K180900 · Uvbiotek, LLC · OLP · Jul 10, 2018 · General, Plastic Surgery
Device Facts
Record ID
K180900
Device Name
LED Light Therapy Device
Applicant
Uvbiotek, LLC
Product Code
OLP · General, Plastic Surgery
Decision Date
Jul 10, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The LED Light Therapy Device is indicated for the treatment of mild to moderate inflammatory acne.
Device Story
Handheld LED light therapy device for home use; manages mild to moderate inflammatory acne. Device utilizes 48 (KN-7000C) or 8 (KN-7000C2) LED cold light sources to emit red (633±10nm) and blue (417±10nm) light. User selects treatment head; holds device against skin area; timer controls duration. Blue light targets acne; red light follows. Powered by rechargeable lithium-ion battery. Provides non-invasive light-based therapy; intended to reduce acne appearance. User operates device; monitors status via indicators (KN-7000C) or OLED display (KN-7000C2). Benefits include convenient, self-administered acne management.
Clinical Evidence
No clinical data provided. Evidence consists of bench testing (biocompatibility per ISO 10993, electrical safety per IEC 60601 series, photobiological safety per IEC 62471) and usability studies confirming user ability to operate the device and understand labeling.
Technological Characteristics
Handheld LED device; ABS polycarbonate lens cover; red (633nm) and blue (417nm) LEDs; rechargeable Li-ion battery (CR18650-22F); IP22 rating; continuous operation mode. Complies with IEC 60601-1, 60601-1-2, 60601-1-11, 60601-2-57, and IEC 62471. Biocompatibility per ISO 10993-1, -5, -10.
Indications for Use
Indicated for treatment of mild to moderate inflammatory acne in patients.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
July 10, 2018
Uvbiotek, LLC % Jinghua Zhou Regulation Control Manager Guangzhou Junyi Information Technology Co., Ltd. Room 215, Huaming Building, Chebei Road Guangzhou, 511660 Cn
Re: K180900
Trade/Device Name: LED Light Therapy Device Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: OLP Dated: April 6, 2018 Received: April 13, 2018
Dear Jinghua Zhou:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Binita S. Ashar -S
Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known)
Device Name LED Light Therapy Device
Indications for Use (Describe) The LED Light Therapy Device is indicated for the treatment of mild to moderate inflammatory acne.
Type of Use (Select one or both, as applicable)
__ Prescription Use (Part 21 CFR 801 Subpart D)
🇿 Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### Section 5 - 510(k) Summary
Date of Summary Preparation: April 3, 2018
#### 1. Submitter's Identifications
Submitter's Name: UVBIOTEK, LLC Address: 3 Depot Street, Hudson Falls, NY 12839, United States Contact Person: Dave Oberhelman Contact Title: Plant Manager Contact E-mail Address: info@uvbiotek.com Telephone: 518-747-3310 Fax: 518-747-2294
### 2. Correspondent's Identifications
Correspondent's Name: Guangzhou Junyi Information Technology Co., Ltd. Address: Room 215, Huaming Building, Chebei Road, Guangzhou, P.R. China ZIP Code: 511660 Contact Person: Jinghua Zhou Contact Title: Regulation Control Manager Contact E-mail Address: kernel 2016@126.com Telephone: +86-20-82329549 Fax: +86-20-82329549
### 3. Name of the Device
Device Classification Name: Over-The-Counter Powered Light Based Laser For Acne Product Name: Over-The-Counter Powered Light Based Laser For Acne Trade Name: LED Light Therapy Device Model: KN-7000C, KN-7000C2 Classification Panel: General & Plastic Surgery Product Code: OLP Device Classification: Class II
### 4. The Predicate Devices
Clear Bi-Light K153081 K081307 OmniLux Clear U
#### 5. Device Description
The LED Light Therapy Device is a lightweight device which uses specific wavelengths of light, produced by light emitting diodes (LEDs), to manage aesthetic conditions. The device produces light in the red light region of the spectrum (633±10nm) and/or in the blue light region of the spectrum (417±10nm), intended to help reduce the appearance of mild to moderate acne.
{4}------------------------------------------------
This device is made up of a main unit and therapy head. The light gathering cover is also referred to as the snoot. The snoot helps to establish the correct treatment distance.
### Description for KN-7000C:
The device uses 48 LED cold light sources, and uses an external removable single cell lithium ion battery for power, with a continuous usage time of 120 minutes. It can be charged using a power adapter.
### The LED information:
| Items | Red head | Blue head |
|--------------------------------|--------------------|--------------------|
| LED numbers | 48 | 48 |
| Total spectrum peak wavelength | 633nm±10nm | 417nm±10nm |
| Total effective irradiance | 45mW/cm2 ± 5mW/cm2 | 25mW/cm2 ± 5mW/cm2 |
| Total luminous power | 1.3W | 0.78W |
| Total energy fluxes | 9J/cm2 | 5.4J/cm2 |
The specification table is as below:
### SPECIFICATIONS
| Model | KN-7000C |
|-----------------------------------|---------------------------------------------------------------------------------------------------------|
| Safety classification | Class II, has internal power source. |
| Operating mode | Continuous operation |
| IP rating | IP22 |
| Adapter model number | LXCP12-005200DEG, Input: 100-240V a.c. 50/60Hz, 0.5A max. Output: 5V d.c. 2A |
| Main unit | input: 5V d.c. 2A/Internal battery: 3.6Vd.c. 2200mAh |
| Internal battery specification | CR18650-22F 3.6V d.c. 2200mAh |
| Working environment | Temperature: 5~40°C, 41-104° F<br>Relative humidity: 15%-90%<br>Barometric pressure: 700hPa~1060hPa |
| Transport and storage environment | Temperature: -25 to 70°C, -13 to 158° F<br>Relative humidity: ≤90 %<br>Barometric pressure: 500~1060hPa |
| Structural configuration | Handheld |
| Display style | No |
| Effective configuration area | 26cm2±10% |
| Irradiance distance | Hold treatment face in contact with skin |
| Spectrum peak wavelength | Red light 633nm±10nm<br>Blue light 417nm±10nm |
| Effective irradiance | Red light 45 ± 5mW/cm2<br>Blue light 25 ± 5mW/cm2 |
| Timing and functions | The device has a timer, with a timing error of less than<br>±2% of the pre-set value |
{5}------------------------------------------------
| | The device has the function of manually stopping |
|-----------------|--------------------------------------------------|
| | irradiation output. |
| Operation noise | Not exceed 60dB(A) |
### The list of accessories
| No. | Name | Specifications | Quantity | Unit |
|-----|--------------------|------------------|----------|--------|
| 1 | Adapter | LXCP12-005200DEG | 1 | Piece |
| 2 | Protective eyewear | YH-8 | 1 | Pieces |
| 3 | Manual | User Manual | 1 | Copy |
| 4 | Certificate | KN-7000C | 1 | Copy |
Description for KN-7000C2:
The device uses 8 LED cold therapy heads and has high stability, strong uniformity, and a long lifespan. It uses an external removable single cell lithium ion battery for power, with a continuous usage time of 4.5 hours. It can be charged using a power adapter.
### The LED information:
| Items | Red head | Blue head |
|--------------------------------|--------------------|--------------------|
| LED numbers | 8 | 8 |
| Total spectrum peak wavelength | 633nm±10nm | 417nm±10nm |
| Total effective irradiance | 65mW/cm2 ± 5mW/cm2 | 35mW/cm2 ± 5mW/cm2 |
| Total luminous power | 0.28W | 0.16W |
| Total energy fluxes | 8.4J/cm2 | 4.8J/cm2 |
The specification table is as below:
### SPECIFICATIONS
| Model | KN-7000C2 |
|-----------------------------------|------------------------------------------------------|
| Safety classification | Class II, has internal power source. |
| Operating mode | Continuous operation |
| IP rating | IP22 |
| Adapter model number | LXCP12-005200DEG, Input: 100-240V a.c. 50/60Hz, |
| | 0.5A max. Output: 5V d.c. 2A |
| Main unit | input: 5V d.c. 2A/Internal battery: 3.6Vd.c. 2200mAh |
| Internal battery specification | CR18650-22F 3.6V d.c. 2200mAh |
| Working environment | Temperature: 5〜40°C , 41-104°F |
| | Relative humidity: 15%-90% |
| | Barometric pressure: 700hPa~1060hPa |
| Transport and storage environment | Temperature: -25 to 70℃, -13 to 158°F |
| | Relative humidity: ≤90 % |
| | Barometric pressure: 500~1060hPa |
{6}------------------------------------------------
| Structural configuration | Handheld |
|------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Display style | OLED display |
| Effective configuration area | 4cm2±10% |
| Irradiance distance | Hold treatment face in contact with skin |
| Spectrum peak wavelength | Red light 633nm±10nm<br>blue light 417nm±10nm |
| Effective irradiance | Red light 65±5mW/cm2<br>blue light 35±5mW/cm2 |
| Timing and functions | The device has a timer, with a timing error of less than<br>±2% of the pre-set value.<br>The device has the function of manually stopping<br>irradiation output. |
| Operation noise | Not exceed 60dB(A) |
The list of accessories
| No. | Name | Specifications | Quantity | Unit |
|-----|--------------------|------------------|----------|--------|
| 1 | Adapter | LXCP12-005200DEG | 1 | Piece |
| 2 | Protective eyewear | YH-8 | 1 | Pieces |
| 3 | Manual | User Manual | 1 | Copy |
| 4 | Certificate | KN-7000C2 | 1 | Copy |
### 6. Intended Use of Device
The LED Light Therapy Device is indicated for the treatment of mild to moderate inflammatory acne.
{7}------------------------------------------------
### 7. Summary of Substantial Equivalence
| | Proposed Device | Predicate device | Predicate device | Comparison |
|-----------------------|-------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|
| 510k Number | ------ | K153081 | K081307 | ------ |
| Product Code | OLP | OLP | OLP | Same |
| Proprietary Name | LED Light Therapy Device | Clear Bi-Light | OmniLux Clear U | ------ |
| Model | KN-7000C | / | / | ------ |
| Manufacturer | Xuzhou Kernel Medical<br>Equipment Co., Ltd. | Michael Todd, LP | Photo Therapeutics Inc | ------ |
| Indications for Use | The LED Light Therapy<br>Device is indicated for the<br>treatment of mild to moderate<br>inflammatory acne. | The Clear Bi-Light is<br>indicated for the treatment of<br>mild to moderate<br>inflammatory acne. | Indicated to treat mild to<br>moderate acne on the face. | Same |
| OTC or Rx | OTC | OTC | OTC | Same |
| Energy source | LED | LED | LED | Same |
| Treatment heads | Separate blue and red,<br>treatment heads | Separate blue and red,<br>treatment heads | Blue and Red in same head,<br>chosen using selector switch | Same |
| Wavelength(s)<br>(nm) | $417\pm10$<br>$633\pm10$ | 405-420<br>630-660 | 415±5<br>633±6 | Substantially equivalent<br>Wavelengths are in the blue<br>and red spectrums. |
| Treatment control | " ▶/■ " button to turn the<br>irradiator on or off for<br>therapy, hold shoot in contact<br>with skin | On/off button puts device in<br>standby mode.<br>Contact sensor in treatment<br>face, touching device to face<br>activates LEDs. | Three position switch: Off –<br>Blue - Red | Substantially equivalent<br>KN-7000C and Clear Bi-Light<br>have on/off button. Contact<br>sensor in proposed device is<br>the same as in the Clear |
### Table 1 Comparison to Predicate Device for KN-7000C
{8}------------------------------------------------
| | | | | Bi-Light. |
|-----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------|
| Treatment indicator | Indicator will light up when<br>treatment; the indicator is<br>yellow when charging; the<br>indicator light is green when it<br>is charged fully. | Indicator light-solid green<br>while treatment is in process;<br>blinking indicates low battery.<br>Audio signal; long beep when<br>power is turned on, short beep<br>every 30 seconds during<br>treatment; long beep at end of<br>3 minute cycle. | No | Indicators light substantially<br>equivalent to the Clear Bi-<br>Light. Indicators light can<br>indicate related status of the<br>device. |
| Treatment timer | The recommended therapy<br>times are 3 minutes per area<br>after which device returns to<br>standby mode. Must press<br>treatment face against skin to<br>reactivate lights. | Up to 3 minutes treatment per<br>area after which device<br>returns to standby mode. Must<br>press treatment face against<br>skin to reactivate lights. | No | Same as the Clear Bi-Light. |
| Treatment regimen | Hold treatment face in contact<br>with skin. Apply blue light for<br>3minutes per skin area,<br>followed by red light for 3<br>minutes per skin area. Can be<br>used daily. | Hold treatment face in contact<br>with skin. Apply blue light for<br>up to 3minutes per skin area,<br>followed by red light for up to<br>3 minutes per skin area. Can<br>be used daily. | Alternate treatments of blue<br>and red light twice a week<br>for 20 minutes per treatment. | Same as the Clear Bi-Light. |
| Dose rate<br>(mW/cm²) | $25\pm5$ (blue)<br>$45\pm5$ (red) | 31.1(blue)<br>54.6(red) | 40(blue)<br>70(red) | Substantially<br>equivalent-within dose rate<br>ranges of predicate devices. |
| Treatment face<br>area(cm²) | $26\pm10$ % | 20 | 28.7 | Substantially<br>equivalent-within size ranges<br>of predicate devices. |
| Max treatment<br>temperature (°C) | 36.8(blue)<br>36.5(red) | 38(blue)<br>40(red) | 39-43 | Substantially<br>equivalent-within temperature<br>ranges of predicate devices |
| Microprocessor<br>controlled | Yes | Yes | Unknown | Same as the Clear Bi-Light. |
| Handheld | Yes | Yes | Yes | Same |
| Patient contacting<br>material | Rigid ABS Clear<br>polycarbonate lens cover | Rigid ABS Clear<br>polycarbonate lens cover | Rigid ABS Clear | Same |
| Power supply | Adapter model number:<br>LXCP12-005200DEG, Input:<br>100-240V a.c. 50/60Hz, 0.5A<br>max. Output: 5V d.c. 2A<br>Main unit input: 5V d.c.<br>2A/Internal battery: 3.6Vd.c.<br>2200mAh<br>Internal battery specifications:<br>CR18650-22F 3.6V d.c.<br>2200mAh | Lithium-ion rechargeable<br>battery AC charger: 100-240V<br>at 50-60 Hz, 500mA | A separate, universal, power<br>supply converts mains AC<br>power to the DC power<br>required | Substantially equivalent to the<br>Clear Bi-Light. All devices<br>have power adapter. |
| Dimensions | 257mm×165mm×70mm<br>(10.1in×6.5in×2.8in) | 6.4cm×3.5cm×14.6cm<br>(2.5in×1.4in×5.7in) | 6cm×4cm×12cm<br>(2.4in×1.6in×4.7in) | Substantially equivalent<br>Both are light-weight, hand<br>held devices. |
| Weight | 174g(6.1 oz) | 107.1g(3.8 oz) | Unknown | Substantially equivalent<br>Both are light-weight, hand<br>held devices. |
| Standard | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-1-11<br>IEC 60601-2-57 | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-1-11<br>IEC 60601-2-57 | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-1-11<br>IEC 60601-2-57 | Same<br>All devices meet the<br>requirements of effectiveness,<br>safety and biocompatibility. |
| IEC 62471 | IEC 62471 | IEC 62471 | | |
| ISO 10993-1 | ISO 10993-1 | ISO 10993-1 | | |
| ISO 10993-5 | ISO 10993-5 | ISO 10993-5 | | |
| ISO 10993-10 | ISO 10993-10 | ISO 10993-10 | | |
{9}------------------------------------------------
{10}------------------------------------------------
### Table 2 Comparison to Predicate Device for KN-7000C2
| | Proposed Device | Predicate device | Predicate device | Comparison | |
|-------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| 510k Number | ------ | K153081 | K081307 | ------ | |
| Product Code | OLP | OLP | OLP | Same | |
| Proprietary Name | LED Light Therapy Device | Clear Bi-Light | OmniLux Clear U | ------ | |
| Model | KN-7000C2 | / | / | ------ | |
| Manufacturer | Xuzhou Kernel Medical<br>Equipment Co., Ltd. | Michael Todd, LP | OmniLux Clear U | ------ | |
| Indications for Use | The LED Light Therapy<br>Device is indicated for the<br>treatment of mild to moderate<br>inflammatory acne. | The Clear Bi-Light is<br>indicated for the treatment<br>of mild to moderate<br>inflammatory acne. | Indicated to treat mild to<br>moderate acne on the face. | Same | |
| OTC or Rx | OTC | OTC | OTC | Same | |
| Energy source | LED | LED | LED | Same | |
| Treatment heads | Separate blue and red,<br>treatment heads | Separate blue and red,<br>treatment heads | Blue and Red in same head,<br>chosen using selector switch | Same as the Clear Bi-Light<br>(both offer separate blue and<br>red light treatment heads). | |
| Wavelength(s)<br>(mm) | $417\pm10$<br>$633\pm10$ | 405-420<br>630-660 | $415\pm5$<br>$633\pm6$ | Substantially equivalent<br>Wavelengths are in the blue | |
| | | | | and red spectrums | |
| Treatment control | “ ▶/■ ” button to turn the<br>irradiator on or off for<br>therapy, The treatment head<br>should rest against the skin<br>area to be treated. | On/off button puts device in<br>standby mode.<br>Contact sensor in treatment<br>face, touching device to face<br>activates LEDs. | Three position switch: Off –<br>Blue - Red | Substantially equivalent<br>KN-7000C and Clear Bi-Light<br>have on/off button. Contact<br>sensor in proposed device is<br>the same as in the Clear Bi-<br>Light. | |
| Treatment indicator | Indicator will light up when<br>treatment; the indicator is<br>yellow when charging; the<br>indicator light is green when<br>it is charged fully. | Indicator light-solid green<br>while treatment is in process;<br>blinking indicates low battery.<br>Audio signal; long beep when<br>power is turned on, short beep<br>every 30 seconds during<br>treatment; long beep at end of<br>3 minute cycle. | No | Indicators light substantially<br>equivalent to the Clear Bi-<br>Light. Indicators light can<br>indicate related status of the<br>device. | |
| Treatment timer | The recommended therapy<br>times are 2 minutes per area<br>after which device returns to<br>standby mode. Must press<br>treatment face against skin to<br>reactivate lights. | Up to 3 minutes treatment per<br>area after which device<br>returns to standby mode.<br>Must press treatment face<br>against skin to reactivate<br>lights. | No | Substantially equivalent to the<br>Clear Bi-Light. Both devices<br>limit treatment to within 3<br>minutes. | |
| Treatment regimen | Hold treatment face in contact<br>with skin. Apply blue light for<br>2minutes per skin area,<br>followed by red light for 2<br>minutes per skin area. Can be<br>used daily. | Hold treatment face in contact<br>with skin. Apply blue light for<br>up to 3minutes per skin area,<br>followed by red light for up to<br>3 minutes per skin area. Can<br>be used daily. | Alternate treatments of blue<br>and red light twice a week for<br>20 minutes per treatment. | Substantially equivalent to the<br>Clear Bi-Light. Both devices<br>limit treatment to within 3<br>minutes. | |
| Dose rate | 35±5(blue) | 31.1(blue) | 40(blue) | Substantially | |
| (mW/cm²) | 65±5(red) | 54.6(red) | 70(red) | equivalent-within dose rate ranges of predicate devices. | |
| Treatment face area(cm²) | 4±10% | 20 | 28.7 | Substantially equivalent-within size ranges of predicate devices | |
| Max treatment temperature(°C) | 37.5(blue)<br>37.0(red) | 38(blue)<br>40(red) | 39-43 | Substantially equivalent-within temperature ranges of predicate devices | |
| Microprocessor controlled | Yes | Yes | Unknown | Same | |
| Handheld | Yes | Yes | Yes | Same | |
| Patient contacting material | Rigid ABS Clear polycarbonate lens cover | Rigid ABS Clear polycarbonate lens cover | Rigid ABS | Same | |
| Power supply | Adapter model number:<br>LXCP12-005200DEG, Input:<br>100-240V a.c. 50/60Hz, 0.5A max. Output: 5V d.c. 2A<br>Main unit input: 5V d.c. 2A/Internal battery: 3.6Vd.c. 2200mAh<br>Internal battery specifications: CR18650-22F 3.6V d.c. 2200mAh | Lithium-ion rechargeable battery AC charger: 100-240V at 50-60 Hz, 500mA | A separate, universal, power supply converts mains AC power to the DC power required | Substantially equivalent to the Clear Bi-Light. All devices have power adapter. | |
| Dimensions | 231mm×159mm×68mm<br>(9.1in×6.3in×2.7in) | 6.4cm×3.5cm×14.6cm<br>(2.5in×1.4in×5.7in) | 6cm×4cm×12cm<br>(2.4in×1.6in×4.7in) | Substantially equivalent All devices are light-weight, hand held devices. | |
| Weight | 122g(4.3 oz) | 107.1g(3.8 oz) | Unknown | Substantially equivalent to the Clear Bi-Light Both are | |
| | | | | light-weight, hand held devices. | |
| Standard | IEC 60601-1 | IEC 60601-1 | IEC 60601-1 | Same<br>All devices meet the<br>requirements of effectiveness,<br>safety and biocompatibility. | |
| | IEC 60601-1-2 | IEC 60601-1-2 | IEC 60601-1-2 | | |
| | IEC 60601-1-11 | IEC 60601-1-11 | IEC 60601-1-11 | | |
| | IEC 60601-2-57 | IEC 60601-2-57 | IEC 60601-2-57 | | |
| | IEC 62471 | IEC 62471 | IEC 62471 | | |
| | ISO 10993-1 | ISO 10993-1 | ISO 10993-1 | | |
| | ISO 10993-5 | ISO 10993-5 | ISO 10993-5 | | |
| | ISO 10993-10 | ISO 10993-10 | ISO 10993-10 | | |
{11}------------------------------------------------
{12}------------------------------------------------
{13}------------------------------------------------
#### 8. Substantial Equivalence discussion:
The indication for use of the KN-7000C LED Light Therapy Device and KN-7000C2 LED Light Therapy Device are the for the predicate devices. Most technical specifications of the KN-7000C LED Light Therapy Device and KN-7000C2 LED Light Therapy Device are either the same or substantially equivalent as compared to the predicate devices that raise new or different questions of safety of effectiveness.
{14}------------------------------------------------
#### 9. Non-Clinical Tests Performed:
The following non-clinical testing was provided in this 510(k):
Biocompatibility Testing - The skin contacting materials of the device were subjected to biocompatibility testing per ISO 10993-1:2009/C1:2010, "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing," for devices in contact with intact skin, including in vitro cytotoxicity, skin sensitization, and skin irritation. All tests passed. The tests comply with the applicable requirements of the following standards:
ISO10993-5:2009, Biological evaluation of medical devices-Part 5: tests for in vitro cytotoxicity.
ISO10993-10:2010, Biological evaluation of medical devices-Part10: tests for irritation and skin sensitization.
Electrical Safety and Electromagnetic Compatibility Testing - The KN-7000C LED Light Therapy Device and KN-7000C2 LED Light Therapy Device have been tested and comply with the applicable requirements of the following standards for medical devices used in the home environment:
● AAMI/ANSI ES60601-1: 2005/(R)2012 and A1:2012., C1:2009/(R)2012 and A2:2010/(R)2012 Medical electrical equipment-Part 1: General requirements for basic safety and essential performance
· IEC 60601-1-2:2014 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility -Requirements and tests
· IEC 60601-1-11:2015 Medical electrical equipment-Part 1-11: General requirements for basic safety and essential performance - Collateral standard: requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
· IEC 60601-2-57:2011 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility -Requirements and tests
Photobiological Safety Testing - The KN-7000C LED Light Therapy Device and KN-7000C2 LED Light Therapy Device have been tested and comply with IEC 62471:2006, Photobiological safety of lamps and lamp systems, 1st edition. This IEC standard incorporates the principles of the following ANSI IESNA recommended practices:
· RP 27.1:2005 Recommended practice for photobiological safety for lamps and lamp systems -General requirements
· RP 27.2:2000 Recommended practice for photobiological safety for lamps and lamp systems -Measurement techniques
· RP-27.3:2007 Recommended practice for photobiological safety for lamps and lamp systems -Risk group classification and labeling
Software Verification and Validation - Software documentation consistent with moderate level of concern was submitted in this 510(k). System validation testing presented in this 510(k) demonstrated that all software requirement specifications are met and all software hazards have been mitigated to acceptable risk levels.
### 10. Usability Study:
{15}------------------------------------------------
The KN-7000C LED Light Therapy Device and KN-7000C2 LED Light Therapy Device were performed usability study separately by similar methods.
#### Usability study summary of KN-7000C LED Light Therapy Device:
This usability Study was started on 1/7/2018 at 3 Depot Street, Hudson Falls, NY, USA.
It was conducted by Dave Oberhelman, Plant/Compliance Manager, UVBioTek
The study was completed on 2/1/2018.
The required number of participants was obtained through networking, social media and advertising in the local newspaper.
All of the participants were able to correctly answer the questions as listed in Appendix 1 and Appendix 2 All the participants were able to self select whether or not the product was correct for them. Some of the participants did not have acne but they would purchase the product for their children who did have acne.
Most of the participants offered suggestions to change the wording to make different parts of the manual easier to read and understand.
The original user manual for this study is in section 5 of this binder. The participant improved version is in section 7 of this binder.
The packaging materials used in the 1st part of this study will be e-mailed to Jonny Chen, of Kernel Medical, because they will not fit in this binder.
The following are samples of some of the suggestions the participants made to make the user manual easier to understand. Reference to page numbers applies to the new updated version of the manual.
State that the treatment heads have color dots on the back to identify what color LED is placed inside the treatment head Page 6 & 7.
List "How to take a treatment" in the table of contents.
Define what a snoot is. Page 9.
Remove 1% mild detergent and replace with diluted, non-corrosive, mild detergent Page 23.
State "Press and Hold" the power button. Page 14.
State at the very beginning that this device has two different color treatment heads. Page 1.
Change "operation mode" to "treatment time selection". Page 10
Change "work mode" button to "treatment time selector". Page 9
State "start treatment with blue LED treatment head". Page 14
State "Install the red treatment head and complete the treatment". Page 15
Define what it means to be insensitive to heat. Add (can not feel heat). Page 1
State that the time setting and treatment start must be completed with in 30 seconds or the unit will
shut off to conserve the battery. Page 14
Move the LED Life statement to page 20.
#### Usability study summary of KN-7000C2 LED Light Therapy Device:
This usability Study was started on 1/20/2018 at 3 Depot Street. Hudson Falls, NY, USA.
It was conducted by Dave Oberhelman, Plant/Compliance Manager, UVBioTek
T he study was completed on 2/4/2018.
The required number of participants was obtained through networking, social media and advertising in the local newspaper.
All of the participants were able to correctly answer the questions as listed in Appendix 1 and
{16}------------------------------------------------
Appendix 2 All the participants were able to self select whether or not the product was correct for them. Some of the participants did not have acne but they would purchase the product for their children who did have acne.
This study was started after the 7000C was started The User Manuals, Warnings, Cautions, Contraindications and package artwork are very similar. Most of the improvements obtained by the 7000C study were incorporated in to the artwork and user manual for the 7000C2, before the 7000C2 study started. The participants did not have any suggestions for improvements because all the material was easy to understand.
Some of the participants noticed that question # 10 in Appendix # 2 did not have a correct answer. Answer 'B" showed "(1)" but is should have shown " (1)". These participants were very good evaluators.
The original user manual for this study is in section 5 of this binder. The participant improved version is in section 7 of this binder.
The packaging materials used in the 1st part of this study will be e-mailed to Jonny Chen, of Kernel Medical, because they will not fit in this binder.
### 11. Conclusion:
Based on comparing to predicate device, the proposed device of KN-7000C and KN-7000C2 are determined to be Substantially Equivalent (SE) to the predicate device, in respect of safety and effectiveness.
--- End of this section ---
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.