samples with high and low levels of anti-HGMCR antibodies. Each sample was tested in five replicates over three days with two reagent lots. LoD was calculated as the LoB + 1.652 x SD (standard deviation) of the replicates for the low level samples. The LoD of the QUANTA Flash HMGCR assay was determined to be 0.2 CU, which is below the AMR of the assay.
**The Limit of Quantitation (LoQ)** was determined by assaying four low level samples prepared by mixing human serum samples with high and low levels of anti-HGMCR antibodies that were run in five replicates over three days with two reagent lots. There were 30 data points per sample. The LoQ of the QUANTA Flash HMGCR assay was determined to be 1.5 CU, which has been set as the lower limit of the analytical measuring range.
e. Analytical specificity:
**Endogenous Interference**: Four human serum specimens, one high positive (257.4 CU), one moderately positive (125.2 CU), one around the cutoff (18.0 CU) and one negative sample (7.4 CU) were tested. Interfering substances were spiked into every specimen in 10% of total specimen volume at the following concentrations: bilirubin (1 mg/ml, 0.5 mg/ml, 0.25 mg/ml); hemoglobin (2 mg/ml, 1 mg/ml, 0.5 mg/ml); triglycerides (1000 mg/dl, 500 mg/dl, 250 mg/dl); cholesterol (332.5 mg/dl, 166.25 mg/dl, 83.125 mg/dl); human IgG (70 mg/dl, 35 mg/dl, 17.5 mg/dl); Rheumatoid Factor IgM (153.4 IU/ml, 76.7 IU/ml, 38.35 IU/mL). The resulting samples were tested in triplicates with the QUANTA Flash HMGCR assay. Recovery of the unit values was calculated compared to control samples spiked with the same volume of
9
{9}
diluents only. No interferences were detected with the aforementioned interferents at the tested concentrations.
Exogenous Interference: Four human serum specimens, one moderately positive (99.5 CU), one low positive (42.6 CU), one around the cutoff (24.1 CU), and one negative sample (9.8 CU) were tested. Interfering substances were spiked into every specimen in 10% of total specimen volume at the following concentrations: atorvastatin (600 mg/ml); PQQ (24 µg/mL); simvastatin (600 ng/ml); methylprednisolone (36 µg/mL); coenzyme Q (0.72 mg/ml). The resulting samples were assessed in triplicates with the QUANTA Flash HMGCR assay. Recovery of the unit values was calculated compared to control samples spiked with the same volume of diluents only. No interferences were detected with the aforementioned interferents at the tested concentrations.
f. Assay cut-off:
The QUANTA HMGCR cut-off was determined by testing a set of samples from a reference population of 157 subjects [23 apparently healthy donors, 22 infectious disease controls (HBV, HCV, syphilis), 18 scleroderma controls, 20 systemic lupus erythematosus controls, 48 end stage renal disease controls, 14 false positive cohort (high reactivity samples) and 12 diagnosed idiopathic inflammatory myopathy (IIM) patient specimens]. The cut-off was established as 20 CU (30,000 RLU) based on the 99th percentile of the results obtained.
| | Positive | Negative |
| --- | --- | --- |
| QUANTA Flash HMGCR | ≥ 20 CU | < 20 CU |
2. Comparison studies:
a. Method comparison with predicate device:
Despite the fact that the two assays aid in the diagnosis of IIM, there is no apparent overlap between the two types of autoantibodies in IIM diagnosed subjects. Consequently, a method comparison study between the QUANTA Flash HMGCR and the QUANTA Flash Jo-1 is not applicable.
b. Matrix comparison:
Not applicable
3. Clinical studies:
a. Clinical Sensitivity and Clinical Specificity:
A total of 723 characterized samples, none of which were used for establishing the reference range, were included in the clinical validation for the QUANTA Flash
10
{10}
HMGCR. The distribution of the samples and the anti-HMGCR positivity rate is in the table below:
| Patient Group | N | N Positive | % Positive |
| --- | --- | --- | --- |
| Infectious (HBV) | 14 | 0 | 0.0% |
| Infectious (HCV) | 13 | 0 | 0.0% |
| Infectious (HIV) | 13 | 0 | 0.0% |
| Infectious (Syphilis) | 12 | 0 | 0.0% |
| Systemic lupus erythematosus (SLE) | 80 | 0 | 0.0% |
| Sjögren's syndrome (SS) | 44 | 0 | 0.0% |
| Scleroderma (SSc) | 59 | 1 | 1.7% |
| Mixed Connective Tissue Disease (MCTD) | 36 | 0 | 0.0% |
| Celiac Disease (CD) | 25 | 0 | 0.0% |
| Rheumatoid Arthritis (RA) | 39 | 0 | 0.0% |
| Fibromyalgia | 13 | 0 | 0.0% |
| Hypothyroidism | 14 | 0 | 0.0% |
| Lyme Disease | 15 | 0 | 0.0% |
| Polymyalgia Rheumatica | 13 | 0 | 0.0% |
| Primary Raynaud's Syndrome | 15 | 0 | 0.0% |
| Sarcoidosis | 15 | 0 | 0.0% |
| Breast cancer | 10 | 0 | 0.0% |
| Colorectal cancer | 10 | 0 | 0.0% |
| Lung cancer | 10 | 0 | 0.0% |
| Ovarian cancer | 10 | 0 | 0.0% |
| Paraneoplastic syndrome | 6 | 0 | 0.0% |
| Total Controls | 466 | 1 | 0.2% |
| Dermatomyositis (DM) | 67 | 0 | 0.0% |
| Amyopathic Dermatomyositis | 8 | 0 | 0.0% |
| Juvenile Dermatomyositis | 13 | 1 | 7.7% |
| Polymyositis (PM) | 88 | 9 | 10.2% |
| Inclusion Body Myositis | 13 | 0 | 0.0% |
| Overlap | 1 | 0 | 0.0% |
| Immune Mediated Necrotizing Myopathy (IMNM) | 67 | 55 | 82.1% |
| Total idiopathic inflammatory myopathy (IIM) | 257 | 65 | 25.3% |
| Total | 723 | - | - |
{11}
Clinical performance data, including sensitivity and specificity, are summarized in the following tables:
| Clinical Performance N = 723 | Diagnosis | | |
| --- | --- | --- | --- |
| | IIM | Controls | Total |
| Positive | 65 | 1 | 66 |
| Negative | 192 | 465 | 657 |
| Total | 257 | 466 | 723 |
| QUANTA Flash HMGCR | Clinical Performance Characteristics (95% Confidence Interval) |
| --- | --- |
| Sensitivity | 25.3% (20.4% – 30.9%) |
| Specificity | 99.8% (98.8% – 100.0%) |
| PPV | 98.5% (90.1% – 99.8%) |
| NPV | 70.8% (69.3% – 72.2%) |
b. Other clinical supportive data (when a. and b. are not applicable):
Not applicable
4. Clinical cut-off:
Not applicable
5. Expected values/Reference range:
The expected value in the normal population is "negative". Anti-HMGCR antibody levels were analyzed using the QUANTA Flash HMGCR on a panel of 100 apparently healthy blood donors (50 females/50 males, ages 17 to 57 years, with an average and median age of 34 years). With a cut-off of 20 CU, all samples were negative with the QUANTA Flash HMGCR. The mean concentration was $< 1.8$ CU, and the values ranged from $< 1.5$ to 8.0 CU.
# N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
# O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.