← Product Code [HRX](/productcode/HRX) · K180876

# DTRAX Spinal System (K180876)

_Providence Medical Technology, Inc. · HRX · May 18, 2018 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K180876

## Device Facts

- **Applicant:** Providence Medical Technology, Inc.
- **Product Code:** [HRX](/productcode/HRX.md)
- **Decision Date:** May 18, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.1100
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

DTRAX Spinal System is a set of instruments indicated to be used to perform posterior cervical fusion in patients with cervical degenerative disc disease.

## Device Story

DTRAX Spinal System is a set of sterile, single-use surgical instruments for posterior cervical fusion. The system includes a facet joint access tool, lateral mass decortication trephine, mallet, cannula, decortication rasp, decortication burr, and tamp. In clinical use, the surgeon uses these instruments to access the posterior cervical spine, decorticate bone surfaces (posterior lateral mass and facet joints), and apply autograft or allograft to facilitate fusion. The device is intended for use by surgeons in a clinical/surgical setting. The instruments provide the mechanical means to prepare the bone site for fusion, directly affecting the surgical procedure's efficacy in achieving spinal stabilization.

## Clinical Evidence

No clinical data provided. Evidence consists of bench testing and simulated use cadaveric testing. Bench testing confirmed mechanical strength and integrity under various loads. Simulated use testing demonstrated the instruments' ability to perform posterior cervical fusion in patients with cervical degenerative disc disease.

## Technological Characteristics

Instruments constructed from stainless steel per ISO 7153-1 and ASTM F899-12b. System includes access tools, trephines, rasps, burrs, and tamps. Manual operation; no energy source. Supplied sterile via radiation (ISO 11137-1/2). Packaging validated per ISO 11607-1/2. Biocompatibility evaluated per ANSI AAMI ISO 10993-1.

## Regulatory Identification

An arthroscope is an electrically powered endoscope intended to make visible the interior of a joint. The arthroscope and accessories also is intended to perform surgery within a joint.

## Predicate Devices

- Vertos Medical mild® Device Kit ([K093062](/device/K093062.md))

## Reference Devices

- Dyonics, Inc. The Arthrofile ([K812506](/device/K812506.md))
- Transcorp, Inc. Transcorp Spinal Access System ([K113352](/device/K113352.md))
- Aesculap, Inc. Hi Line XS disposable Rosen Burr III
- Medtronic Cervical FacetLift ID/S Instruments

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

May 18, 2018

Providence Medical Technology, Inc. Edward Liou Chief Operating Officer 3857 Hopyard Rd. Suite 300 Pleasanton, California 94588

Re: K180876

Trade/Device Name: DTRAX Spinal System Regulation Number: 21 CFR 888.1100 Regulation Name: Arthroscope Regulatory Class: Class II Product Code: HRX Dated: April 2, 2018 Received: April 3, 2018

Dear Edward Liou:

We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good

{1}------------------------------------------------

manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

> Sincerely, Jennifer R. Stevenson -S For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation

Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known)

#### K180876

Device Name DTRAX® Spinal System

Indications for Use (Describe)

DTRAX Spinal System is a set of instruments indicated to be used to perform posterior cervical fusion in patients with cervical degenerative disc disease.

Type of Use (Select one or both, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# Section 6. 510(k) Summary

| 510(k) Owner:                     | Providence Medical Technology, Inc.<br>3875 Hopyard Rd., Suite 300<br>Pleasanton, CA 94588<br>T: 415-923-9376<br>F: 415-923-9377                                                                                                             |
|-----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person:                   | Edward Liou<br>Email: ed@providencemt.com<br>T: 415-754-8593                                                                                                                                                                                 |
| Date Summary Prepared:            | May 16th, 2018                                                                                                                                                                                                                               |
| Trade Name:                       | DTRAX® Spinal System                                                                                                                                                                                                                         |
| Common Name:                      | Arthroscopic Accessories                                                                                                                                                                                                                     |
| Device Classification Regulation: | 21 CFR 888.1100                                                                                                                                                                                                                              |
| Device Product Code & Panel:      | HRX<br>Orthopaedic and Rehabilitation Devices Panel                                                                                                                                                                                          |
| Primary Predicate Device:         | Vertos Medical, Inc. Vertos Medical mild® Device Kit (K093062)                                                                                                                                                                               |
| Reference Devices:                | Dyonics, Inc. The Arthrofile (K812506)<br>Transcorp, Inc. Transcorp Spinal Access System (K113352)<br>Aesculap, Inc. Hi Line XS disposable Rosen Burr III (Class I exempt)<br>Medtronic Cervical FacetLift ID/S Instruments (Class I exempt) |

### Device Information

#### A. Device Description

The DTRAX® Spinal System is a set of surgical instruments indicated for performing posterior cervical fusion. The instruments will be supplied sterile and single use only. The system consists of a facet joint access tool, lateral mass decortication trephine, mallet, cannula, decortication rasp, decortication burr, and tamp. In combination, these instruments allow the user to access the posterior cervical spine to perform posterior cervical fusion by decortication of bone surfaces, including the posterior lateral mass and facet joints, combined with application of autograft or allograft.

{4}------------------------------------------------

| DTRAX® Spinal System<br>Catalog# DX-22-100 |           |                        |
|--------------------------------------------|-----------|------------------------|
| 1                                          | PT-01-100 | Access Chisel          |
| 2                                          | PD-34-100 | Access Chisel Handle   |
| 3                                          | PD-05-200 | Decortication Trephine |
| 4                                          | PD-04-000 | Guide Tube             |
| 5                                          | PD-02-101 | Decortication Rasp     |
| 6                                          | PD-02-300 | Decortication Burr     |
| 7                                          | PD-12-000 | Bone Graft Tamp        |
| 8                                          | PD-05-004 | Fork Mallet            |

#### B. Indications for Use

DTRAX® Spinal System is a set of instruments indicated to be used to perform posterior cervical fusion in patients with cervical degenerative disc disease.

#### C. Summary of Technological Characteristics

The fundamental operational principles, surgical approach, design, materials, and performance of the DTRAX® Spinal System and the predicate device are essentially the same. Therefore, the technological characteristics of the DTRAX® Spinal System do not raise any new safety and effectiveness questions not addressed in the predicate device.

### D. Summary of Performance Testing

Non-clinical testing has demonstrated the DTRAX® Spinal System meets the performance requirements to safely distribute and use in accordance with the Indication for Use.

#### Premarket Notification

{5}------------------------------------------------

- Simulated use cadaveric testing demonstrated the instruments can be used to perform posterior cervical fusion in patients with cervical degenerative disc disease.
- The bench testing demonstrated the safety and efficacy of the instruments and the strength . and integrity to resist impaction, insertion, removal, and rotational loads to perform the stated intend use.
- Additionally, the DTRAX® Spinal System demonstrated that it is in compliance with the following FDA consensus standards:
	- FDA Recognition Number 14-428: ISO 11137-1 First Edition 2006-04-15 Sterilization о Of Health Care Products - Radiation - Part 1: Requirements For Development, Validation And Routine Control Of A Sterilization Process For Medical Devices [Including: Amendment 1 (2013)]
	- FDA Recognition Number 14-438: ISO 11137-2 Third Edition 2013-06-01 Sterilization o Of Medical Devices - Microbiological Methods - Part 2: Tests Of Sterility Performed In The Definition, Validation And Maintenance Of A Sterilization Process
	- FDA Recognition Number 14-407: ISO 11737-1 Second Edition 2006-04-01 o Sterilization of Medical Devices – Microbiological Methods – Part 1: Determination of a Population of Microorganisms on Products
	- o FDA Recognition Number 14-287: ISO 11737-2: 2009/(R) 2014 Sterilization of medical devices – Microbiological methods – Part 2: Tests of Sterility Performed in the Definition, Validation and Maintenance of Sterilization Process.
	- FDA Recognition Number 14-457: ISO 11607-1:2006/(R)2010 Packaging For o Terminally Sterilized Medical Devices - Part 1: Requirements For Materials, Sterile Barrier Systems And Packaging [Including: Amendment 1 (2014)]
	- FDA Recognition Number 14-458: ISO 11607-2:2006/(R)2010 Packaging for o Terminally Sterilized Medical Devices – Part 2: Validation Requirements for Forming, Sealing and Assembly Processes [Including: Amendment 1 (2014)
	- O FDA Recognition Number 2-156: ANSI AAMI ISO10993-1:2009/(R)2013 Biological Evaluation Of Medical Devices - Part 1: Evaluation And Testing Within A Risk Management Process
	- FDA Recognition Number 8-344: ISO 7153-1 Second Edition 1991-04-01 Surgical o Instruments -- Metallic Materials -- Part 1: Stainless Steel [Including: Amendment 1 (1999)]
	- FDA Recognition Number 8-343: ASTM F899-12b Standard Specification for Wrought o Stainless Steels for Surgical Instruments

## E. Basis of Substantial Equivalence

The DTRAX® Spinal System is substantially equivalent compared to the cleared primary predicate device, Vertos Medical, Inc., Vertos Medical Mild® Device Kit (K093062) and the reference devices, on the basis the devices have the same technological characteristics, similar design features, same materials and principles of operation. Performance data demonstrate that the DTRAX® Spinal System is as safe and effective as the predicate devices.

---

**Source:** [https://fda.innolitics.com/device/K180876](https://fda.innolitics.com/device/K180876)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
