← Product Code [GBM](/productcode/GBM) · K180801

# Actreen Hi-Lite Cath, Actreen Hi-Lite Set (K180801)

_B.Braun Medical, Inc. · GBM · Nov 5, 2018 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K180801

## Device Facts

- **Applicant:** B.Braun Medical, Inc.
- **Product Code:** [GBM](/productcode/GBM.md)
- **Decision Date:** Nov 5, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.5130
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

Actreen® Hi-Lite Intermittent Urinary Catheters are indicated for intermittent urinary catheterization by adult and pediatric patients with chronic urine retention or voiding dysfunction.

## Device Story

Flexible tubular catheter for intermittent urinary drainage; inserted through urethra to pass fluids from urinary tract. Available in 14.5-inch length with straight or curved tip configurations; French gauge sizes 8-18. Features hydrophilic lubrication and smooth oval eyelets to minimize friction. Set includes pre-attached 1000mL urine collection bag with anti-reflux valve. Designed for single-use; packaging facilitates no-touch access for users with limited dexterity. Used in clinical or home settings by patients or caregivers to manage urine retention or voiding dysfunction.

## Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing and biocompatibility assessment. Biocompatibility testing performed per ISO 10993-1 for mucosal membrane contact (prolonged duration). Performance testing conducted per EN 1616:1997 (Sterile Urethral Catheters for Single Use).

## Technological Characteristics

Materials: Thermoplastic polyolefin (TPO) tubing, hydrophilic lubricant, EVA connector, acrylate/cyanoacrylate adhesive. Packaging: Polyethylene/polypropylene. Design: 14.5-inch length, straight/curved tips, French sizes 8-18. Includes anti-reflux valve in set configuration. Sterilization: Beta irradiation (E-beam).

## Regulatory Identification

A urological catheter and accessories is a flexible tubular device that is inserted through the urethra and used to pass fluids to or from the urinary tract. This generic type of device includes radiopaque urological catheters, ureteral catheters, urethral catheters, coudé catheters, balloon retention type catheters, straight catheters, upper urinary tract catheters, double lumen female urethrographic catheters, disposable ureteral catheters, male urethrographic catheters, and urological catheter accessories including ureteral catheter stylets, ureteral catheter adapters, ureteral catheter holders, ureteral catheter stylets, ureteral catheterization trays, and the gastro-urological irrigation tray (for urological use).

## Predicate Devices

- Actreen® Mini Intermittent Urinary Catheters ([K151772](/device/K151772.md))

## Reference Devices

- VaPro TM Intermittent Catheter ([K141642](/device/K141642.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, featuring a stylized human figure. To the right of that is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

November 5, 2018

B. Braun Medical, Inc. Anita J. Nemeth Senior Regulatory Affairs Analyst 901 Marcon Boulevard Allentown, PA 18109

Re: K180801

> Trade/Device Name: Actreen® Hi-Lite Intermittent Urinary Catheters Regulation Number: 21 CFR§ 876.5130 Regulation Name: Urological Catheter and Accessories Regulatory Class: II Product Code: GBM Dated: October 15, 2018 Received: October 16, 2018

Dear Anita J. Nemeth:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be avare that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see

https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Mark R. Kreitz -S

for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known)

K180801

Device Name

Actreen® Hi-Lite Intermittent Urinary Catheters

Indications for Use (Describe)

Actreen® Hi-Lite Intermittent Urinary Catheted for intermittent urinary catheterization by adult and pediatric patients with chronic urine retention or voiding dysfunction.

Type of Use (Select one or both, as applicable)

|--|--|

X Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

#### 5. 510(k) SUMMARY

| DATE:                     | March 23, 2018                                                                                                                                                                                        |
|---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| SUBMITTER:                | B. Braun Medical Inc.<br>901 Marcon Boulevard<br>Allentown, PA 18109-9341<br>610-266-0500                                                                                                             |
|                           | Contact: Anita J. Nemeth<br>Phone: (610) 596-2581<br>Fax: (610) 266-4962<br>E-mail: anita.nemeth@bbraunusa.com                                                                                        |
| DEVICE NAME:              | Actreen® Hi-Lite Intermittent Urinary Catheters                                                                                                                                                       |
| COMMON NAME:              | Intermittent Urinary Catheter                                                                                                                                                                         |
| DEVICE<br>CLASSIFICATION: | 21 CFR §876.5130, Class II<br>Urological Catheter and Accessories<br>Classification Product Code: GBM                                                                                                 |
| PREDICATE DEVICE:         | 510(k) Number: K151772<br>Device Name: Actreen® Mini Intermittent Urinary Catheters<br>Classification Product Code: GBM<br>Regulation Number: §876.5130, Class II<br>Applicant: B. Braun Medical Inc. |
| REFERENCE DEVICE:         | 510(k) Number: K141642<br>Device Name: VaPro <sup>TM</sup> Intermittent Catheter<br>Classification Product Code: GBM<br>Regulation Number: §876.5130, Class II<br>Applicant: Hollister, Incorporated  |

### DESCRIPTION OF THE DEVICE SUBJECT TO PREMARKET NOTIFICATION

The B. Braun Actreen® Hi-Lite Intermittent Urinary Catheters include the Actreen® Hi-Lite Cath and Actreen® Hi-Lite Set. The Actreen® Hi-Lite Set is identical to the Actreen® Hi-Lite Cath, except that the Actreen® Hi-Lite Set includes a urine collection bag which is pre-attached to the catheter. The proposed devices are flexible tubular devices that are inserted through the urethra and used to pass fluids from the urinary tract. The Actreen® Hi-Lite Intermittent Urinary Catheters are offered in a 14.5 inch length, a variety of gauge sizes and straight or curved tip configurations to accommodate the individual anatomy of both male and female users. The outer packaging was designed to facilitate easier, no touch access for those with limited dexterity, while the hydrophilic lubrication makes this single use catheter ready to use.

#### CONFIDENTIAL

{4}------------------------------------------------

## INDICATIONS FOR USE

Actreen® Hi-Lite Intermittent Urinary Catheters are indicated for intermittent urinary catheterization by adult and pediatic patients with chronic urine retention or voiding dysfunction.

## SUBSTANTIAL EQUIVALENCE

B. Braun Medical Inc. Actreen® Hi-Lite Intermittent Urinary Catheters are substantially equivalent to the Actreen® Mini Intermittent Urinary Catheters (K151772) predicate devices having similar intended use, technological properties, and performance.

The Actreen® Hi-Lite Intermittent Urinary Catheter differs from the predicate device in that the Actreen® Hi-Lite Intermittent Urinary Catheter is available in a longer length, additional gauge sizes and straight or curved tip configurations to accommodate the individual anatomy of both male and female users. The predicate device is offered in a 3.5 inch straight catheter tube length in French gauge sizes ranging from 10-14, whereas the proposed device is offered in a 14.5 inch catheter tube length with straight or curved tip options and French gauge sizes ranging from 8-18 for the catheter and French gauge sizes from 10-18 for the set. Additionally, the pre-attached urine collection bag on the predicate device is designed to hold a volume up to 700mL, while the proposed device collection bag holds a volume up to 1000mL.

### Intended Use

The Actreen® Hi-Lite Intermittent Urinary Catheters have similar intended use to the predicate devices. Both the proposed device and the predicate device are intended for intermittent urinary catheterization.

### Technical Characteristics

The predicate devices and proposed are comprised of the same materials, with the exception of one material of the primary packaging of the catheter. They include the same components, the same general design and have similar technological characteristics. Both are made of thermoplastic polyolefin (TPO) catheter tubing which is pre-lubricated with a hydrophilic lubricant, and include smooth, oval eyelets designed to minimize friction. The proposed Actreen® Hi-Lite Intermittent Urinary Catheters are available in similar gauge sizes as the Actreen® Mini Intermittent Urinary Catheters. The Actreen® Hi-Lite Set and Actreen® Mini Set both come with a pre-attached polyethylene/polypropylene urine collection bag and both contain an "anti-reflux valve" that allows urine to flow into the bag, but not back out of the bag into the catheter.

### Performance Data

Biocompatibility and performance testing were performed to support substantial equivalence of the subject devices to the predicate devices. Biocompatibility testing was performed in accordance with ISO 10993-1. Performance testing was performed according to EN 1616: 1997 Sterile Urethral Catheters for Single Use.

Table 5.1 summarizes the comparison between the Actreen® Hi-Lite Intermittent Urinary Catheters and the Actreen® Mini Intermittent Urinary Catheters.

#### CONFIDENTIAL

{5}------------------------------------------------

| Table 5.1 Device Comparison Table      |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                                                                                                                                                                                                                                                             |
|----------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Item                                   | Proposed Device: KXXXXXX<br>Product code: GBM<br>Actreen® Hi-Lite Cath<br>Actreen® Hi-Lite Set                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Predicate Device: K151772<br>Product code: GBM<br>Actreen® Mini Cath<br>Actreen® Mini Set                                                                                                                                                                                                                                                                                                   |
| Intended Use                           | Intermittent urinary catheterization                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Intermittent urinary catheterization                                                                                                                                                                                                                                                                                                                                                        |
| Indications for<br>Use                 | Actreen® Hi-Lite Intermittent Urinary Catheters<br>are indicated for intermittent urinary<br>catheterization by adult and pediatric patients with<br>chronic urine retention or voiding dysfunction.                                                                                                                                                                                                                                                                                                                                                                                                                             | Actreen® Mini Intermittent Urinary Catheters are<br>indicated for intermittent urinary catheterization by<br>female patients with chronic urine retention or<br>voiding dysfunction.                                                                                                                                                                                                        |
| Description                            | The Actreen® Hi-Lite Cath and Actreen® Hi-Lite<br>Set are sterile, single use, disposable catheters<br>designed for use by male or female patients.<br>Once removed from the package, the catheters<br>are ready to use and are pre- lubricated with a<br>hydrophilic lubrication. They are available in<br>one length, with a straight or curved tip<br>configuration to accommodate the individual<br>anatomy of both male and female users. The<br>catheter is available in six gauges sizes, while<br>the set is available in five gauge sizes. The<br>Actreen® Hi-Lite Set includes a pre-attached<br>urine collection bag. | The Actreen® Mini Cath and Actreen® Mini Set are<br>sterile, single use, disposable catheters designed for<br>use by females.<br>Once removed from the package the catheters are<br>ready to use and are pre- lubricated with a<br>hydrophilic lubrication. They are available in one<br>length with a straight tip. The Actreen® Mini Set<br>includes a pre-attached urine collection bag. |
| Dimensions                             | Catheter: 14.5 inch straight or curved tip<br>catheter tube length,<br>French sizes 08, 10, 12, 14, 16, 18<br><br>Set: 14.5 inch straight or curved tip catheter<br>tube length,<br>French sizes 10, 12, 14, 16, 18<br>Actreen® Hi-Lite Set urine collection bag<br>volume: 1000mL                                                                                                                                                                                                                                                                                                                                               | Catheter/Set: 3.5 inch straight catheter tube length,<br>French sizes 10, 12, 14<br>Actreen® Mini Set urine collection bag volume:<br>700mL                                                                                                                                                                                                                                                 |
| Materials                              | Catheter tube : Thermoplastic Polyolefin (TPO)<br>Lubricant : hydrophilic lubricant<br>Connector : EVA<br>Connector glue : Acrylate UV or<br>Cyanoacrylate<br>Actreen® Hi-Lite Cath primary packaging :<br>Polyethylene, Polypropylene<br>Actreen® Hi-Lite Set primary packaging and<br>collection bag : Polyethylene, Polypropylene                                                                                                                                                                                                                                                                                             | Catheter tube : Thermoplastic Polyolefin (TPO)<br>Lubricant : hydrophilic lubricant<br>Connector : EVA<br>Connector glue : Acrylate UV or<br>Cyanoacrylate<br>Actreen® Mini Cath primary packaging :<br>Polyethylene<br>Actreen® Mini Set primary packaging and<br>collection bag : Polyethylene, Polypropylene                                                                             |
| General<br>Performance<br>Requirements | EN 1616 : 1997 - Sterile Urethral Catheters for<br>Single Use.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | EN 1616 : 1997 - Sterile Urethral Catheters for<br>Single Use.                                                                                                                                                                                                                                                                                                                              |
| Biocompatibility                       | In accordance with ISO 10993-1<br>Classification: surface-contacting devices -<br>mucosal membrane<br>Contact Duration: prolonged exposure and a<br>duration of use greater than 24 hours but less<br>than 30 days.                                                                                                                                                                                                                                                                                                                                                                                                              | In accordance with ISO 10993-1<br>Classification: surface-contacting devices -<br>mucosal membrane<br>Contact Duration: prolonged exposure and a<br>duration of use greater than 24 hours but less<br>than 30 days.                                                                                                                                                                         |
| Sterilization                          | Beta Irradiation (E-beam)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Beta Irradiation (E-beam)                                                                                                                                                                                                                                                                                                                                                                   |

{6}------------------------------------------------

## CONCLUSION

Based on the results of biocompatibility and performance testing, the proposed B. Braun Medical Actreen® Hi-Lite Intermittent Urinary Catheters are considered substantially equivalent to the predicate devices.

---

**Source:** [https://fda.innolitics.com/device/K180801](https://fda.innolitics.com/device/K180801)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
