Indigo Aspiration System - Modified 110 Aspiration Tubing
K180939 · Penumbra, Inc. · QEW · May 3, 2018 · Cardiovascular
Device Facts
Record ID
K180939
Device Name
Indigo Aspiration System - Modified 110 Aspiration Tubing
Applicant
Penumbra, Inc.
Product Code
QEW · Cardiovascular
Decision Date
May 3, 2018
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.5150
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
INDIGO Aspiration Catheters and Separators: As part of the INDIGO Aspiration System, the INDIGO Aspiration Catheters and Separators are indicated for the removal of fresh, soft emboli and thrombi from vessels of the peripheral and venous systems. INDIGO Aspiration Tubing: As part of the INDIGO Aspiration System, the INDIGO Sterile Aspiration Tubing is indicated to connect the INDIGO Aspiration Catheters to the Penumbra Aspiration Pump. Penumbra Aspiration Pump: The Penumbra Aspiration Pump is indicated as a vacuum source for Penumbra Aspiration Systems.
Device Story
Indigo Aspiration System removes thrombus from peripheral/venous vasculature via continuous aspiration. System components: Aspiration Catheter, Separator, Aspiration Tubing, and Penumbra Aspiration Pump. Physician introduces catheter through guide catheter/sheath to occlusion site; pump provides vacuum source connected via tubing. Separator may be deployed through catheter to clear lumen blockages. Tubing includes flow switch for physician control. Used in clinical settings by physicians. System benefits patients by facilitating mechanical removal of occlusive thrombi, restoring vessel patency.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench-top testing, including dimensional/visual inspection, suction connector/canister lid compatibility, rotating luer/RHV compatibility, lumen ovalization under vacuum, joint leak testing, flow control switch function, and system compatibility testing. Biocompatibility testing per EN ISO 10993-1 performed for modified tubing.
Technological Characteristics
System comprises aspiration catheter, separator, aspiration tubing, and vacuum pump. Tubing materials are biocompatible interventional grade. Connectivity: Mechanical vacuum connection via tubing. Sterilization: EO (catheters/separators/tubing). Software: None (mechanical/fluidic system).
Indications for Use
Indicated for removal of fresh, soft emboli and thrombi from peripheral arterial and venous vessels. Used with Penumbra Aspiration Pump as vacuum source.
Regulatory Classification
Identification
An embolectomy catheter is a balloon-tipped catheter that is used to remove thromboemboli, i.e., blood clots which have migrated in blood vessels from one site in the vascular tree to another.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
May 3, 2018
Penumbra, Inc. Mr. Richard Kimura Senior Regulatory Specialist One Penumbra Place Alameda, California 94502
Re: K180939
Trade/Device Name: Indigo Aspiration System Regulation Number: 21 CFR 870.5150 Regulation Name: Embolectomy Catheter Regulatory Class: Class II Product Code: DXE Dated: April 9, 2018 Received: April 10, 2018
Dear Mr. Kimura:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
## Page 2 - Mr. Richard Kimura
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Kenneth J. Cavanaugh -S
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K180939
Device Name Indigo Aspiration System
### Indications for Use (Describe)
INDIGO Aspiration Catheters and Separators:
As part of the INDIGO Aspiration System, the INDIGO Aspiration Catheters and Separators are indicated for the removal of fresh, soft emboli and thrombi from vessels of the peripheral and venous systems.
INDIGO Aspiration Tubing:
As part of the INDIGO Aspiration System, the INDIGO Sterile Aspiration Tubing is indicated to connect the INDIGO Aspiration Catheters to the Penumbra Aspiration Pump.
Penumbra Aspiration Pump:
The Penumbra Aspiration Pump is indicated as a vacuum source for Penumbra Aspiration Systems.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the word "Penumbra" in a bold, red font. To the right of the word is a circular logo, also in red, with a white "P" inside. The logo is a stylized "P" formed by a white line that creates the shape of the letter within the red circle.
#### 1 510(k) Summary
(as required by 21 CFR 807.92)
Pursuant to Section 12, Part (a)(i)(3A) of the Safe Medical Devices Act of 1990, Penumbra, Inc. is providing the summary of Substantial Equivalence for the Indigo Aspiration System® with modified 110 Aspiration Tubing.
#### 1.1 Sponsor/Applicant Name and Address
Penumbra, Inc. One Penumbra Place Alameda, CA 94502 USA
#### 1.2 Sponsor Contact Information
Richard Kimura Senior Regulatory Affairs Specialist Phone: (510) 995-2034 FAX: (510) 217-6414 Email: rkimura@penumbrainc.com
#### 1.3 Date of Preparation of 510(k) Summary
April 9, 2018
#### 1.4 Device Trade or Proprietary Name
Indigo Aspiration System®
#### 1.5 Device Classification
| Regulatory Class: | II |
|-----------------------|-----------------------|
| Classification Panel: | Cardiovascular |
| Classification Name: | Catheter, Embolectomy |
| Regulation Number: | 21 CFR §870.5150 |
| Product Code: | DXE |
#### 1.6 Predicate Device
| 510(k) Number / Clearance Date | Name of Device | Name of Manufacturer |
|----------------------------------|--------------------------|---------------------------------------------------------------|
| Predicate Device | | |
| K180105 cleared on March 8, 2018 | Indigo Aspiration System | Penumbra, Inc.<br>One Penumbra Place<br>Alameda, CA 94502 USA |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the word "Penumbra" in a bold, red font. To the right of the word is a circular logo, also in red. The logo features a white, stylized "P" shape within the circle.
#### 1.7 Predicate Comparison
| | Predicate Device | Subject Device1 |
|-------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------|
| Trade Name | Indigo Aspiration System | SAME |
| 510(k) No. | K180105 | To Be Determined |
| Classification | Class II, DXE | SAME |
| Indication for Use | INDIGO Aspiration Catheters and<br>Separators<br>As part of the INDIGO Aspiration System,<br>the INDIGO Aspiration Catheters and<br>Separators are indicated for the removal of<br>fresh, soft emboli and thrombi from vessels<br>of the peripheral arterial and venous<br>systems.<br><br>INDIGO Aspiration Tubing<br>As part of the INDIGO Aspiration System,<br>the INDIGO Sterile Aspiration Tubing is<br>indicated to connect the INDIGO Aspiration<br>Catheters to the Penumbra Aspiration<br>Pump.<br><br>Penumbra Aspiration Pump<br>The Penumbra Aspiration Pump is indicated<br>as a vacuum source for Penumbra<br>Aspiration Systems. | SAME |
| Aspiration Tubing | | |
| | Current 110 Aspiration Tubing | Modified 110 Aspiration Tubing |
| Tubing Inner<br>Diameter (ID) | 0.110 in | SAME |
| Tubing Outer<br>Diameter (OD) | 0.188 in | SAME |
| Overall Length | 112.0 in | 100.0 in |
| Distal Length | 7.0 in | N/A (Single Piece Construction) |
| Materials | Biocompatible, commonly utilized for<br>interventional devices | SAME |
| Packaging<br>Configuration | Individually packaged | SAME |
| Sterilization | EO | SAME |
| Shelf Life | 36 months | SAME |
<sup>1</sup> The Indigo System Aspiration Catheters, Separators, and Aspiration Pumps are unchanged and remain identical to those of the currently cleared Indigo System (K180105).
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Penumbra, Inc., a global healthcare company focused on innovative therapies. The word "Penumbra" is written in a bold, red font. To the right of the word is a red circle with a stylized white "P" inside. The logo is simple and modern, and it is likely used on the company's website, marketing materials, and products.
#### 1.8 Device Description
The Indigo Aspiration System ("Indigo System") is intended for the removal of fresh, soft emboli and thrombi from vessels of the peripheral and venous systems using the Indigo Aspiration Catheter, Indigo Separator, Indigo Aspiration Tubing, and Penumbra Aspiration Pump. The Indigo System was most recently cleared under K180105.
The Indigo System is designed to remove thrombus from the vasculature using continuous aspiration. The Aspiration Catheter targets aspiration from the Aspiration Pump directly to the thrombus. The Separator may be used to clear the lumen of the Aspiration Catheter should it become blocked with thrombus. The use of the Separator may not be necessary when using an Aspiration Catheter with an I.D. of 0.054in or larger. The Aspiration Catheter is introduced through a guide catheter or long femoral sheath and into the site of the primary occlusion. The Aspiration Catheter is used with the Aspiration Pump to aspirate thrombus from an occluded vessel. As needed, a Separator may be deployed from the Aspiration Catheter to assist with thrombus removal. The Separator is advanced and retracted through the Aspiration Catheter at the proximal margin of the primary occlusion to facilitate clearing of the thrombus from the Aspiration Catheter tip. For the aspiration source, the Aspiration Catheter is used in conjunction with the Aspiration Pump, which is connected using the Aspiration Tubing and Canister. The Aspiration Catheter may be provided with a steam shaping mandrel and rotating hemostasis valve, and a peelable sheath. The Separator is provided with an introducer and torque device. The Aspiration Catheters and Separators are visible under fluoroscopy.
## The Indigo Aspiration Tubing
The Indigo Aspiration Tubing connects the Penumbra Aspiration Pump Canister to the Indigo Aspiration Catheter within the sterile field, providing a means for introducing vacuum during procedures. The Aspiration Tubing has a flow switch that allows the physician to start and stop the flow of aspiration. The Aspiration Tubing is available in three inner diameters: 0.071 in [1.80mm] for IST1, 0.088 in [2.24mm] for IST2, and 0.110 in [2.79mm] for IST3 and IST4. The Aspiration Tubing is designed and composed of materials which are commonly used in interventional devices.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the Penumbra logo. The word "Penumbra" is written in a sans-serif font in a dark red color. To the right of the word is a circular logo, which is also dark red with a white "P" inside. The logo is simple and modern.
#### 1.9 Indications for Use
## INDIGO Aspiration Catheters and Separators
As part of the INDIGO Aspiration System, the INDIGO Aspiration Catheters and Separators are indicated for the removal of fresh, soft emboli and thrombi from vessels of the peripheral arterial and venous systems.
## INDIGO Aspiration Tubing
As part of the INDIGO Aspiration System, the INDIGO Sterile Aspiration Tubing is indicated to connect the INDIGO Aspiration Catheters to the Penumbra Aspiration Pump.
### Penumbra Aspiration Pump
The Penumbra Aspiration Pump is indicated as a vacuum source for Penumbra Aspiration Systems.
## 1.10 Summary of Non-Clinical Data
As required under Section 12, Part (a)(i)(3A) of the Safe Medical Devices Act of 1990, a summary of any information regarding substantial equivalence of the device follows.
Included in this section are summary descriptions of the testing which substantiates the performance of the subject Indigo System with modified 110 Aspiration Tubing as well as its substantial equivalence to the predicate device:
- Biocompatibility o
- Design Verification (Bench-Top Testing) ●
- Shelf Life o
- Sterilization ●
- o Packaging
The subject Indigo System with modified 110 Aspiration Tubing met all established requirements.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the word "Penumbra" in a red sans-serif font, followed by a red circle with a white "P" inside. The "P" is stylized with a horizontal line extending from the top of the "P" to the edge of the circle. The logo is simple and modern, and the red color gives it a sense of energy and excitement. The word "Penumbra" is likely the name of a company or product, and the logo is likely used to identify the company or product.
## 1.10.1 Biocompatibility
Biocompatibility testing was performed on the modified 110 Aspiration Tubing and the results are summarized in the table below. The studies were selected in accordance with EN ISO 10993-1 guidelines (Biological Evaluation of Medical Devices) for a limited exposure (≤24 hours), surface device, with skin contact. All studies were conducted pursuant to 21 CFR, Part 58, Good Laboratory Practices (GLP). The modified 110 Aspiration Tubing met all acceptance criteria and successfully passed all biocompatibility tests per EN ISO 10993-1 guidelines.
| Test / Standard | Acceptance Criteria | Results | Pass / Fail |
|--------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------|-------------|
| Cytotoxicity (MEM Elution) / EN ISO 10993-5 | Sample extracts must yield a cell lysis grade of 2 or lower | Grade 2: Mild | Pass |
| Sensitization / EN ISO 10993-10 | Test Group shall yield Grade<br>1 score on Magnusson and<br>Kligman scale (provided<br>Control Grade < 1) | Grade 0: No visible change | Pass |
| Irritation (Intracutaneous<br>Reactivity Irritation Test) /<br>EN ISO 10993-10 | The difference in the mean test<br>article and mean control score<br>must be grade 1.0 or lower | Grade 0.0 difference (saline<br>extract) and Grade 0.1<br>difference (sesame oil extract) | Pass |
### Summary of Biocompatibility Testing for Modified 110 Aspiration Tubing
The subject and predicate Indigo System Aspiration Catheters and Separators are identical (most recently cleared under K180105). There are no changes to the materials and processes of the Indigo System Aspiration Catheters and Separators, the biocompatibility data of which were reviewed and cleared under previous premarket notifications for the Indigo System. Therefore, no additional biocompatibility testing is required or was performed for the Indigo System Aspiration Catheters and Separators.
The Penumbra Aspiration Pumps (Pump MAX and Engine Pump) are non-sterile reusable capital equipment and their associated canisters are non-sterile, single use only. The pumps and canisters do not contact the patient, nor are they introduced into the sterile field. As such, biocompatibility testing is not required and was not performed for the Aspiration Pumps and canisters.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Penumbra, Inc., a global healthcare company focused on innovative therapies. The logo consists of the word "Penumbra" in a bold, red font, followed by a circular emblem. The emblem is also red and features a stylized "P" shape in white, creating a distinctive visual identity for the company.
# 1.10.2 Design Verification (Bench-top Testing)
With the exception of the modified flow switch assembly and the single piece of continuous tubing, the modified 110 Aspiration Tubing will have the same specifications as the current 110 Aspiration Tubing. Bench-top testing was conducted to evaluate the physical and mechanical properties of the modified 110 Aspiration Tubing. All benchtop studies were conducted using good scientific practices and statistical sampling methods as required by the Penumbra Design Control procedures. Performance testing was based on the design specifications, risk analysis, performance standards, and guidance documents. All testing was performed using units which were 2x EO-sterilized.
A summary of all testing performed on the modified 110 Aspiration Tubing is provided in the table below. The modified 110 Aspiration Tubing met all acceptance criteria and passed all tests.
| Test/<br>Test Subject | Attribute | Acceptance Criteria | Result |
|----------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------|--------|
| Dimensional/<br>Visual Inspection | These evaluations confirm that the test<br>units used in Design Verification<br>testing meet all dimensional and visual<br>specifications. | 100% Must meet Specification | Pass |
| Suction<br>Connector /<br>Canister Lid<br>Compatibility | Suction Connector of Aspiration<br>Tubing Assembly securely attaches to<br>Pump Canister lid via press fit. | 100% Must meet Specification | Pass |
| Rotating Luer/<br>RHV<br>Compatibility | Rotating Luer of Aspiration Tubing<br>Assembly securely connects to RHV<br>port. | 100% Must meet Specification | Pass |
| Aspiration<br>Tubing Lumen<br>Ovalization<br>under Vacuum | Aspiration Tubing Assembly maintains<br>functionality and maintains an open<br>lumen at vacuum pressure per product<br>specification. | 100% Must meet Specification | Pass |
| Aspiration<br>Tubing Joint<br>Leak under<br>Vacuum | Aspiration Tubing Assembly maintains<br>functionality with no leaks at vacuum<br>pressure per product specification. | 100% Must meet Specification | Pass |
| Flow Control<br>Switch Function | Flow Control Switch completely and<br>immediately stops fluid flow after a<br>specified number of ON/OFF cycles. | 100% Must meet Specification | Pass |
| Penumbra<br>Aspiration<br>System<br>Compatibility<br>with Aspiration | The Aspiration Tubing Assembly is<br>compatible with Penumbra Aspiration<br>System (Clot can be removed under<br>minimum vacuum pressure per product<br>specification). | 100% Must meet Specification | Pass |
## Summary of Bench-Top Testing for Modified 110 Aspiration Tubing
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the Penumbra logo. The logo consists of the word "Penumbra" in a red sans-serif font, followed by a red circle with a white "P" inside. The "P" is stylized with a horizontal line extending from the top of the letter to the edge of the circle. The logo is simple and modern, and the red color gives it a sense of energy and excitement.
| Test/<br>Test Subject | Attribute | Acceptance Criteria | Result |
|----------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------|--------|
| Catheter and<br>Separator | | | |
| Penumbra<br>Aspiration<br>System<br>Compatibility<br>with Aspiration<br>Catheter | The Aspiration Tubing Assembly is<br>compatible with Penumbra Aspiration<br>System (Clot can be removed under<br>minimum vacuum pressure per product<br>specification). | 100% Must meet Specification | Pass |
| Suction<br>Connector /<br>Tubing Joint<br>Tensile | Break force per product specification. | 100% Must meet Specification | Pass |
| Rotating Male<br>Luer / Tubing<br>Joint Tensile | Break force per product specification. | 100% Must meet Specification | Pass |
## 1.10.3 Shelf Life
The modified 110 Aspiration Tubing has demonstrated device stability for 36 months based on accelerated aging and may be labelled with a 36 month shelf life. Test units were 2x EO-sterilized and underwent transportation conditioning per ASTM D4169, Distribution Cycle 3, Assurance Level 2.
The subject and predicate Indigo System Aspiration Catheters and Separators are identical. Therefore, there are no changes to the previously determined stability of the Aspiration Catheters and Separators, the data for which were reviewed and cleared under previous Indigo System premarket notifications. No additional shelf life testing is required or was performed for the Indigo System Aspiration Catheters and Separators.
The subject and predicate Penumbra Aspiration Pumps (Pump MAX and Engine Pump) and their associated canisters are identical. Furthermore, the Aspiration Pumps are nonsterile reusable capital equipment. Therefore, shelf life testing is not applicable to the Aspiration Pumps. Both the Pump MAX and Engine Pump have established a 500 hour minimum operating life based on completed testing, the data for which were reviewed and cleared under previous Indigo System premarket notifications.
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the logo for Penumbra, Inc., a global healthcare company focused on innovative therapies. The logo consists of the word "Penumbra" in a red, sans-serif font, followed by a circular emblem. The emblem is red with a white "P" shape inside, creating a modern and recognizable brand identity. The logo is clean and professional, reflecting the company's focus on medical technology.
## 1.10.4 Sterilization
The modified 110 Aspiration Tubing has demonstrated sterility by EO in accordance with EN ISO 11135. All test samples met the acceptance criteria for EO residual testing per EN ISO 10993-7, Comparative Resistance Testing per AAMI TIR 28, and Endotoxin (LAL) Testing per ANSI/AAMI ST72.
The subject and predicate Indigo System Aspiration Catheters and Separators (most recently cleared under K180105) are identical. There are no changes to the previously provided sterilization data of these devices, which were reviewed and cleared under previous Indigo System premarket notifications. No additional sterilization testing is required or was performed for these devices.
Sterilization testing is not applicable to the Penumbra Aspiration Pumps (Pump MAX/Canister Tubing and Engine Pump/Canister). Both are supplied non-sterile and are not intended to be sterilized.
## 1.10.5 Packaging
The packaging for the modified 110 Aspiration Tubing utilizes materials that are commonly used in other Penumbra device packaging. The modified 110 Aspiration Tubing is placed in a Nylon/ADH/HDPE/ Tyvek® pouch which is heat-sealed to maintain sterility post EO sterilization. The inner pouch is then placed in a display box for further protection. Both the inner pouch and display box receive a product label. An Instruction for Use (IFU) is placed inside the display box of each unit.
The packaging materials and process of the subject and predicate Indigo System Aspiration Catheters, Separators, and Aspiration Pumps (Pump MAX/Canister Tubing and Engine Pump/Canister) are identical (most recently cleared under K180105). There are no changes to the previously provided packaging material listing or the packaging process for these devices, which were reviewed and cleared under previous Indigo System premarket notifications. No additional packaging testing is required or was performed for these devices.
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the word "Penumbra" in a bold, red font. To the right of the word is a circular logo, also in red, with a white stylized "P" inside. The logo appears to be a stylized representation of the company's name.
# 1.11 Summary of Substantial Equivalence
The subject Indigo System with modified 110 Aspiration Tubing is substantially equivalent to the predicate Indigo System with regard to indications, intended use, design, performance, materials, sterilization, and packaging.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.