K180918 · Planmed OY · OAS · Nov 30, 2018 · Radiology
Device Facts
Record ID
K180918
Device Name
Planmed Verity
Applicant
Planmed OY
Product Code
OAS · Radiology
Decision Date
Nov 30, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1750
Device Class
Class 2
Indications for Use
Planmed Verity is intended to be used for X-ray computed cone beam tomography imaging of anatomies within upper and lower extremities, head and neck.
Device Story
Planmed Verity is a cone beam computed tomography (CBCT) system for 3D imaging of extremities, head, and neck. It uses an amorphous silicon flat panel detector to convert X-ray photons into digital data. Operated by clinicians via an integrated workstation and Verity Manager software, the system acquires, processes, and displays images for diagnostic review. It supports connectivity with hospital information systems (HIS/RIS/MIS) via worklists. New features include ULD (Ultra Low Dose) noise filtration and CALM (Correction Algorithm for Latent Movement) for motion blur reduction. An optional patient support tray facilitates head and neck imaging. The device aids clinical decision-making by providing high-quality 3D anatomical images while reducing radiation dose and potential re-takes due to patient movement, benefiting patients through improved diagnostic accuracy and safety.
Clinical Evidence
Clinical image evaluation performed by experienced radiologists using representative sample scans. Overall image quality was deemed acceptable for all cases. Effectiveness of the CALM motion blur reduction algorithm was verified using a skull phantom with simulated motion blur, showing clear reduction in blur. The ULD (Ultra Low Dose) protocol was evaluated in a clinical study in Helsinki, Finland, where radiologists confirmed clinical suitability for head and neck imaging. No new physical laboratory testing was required due to the use of the same imaging chain and detector performance specifications as the predicate.
Technological Characteristics
CBCT system; amorphous silicon flat panel detector (127 um pixel size); integrated PC workstation; X-ray generator/tube same as predicate. Connectivity: HIS/RIS/MIS integration. Software: Verity Manager with ULD noise filtration and CALM motion blur reduction. Standards: IEC 60601-1, IEC 60601-1-3, IEC 60601-1-6, IEC 60601-2-28, IEC 60601-2-54, IEC 60601-1-2 (EMC), ISO 10993-1 (biocompatibility).
Indications for Use
Indicated for X-ray computed cone beam tomography imaging of upper and lower extremities, head, and neck anatomies.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Predicate Devices
Planmed Verity cone beam computed tomography X-ray System (K143435)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font size below that.
November 30, 2018
Planmed Oy Lars Moring Regulatory Affairs Manager Sorvaajankatu 7 HELSINKI, 00880 FINLAND
Re: K180918
Trade/Device Name: Planmed Verity Regulation Number: 21 CFR 892.1750 Regulation Name: Computed Tomography X-Ray System Regulatory Class: Class II Product Code: OAS, JAK Dated: April 5, 2018 Received: April 9, 2018
Dear Lars Moring:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Hol 2. Mild
for Robert A. Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K180918
Device Name
Planmed Verity
Indications for Use (Describe)
Planmed Verity is intended to be used for X-ray computed cone beam tomography imaging of anatomies within upper and lower extremities, head and neck.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------|
| <div> <span> X Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
#### I. SUBMITTER
### Manufacturer
Planmed Oy Sorvaajankatu 7 FI-00880 Helsinki, Finland Phone: +358 20 7795 300 Fax: +358 20 7795 396 Contact person: Lars Moring
# U.S. designated agent
Planmed USA Inc. 100 North Gary Avenue, Suite A Roselle, IL 60172 Phone: (630) 894 2200 Fax: (630) 894 4271 Contact person : Ed McDonough
Date Prepared: November 28, 2018
#### II. DEVICE
| Name of Device: | Planmed Verity |
|-----------------------|------------------------------------------------------------|
| Common or Usual Name: | Cone Beam Computed Tomography (CBCT) System |
| Classification Name: | Computed Tomography X-ray System (CT)<br>(21 CFR 892.1750) |
| Regulatory Class: | II |
| Product Code: | OAS, JAK |
#### III. PREDICATE DEVICE
Planmed Verity cone beam computed tomography X-ray System, K143435. This predicate has not been subject to a design-related recall. No reference devices were used in this submission.
#### DEVICE DESCRIPTION IV.
The Planmed Verity is a cone beam computed tomography x-ray system for generating 3D imaging scans of extremity, head and neck anatomies. The Planmed Verity utilizes an amorphous silicon based digital image receptor to capture digital images. The receptor directly converts the incoming X-ray photons to digital image data.
The workflow with Planmed Verity is controlled from the integrated acquisition workstation and Planmed Verity Manager image acquisition and communications software. The patient information is entered manually or received from the hospital. radiology, or x-ray modality information systems (HIS, RIS, or MIS, respectively), as a format of modality worklist. Subsequently, the images are acquired, processed, and displayed for preview. After initial evaluation by the operator, the images are either printed or transferred for soft-copy review.
{4}------------------------------------------------
#### V. INDICATIONS FOR USE
Planmed Verity is intended to be used for X-ray computed cone beam tomography imaging of anatomies within upper and lower extremities, head and neck.
#### VI. COMPARISON OF INDICATIONS FOR USE
An added imaging protocol has been introduced to the system compared with the predicative device version.
The Indications for Use (IFU) statements are similar for both systems. IFU of the predicate system refers to the currently cleared system version with the indications for use covering extremities and head anatomies (mainly maxillofacial and sinus). The new version of the system may also be used for imaging of other head anatomies and the neck. This protocol uses a new patient support tray.
#### VII. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
### General
Both the predicate device and the subject device are using the same basic operating principles and are technically almost identical. The subject device is an improved version of the predicate device. It has the same outer design and no changes to the physical construction visible from the outside with the exception of one added optional patient support tray. X-ray generation and control are the same. The controlling PC workstation of the subject device is integrated into the x-ray device like in the predicate system. The user interface is still the same touch screen like in the predicate device and the managing software is essentially unchanged in its controls and functionality and from the operator's perspective. All labeling material is mainly unchanged. Only the added software features are updated to the user's manual and to the technical manual.
### Integrated detector
The system uses an integrated flat panel detector. The type of the detector has changed slightly to a new version of the same detector. Pixel matrixes are identical, pixel size is 127 um by 127um. Quality assurance with pixel defect acceptance criteria comparison is unchanged.
### X-ray unit
Physical dimensions of the units are the same.
### X-ray field size and SID
X-ray field size, SID, cone beam geometry and scanning angle are unchanged.
### X-ray tube
Units use the same X-ray tube.
### X-ray generator
Units use the same X-ray generator.
### X-ray collimator
Collimator mechanics has been simplified with same functionality and filtration.
### Differences and added features
The complete list of differences of the subject device to the predicate device is as follows:
The integrated PC's brand has been changed from HP to Dell. The Microsoft Windows operating system has been upgraded from Windows 7 to Windows 10.
The Verity Manager software newly includes image enhancement options like ULD (Ultra Low Dose) noise filtration and CALM (Correction Algorithm for Latent Movement) motion blur reduction protocols.
{5}------------------------------------------------
One optional patient supporting tray has been added for the newly introduced head and neck imaging protocols. All previously used patient support options of the predicate system are available and supported also with the subject system.
The amorphous silicon detector is replaced by a new version of the same detector type. The new version of the detector has slightly better imaging characteristics (MTF, DQE and noise performance) as the predicate device. Yet its physical construction, specifications and software interfacing are essentially the same as before.
There are no other differences of the subject system to the predicate system.
The above listed differences are deemed to be non-critical for the subject device compared to the predicate since the system is mainly unchanged. The safety of the system and the imaging performance for the intended use are maintained or improved for the new system version in comparison to the predicate. The added features have been clinically tested and approved by experienced operators and radiologists and found valuable for diagnostic use without adding risks to patient or operator safety or health. Labeling material has been updated to include the new functionality. Dosimetry information has been added to aid in judgement of the radiation safety of intended diagnostic procedures.
#### VIII. PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination.
# Biocompatibility testing
To evaluate biocompatibility of the Planmed Verity CBCT device, we performed safety evaluation of acute and repeat toxicity. As conclusion, none of the compounds measured in the extracts were present at levels which could adversely affect the patient. Thus, there is no risk or concern to the patient's safety from contact with the materials of construction of this X-ray unit.
# Electrical safety and electromagnetic compatibility (EMC)
Electrical safety and EMC testing were conducted on the Planmed Verity CBCT system. The system complies with standards IEC 60601-1, IEC 60601-1-3, IEC 60601-1-6, IEC 60601-2-28, IEC 60601-2-54, ISO 10993-1 for safety, and IEC 60601-1-2 for EMC.
# Software Verification and Validation Testing
Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "Moderate" level of concern.
# Physical laboratory testing
The performance of the system has not changed from the predicate device. Image quality is not affected by the change of the detector type from 2520D to 2520 DX version. Thus no new physical laboratory testing was deemed necessary. Instead a comparison of the manufacturer's (Varex) non-clinical physics data was performed. Both systems use the same software algorithms and interfacing software to control the flat panel detector and to compute the reconstruction. The image acquisition chain or the imaging protocols' number of frames and scan angles have not been altered in the subject system compared to the predicate device.
# Clinical image evaluation
A clinical image evaluation has been performed where an experienced radiologist scored representative sample scans and diagnostic images. The overall image quality was acceptable for all cases and image types.
{6}------------------------------------------------
Parts of the clinical evaluations were performed with a system not being in final finished form of the subject device. More precisely the flat panel detector used has been of type 'D' not 'DX'. However since the imaging performance of both types is essentially equal and would not alter the scoring of the image quality seen by the radiologist we believe that all clinical testing is relevant and acceptable for the verification of the subject device.
It may also be noted that the new software features CALM and ULD can be upgraded to the predicate Verity systems. Once cleared for marketing, these features will benefit also the currently installed base of units and the patient diagnostics and radiation safety.
#### IX. CONCLUSIONS
The renewed UL/CB/CE testing and certification process support the safe use and imaging performance and usability of the device. Both hardware and software verification and validation demonstrate that the Planmed Verity system performs as intended in the specified use conditions. The clinical image evaluation also shows that the device performance is equal to the predicate device that is currently marketed for the same intended use. The added software features CALM and ULD improve image quality and lower the patient dose while maintaining and/or improving diagnostic image quality.
Summary of conclusions drawn from the clinical and non-clinical testing:
Since the overall construction, x-ray generation or control software have not changed significantly there are no new added risks to the subject device. The integrated flat panel detector has changed to a newer version with the same imaging performance specifications. Hence no new physical laboratory testing was deemed necessary. Only the new revision of the flat panel imaging detector performance has been compared for its non-clinical testing data and was found equal or slightly better.
The clinical tests included imaging performance scoring with use of the CALM motion blur reduction software feature, as well as use of the ULD ultra low dose imaging protocol for patient radiation dose reduction.
The clinical test report describes the CALM feature in detail. The scoring using CALM compared to not using it is presented. Based on this evaluation the CALM feature improves the imaging quality in most cases and hence it reduces the risk of re-takes because of patient movement. An added section with results of imaging performance evaluation with CALM is presented.
The effectiveness of the CALM feature was additionally verified with imaging a scull phantom with suitable test inserts. Simulated small and large amplitude motion blur as well as step-like distortion was added to the projection images and the effectiveness of the CALM feature to correct the motion blur was verified. Results show that the CALM algorithm clearly reduces motion blur.
ULD protocol was verified in a separate study at a clinic in Helsinki, Finland. The radiologists evaluated the clinical image quality of the ULD feature to be clinically suitable for imaging head and neck anatomies.
Based on the clinical tests performed it can be concluded that the subject device is as safe and effective for its indicated use as the predicate system when used as instructed.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.