SmartGrid feature is a software option that provides, upon request, a diagnostic radiograph image with a reduction in visible x-ray scatter similar to the effect of an anti-scatter grid.
Device Story
SmartGrid II is a software feature for Carestream digital radiography systems; it processes radiographic images acquired without a physical anti-scatter grid. The software uses an algorithm to estimate and remove scatter, improving image contrast and reducing scatter appearance. Users select the processing via the DirectView application either before or after image acquisition. The software operates on the system's workstation; it does not require external cloud processing. Radiologists view the processed images to perform clinical diagnoses. By simulating the effect of a physical anti-scatter grid, the device allows for diagnostic-quality imaging without the physical hardware, potentially simplifying workflows and reducing patient dose associated with grid usage.
Clinical Evidence
Bench testing only. Verification and validation included installation testing, image quality assessment (noise reduction, contrast/brightness), regression testing, integration testing, and error handling. A clinical evaluation was performed using existing research images processed by both the predicate and SmartGrid II; a board-certified radiologist confirmed that images processed with SmartGrid II were equivalent in diagnostic quality to those processed with the predicate device.
Technological Characteristics
Optional software component for digital radiography systems. Functions via scatter factor estimation, scatter correction, noise reduction, and image rendering. Operates on Carestream DirectView application software. No physical materials or energy sources; purely algorithmic image processing.
Indications for Use
Indicated for use in diagnostic radiography to reduce visible x-ray scatter in images acquired without a physical anti-scatter grid.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which consists of a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG" in blue, and below that, the word "ADMINISTRATION" in a smaller font, also in blue.
Carestream Health, Inc. % Ms. Diane Koetter Sr. Manager, Regulatory Affairs and Quality Systems 150 Verona Street ROCHESTER NY 14608
April 13, 2018
Re: K180667
Trade/Device Name: SmartGrid Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: MOB Dated: March 12, 2018 Received: March 14, 2018
Dear Ms. Koetter:
We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
(1-800-638-2041 or 301-796-7100).
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone
Sincerely,
Michael D'Hara For
Robert Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K180667
Device Name SmartGrid
Indications for Use (Describe)
"SmartGrid feature is a software option that provides, upon request by user, a diagnostic radiograph image with a reduction in visible x-ray scatter similar to the effect of an anti-scatter grid."
Type of Use (Select one or both, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Carestream
# "510(k) Summary"
| 510(k) Owner Name: | Carestream Health, Inc. |
|------------------------|-----------------------------------------------------|
| 510(k) Owner Address: | 150 Verona Street |
| | Rochester, New York 14608 |
| 510(k) Owner Phone: | 585 627-6505 |
| 510(k) Owner Fax: | 585 627-8802 |
| Contact Person & Info: | Diane Koetter |
| | Sr. Manager, Regulatory Affairs and Quality Systems |
| | diane.koetter@carestream.com |
| | 585-627-6505 |
| Date Summary Prepared: | April 11, 2018 |
| Device Trade Name: | SmartGrid |
| Device Common Name: | Flat Panel Digital Detector System |
| Classification Name: | Stationary x-ray system |
| Device Class: | II |
| Device Code: | MQB |
| Regulation Number: | 21 CFR 892.1680 |
| Predicate Device: | SmartGrid |
| | Manufactured by: Carestream Health, Inc. |
| | 510(k) No.: K163157 (March 21, 2017) |
Please Note: The marketing name for the modified SmartGrid device is not expected to change. The modified device is referred to as "SmartGrid II" within this submission for clarity.
#### Device Description:
The SmartGrid II software is designed to improve contrast and reduce the appearance of scatter in radiographic images that have been acquired without a physical grid. SmartGrid II encapsulates an algorithm for estimating and removing scatter from radiographic images.
The SmartGrid II software feature is accessible through the DirectView application software. Users will be able to select SmartGrid II processing before an image is acquired, or to change whether SmartGrid II processing is applied to a previously acquired image.
{4}------------------------------------------------
#### Indications for Use / Intended Use:
The Indications for Use for the device, as described in its labeling, are:
"SmartGrid feature is a software option that provides, upon request, a diagnostic radiograph image with a reduction in visible x-ray scatter similar to the effect of an anti-scatter grid."
The intended use for this device, as determined by descriptions and the proposed labeling contained in this submission, is similar to the Indications for Use statement provided above.
The Indications for Use for the subject device is the same as for the predicate device and the intended use remains unchanged. Any variation in features or technical specifications have been identified and addressed through testing (described below) to support a substantial equivalence determination.
#### Comparison of Technological Characteristics:
Based upon information provided within this submission, we believe that the SmartGrid II software is substantially equivalent to the legally marketed SmartGrid (predicate device).
SmartGrid II has the same technological characteristics as SmartGrid (predicate device). Both are optional software components that allow for reduction in visible x-ray scatter in radiographic images.
Both devices use:
- Scatter Factor Estimation ●
- Scatter Correction
- Noise Reduction and Image Rendering .
The predicate device differs from the investigational device in the methods used for Scatter Estimation and Scatter Correction. For instance, the original SmartGrid feature estimated and corrected scatter based on the exposed area within the collimation field. SmartGrid II estimates and corrects scatter based on the full detector field rather than the collimated subregion. In addition, the SmartGrid and SmartGrid II algorithms differ slightly to further improve the scatter estimation and correction.
#### Discussion of Testing:
The performance characteristics and operation / usability of the SmartGrid II software were evaluated in non-clinical (bench) testing. These studies demonstrated the intended workflow, related performance, overall function, verification and validation of requirements for intended use, and reliability of the software. Non-clinical test results have demonstrated that the device conforms to its specifications. Predefined acceptance criteria were met and demonstrated that the device is as safe, as effective, and performs as well as or better than the predicate device.
{5}------------------------------------------------
Specifically, the following "bench" testing was performed as described below:
- Installation testing was performed to verify successful installation of the SmartGrid ● II software component onto the representative system.
- Specific test cases were implemented to verify that SmartGrid II image processing . consistently produces a diagnostic quality image
- . Specific test cases were implemented to verify that image noise was sufficiently reduced and image contrast/brightness were sufficient for clinical diagnosis
- . Regression testing was performed using manual and automated test case execution
- Integration tests were performed to ensure proper performance of the SmartGrid II ● software feature when integrated into the Carestream application software
- Error handling was observed for proper function through monitoring of event logs
- Ad Hoc testing was performed as considered applicable to help ensure no . unexpected negative effects
In addition to bench testing, existing clinical research images processed with both the original SmartGrid and SmartGrid II software were evaluated by a board certified radiologist. This clinical evaluation was intended to confirm no unexpected negative effects with respect to diagnostic image quality associated with the use of the improved algorithm. The reader study demonstrated that images processed with SmartGrid II were equivalent in diagnostic quality to images processed with the predicate device, SmartGrid.
#### Summary:
The differences between SmartGrid II and the predicate device do not affect the intended use of the device or alter the fundamental scientific technology of the device. Performance testing results support a substantial equivalence determination of SmartGrid II to the predicate device.
### FDA Reference Materials:
The following FDA guidance documents were consulted in the development of the SmartGrid II software feature:
- . General Principles of Software Validation – Final Guidance for Industry and FDA Staff; issued 01/11/2002
- Off-The-Shelf Software Use in Medical Devices – Guidance for Industry and FDA Reviewers, and Compliance; issued 09/09/1999
- Content of Premarket Submissions for Management of Cybersecurity in Medical ● Devices - Guidance for Industry and Food and Drug Administration Staff; issued 10/02/2014
- . Deciding When to submit a 510(k) for a Software Change to an Existing Device -Guidance for Industry and Food and Drug Administration Staff; issued 10/25/2017
- Guidance for the Content of Premarket Submissions for Software Contained in . Medical Devices: issued 05/11/2005
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.