The WATCHMAN™ TruSeal™ Access System is intended to provide vascular and transseptal access for all WATCHMAN Left Atrial Appendage Closure Devices with Delivery Systems.
Device Story
The WATCHMAN TruSeal Access System consists of an access sheath and dilator used by interventional cardiologists or electrophysiologists during percutaneous transseptal procedures. It provides a conduit for the delivery of WATCHMAN Left Atrial Appendage Closure devices. The system is a mechanical catheter-based tool; it does not utilize electronic sensors or software. It facilitates the navigation and placement of the closure device within the heart. The device is intended for single-use and is provided sterile.
Clinical Evidence
No clinical data. Substantial equivalence is supported by bench testing, including dimensional characterization, tip deflection, torqueability, tensile strength, kink resistance, pressure retention, and biocompatibility testing (cytotoxicity, sensitization, reactivity, systemic toxicity, complement activation, pyrogenicity, and hemolysis).
Technological Characteristics
Percutaneous catheter system comprising an access sheath and dilator. Conforms to EN ISO 10555-1:2013. Features include radiopaque markers, luer lock connections, and specific curve shapes for transseptal navigation. Sterile, single-use device.
Indications for Use
Indicated for interventional cardiologists and/or electrophysiologists requiring vascular and transseptal access for WATCHMAN Left Atrial Appendage Closure Devices with Delivery Systems.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
July 20, 2018
Boston Scientific Nikki Ibis Principal Regulatory Affairs Specialist Three Scimed Place Maple Grove, Minnesota 55311-1566
Re: K180864
Trade/Device Name: WATCHMAN TruSeal Access System Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DOY Dated: June 27, 2018 Received: June 28, 2018
Dear Ms. Ibis:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Rachel E. Neubrander -S
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K180864
Device Name WATCHMAN™ TruSeal™ Access System
Indications for Use (Describe)
The WATCHMAN™ TruSeal™ Access System is intended to provide vascular and transseptal access for all WATCHMAN Left Atrial Appendage Closure Devices with Delivery Systems.
Type of Use (Select one or both, as applicable)
| ☑ Residential Use (Part 91, CFR 901 Subpart D) |
|-------------------------------------------------------|
| ☐ On-The-Scene Coordinator Use (91 CFR 901 Subpart E) |
> Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary per 21 CFR §807.92
| Sponsor: | Boston Scientific Corporation<br>300 Boston Scientific Way<br>Marlborough, Massachusetts 01752<br>USA | | |
|---------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------|---------|---------------|
| Contact Name and<br>Information | Nikki M Ibis<br>Three Scimed Place<br>Maple Grove, MN 55311-1566<br>Phone: 763-494-2381<br>Fax: 763-494-2981<br>e-mail: Nikki.Ibis@bsci.com | | |
| Prepared | 30 March 2018 | | |
| Proprietary Name | WATCHMAN™ TruSeal™ Access System | | |
| Common Name | Catheter, Percutaneous | | |
| Product Code | DQY | | |
| Classification | Class II, 21 CFR Part 870.1250 | | |
| Primary Predicate<br>Device | Amplatzer® TorqVue®<br>45°x45° Delivery<br>Sheath | K083214 | 12 May 2009 |
| Reference<br>Device(s) | WATCHMAN® Access<br>System | P130013 | 13 March 2015 |
#### Device Description
The Boston Scientific WATCHMAN™ TruSeal™ Access System is composed of an Access Sheath and Dilator. The TruSeal Access System is used to provide vascular and transseptal access for all WATCHMAN Left Atrial Appendage Closure Devices with Delivery Systems. The TruSeal Access System will be used by interventional cardiologists and/or electrophysiologists who are trained in percutaneous and transseptal procedures.
## Indications for Use / Intended Use
The WATCHMAN™ TruSeal™ Access System is intended to provide vascular and transseptal access for all WATCHMAN Left Atrial Appendage Closure Devices with Delivery Systems.
{4}------------------------------------------------
## Comparison of Technological Characteristics
The WATCHMAN™ TruSeal™ Access System incorporates substantially equivalent design, packaging, fundamental technology, and intended use as those featured in the predicate, Amplatzer® TorqVue® 45°x45° Delivery Sheath, K083214.
### Performance Data
Design verification testing was performed to support a determination of substantial equivalence according to Class II Special Controls Guidance Document for Certain Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheters, September 2010. The WATCHMAN™ TruSeal™ Access System also conforms to relevant sections of EN ISO 10555-1: 2013, Sterile, single-use intravascular catheters; Part 1: General Requirements. The results of the tests provide reasonable assurance that the proposed device has been designed and tested to assure conformance to the requirements for its intended use. No new safety or performance issues were raised during the testing.
The following device performance tests were completed:
| • | Dimensional Characterization | • | Tip Deflection |
|---|-------------------------------------------|---|---------------------------------|
| • | Distal and Proximal Marker<br>Location | • | Torqueability |
| • | Curve Shape | • | Snap Fit Attachment Force |
| • | TruSeal Access System Guidance / Crossing | • | Access Sheath Compression Force |
| • | Contrast Flow Rate | • | Dilator Stiffness |
| • | Leak-Free Conduits | • | Tensile |
| • | Pressure Retention | • | Radiopacity |
| • | Corrosion Resistance | • | Luer Lock Testing |
| • | Sheath Force Transmission | • | Particulates |
| • | Kink Resistance | • | Surface |
The following package performance tests were completed:
| • | Master Shipper Carton Integrity | • | TruSeal Access System Containment |
|---|---------------------------------|---|-----------------------------------|
| • | Shelf Carton Condition | • | Tray Condition |
| • | Sterile Barrier Integrity | • | Removal From Packaging |
| • | Sterile Barrier Seal Strength | • | DFU and eDFU Card Print Quality |
| • | Ease of Opening Pouch | • | Label Adhesion and Print Quality |
{5}------------------------------------------------
The following biocompatibility tests were completed:
- Cytotoxicity ●
- Sensitization
- Intracutaneous Reactivity .
- Acute Systemic Toxicity .
- Complement Activation .
### Clinical Testing
Clinical evaluation was not required for this device.
### Conclusion
Based on the intended use use, technological characteristics, safety and performance testing, the WATCHMAN™ TruSeal™ Access System is appropriate for the stated intended use and is considered to be substantially equivalent to the Amplatzer® TorqVue® 45°x45° Delivery Sheath, K083214.
- Materials Mediated Pyrogenicity ●
- . Hemolysis Direct Contact
- Hemolysis Extract Method .
- . In Vitro Platelet and Leukocyte Count
- Partial Thromboplastin Time