AMBIENT Super MULTIVAC 50 Wand with Integrated Finger Switches
K180848 · ArthroCare Corporation · GEI · May 1, 2018 · General, Plastic Surgery
Device Facts
Record ID
K180848
Device Name
AMBIENT Super MULTIVAC 50 Wand with Integrated Finger Switches
Applicant
ArthroCare Corporation
Product Code
GEI · General, Plastic Surgery
Decision Date
May 1, 2018
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The AMBIENT Super MULTIVAC 50 Wand with Integrated Finger Switches is indicated for resection, ablation, and coagulation of soft tissue and hemostasis of blood vessels in the following arthroscopic and orthopedic procedures. All Joints (Hip, Knee, Shoulder, Wrist, Ankle, Elbow) Ablation/Debridement Articular Cartilage Bursectomy Chondroplasty Fascia Ligament Scar Tissue Soft Tissue Synovectomy Tendon Knee ACL/PCL Notchplasty Shoulder Acromioplasty Subacromial Decompression Excision/Resection All Joints (Hip, Knee, Shoulder, Wrist, Ankle, Elbow) Articular Labrum Capsule Cysts Ligament Loose Bodies Plica Removal Scar Tissue Soft Tissue Synovial Membrane Tendon Hip Acetabular Labrum Knee Capsular Release Cartilage Flaps Discoid Meniscus Lateral Release Meniscal Cystectomy Meniscectomy Villusectomy Shoulder Frozen Shoulder Release Glenoidale Labrum Wrist Triangular Fibrocartilage (TFCC) Coagulation All Joints (Hip, Knee, Shoulder, Wrist, Ankle, Elbow) Articular Cartilage Ligament Tendon Knee ACL/PCL Medial Retinaculum Shoulder Glenohumeral Capsule Rotator Cuff Wrist Carpal Ligaments Wrist Tendons
Device Story
Bipolar, high-frequency electrosurgical wand for arthroscopic/orthopedic surgery; used with QUANTUM 2 Controller. Device inputs: RF power from controller; user-activated finger switches. Device transforms inputs via tungsten active electrode and stainless steel return electrode to perform soft tissue resection, ablation, and coagulation in saline environment. Features AMBIENT technology for real-time temperature monitoring (20-60°C, ±3°C) of irrigating fluid with user-adjustable alarms. Used by surgeons in OR; provides real-time thermal feedback to prevent tissue damage. Output: thermal/electrical energy for tissue modification. Benefits: precise tissue management, hemostasis, and thermal monitoring during minimally invasive procedures.
Clinical Evidence
Bench testing only. Testing included transit conditioning, visual inspection, IEC testing, Ambient accuracy, ablation performance, coagulation testing, and use-limiting feature verification. No clinical data included.
Technological Characteristics
Bipolar electrosurgical wand; tungsten active electrode; polyolefin insulation; alumina spacer; stainless steel shaft; polycarbonate/polyolefin handle. Dimensions: 11-inch length, 43-degree distal bend. Connectivity: proprietary connector with embedded PCBA for communication with QUANTUM 2 Controller. Sterilization: irradiation. Single-use disposable.
Indications for Use
Indicated for resection, ablation, and coagulation of soft tissue and hemostasis of blood vessels in arthroscopic and orthopedic procedures involving hip, knee, shoulder, wrist, ankle, and elbow joints.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
Predicate Devices
AMBIENT Super MULITVAC 50 Wand with Integrated Finger Switches (K083306)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
May 1, 2018
ArthroCare Corporation Ms. Shruthi Bhat Regulatory Affairs Specialist II 7000 West William Cannon Drive Austin, Texas 78735
Re: K180848
Trade/Device Name: AMBIENT Super MULTIVAC 50 Wand with Integrated Finger Switches Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI Dated: March 30, 2018 Received: April 2, 2018
Dear Ms. Bhat:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
> Sincerely. Jennifer R. Stevenson For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices
Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known)
K180848
Device Name
AMBIENT™ Super MULITVAC™ 50 Wand with Integrated Finger Switches
Indications for Use (Describe) See attached below
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for ArthroCare. The logo consists of a red circular design on the left and the word "ArthroCare" in gray on the right. The circular design is made up of several curved lines that create a sense of movement. The word "ArthroCare" is written in a sans-serif font, and there is a registered trademark symbol next to the word.
# Indications for Use:
The AMBIENT Super MULTIVAC® 50 Wand with Integrated Finger Switches is indicated for resection,
ablation, and coagulation of soft tissue and hemostasis of blood vessels in the following arthroscopic and orthopedic procedures.
| | | Ablation/Debridement |
|-------------------------------------------------------------|----------------------------------------------------------------------|---------------------------------------------------------------|
| All Joints (Hip, Knee,<br>Shoulder, Wrist,<br>Ankle, Elbow) | ArticularCartilageBursectomyChondroplastyFasciaLigament | Scar TissueSoft TissueSynovectomyTendon |
| Knee | ACL/PCLNotchplasty | |
| Shoulder | AcromioplastySubacromial Decompression | |
| | | Excision/Resection |
| All Joints (Hip, Knee,<br>Shoulder, Wrist,<br>Ankle, Elbow) | ArticularLabrumCapsuleCystsLigamentLoose Bodies | Plica RemovalScar TissueSoft TissueSynovial<br>MembraneTendon |
| Hip | Acetabular Labrum | |
| Knee | Capsular<br>ReleaseCartilage FlapsDiscoid<br>MeniscusLateral Release | Meniscal<br>CystectomyMeniscectomyVillusectomy |
| Shoulder | Frozen Shoulder ReleaseGlenoidale Labrum | |
| Wrist | Triangular Fibrocartilage (TFCC) | |
| | | Coagulation |
| All Joints (Hip, Knee,<br>Shoulder, Wrist,<br>Ankle, Elbow) | Articular CartilageLigamentTendon | |
| Knee | ACL/PCLMedial Retinaculum | |
| Shoulder | Glenohumeral CapsuleRotator Cuff | |
| Wrist | Carpal LigamentsWrist Tendons | |
7000 W. William Cannon Drive | Building One | Austin, Texas 78735 | ۾/ 2012.391.3901 | www.arthrocare.com
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for ArthroCare. The logo consists of a circular design on the left, made up of several curved red lines. To the right of the design is the word "ArthroCare" in a dark gray, sans-serif font. A registered trademark symbol is located to the upper right of the word "Care".
# 510(k) Summary
# AMBIENT® Super MULITVAC° 50 Wand with Integrated Finger Switches
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
# General Information
| Submitter Name: | ArthroCare Corporation |
|-------------------|---------------------------------------------------------------|
| Address | 7000 West William Cannon Drive |
| | Austin, TX 78735 |
| Contact Person: | Mrs. Shruthi Bhat M.Sc. (Biosciences) MS (Regulatory Affairs) |
| | Regulatory Affairs Specialist II |
| | ArthroCare Corporation |
| | Phone: (512) 895-1295 |
| | Fax: 512-895-1489 |
| | Email: shruthi.bhat@smith-nephew.com |
| | Address: 7000 West William Cannon Drive |
| | Austin, TX 78735 |
| Submitter Person: | Mrs. Shruthi Bhat M.Sc. (Biosciences) MS (Regulatory Affairs) |
| | Regulatory Affairs Specialist II |
| | ArthroCare Corporation |
| | Phone: (512) 895-1295 |
| | Fax: 512-895-1489 |
| | Email: shruthi.bhat@smith-nephew.com |
| | Address: 7000 West William Cannon Drive |
| | Austin, TX 78735 |
| Date Prepared: | 30 March 2018 |
#### Device Name
Proprietary Name: AMBIENT Super MULTIVAC° 50 Wand with Integrated Finger Switches 7000 W. William Cannon Drive | Building One | Austin, Texas 78735 | ρέας 512.391.3901 | www.arthrocare.com
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for ArthroCare. The logo consists of a circular design on the left, made up of several curved red lines. To the right of the design is the word "ArthroCare" in a dark gray sans-serif font. A registered trademark symbol is located to the upper right of the word "Care".
| Common Name: | Electrosurgical devices and accessories |
|----------------------|----------------------------------------------------------------|
| Classification Name: | Electrosurgical cutting and coagulation device and accessories |
| Device Class: | Class II |
| Product Code: | GEI |
| CFR Section: | 21 CFR 878.4400 |
#### Predicate Device
AMBIENT Super MULITVAC® 50 Wand with Integrated Finger Switches cleared under 510(k) K083306 (Modification to ArthroCare ArthroWands) on December 10, 2008.
#### Description
The AMBIENT Super MULTIVAC° 50 Wand with Integrated Finger Switches ("Wand") is a bipolar, high frequency electrosurgical device used for specific indications in orthopedic and arthroscopic surgeries. The AMBIENT feature provides accurate (± 3° C) real-time temperature monitoring of the circulating irrigating fluid between 20° C and includes a user adjustable alarm set point. The Wand is designed for use up to five minutes of cumulative active ablation time at the default set point of 7 and is to be used only with the QUANTUM® 2 Controller ("Controller").
## Intended Use/Indications for Use
The AMBIENT Super MULTIVAC* 50 Wand with Integrated Finger Switches is indicated for resection, ablation, and coagulation of soft tissue and hemostasis of blood vessels in the following arthroscopic and orthopedic procedures.
| | Ablation/Debridement | |
|-------------------------------------------------------------|-----------------------------------------------------|--------------------------------------------|
| All Joints (Hip, Knee,<br>Shoulder, Wrist,<br>Ankle, Elbow) | Articular Cartilage Bursectomy Chondroplasty Fascia | Scar Tissue Soft Tissue Synovectomy Tendon |
7000 W. William Cannon Drive | Building One | Austin, Texas 78735 | phone 512.391.3901 | www.arthrocare.com
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image contains the logo for ArthroCare. The logo consists of a circular graphic on the left, composed of several curved red lines. To the right of the graphic is the company name, "ArthroCare," in a dark gray, sans-serif font. A registered trademark symbol is present next to the word "Care".
| | Ligament | |
|-------------------------------------------------------------|-------------------------------------------------------------------|----------------------------------------------------------------|
| Knee | ACL/PCL Notchplasty | |
| Shoulder | Acromioplasty Subacromial Decompression | |
| | Excision/Resection | |
| All Joints (Hip, Knee,<br>Shoulder, Wrist,<br>Ankle, Elbow) | Articular Labrum Capsule Cysts Ligament Loose Bodies | Plica Removal Scar Tissue Soft Tissue Synovial Membrane Tendon |
| Hip | Acetabular Labrum | |
| Knee | Capsular Release Cartilage Flaps Discoid Meniscus Lateral Release | Meniscal Cystectomy Meniscectomy Villusectomy |
| Shoulder | Frozen Shoulder Release Glenoidale Labrum | |
| Wrist | Triangular Fibrocartilage (TFCC) | |
| | Coagulation | |
| All Joints (Hip, Knee,<br>Shoulder, Wrist,<br>Ankle, Elbow) | Articular Cartilage Ligament Tendon | |
| Knee | ACL/PCL Medial Retinaculum | |
| Shoulder | Glenohumeral Capsule | |
| | Rotator Cuff | |
| Wrist | Carpal Ligaments Wrist Tendons | |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for ArthroCare. The logo consists of a stylized red circular design on the left and the name "ArthroCare" in gray text on the right. The circular design is made up of several curved lines that create a sense of movement. The logo is simple and modern.
# Summary of Technological Characteristics
| Parameter/ Characteristics | Predicate Device | Subject Device |
|---------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------|
| Indications for Use | Resection, ablation, and coagulation<br>of soft tissue and hemostasis of blood<br>vessels in arthroscopic and orthopedic<br>procedures. | Same |
| Arthrocare Controller(s) | Quantum 2 | Same |
| Wand Materials | | |
| Materials Biocompatible | Yes | Same |
| Active Electrode | Tungsten | Same |
| Electrode Insulation | Polyolefin | Same |
| Spacer | Alumina | Same |
| Shaft (Return Electrode) | Stainless Steel | Same |
| Suction Tubing/ Saline Delivery<br>Tubing | PVC & Nylon | Same |
| Suction / Irrigation Line<br>Connector | PVC | Same |
| Adhesive | Epoxy, Cyanoacrylate, UV Adhesive | Same |
| Cable Assembly Design | The cable assembly is composed of a<br>cable that at one end contains a unique<br>connector with embedded PCBA to<br>mate with the RF generator/controller,<br>and at the other end contains a PCBA<br>that is housed within the handle of the | Same |
| | | |
| | wand. The cable contains 8<br>conductors. Two (2) of the conductors<br>transmit the RF power from the RF<br>generator. Six (6) of the conductors<br>are used for signal transmission. The<br>PCBAs contain wand operating<br>parameters and process fingerswitch<br>and thermistor inputs and<br>communicate with the RF generator<br>via the 6 signal transmission<br>conductors. | |
| Resistor R1 in the connector of<br>the Integrated Cable | 576 ohms | 1210 ohms |
| Resistor R2 in the connector of<br>the Integrated Cable | 1050, 1210, 1300, 1400, 1500, 1620,<br>1740, or 1870 Ohms. | Same |
| External Insulation (Return<br>Insulation) | Polyester | Same |
| Handle Material | Polycarbonate, Polyolefin (finger<br>switch material) | Same |
| Handle Length | 5.75 | Same |
| Wand Temperature Sensor (non-<br>body contacting) | Copper, Constantan | Same |
| Packaging | Tyvek Lid 09570 | Same |
| Cable Wrap for Packaging | Paper | Same |
| <b>Physical Specifications (Wands)</b> | | |
| Total Length | 11 inches | Same |
| Distal Bend Angle | 43 degrees | Same |
| Number of Electrodes | 1 | Same |
| Number of Internal Suctions<br>Channels | 1 | Same |
| <b>Wand Features</b> | | |
| Electrode Configurations | Screen | Same |
| Electrode Articulation | 0 degrees | Same |
| Shaft Rotation | 0 degrees | Same |
| Spacer Configurations | Single Lumen | Same |
| Rigid Construction | Yes | Same |
| Suction and/ or Irrigation | Suction | Same |
| Packaged Sterile | Yes | Same |
| Single Use Disposable | Yes | Same |
| Operates in saline environment | Yes | Same |
| Bipolar/ monopolar | Bipolar | Same |
| Activation | Foot Control or Wand with Integrated<br>Finger Switch | Same |
| | Coagulation Voltage Setting | Adjustable (3 set points),<br>2 Active Set Points |
| Use Limiting Scheme | Activation of fuse allows up to 40<br>minutes of use and an overall<br>connection time up to 8 hours | Activation of fuse allows up to 7<br>minutes of use and an overall<br>connection time up to 8 hours |
| Performance Specifications | | |
| Environmental Exposure | 2 hours in Normal Saline or blood | Same |
| Sterilization | Irradiation | Same |
| Wand Temperature Sensor | T-Type thermocouple with range of<br>20-60° C + 3 ° C | Same |
7000 W. William Cannon Drive | Building One | Austin, Texas 78735 | ρέος 512.391.390 | www.arthrocare.com
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for ArthroCare. The logo consists of a circular design on the left, made up of several curved red lines. To the right of the design is the word "ArthroCare" in a dark gray, sans-serif font. A small registered trademark symbol is located to the upper right of the word "Care".
7000 W. William Cannon Drive | Building One | Austin, Texas 78735 | رمان 2012.391.3901 | www.arthrocare.com
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for ArthroCare. The logo consists of a stylized, circular graphic on the left and the word "ArthroCare" on the right. The graphic is made up of several curved, red lines that converge towards the center, creating a sense of movement. The word "ArthroCare" is written in a bold, sans-serif font, and the registered trademark symbol is located to the right of the word.
## Clinical Data
No clinical data are included in this submission.
#### Performance Data
Performance bench testing, including functional testing, Ambient accuracy, ablation performance, use-limiting and coagulation testing was performed on the subject device which demonstrates that the subject device continues to meet the design requirements and supports substantial equivalence with the predicate device. The tests are listed below:
- . Transit Conditioning
- . Visual Inspection Post-Transit
- . IEC Testing
- Ambient Accuracy Testing .
- Ablation Performance Testing at Default Set Point and Suction ●
- . Coagulation Testing
7000 W. William Cannon Drive | Building One | Austin, Texas 78735 | ρόνο 512.391.3901 | www.arthrocare.com
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image features the logo for "ArthroCare." To the left of the text is a circular graphic made of several curved lines. The text "ArthroCare" is in a sans-serif font and is dark gray. A registered trademark symbol is to the right of the word "Care."
- Ablation Performance Testing at Default Set Point & Suction & Use Limiting Feature .
- . Use-Limiting Feature
All devices used for verification testing were sent to an ArthroCare approved supplier for sterilization.
## Conclusion
All testing demonstrates that the AMBIENT® Super MULTIVAC° 50 Wand with Integrated Finger Switches performs as intended and has acceptable performance when used in accordance with its labeling. As the indications for use, contraindications, principle of operation and technology are unchanged from the predicate device, except for the Integrated Cable, the subject device, AMBIENT^ Super MULTIVAC° 50 Wand with Integrated Finger Switches, is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.