Voyant Open Fusion Device

K180699 · Applied Medical Resources Corporation · GEI · Apr 13, 2018 · General, Plastic Surgery

Device Facts

Record IDK180699
Device NameVoyant Open Fusion Device
ApplicantApplied Medical Resources Corporation
Product CodeGEI · General, Plastic Surgery
Decision DateApr 13, 2018
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 878.4400
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Voyant Open Fusion Device is a bipolar, electrosurgical device indicated for use with the Voyant electrosurgical generator in open procedures where the ligation and division of vessels and tissue bundles is desired. The device can seal and divide vessels up to and including 7mm in diameter and tissue bundles that can be captured in the jaws of the device. The device has not been shown to be effective for tubal coagulation for sterilization procedures, and should not be used for these procedures.

Device Story

Bipolar electrosurgical instrument; uses RF energy from Voyant Electrosurgical Generator; seals and divides vessels up to 7mm and tissue bundles. Used in open surgery by physicians. Features pistol-grip handle, trigger for jaw closure, energy activation button, and manual blade for tissue division. Device software manages energy delivery. Benefits include efficient hemostasis and tissue division. Bench and preclinical testing confirm performance equivalent to predicate.

Clinical Evidence

No human clinical data. Evidence includes bench testing (burst pressure, durability) and preclinical testing (thermal spread, chronic survival study). Chronic study evaluated long-term seal quality, hemostasis, and adjacent structure safety in animal models. All testing met predetermined acceptance criteria.

Technological Characteristics

Bipolar electrosurgical sealer-divider; RF energy source; pistol-grip form factor; manual blade for tissue division. Complies with IEC 60601-1 (electrical safety), IEC 60601-2-2 (electrosurgical safety), and IEC 60601-1-2 (EMC). Software-controlled energy delivery.

Indications for Use

Indicated for patients undergoing open surgical procedures requiring ligation and division of vessels up to 7mm in diameter and tissue bundles. Contraindicated for tubal coagulation or sterilization procedures.

Regulatory Classification

Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. April 13, 2018 Applied Medical Resources Corporation Patricia Villarreal, Regulatory Affairs Analyst 22872 Avenida Empresa Rancho Santa Margarita, California 92688 Re: K180699 Trade/Device Name: Voyant Open Fusion Device Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI Dated: March 15, 2018 Received: March 16, 2018 Dear Patricia Villarreal: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing, practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Jennifer R. Stevenson -S For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K180699 Device Name Voyant Open Fusion Device Indications for Use (Describe) The Voyant Open Fusion Device is a bipolar, electrosurgical device indicated for use with the Voyant electrosurgical generator in open procedures where the ligation and division of vessels and tissue bundles is desired. The device can seal and divide vessels up to and including 7mm in diameter and tissue bundles that can be captured in the jaws of the device. The device has not been shown to be effective for tubal coagulation for sterilization procedures, and should not be used for these procedures. | Type of Use (Select one or both, as applicable) | |-------------------------------------------------| |-------------------------------------------------| > Prescription Use (Part 21 CFR 801 Subpart D) _ Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary | 510(k) Submitter: | Applied Medical Resources Corporation<br>22872 Avenida Empresa<br>Rancho Santa Margarita, CA 92688<br>(949) 713 - 8000 | |------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Contact Person: | Patricia Villarreal<br>Regulatory Affairs Analyst<br>Applied Medical Resources Corporation<br>Tel: (949) 713-6235<br>Fax: (949) 713-8200<br>Email: patricia.villarreal@appliedmedical.com | | Preparation Date: | March 15, 2018 | | Trade Name: | Voyant® Open Fusion Device | | Common Name: | Bipolar Electrosurgical Sealer-Divider | | Classification: | General and Plastic Surgery - Electrosurgical Cutting and Coagulation Device<br>and Accessories<br>Regulation: 21 CFR 878.4400<br>Device Class: Class II<br>Product Code: GEI | | Predicate Device: | Voyant® Open Fusion Device<br>510(k)#: K162676<br>Product Code: GEI | | Device<br>Description: | The Voyant® Open Fusion device is a bipolar electrosurgical instrument that<br>uses RF energy, provided by the Voyant Electrosurgical Generator (K141288),<br>in open procedures where the ligation and division of vessels up to and including<br>7 mm in diameter and tissue bundles are desired. | | Intended Use: | The Voyant Open Fusion Device is a bipolar, electrosurgical device indicated for<br>use with the Voyant electrosurgical generator in open procedures where the<br>ligation and division of vessels and tissue bundles is desired. | | | The device can seal and divide vessels up to and including 7mm in diameter and<br>tissue bundles that can be captured in the jaws of the device. | | | The device has not been shown to be effective for tubal sterilization or tubal<br>coagulation for sterilization procedures, and should not be used for these<br>procedures. | {4}------------------------------------------------ ## Summary of Technological Characteristics between Subject and Predicate Devices: The subject device design is the same as the predicate, with the exception of the change in software of the device. The subject device has a similar design as the predicate. The fundamental technological features, and intended use of the subject device are the same as the predicate. The subject and predicate instruments are designed to deliver RF energy to vessels and tissue captured between the jaws of the device for tissue fusion. Both instruments feature pistol-grip style handles with a trigger for jaw closure and a button on the back of the handle for energy activation. The instruments are equipped with a mechanical, user-actuated blade for the division of sealed tissue. ## Discussion of Performance Testing: The 2016 FDA Guidance, Premarket Notification (510(k)) Submissions for Bipolar Electrosurgical Vessel Sealers for General Surgery, was considered in evaluating the subject device's electrical, mechanical and functional capabilities. The tests addressed below were utilized to demonstrate safety and efficacy of the subject device and substantial equivalence to the predicate device. ## EMC, Electrical Safety, and Mechanical Testing The Voyant Open Fusion Device complies with IEC 60601-1;ed.3.1;2012 and IEC 60601-2-2;ed.6;2017 for electrical safety testing and IEC 60601-1-2;ed.3;2007 for electromagnetic compatibility testing. ## Bench Burst pressure testing was conducted on the subject and predicate Voyant Open Fusion Devices. Vessels representative of the devices' indications were sealed and the burst pressure for each vessel was recorded. The results of the study demonstrated that the subject device met the predetermined acceptance criteria. Durability testing was performed to verify the continued performance of the subject device over multiple device activations. The result of the study demonstrated that the subject device met the predetermined acceptance criteria. ## Preclinical Thermal spread testing was performed to evaluate the thermal spread damage produced by the subject and predicate Voyant Open Fusion Devices. Vessels representative of the devices' indications were sealed and measurements of the fused areas were taken. Analysis of the measurements demonstrated that the subject device met the predetermined acceptance criteria. A chronic survival study was performed to evaluate long-term seal quality but also to evaluate device performance and the potential for an adverse effect on adjacent structures. Vessels representative of the devices' indications were sealed and evaluated for hemostasis and signs of hematoma. The result of the study demonstrated that the subject device met the predetermined acceptance criteria. ## Software Verification Unit, integration, and system level verification were conducted to evaluate the implementation and performance of the device software script. {5}------------------------------------------------ ## Conclusion: The subject Voyant Open Fusion Device is substantially equivalent in performance to the predicate Voyant Open Fusion Device with respect to intended use (i.e. vessel sealing performance and local tissue effects) and does not raise any new issues of safety and efficacy.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...