← Product Code [DQO](/productcode/DQO) · K180846

# Arterial Pressure Monitoring Set/Tray (K180846)

_Cook Incorporated · DQO · Dec 19, 2018 · Cardiovascular · SESK_

**Canonical URL:** https://fda.innolitics.com/device/K180846

## Device Facts

- **Applicant:** Cook Incorporated
- **Product Code:** [DQO](/productcode/DQO.md)
- **Decision Date:** Dec 19, 2018
- **Decision:** SESK
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1200
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Pediatric

## Indications for Use

The Arterial Pressure Monitoring Set/Tray is intended for arterial blood pressure monitoring and blood sampling in adult and pediatric patients. - 2.5 French catheters are intended for patients from birth and older - 3.0 and 4.0 French catheters are intended for patients 1 year and older

## Device Story

Arterial Pressure Monitoring Set/Tray facilitates percutaneous arterial access for blood pressure monitoring and blood sampling. System includes pressure monitoring catheter, wire guide, entry access needle, syringe, and optional dilator/convenience components. Clinician inserts catheter into radial, brachial, dorsalis pedis, femoral, or posterior tibial arteries using Seldinger technique. Catheter provides direct intravascular access for pressure transduction; output is displayed on external patient monitoring equipment. Device is single-use, sterile, and intended for clinical settings.

## Clinical Evidence

Bench testing only. Performance testing included catheter shaft/hub tensile strength (BS EN ISO 10555-1:2013), liquid/air leakage, kink resistance (BS EN 13868), flow rate, and wire guide corrosion/fracture/flex/tensile testing (BS EN ISO 11070:2014). Biocompatibility testing performed per ISO 10993-1 (cytotoxicity, sensitization, irritation, systemic toxicity, pyrogenicity, genotoxicity, implantation, hemocompatibility). Sterilization residuals verified per ANSI/AAMI/ISO 10993-7.

## Technological Characteristics

Catheters: polyurethane, ethylene-vinyl acetate, or polyethylene; 2.5-4.0 Fr. Wire guide: stainless steel coils, safety wire, mandril wire, All-State solder. Access needle: stainless steel. Dilator: polyethylene. Sterilization: ethylene oxide. Connectivity: none (passive diagnostic catheter).

## Regulatory Identification

An intravascular diagnostic catheter is a device used to record intracardiac pressures, to sample blood, and to introduce substances into the heart and vessels. Included in this generic device are right-heart catheters, left-heart catheters, and angiographic catheters, among others.

## Predicate Devices

- Cook Pressure Monitoring Catheter ([K002254](/device/K002254.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The FDA logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The Department of Health & Human Services logo features a stylized human figure. The FDA part of the logo has the acronym "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

December 19, 2018

Cook Incorporated Johnathan Liu Regulatory Affairs Specialist 750 Daniels Wav Bloomington, Indiana 47404

Re: K180846

Trade/Device Name: Arterial Pressure Monitoring Set/Tray Regulation Number: 21 CFR 870.1200 Regulation Name: Diagnostic intravascular catheter Regulatory Class: Class II Product Code: DQO Dated: November 26, 2018 Received: November 27, 2018

Dear Johnathan Liu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. However, you are responsible to determine that the medical devices you use as components in the tray have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were legally on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit/tray. The general controls of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Stephen C. Browning -S5

for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

#### Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

#### 510(k) Number (if known)

#### K180846

Device Name Arterial Pressure Monitoring Set/Tray

Indications for Use (Describe)

The Arterial Pressure Monitoring Set/Tray is intended for arterial blood pressure monitoring in adult and pediatric patients.

- 2.5 French catheters are intended for patients from birth and older
- 3.0 and 4.0 French catheters are intended for patients 1 year and older

| Type of Use (Select one or both, as applicable)                                                                 | <table><tr><td><span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D)             </span></td><td><span>☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span></td></tr></table> | <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D)             </span> | <span>☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
|-----------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------|------------------------------------------------------------|
| <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D)             </span> | <span>☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span>                                                                                                                                                          |                                                                                                                 |                                                            |

CONTINUE ON A SEPARATE PAGE IF NEEDED.

 This section applies only to requirements of the Paperwork Reduction Act of 1995.

***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***

 The burden time for this collection of information is estimated to average 79 hours per response, including the
 time to review instructions, search existing data sources, gather and maintain the data needed and complete
 and review the collection of information. Send comments regarding this burden estimate or any other aspect
 of this information collection, including suggestions for reducing this burden, to:

 Department of Health and Human Services  

 Food and Drug Administration  

 Office of Chief Information Officer  

 Paperwork Reduction Act (PRA) Staff

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

PRAStaff(@fda.hhs.gov

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image is a logo for Cook Medical. The logo is set against a red background. The word "COOK" is written in large, white, sans-serif font. Below "COOK" is the word "MEDICAL" in a smaller, white, sans-serif font, set against a black background.

COOK INCORPORATED 750 DANIELS WAY, P.O. BOX 489 BLOOMINGTON, IN 47402-0489 U.S.A. HONE: 812.339.2235 TOLL FREE: 800.457.4500 WWW.COOKMEDICAL.COM

> K180846 Page 1 of 6

#### 2.0 510(k) SUMMARY

#### K180846 Arterial Pressure Monitoring Set/Tray 21 CFR §807.92 Date Prepared: December 17, 2018

#### Submitted By:

| Submission:           | Traditional 510(k) Premarket Notification                     |
|-----------------------|---------------------------------------------------------------|
| Applicant:            | Cook Incorporated                                             |
| Applicant Address:    | Cook Incorporated<br>750 Daniels Way<br>Bloomington, IN 47404 |
| Primary Contact:      | Johnathan Liu                                                 |
| Secondary Contact:    | Chelsea Woods                                                 |
| Email:                | regsubmissions@cookmedical.com                                |
| Contact Phone Number: | (812) 335-3575 x104509                                        |
| Contact Fax Number:   | (812) 332-0281                                                |

#### Device Information:

| Trade Name:           | Arterial Pressure Monitoring Set/Tray |
|-----------------------|---------------------------------------|
| Common Name:          | Catheter, Intravascular, Diagnostic   |
| Classification Name:  | Diagnostic intravascular catheter     |
| Regulation:           | 21 CFR §870.1200                      |
| Product Code:         | DQO                                   |
| Device Class:         | II                                    |
| Classification Panel: | Cardiovascular                        |

### Predicate Device:

The Arterial Pressure Monitoring Set/Tray is substantially equivalent to the following device: Cook Pressure Monitoring Catheter (K002254, Cook Incorporated) cleared on March 16, 2001.

{4}------------------------------------------------

Cook Incorporated - Traditional 510(k) Arterial Pressure Monitoring Set/Tray December 17, 2018

#### Device Description:

The Arterial Pressure Monitoring Set/Tray is comprised of a pressure monitoring catheter, wire guide, entry access needle, and syringe. Certain sets are also available with a dilator and/or other convenience components. The catheter is manufactured from polyurethane, ethylene-vinyl acetate, or polyethylene tubing and is designed with a pre-molded winged hub. Catheters are manufactured 2.5, 3.0, and 4.0 French in size with endhole diameters of 0.015, 0.018, and 0.021 inches, respectively. The wire guide is manufactured from stainless steel coils, a safety wire, a mandril wire, and All-State solder. The access needle is manufactured from stainless steel. The dilator is manufactured from polyethylene and has a pre-molded proximal hub. Other convenience accessories include syringes, lidocaine solution, PVP ointment, antiseptic, Monoject needles, a scalpel, gauze, drapes, and needle holders. The Arterial Pressure Monitoring Set/Tray is sterilized by ethylene oxide and intended for one-time use.

#### Indications for Use:

The Arterial Pressure Monitoring Set/Tray is intended for arterial blood pressure monitoring and blood sampling in adult and pediatric patients.

- -2.5 French catheters are intended for patients from birth and older
- -3.0 and 4.0 French catheters are intended for patients 1 year and older

### Comparison to Predicate Device:

The Arterial Pressure Monitoring Set/Tray and the predicate device, Cook Pressure Monitoring Catheter (K002254), are substantially equivalent in that these devices are identical in intended use, principle of operation, and fundamental technologies. Additionally, the subject device has similar materials and dimensions as the predicate device. The differences between the subject device and the predicate device, including pediatric indication, materials, and other dimensions do not raise any new issues of safety and/or effectiveness. The substantial equivalence comparison of the subject device to the predicate is provided in the table below.

{5}------------------------------------------------

|                               | PREDICATE DEVICE                                                                                                                                                                            | SUBJECT DEVICE                                                                                                                                                                                                                                                                                                        |
|-------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                               | Cook Pressure Monitoring Catheter<br>(K002254)                                                                                                                                              | Arterial Pressure Monitoring Set/Tray                                                                                                                                                                                                                                                                                 |
| Regulation                    | 21 CFR §870.1200                                                                                                                                                                            | IDENTICAL TO PREDICATE                                                                                                                                                                                                                                                                                                |
| Product Code                  | DQO - Catheter, Intravascular, Diagnostic                                                                                                                                                   | IDENTICAL TO PREDICATE                                                                                                                                                                                                                                                                                                |
| Classification                | II                                                                                                                                                                                          | IDENTICAL TO PREDICATE                                                                                                                                                                                                                                                                                                |
| Indications for<br>Use        | The Cook Pressure Monitoring Catheter is<br>used for arterial pressure monitoring and<br>blood sampling. It is provided sterile in<br>peel-open packages and intended for one-<br>time use. | The Arterial Pressure Monitoring Set/Tray is<br>intended for arterial blood pressure<br>monitoring and blood sampling in adult and<br>pediatric patients.<br>- 2.5 French catheters are intended for<br>patients from birth and older<br>- 3.0 and 4.0 French catheters are<br>intended for patients 1 year and older |
| Device is for<br>One-time use | Yes                                                                                                                                                                                         | IDENTICAL TO PREDICATE                                                                                                                                                                                                                                                                                                |
| Catheter<br>Insertion Method  | Percutaneously via Seldinger technique                                                                                                                                                      | IDENTICAL TO PREDICATE                                                                                                                                                                                                                                                                                                |
| Catheter<br>Insertion Site    | Not specified                                                                                                                                                                               | Radial, Brachial, Dorsalis Pedis, Femoral,<br>Posterior Tibial Arteries                                                                                                                                                                                                                                               |
| Catheter Shaft<br>Material    | Polyurethane                                                                                                                                                                                | Polyethylene<br>Ethylene-Vinyl Acetate<br>Polyurethane                                                                                                                                                                                                                                                                |
| Catheter Hub<br>Material      | Polyurethane                                                                                                                                                                                | Polyethylene<br>Polyurethane                                                                                                                                                                                                                                                                                          |
| Catheter Outer<br>Diameter    | 2.5, 3.0, 4.0, 5.0 French                                                                                                                                                                   | 2.5, 3.0, 4.0 French                                                                                                                                                                                                                                                                                                  |
| Catheter Length               | 2.5, 5.0, 7.5, 8.0, 12.0, 22.0 cm                                                                                                                                                           | 2.5, 4.0, 5.0, 6.0, 8.0, 12.0 cm                                                                                                                                                                                                                                                                                      |
| Catheter Lumen<br>Design      | Single Lumen                                                                                                                                                                                | IDENTICAL TO PREDICATE                                                                                                                                                                                                                                                                                                |
| Catheter Distal<br>End        | Straight Tip                                                                                                                                                                                | IDENTICAL TO PREDICATE                                                                                                                                                                                                                                                                                                |
| Dilator Shaft<br>Material     | Polypropylene                                                                                                                                                                               | Polyethylene                                                                                                                                                                                                                                                                                                          |
| Wire Guide<br>Material        | Stainless Steel                                                                                                                                                                             | Stainless Steel<br>All-State Solder                                                                                                                                                                                                                                                                                   |
| Wire Guide<br>Diameter        | 0.015, 0.018, 0.021, 0.035 inch                                                                                                                                                             | IDENTICAL TO PREDICATE                                                                                                                                                                                                                                                                                                |
| Packaging                     | Tray w/ Tyvek Lidstock                                                                                                                                                                      | IDENTICAL TO PREDICATE                                                                                                                                                                                                                                                                                                |
| Sterilization                 | EtO                                                                                                                                                                                         | IDENTICAL TO PREDICATE                                                                                                                                                                                                                                                                                                |
| Sterility<br>Assurance Level  | 10-6                                                                                                                                                                                        | IDENTICAL TO PREDICATE                                                                                                                                                                                                                                                                                                |

{6}------------------------------------------------

Cook Incorporated - Traditional 510(k) Arterial Pressure Monitoring Set/Tray December 17, 2018

### Technological Characteristics:

The subject device, Arterial Pressure Monitoring Set/Tray, was subjected to applicable testing to assure reliable design and performance under the testing parameters. The following tests have been conducted to ensure reliable design and performance under the specified testing parameters:

# Performance Testing:

- Catheter Shaft Tensile (Time Zero and Aged) The peak tensile load for the shaft . section of the catheter shall be greater than or equal to 5 N for 2.5 and 3.0 Fr catheters, and 10 N for 4.0 and 5.0 Fr catheters in accordance with BS EN ISO 10555-1:2013. The acceptance criteria were met.
- . Catheter Hub-to-Shaft Tensile (Time Zero and Aged) - The peak tensile load for the hub-to-shaft section of the catheter shall be greater than or equal to 5 N for 2.5 and 3.0 Fr catheters, and 10 N for 4.0 and 5.0 Fr catheters in accordance with BS EN ISO 10555-1:2013. The acceptance criteria were met.
- Catheter Liquid Leakage (Time Zero and Aged) No part of the catheter shall leak . liquid when tested in accordance with Annex C of BS EN ISO 10555-1. The acceptance criterion was met.
- . Catheter Air Leakage (Time Zero and Aged) - No air shall enter the hub when tested in accordance with Annex D of BS EN ISO 10555-1. The acceptance criterion was met.
- Catheter Kink Length (Time Zero and Aged) The catheter shall not kink (flowrate . reduced by 50%) at a specified kink (circumferential) length when tested in accordance with Annex B of BS EN 13868. The acceptance criterion was met.
- Catheter Flow Rate (Time-Zero) The flow rate shall be a minimum of 90% of the . flow rate stated by the manufacturer when tested in accordance with Annex E of BS EN ISO 10555-1. The acceptance criterion was met.
- Visual Inspection, Dimensional Verification, Compatibility (Time-Zero, Aged) Test . was conducted to demonstrate that the catheter dimensions are within the specified tolerances and the catheter and wire guide are compatible.
- . Wire Guide Corrosion (Time-Zero) – The wire guide shall show no evidence of corrosion that could affect the functional performance when tested in accordance with Annex B of BS EN ISO 11070:2014. The acceptance criterion was met.

{7}------------------------------------------------

- · Wire Guide Fracture (Time-Zero) The wire guide shall not fracture when wound around an approximate former in accordance with Annex F of BS EN ISO 11070. The acceptance criterion was met.
- . Wire Guide Flex (Time-Zero) – The wire guide shall show no signs of defects or damage, including flaking or material loss when tested in accordance with Annex G of BS EN ISO 11070. The acceptance criterion was met.
- . Wire Guide Tensile (Time-Zero) – The peak tensile load to failure shall be greater than or equal to 5 N for wire guides with an outside diameter greater than or equal to 0.55 mm and less than 0.75 mm, and 10 N for wire guides with an outside diameter greater than or equal to 0.75 mm in accordance with Annex H of BS EN ISO 11070. The acceptance criterion was met.
- . Dilator Hub-to-Shaft Tensile (Time Zero and Aged) – The peak tensile load for the hub-to-shaft section of the dilator shall be greater than or equal to 5 N for 3.0 Fr dilators and 10 N for 3.5, 4.0, 5.0, and 5.5 Fr dilators, in accordance with BS EN ISO 11070. The acceptance criterion was met.

Biocompatibility Testing:

- Per ISO 10993-1 and FDA guidance, testing for cytotoxicity, sensitization, . intracutaneous irritation, acute systemic toxicity, material-mediated pyrogenicity, subacute/subchronic toxicity, genotoxicity, implantation, and hemocompatibility were performed to ensure the biocompatibility of the subject device set.
Sterilization Testing:

- Per ISO 10993-1 Residual levels of ethylene oxide (EO) and ethylene chlorohydrin . (ECH) have been evaluated and verified to be no greater than the maximum limits as specified in ANSI/AAMI/ISO 10993-7:2008(R)2012: Biological evaluation of medical devices - Part 7: Ethylene oxide sterilization residuals.
Packaging Testing following simulated distribution:

- . Visual Inspection - No package shall display damage that will compromise sterility of its contents. The acceptance criterion was met.
- . Destructive Inspection – The components inside the package shall show no signs of damage that would interfere with the performance of the product. The acceptance criterion was met.
- Bubble Test The package shall show no signs of pin holes or imperfect seals . indicated by bubbles created from the pressurized air leaking into the surrounding solution. The acceptance criterion was met.

{8}------------------------------------------------

Cook Incorporated - Traditional 510(k) Arterial Pressure Monitoring Set/Tray December 17, 2018

- Seal Strength Test Per EN 868-5:1999, the seal strength shall not be less than 1.2N ● per 15 mm (0.46 lbf/in). For additional confidence, the mean seal force - 3 σ shall meet or exceed 0.50 lbf/in. The acceptance criterion was met.
# Conclusion:

The results of these tests confirm that the Arterial Pressure Monitoring Set/Tray meets the design input requirements based on the intended use and support the conclusion that this device does not raise different questions of safety and/or effectiveness and is substantially equivalent to the predicate device, the Cook Pressure Monitoring Catheter (K002254).

---

**Source:** [https://fda.innolitics.com/device/K180846](https://fda.innolitics.com/device/K180846)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
