← Product Code LFD · K180680

# Salivea Dry Mouth Mouthwash, Salivea Dry Mouth Mouthspray (K180680)

_Laclede, Inc. · LFD · Oct 11, 2018 · DE · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K180680

## Device Facts

- **Applicant:** Laclede, Inc.
- **Product Code:** LFD
- **Decision Date:** Oct 11, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Device Class:** Class U
- **Review Panel:** DE
- **Attributes:** Therapeutic

## Intended Use

Relieve symptoms of dry mouth, refresh, moisturize, clean and sooth oral irritation, and oral dryness.

## Device Story

Salivea Dry Mouth Mouthwash and Mouthspray are saliva substitutes formulated with moisturizers, humectants, and salivary enzymes. The device functions by lubricating, moisturizing, and soothing oral tissues to treat xerostomia symptoms. The mouthwash is supplied in a polyethylene terephthalate (PET) bottle; the mouthspray is supplied in a non-pressurized pump-action PET bottle. Used in the oral cavity, the product is applied by the patient as needed. The device provides symptomatic relief of oral dryness and irritation, benefiting patients by mimicking natural saliva functions.

## Clinical Evidence

No clinical data. Evidence consists of bench testing including biocompatibility (Mucosal Irritation per ANSI/ADA/ISO 10993-10, Sensitization per M&K/GPMT/ISO 10993-10, and Oral toxicity per ANSI/ADA/ISO 10993-5) and physical property testing (pH, viscosity, specific gravity) compared against predicate devices. Real-time stability testing confirmed physical properties met acceptance criteria.

## Technological Characteristics

Saliva substitute containing moisturizers, humectants, and salivary enzymes. Packaging: Polyethylene terephthalate (PET) bottles. Non-sterile. Biocompatibility testing conducted per ISO 10993 standards. Physical properties include pH, viscosity, and specific gravity.

## Predicate Devices

- BIOTENE MOISTURIZING MOUTHSPRAY, GSKCH ([K103745](/device/K103745.md))
- BIOTENE DRY MOUTH MOUTHWASH, GSKCH ([K101477](/device/K101477.md))

## Reference Devices

- Oral 7 Moisturizing Spray and Mouthwash, J.C.E.C Company Inc. ([K142549](/device/K142549.md))
- Oral Balance Gel, Laclede, Inc. ([K061331](/device/K061331.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

October 11, 2018

Laclede, Inc. Rajvinder Atwal QA Manager 2103 E. University Dr. Rancho Dominguez, California 90220

Re: K180680

Trade/Device Name: Salivea Dry Mouth Mouthwash, Salivea Dry Mouth Mouthspray Regulatory Class: Unclassified Product Code: LFD Dated: September 5, 2018 Received: September 6, 2018

Dear Rajvinder Atwal:

We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you. however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see

{1}------------------------------------------------

https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Mary S. Runner -S3

forTina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known)

Device Name

Salivea Dry Mouth Mouthwash and Salivea Dry Mouth Mouth Spray.

#### Indications for Use (Describe)

Relieve symptoms of dry mouth, refresh, moisturize, clean and sooth oral irritation, and oral dryness.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510 (k) Summary

| Submission By: | Laclede, Inc.                                          |
|----------------|--------------------------------------------------------|
| Address:       | 2103 E. University Drive<br>Rancho Dominguez, Ca 90220 |
| Telephone:     | 310-605-4280                                           |
| Facsimile:     | 310-605-4288                                           |
| Contact Name:  | Michael A. Pellico, President                          |
| Email:         | mpellico@laclede.com                                   |

Date 510(k) Summary Prepared: September 13, 2018

| Trade Name/Device Name: | SALIVEA DRY MOUTH MOUTHWASH<br>SALIVEA DRY MOUTH MOUTHSPRAY |
|-------------------------|-------------------------------------------------------------|
| Regulation Number:      | Unclassified                                                |
| Regulation Name:        | Saliva, Artificial                                          |
| Product Code:           | LFD                                                         |

#### Primary Predicate Device:

BIOTENE MOISTURIZING MOUTHSPRAY, GSKCH, K103745 BIOTENE DRY MOUTH MOUTHWASH, GSKCH, K101477

#### Reference Predicate Device:

Oral 7 Moisturizing Spray and Mouthwash, J.C.E.C Company Inc., K.142549 Oral Balance Gel, Laclede, Inc., K061331

#### Device Description:

Salivea Dry Mouth Mouthwash and Salivea Dry Mouth Mouthspray are specially formulated saliva substitute which contain moisturizers, humectants and salivary enzymes that collectively have lubricating, moisturizing, soothing and refreshing properties to relieve and treat the symptoms of dry mouth.

The Salivea Dry Mouth MouthSpray is supplied in non-pressurized pump action spray white Polyethylene terephthalate bottle.

The Salivea Dry Mouth Mouthwash is supplied in white Polyethylene terephthalate bottle.

## Indications for Use:

Relieve symptoms of dry mouth, refresh, moisturize, clean and sooth oral irritation, and oral dryness.

### Substantial Equivalence:

Salivea products have the same intended use, technical characteristics and physiological purpose. Any minor variations in formula/composition are to allow proper dispensing and use of product and do not affect the function, indications, or equivalency of proposed product.

{4}------------------------------------------------

| Product                  | Salivea Dry<br>Mouthwash                                                                                                             | Biotene Dry<br>Mouthwash<br>(K101477)                                                                                                            | Salivea Dry<br>MouthSpray                                                                                                                                 | Biotene<br>Moisturizing<br>Mouthspray<br>(K103745)                                                                                               |
|--------------------------|--------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|
| Method of Use            | Ready to use<br>liquid                                                                                                               | Ready to use<br>liquid                                                                                                                           | Ready to use<br>spray                                                                                                                                     | Ready to use<br>spray                                                                                                                            |
| Applications<br>per Day  | As needed                                                                                                                            | As needed                                                                                                                                        | As needed                                                                                                                                                 | As needed                                                                                                                                        |
| Disease State            | Xerostomia                                                                                                                           | Xerostomia                                                                                                                                       | Xerostomia                                                                                                                                                | Xerostomia                                                                                                                                       |
| Area of Use              | Oral Cavity                                                                                                                          | Oral Cavity                                                                                                                                      | Oral Cavity                                                                                                                                               | Oral Cavity                                                                                                                                      |
| Type<br>of<br>Product    | Liquid                                                                                                                               | Liquid                                                                                                                                           | Liquid                                                                                                                                                    | Liquid                                                                                                                                           |
| Indications<br>of<br>Use | Supports<br>Saliva's Natural<br>Defense,<br>Refresh Breath,<br>moisturizes,<br>gently cleans<br>and soothes Dry<br>Oral Irritations. | Relieves the<br>symptoms of<br>dry mouth,<br>refresh,<br>moisturize,<br>clean, soothe<br>oral<br>irritation<br>and<br>lubricate<br>oral dryness. | Relieve the<br>symptoms of<br>Dry Mouth,<br>instantly<br>moisturizes<br>&<br>hydrates dry<br>tissue, refreshes<br>breath<br>and<br>soothes<br>irritation. | Relieves the<br>symptoms of<br>dry mouth,<br>refresh,<br>moisturize,<br>clean, soothe<br>oral<br>irritation<br>and<br>lubricate<br>oral dryness. |
| Presentation             | Non-Sterile,<br>PET bottle                                                                                                           | Non-Sterile,<br>PET bottle                                                                                                                       | Non-Sterile,<br>PET bottle                                                                                                                                | Non-Sterile,<br>PET bottle                                                                                                                       |

## Discussion and conclusions from the Nonclinical Testing:

Biocompatibility was established through testing generally following ISO 10993, and includes Mucosal Irritation, Sensitization, and Oral toxicity testing. The results met the acceptance criteria of testing.

- · Mucosal irritation (ANSI/ADA/ISO993-10)
- · Sensitization (M&K, GPMT / ISO993-10)
- · Oral toxicity (ANSI/ADA/ISO10993-5)

Physical testing includes pH, Viscosity and Specific gravity. Comparative physical properties testing with predicate device was performed and results were comparable.

Real time stability testing on physical properties i.e., Appearance, pH, viscosity and specific gravity were examined and met acceptance criteria.

#### Conclusion:

The subject Salivea Products has same materials (i.e., moisturizers, humectants and salivary enzymes) as the primary predicate devices. All ingredients in the Salivea products are commonly used for their intended functions. Salivea products have the same intended use and the same technological properties as primary and reference predicate devices. The performance data demonstrate that the subject device is substantially equivalent to the predicate device.

---

**Source:** [https://fda.innolitics.com/device/K180680](https://fda.innolitics.com/device/K180680)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
