K180799 · Clarius Mobile Health Corp. · IYN · May 14, 2018 · Radiology
Device Facts
Record ID
K180799
Device Name
Clarius Ultrasound Scanner
Applicant
Clarius Mobile Health Corp.
Product Code
IYN · Radiology
Decision Date
May 14, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1550
Device Class
Class 2
Attributes
AI/ML, Pediatric, 3rd-Party Reviewed
Indications for Use
Diagnostic ultrasound imaging and fluid flow analysis.
Device Story
Portable, software-controlled ultrasound system; acquires/displays high-resolution, real-time ultrasound data. System comprises wireless transducers (convex, linear, phased array, endocavitary) communicating via Wi-Fi Direct to off-the-shelf iOS/Android tablets/smartphones. Piezoelectric material transmits high-frequency sound waves; converts echoes into electronic signals; constructs images. Transducer houses battery, beamformer, pre-scan converter, Wi-Fi components. Used by trained healthcare professionals in clinics, EMS, ambulances, or aircraft. Output displayed on mobile device; assists in diagnostic imaging and needle guidance; facilitates clinical decision-making through real-time visualization of internal structures.
Clinical Evidence
No clinical data. Substantial equivalence established via bench testing, including electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), acoustic output (NEMA UD 2), and biocompatibility (ISO 10993 series).
Technological Characteristics
Wireless ultrasound transducers (convex, linear, endocavitary) with Li-ion battery. Sensing via piezoelectric material. Connectivity: Wi-Fi Direct (Ad-Hoc) to iOS/Android devices. Software-controlled beamforming and processing. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-37, NEMA UD 2, ISO 10993, ISO 62304, ISO 14971.
Indications for Use
Indicated for diagnostic ultrasound imaging and fluid flow analysis in ophthalmic, fetal, abdominal, intra-operative (non-neurological), pediatric, small organ, cephalic (adult), trans-rectal, trans-vaginal, musculo-skeletal (conventional, superficial), urology, gynecology, cardiac (adult, pediatric), peripheral vessel, carotid, and procedural guidance of needles into the body. Intended for use by trained healthcare professionals in professional healthcare facilities, emergency medical services, ambulances, or rotary aircraft.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
ACUSON S3000, S2000, S1000 Diagnostic Ultrasound Systems (K172162)
Submission Summary (Full Text)
{0}------------------------------------------------
May 14, 2018
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The FDA text logo is in blue and includes the words "FDA U.S. FOOD & DRUG" on the top line and "ADMINISTRATION" on the bottom line.
Clarius Mobile Health Corp. % Mr. Mark Job Responsible Third Party Official Regulatory Technology Services, LLC 1394 25th Street. NW BUFFALO MN 55313
Re: K180799
Trade/Device Name: Clarius Ultrasound Scanner Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, ITX Dated: May 8, 2018 Received: May 9, 2018
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice
(https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Michael D. O'Hara For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known)
K180799
Device Name Clarius Ultrasound Scanner
#### Indications for Use (Describe)
The Clarius Ultrasound Scanner is a software-based ultrasound imaging system and accessories, intended for diagnostic imaging. It is indicated for diagnostic ultrasound imaging and fluid flow analysis in the following applications: ophthalmic, fetal, abdominal, intra-operative (non-neurological), pediatric, small organ, cephalic (adult), trans-rectal, trans-vaginal, musculo-skeletal (conventional, superficial), urology, cardiac (adult, pediatric), peripheral vessel, carotid, and procedural guidance of needles into the body.
The system is a transportable ultrasound system intended for use in environments where is provided by trained healthcare professionals.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### Indications for Use Tables
Clarius Ultrasound Scanner
| Clinical Application | | Mode of Operation | | | | | | |
|---------------------------|-----------------------------------------------------|-------------------|---|------------------|------------------|---------------|---------------------------|--------|
| General<br>(Track 1 Only) | Specific<br>(Tracks 1 & 3) | B | M | Color<br>Doppler | Power<br>Doppler | PW<br>Doppler | Combined<br>(Specify) | Other* |
| Ophthalmic | Ophthalmic | P | | | | | | |
| Fetal Imaging<br>& Other | Fetal | P | P | P | P | N | B+M; B+CD;<br>B+PD; B+PWD | |
| | Abdominal | P | P | P | P | N | B+M; B+CD;<br>B+PD; B+PWD | Note 1 |
| | Intra-operative (Abdominal<br>organs & vascular) | P | P | P | P | N | B+M; B+CD;<br>B+PD; B+PWD | Note 1 |
| | Laparoscopic | | | | | | | |
| | Pediatric | P | P | P | P | N | B+M; B+CD;<br>B+PD; B+PWD | Note 1 |
| | Small Organ (Thyroid, Prostate,<br>Scrotum, Breast) | P | P | P | P | N | B+CD; B+PD;<br>B+PWD | Note 1 |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | P | P | P | P | N | B+M; B+CD;<br>B+PD; B+PWD | |
| | Trans-rectal | N | N | N | N | N | B+M; B+CD;<br>B+PD; B+PWD | |
| | Trans-vaginal | N | N | N | N | N | B+M; B+CD;<br>B+PD; B+PWD | |
| | Trans-urethral | | | | | | | |
| | Trans-esophageal (non-Cardiac) | | | | | | | |
| | Musculo-skeletal (Conventional) | P | P | P | P | N | B+CD; B+PD;<br>B+PWD | Note 1 |
| | Musculo-skeletal (Superficial) | P | P | P | P | N | B+CD; B+PDN | Note 1 |
| | Intravascular | | | | | | | |
| | Other (Urology, Gynecology) | P | P | P | P | N | B+M; B+CD;<br>B+PD; B+PWD | |
| Cardiac | Cardiac Adult | P | P | P | | N | B+M; B+CD;<br>B+PWD | |
| | Cardiac Pediatric | P | P | P | | N | B+M; B+CD;<br>B+PWD | |
| | Intravascular (Cardiac) | | | | | | | |
| | Trans-esophageal (Cardiac) | | | | | | | |
| | Intra-cardiac | | | | | | | |
| | Other | | | | | | | |
| Peripheral<br>Vessel | Peripheral Vessel | P | P | P | P | N | B+M; B+CD;<br>B+PD; B+PWD | Note 1 |
| | Other (Carotid) | P | P | P | P | N | B+M; B+CD;<br>B+PD; B+PWD | Note 1 |
#### System: Clarius Ultrasound Scanner Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
{4}------------------------------------------------
### System: Clarius Ultrasound Scanner Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
{5}------------------------------------------------
#### C3 Convex Scanner
#### Clinical Application Mode of Operation PW B M Power Combined Other* General Specific Color (Track 1 Only) (Tracks 1 & 3) Doppler Doppler Doppler (Specify) Ophthalmic Ophthalmic Fetal Imaging Fetal P P P P N B+M; B+CD; & Other B+PD; B+PWD P P P P Abdominal N B+M; B+CD; B+PD; B+PWD P P P Intra-operative (Abdominal P N B+M; B+CD; B+PD; B+PWD organs & vascular) Laparoscopic Pediatric P P P P N B+M; B+CD; B+PD; B+PWD Small Organ (Thyroid, Prostate, Scrotum, Breast) Neonatal Cephalic P P Adult Cephalic P P N B+M; B+CD; B+PD; B+PWD Trans-rectal Trans-vaginal Trans-urethral Trans-esophageal (non-Cardiac) P P P P Musculo-skeletal (Conventional) N B+M; B+CD; B+PD; B+PWD Musculo-skeletal (Superficial) Intravascular Other (Urology, Gynecology) P P P P N B+M; B+CD; B+PD; B+PWD
#### Device Name: C3 Convex Scanner Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
{6}------------------------------------------------
#### Device Name: C3 Convex Scanner
| Cardiac | Cardiac Adult | P | P | P | | N | B+M; B+CD;<br>B+PWD |
|-------------------------------------------------------|----------------------------|---|---|---|---|---|---------------------------|
| | Cardiac Pediatric | P | P | P | | N | B+M; B+CD;<br>B+PWD |
| | Intravascular (Cardiac) | | | | | | |
| | Trans-esophageal (Cardiac) | | | | | | |
| | Intra-cardiac | | | | | | |
| | Other | | | | | | |
| Peripheral<br>Vessel | Peripheral Vessel | P | P | P | P | N | B+M; B+CD;<br>B+PD; B+PWD |
| | Other (Carotid) | | | | | | |
| N = new indication; P = previously cleared by K172385 | | | | | | | |
| Note 1: Needle Enhancement in B-Mode. | | | | | | | |
{7}------------------------------------------------
#### C7 Convex Scanner
#### Clinical Application Mode of Operation M PW B Combined Other* General Specific Color Power (Track 1 Only) (Tracks 1 & 3) Doppler Doppler Doppler (Specify) Ophthalmic Ophthalmic Fetal Imaging Fetal P P P P N B+M; B+CD; & Other B+PD; B+PWD P P P P N Abdominal B+M; B+CD; B+PD; B+PWD P P Intra-operative (Abdominal P P N B+M; B+CD; B+PD; B+PWD organs & vascular) Laparoscopic Pediatric P P P P N B+M; B+CD; B+PD; B+PWD Small Organ (Thyroid, Prostate, P P P P N B+M; B+CD; Scrotum, Breast) B+PD; B+PWD Neonatal Cephalic Adult Cephalic Trans-rectal Trans-vaginal Trans-urethral Trans-esophaqeal (non-Cardiac) Musculo-skeletal (Conventional) P P P P N B+M; B+CD; B+PD; B+PWD Musculo-skeletal (Superficial) Intravascular Other (Urology, Gynecology) P P P P N B+M; B+CD; B+PD; B+PWD
#### Device Name: C7 Convex Scanner Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
{8}------------------------------------------------
#### Device Name: C7 Convex Scanner
| Cardiac | Cardiac Adult | P | P | P | P | N | B+M; B+CD;<br>B+PD; B+PWD | |
|------------------------------------------------------------------------------------------------|----------------------------|---|---|---|---|---|---------------------------|--|
| | Cardiac Pediatric | P | P | P | | N | B+M; B+CD;<br>B+PWD | |
| | Intravascular (Cardiac) | | | | | | | |
| | Trans-esophageal (Cardiac) | | | | | | | |
| | Intra-cardiac | | | | | | | |
| | Other | | | | | | | |
| Peripheral<br>Vessel | Peripheral Vessel | P | P | P | P | N | B+M; B+CD;<br>B+PD; B+PWD | |
| | Other (Carotid) | | | | | | | |
| N = new indication; P = previously cleared by K172385<br>Note 1: Needle Enhancement in B-Mode. | | | | | | | | |
{9}------------------------------------------------
#### L7 Linear Scanner
#### Clinical Application Mode of Operation PW Specific B M Power Combined Other* General Color (Track 1 Only) (Tracks 1 & 3) Doppler Doppler Doppler (Specify) P Ophthalmic Ophthalmic Fetal Imaging Fetal & Other Abdominal P P P P N B+M; B+CD; Note 1 B+PD; B+PWD P P Intra-operative (Abdominal P P N B+M; B+CD; Note 1 organs & vascular) B+PD; B+PWD Laparoscopic P P Pediatric P P N B+M; B+CD; Note 1 B+PD; B+PWD Small Organ (Thyroid, Prostate, P P P N B+CD; B+PD; Note 1 Scrotum, Breast) B+PWD Neonatal Cephalic Adult Cephalic Trans-rectal Trans-vaginal Trans-urethral Trans-esophageal (non-Cardiac) Musculo-skeletal (Conventional) P P P P N B+M; B+CD; Note 1 B+PD; B+PWD Musculo-skeletal (Superficial) P P P P B+M; B+CD; N Note 1 B+PD; B+PWD Intravascular Other (Urology, Gynecology) Cardiac Cardiac Adult Cardiac Pediatric Intravascular (Cardiac) Trans-esophageal (Cardiac) Intra-cardiac Other
#### Device Name: L7 Linear Scanner Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
{10}------------------------------------------------
#### Device Name: L7 Linear Scanner
| Peripheral<br>Vessel | Peripheral Vessel | P | P | P | P | N | B+M; B+CD;<br>B+PD; B+PWD | Note 1 |
|------------------------------------------------------------------------------------------------|-------------------|---|---|---|---|---|---------------------------|--------|
| | Other (Carotid) | P | P | P | P | N | B+M; B+CD;<br>B+PD; B+PWD | Note 1 |
| N = new indication; P = previously cleared by K172385<br>Note 1: Needle Enhancement in B-Mode. | | | | | | | | |
{11}------------------------------------------------
#### EC7 Scanner
#### Device Name: EC7 Scanner
| Clinical Application | | Mode of Operation | | | | | | |
|---------------------------|-----------------------------------------------------|-------------------|---|------------------|------------------|---------------|---------------------------|--------|
| General<br>(Track 1 Only) | Specific<br>(Tracks 1 & 3) | B | M | Color<br>Doppler | Power<br>Doppler | PW<br>Doppler | Combined<br>(Specify) | Other* |
| Ophthalmic | Ophthalmic | | | | | | | |
| Fetal Imaging &<br>Other | Fetal | N | N | N | N | N | B+M; B+CD;<br>B+PD; B+PWD | |
| | Abdominal | N | N | N | N | N | B+M; B+CD;<br>B+PD; B+PWD | |
| | Intra-operative (Abdominal<br>organs & vascular) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | | | | | | | |
| | Small Organ (Thyroid, Prostate,<br>Scrotum, Breast) | N | N | N | N | N | B+M; B+CD;<br>B+PD; B+PWD | |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | N | N | N | N | N | B+M; B+CD;<br>B+PD; B+PWD | |
| | Trans-vaginal | N | N | N | N | N | B+M; B+CD;<br>B+PD; B+PWD | |
| | Trans-urethral | | | | | | | |
| | Trans-esophageal (non-Cardiac) | | | | | | | |
| | Musculo-skeletal (Conventional) | | | | | | | |
| | Musculo-skeletal (Superficial) | | | | | | | |
| | Intravascular | | | | | | | |
| | Other (Urology, Gynecology) | N | N | N | N | N | B+M; B+CD;<br>B+PD; B+PWD | |
{12}------------------------------------------------
#### Device Name: EC7 Scanner
| Cardiac | Cardiac Adult | | | | |
|----------------------|----------------------------|--|--|--|--|
| | Cardiac Pediatric | | | | |
| | Intravascular (Cardiac) | | | | |
| | Trans-esophageal (Cardiac) | | | | |
| | Intra-cardiac | | | | |
| | Other | | | | |
| Peripheral<br>Vessel | Peripheral Vessel | | | | |
| | Other (Carotid) | | | | |
{13}------------------------------------------------
Image /page/13/Picture/0 description: The image shows the Clarius logo. The logo consists of an orange symbol that looks like three concentric letter C's, followed by the word "clarius" in dark blue, sans-serif font. The logo is simple and modern, with a focus on the company name.
# 510(k) Summary
### 1. Submitter's name, address, telephone number, and contact person:
| Submitter: | Brendan Seward<br>QA/RA Manager<br>Clarius Mobile Health Corp.<br>350 – 3605 Gilmore Way, Burnaby, B.C., Canada, V5G 4X5<br>Email: brendan.seward@clarius.me<br>Tel: (+1) 778-800-9975<br>Fax: N/A |
|----------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date Prepared: | February 20, 2018 |
- 2. Name of the device, including the trade or proprietary name, if applicable, the common or usual name, and the classification, if known:
| Device Name: | Clarius Ultrasound Scanner |
|-----------------------|----------------------------------------------|
| Common Name: | Diagnostic Ultrasound System and Accessories |
| Classification: | Class II |
| Classification Names: | |
| 21 CFR Section | Classification Name | Product Code |
|----------------|------------------------------------------|--------------|
| 892.1550 | Ultrasonic Pulsed Doppler Imaging System | 90 IYN |
| 892.1560 | Ultrasonic Pulsed Echo Imaging System | 90 IYO |
| 892.1570 | Diagnostic Ultrasound Transducer | 90 ITX |
#### 3. Substantially Equivalent Devices:
| Device Name | 510(k) Number |
|----------------------------------------------------------|---------------|
| Clarius Ultrasound Scanner | K172385 |
| ACUSON S3000, S2000, S1000 Diagnostic Ultrasound Systems | K172162 |
### Device Description
The Clarius Ultrasound Scanner is a portable, general-purpose, software-controlled, diagnostic ultrasound system used to acquire and display high-resolution, real-time ultrasound data through an off-the-shelf iOS or Android device. The Clarius Ultrasound Scanner comprises a series of wireless transducers employing Bluetooth and Wi-Fi-based technology to communicate with traditional tablet/smartphone devices via direct Wi-Fi. This allows the user to export ultrasound images and display them across a range of portable personal devices.
{14}------------------------------------------------
Image /page/14/Picture/0 description: The image shows the Clarius logo. The logo consists of an orange symbol that looks like three concentric letter C's. To the right of the symbol is the word "clarius" in dark blue, with the first letter in lowercase.
Image /page/14/Figure/1 description: The image shows a diagram of an ultrasound system. On the left side of the image, there is a device that is labeled with the following components: Battery, Wi-Fi, Power Circuitry, Ultrasound System-on-Chip Beamforming & Processing, Transmit/Receive Analog Interface, and Transducer Stack. The device is scanning a patient, and the ultrasound image is being displayed on a Wi-Fi enabled smartphone or tablet on the right side of the image. The smartphone or tablet also has a user interface.
Ultrasound Probe
User Interface / Display
Piezoelectric material in the systems transmits high frequency, non-ionizing sound waves to the designated region of the body and converts the subsequent echoes detected into electronic signals in order to construct an image of the internal structures of an anatomical field. This image is sent wirelessly to an external iOS or Android viewing device on which the image can be displayed. The transducer houses a battery and power generator, multichannel beam former, pre-scan converter, and Wi-Fi components. The battery is removable and comes with a separate charger. Communication between the transducer and the compatible viewing device is performed via Wi-Fi Direct® (Ad-Hoc mode with security) for easy pairing.
The Clarius Ultrasound System product/package components include:
- 1. Software:
- . The Clarius Ultrasound App (Clarius App) for iOS; OR
- The Clarius Ultrasound App (Clarius App) for Android. ●
- 2. Transducers/Scanners:
- Clarius Scanner C3 (C3 Convex Transducer);
- . Clarius Scanner EC7 (EC7 Convex Endocavitary Transducer);
- Clarius Scanner C7 (C7 Convex Transducer); OR ●
- Clarius Scanner L7 (L7 Linear Transducer).
{15}------------------------------------------------
Image /page/15/Picture/0 description: The image shows the logo for Clarius Mobile Health. The logo consists of an orange symbol on the left and the word "clarius" in dark blue on the right. The symbol is made up of three curved lines that form a stylized "C". The word "clarius" is in a sans-serif font and is all lowercase.
- 3. Accessories:
- a. Clarius-Built:
- o Battery Pack (Li-ion);
- Battery Charger; O
- Clarius Dock (Optional; charging station); and O
- Clarius Fan (Optional). O
- b. OEM/Off-the-Shelf Product(s):
- Medical Power Supply (Off-the-shelf power adaptor from SL Power Electronics; O Model Number ME20A1203B02; approved in the US); and
- Medical Power Supply (Optional for use with Clarius Dock; off-the-shelf power o adaptor from MEAN WELL Enterprises Co., Ltd.; Model Number GSM160B12-R7B; approved in the US).
The concept of the Clarius Ultrasound System transducers and software is to provide an easy to use, highperformance, low-cost, ultrasound platform for teaching and clinical applications. The Clarius Ultrasound Scanner is intended for use in professional healthcare facilities where healthcare is provided by trained medical professionals. The device is also intended for use in Emergency Medical Service, ambulance, or rotary aircraft environments.
### Changes Implemented
The present submission is an expansion based upon a previous clearance – K172385. The subject changes of this new submission are the added mode of operation of Power Doppler and the addition of a new transducer model, Clarius Scanner EC7, which functions as an endocavitary scanner. This new scanner adds trans-rectal and trans-vaginal to the indications for use of the system. Appropriate changes have been made to mitigate risks and ensure effectiveness of these new features.
### Intended Use
Diagnostic ultrasound imaging and fluid flow analysis.
### Indications for Use
The Clarius Ultrasound Scanner is a software-based ultrasound imaging system and accessories, intended for diagnostic imaging. It is indicated for diagnostic ultrasound imaging and fluid flow analysis in the following applications: ophthalmic, fetal, abdominal, intra-operative (non-neurological), pediatric, small organ, cephalic (adult), trans-rectal, trans-vaginal, musculo-skeletal (conventional, superficial), urology, gynecology, cardiac (adult, pediatric), peripheral vessel, carotid, and procedural guidance of needles into the body.
The system is a transportable ultrasound system intended for use in environments where is provided by trained healthcare professionals.
### Predicate Devices
Equivalent devices are referred to as predicate devices in alignment with the FDA's standard terminology for comparable devices. The predicate devices selected to demonstrate equivalence are:
- 1. Device Name FDA 510(k) Number Clarius Ultrasound Scanner (K172385) FDA 510(k) Number 2. Device Name ACUSON S3000, S2000, S1000 Diagnostic Ultrasound Systems K172162
{16}------------------------------------------------
Image /page/16/Picture/0 description: The image shows the Clarius logo. The logo consists of an orange icon on the left and the word "clarius" in dark blue on the right. The icon is made up of three curved lines that form a stylized "C". The word "clarius" is written in a sans-serif font.
## Determination of Substantial Equivalence
The Clarius Ultrasound system is a Track 3 system that employs the same fundamental scientific technology as that cleared with K172385 and K172162. All indications for use introduced by the Clarius Ultrasound Scanner are similar to at least one of the predicate devices. A comparison table is provided below:
| Criteria for Comparison | Clarius Ultrasound Scanner | Clarius Ultrasound System<br>(K172385) | ACCUSON S3000, S2000, S1000<br>Ultrasound Systems<br>(K172162) |
|-------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Portability | Portable ultrasound system | Portable ultrasound system | Mobile ultrasound system |
| Power Source | Removable battery (Li-ion) | Removable battery (Li-ion) | Battery or AC power |
| Display | iOS or Android mobile device | iOS or Android mobile device | Inbuilt digital display |
| Wireless Capability | Communicates wirelessly via Wi-Fi and Bluetooth | Communicates wirelessly via Wi-Fi and Bluetooth | Communicates via cable |
| 510(k) Track | Track 3 | Track 3 | Track 3 |
| Transducer Types | – Convex Array<br>– Linear Array<br>– Phased Array<br>– Intracavity | – Convex Array<br>– Linear Array<br>– Phased Array | – Convex Array<br>– Linear Array<br>– Phased Array<br>– Intracavity Array<br>– Trans-esophageal |
| Intended Use | Diagnostic ultrasound imaging and fluid flow analysis | Diagnostic ultrasound imaging and fluid flow analysis | Diagnostic ultrasound imaging and fluid flow analysis |
| Indications for Use | – Ophthalmic<br>– Fetal<br>– Abdominal<br>– Intraoperative (Ab/Vasc)<br>– Pediatric<br>– Small organ<br>– Adult cephalic<br>– Trans-rectal<br>– Trans-vaginal<br>– Musculo-skel. (Conv.)<br>– Musculo-skel. (Superfic.)<br>– Urology<br>– Gynecology<br>– Cardiac adult<br>– Cardiac pediatric<br>– Peripheral vessel<br>– Carotid<br>– Needle guidance | – Ophthalmic<br>– Fetal<br>– Abdominal<br>– Intraoperative (Ab/Vasc)<br>– Pediatric<br>– Small organ<br>– Adult cephalic<br>– Musculo-skel. (Conv.)<br>– Musculo-skel. (Superfic.)<br>– Urology<br>– Gynecology<br>– Cardiac adult<br>– Cardiac pediatric<br>– Fetal echo<br>– Peripheral vessel<br>– Carotid<br>– Needle guidance | – Fetal<br>– Abdominal<br>– Intraoperative (Ab/Vasc)<br>– Intraoperative (Neuro)<br>– Pediatric<br>– Small organ<br>– Neonatal cephalic<br>– Adult cephalic<br>– Trans-rectal<br>– Trans-vaginal<br>– Musculo-skel. (Conv.)<br>– Musculo-skel. (Superfic.)<br>– Gynecology<br>– Cardiac adult<br>– Cardiac pediatric<br>– Trans-esophageal (Card.)<br>– Peripheral vessel<br>– Needle guidance |
| Modes of Operation | – B-mode<br>– M-mode<br>– Color Doppler<br>– Power Doppler<br>– PWD<br>– CWD<br>– Combined (B+M; B+CD; B+PD;<br>B+PWD) | – B-mode<br>– M-mode<br>– Color Doppler<br>– Power Doppler<br>– CWD<br>– Combined (B+M, B+CD, B+PD) | – B-Mode<br>– M-Mode<br>– Color Doppler<br>– Color M Doppler<br>– Power Doppler<br>– PWD<br>– CWD<br>– Harmonic Imaging<br>– Combined (B+M, B+PWD,<br>B+CD, B+PD, B+CWD) |
{17}------------------------------------------------
Image /page/17/Picture/0 description: The image shows the Clarius logo. The logo consists of an orange icon on the left and the word "clarius" in dark blue on the right. The icon is made up of three concentric circles that are not fully closed, creating a stylized "C" shape.
## Equivalency Conclusion
The subject device does not introduce any new technology or indications for use; therefore, the system is substantially equivalent to the predicate devices.
### Nonclinical Performance Data
Nonclinical performance tests show compliance to the following standards:
| Reference No. | Year | Title |
|------------------------|-------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| AAMI/ANSI<br>ES60601-1 | 2012 | Medical Electrical Equipment - Part 1: General Requirements for Basic Safety and Essential<br>Performance |
| AAMI ANSI<br>ES60601-1 | 2012 | AAMI / ANSI ES60601-1:2005/(R)2012 And A1:2012, C1:2009/(R)2012 And A2:2010/(R)2012<br>(Consolidated Text) Medical Electrical Equipment - Part 1: General Requirements for Basic Safety<br>and Essential Performance (IEC 60601-1:2005, Mod). (General II (ES/EMC)) |
| IEC 60601-1-2 | 2007<br>&<br>2014 | Medical electrical equipment - Part 1-2: General requirements for basic safety and essential<br>performance - Collateral Standard: Electromagnetic Capability - Requirements and tests. (4th<br>Edition) |
| IEC 60601-1-6 | 2013 | Medical electrical equipment - Part 1-6: General requirements for basic safety and essential<br>performance -Collateral standard: Usability |
| IEC 60601-1-12 | 2014 | Medical Electrical Equipment - Part 1-12: General requirements for basic safety and essential<br>performance - Collateral Standard: Requirements for medical electrical equipment and medical<br>electrical systems used in the emergency medical services environment |
| IEC 60601-2-37 | 2004 | Medical electrical equipment - Part 2-37: Particular requirements for the basic safety and essential<br>performance of ultrasonic medical diagnostic and monitoring equipment. (3rd and 4th Edition) |
| IEC 62133 | 2012 | Secondary Cells and Batteries Containing Alkaline or Other Non-Acid Electrolytes - Safety<br>Requirements for Portable Sealed Secondary Cells, And for Batteries Made from Them, For Use in<br>Portable Applications [Including: Corrigendum 1 (2013)] |
| IEC 62366 | 2014 | Consolidated Version Medical Devices - Application of Usability Engineering to Medical Devices |
| ISO 10993-1 | 2009 | Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management<br>process |
| ISO 10993-5 | 2014 | Biological Evaluation of Medical Devices - Part 5: Tests for In Vitro Cytotoxicity |
| ISO 10993-10 | 2014 | Biological Evaluation of Medical Devices - Part 10: Tests for Irritation and Skin Sensitization |
| ISO 10993-11 | 2017 | Biological evaluation of medical devices -- Part 11: Tests for systemic toxicity |
| ISO-10993-12 | 2014 | Biological Evaluation of Medical Devices - Part 12: Sample Preparation and Reference Materials |
| ISO 62304 | 2006 | Medical Device Software - Software Life Cycle Processes |
| ISO 15223-1 | 2012 | Medical Devices - Symbols to be Used with Medical Devices Labels, Labeling, and Information to be<br>Supplied - Part 1: General Requirements |
| ISO 14971 | 2007 | Medical Devices - Applications of Risk Management to Medical Devices |
| NEMA UD 2 | 2009 | Acoustic Output Measurement Standard for Diagnostic Ultrasound Equipment - Revision 3 |
## Clinical Testing
The Clarius Ultrasound Scanner did not require clinical testing to establish substantial equivalence to the predicate devices indicated.
### Quality Assurance Measures
Quality assurance measures applied to the system design and development include, but were not limited to: risk analysis, verification and validation, product specifications, and design reviews.
{18}------------------------------------------------
Image /page/18/Picture/0 description: The image shows the logo for Clarius. The logo consists of an orange symbol that looks like three curved lines forming a C shape. To the right of the symbol is the word "clarius" in a dark blue sans-serif font.
## Conclusion
This device is a modification of an existing licensed device using technologies that exist on the market today. The development and testing conducted on the device ascertain that it is safe for use by physicians. The Clarius Ultrasound Scanner does not introduce indications for use, technological features, or system characteristics that are not seen in its predicate devices; the device is substantially equivalent in safety and effectiveness to these predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.