Halyard Lavender Nitrile Powder-Free Exam Glove Tested for Use with Chemotherapy Drugs
K180646 · Halyard Health · LZC · Jun 27, 2018 · General Hospital
Device Facts
Record ID
K180646
Device Name
Halyard Lavender Nitrile Powder-Free Exam Glove Tested for Use with Chemotherapy Drugs
Applicant
Halyard Health
Product Code
LZC · General Hospital
Decision Date
Jun 27, 2018
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 880.6250
Device Class
Class 1
Indications for Use
The Halyard Lavender Nitrile Powder-Free Exam Glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. These gloves were tested for use with the following chemotherapy drugs: Cyclophosphamide (20.0 mg/ml) No breakthrough up to 240 minutes Doxorubicin HCl (2.0 mg/ml) No breakthrough up to 240 minutes Etoposide (20.0 mg/ml) No breakthrough up to 240 minutes 5-Fluorouracil (50.0 mg/ml) No breakthrough up to 240 minutes Paclitaxel (Taxol) (6.0 mg/ml) No breakthrough up to 240 minutes Cisplatin (1.0 mg/ml) No breakthrough up to 240 minutes Dacarbazine (10.0 mg/ml) No breakthrough up to 240 minutes Ifosfamide (50.0 mg/ml) No breakthrough up to 240 minutes Mitoxantrone (2.0 mg/ml) No breakthrough up to 240 minutes Vincristine sulfate (1.0 mg/ml) No breakthrough up to 240 minutes Carmustine (3.3 mg/ml) No breakthrough up to 0.3 minutes ThioTEPA (10.0 mg.ml) No breakthrough up to 30.9 minutes Warning: Not for Use With : Carmustine, ThioTEPA
Device Story
Disposable, lavender-colored, chlorinated, nitrile, powder-free, textured fingertip, ambidextrous, non-sterile patient examination glove; packed in cardboard dispenser box. Used in clinical settings by healthcare professionals to prevent cross-contamination between patient and examiner. Provides barrier protection; tested for permeation resistance against specific chemotherapy drugs per ASTM D6978-05. Output is physical barrier protection; aids in maintaining sterile/clean field during patient examination and chemotherapy handling.
Clinical Evidence
Bench testing only. No clinical data. Performance testing included ASTM D5151-06 (leakage, 1.0% AQL), ASTM D6124-06 (residual powder 0.04 mg/glove), ASTM D6319-10 (tensile strength/elongation), and ASTM D6978-05 (chemotherapy permeation). Biocompatibility testing per ISO 10993-10 (irritation/sensitization) and ISO 10993-11 (systemic toxicity) showed no adverse responses.
Technological Characteristics
Material: Nitrile. Form: 9.5-inch, powder-free, textured fingertip, ambidextrous, non-sterile. Standards: ASTM D5151-06, ASTM D6319-10, ASTM D6124-06, ASTM D6978-05, ISO 10993-10, ISO 10993-11. No software or electronic components.
Indications for Use
Indicated for use as a disposable medical glove worn on the examiner's hand to prevent contamination between patient and examiner. Tested for resistance to specific chemotherapy drugs. Not for use with Carmustine or ThioTEPA.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
Predicate Devices
Blue Non-Sterile Powder Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs (K172525)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, featuring a stylized human figure. To the right of that is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Halyard Health Christine Macauley Regulatory Affairs Technical Leader 5405 Windward Parkway Alpharetta. Georgia 30004
Re: K180646
Trade/Device Name: Halyard Lavender Nitrile Powder-Free Exam Glove Tested for Use with Chemotherapy Drugs Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: Class I Product Code: LZC. LZA Dated: Jun 11, 2018 Received: June 12, 2018
Dear Christine Macauley:
We have reviewed your Section 510/k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, Isting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820);
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Clarence W. Murray lii III -S
For Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K180646
#### Device Name
Halyard Lavender Nitrile Powder-Free Exam Gloves Tested for Use with Chemotherapy Drugs
Indications for Use (Describe)
The Halyard Lavender Nitrile Powder-Free Exam Glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. These gloves were tested for use with the following chemotherapy drugs:
- Cyclophosphamide (20.0 mg/ml) No breakthrough up to 240 minutes Doxorubicin HCl (2.0 mg/ml) No breakthrough up to 240 minutes Etoposide (20.0 mg/ml) No breakthrough up to 240 minutes 5-Fluorouracil (50.0 mg/ml) No breakthrough up to 240 minutes Paclitaxel (Taxol) (6.0 mg/ml) No breakthrough up to 240 minutes Cisplatin (1.0 mg/ml) No breakthrough up to 240 minutes Dacarbazine (10.0 mg/ml) No breakthrough up to 240 minutes Ifosfamide (50.0 mg/ml) No breakthrough up to 240 minutes Mitoxantrone (2.0 mg/ml) No breakthrough up to 240 minutes Vincristine sulfate (1.0 mg/ml) No breakthrough up to 240 minutes Carmustine (3.3 mg/ml) No breakthrough up to 0.3 minutes ThioTEPA (10.0 mg.ml) No breakthrough up to 30.9 minutes Warning: Not for Use With : Carmustine, ThioTEPA
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------|
| <label><input type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> | <label><input checked="" type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary-K180646
| Date Summary<br>was Prepared | June 26, 2018 |
|------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Submitter | Christine L. Macauley<br>Regulatory Affairs Technical Leader<br>Halyard Health, Inc.<br>5405 Windward Parkway<br>Alpharetta, GA 30004<br>Email: Chris.Macauley@hyh.com<br>Ph: 470-448-5158 |
| Primary Contact for<br>this 510(k) Submission | Same as above |
| Device Trade Name | Halyard® Lavender Nitrile Powder-Free Exam Glove Tested for Use with<br>Chemotherapy Drugs |
| Device Common Name | Medical Exam Gloves |
| Device Product Code<br>and Classification Name | LZC<br>Class I, 21 CFR §880.6250<br>Patient Examination Glove |
| Predicate Device | K172525<br>Blue Non-Sterile Powder Free Nitrile Examination Gloves Tested for Use<br>with Chemotherapy Drugs |
| Reference Device | K081260<br>Kimberly-Clark Lavender Nitrile Powder-Free Exam Gloves |
| Subject Device Description | Halyard® Lavender Nitrile Powder-Free Exam Gloves Tested for Use with<br>Chemotherapy Drugs are disposable, lavender-colored, chlorinated, nitrile,<br>powder-free, textured fingertip, ambidextrous, non-sterile patient<br>examination gloves that are packed in a cardboard dispenser box. Device<br>follows consensus standards:<br>• ASTM D5151-06 Standard Test Method for Detection of Holes in<br>Medical Gloves<br>• ASTM D6319-10 Standard Specification for Nitrile Examination<br>Gloves for Medical Applications<br>• ASTM D6124-06 Standard Test Method for Residual Powder on<br>Medical Gloves<br>• ASTM D6978-05 Standard Practice for Assessment of Resistance of<br>Medical Gloves to Permeation by Chemotherapy Drugs<br>• ISO 10993-11:2006, Biological evaluation of medical devices - Part<br>11: Tests for Systemic Toxicity<br>• ISO 10993-10: 2010: Biological evaluation of medical devices - Part<br>10: Tests for Irritation and Skin Sensitization |
{4}------------------------------------------------
| Indications for Use | The Halyard® Lavender Nitrile Powder-Free Exam Glove Tested for Use with<br>Chemotherapy Drugs is a disposable device intended for medical purposes that<br>is worn on the examiner's hand to prevent contamination between patient and<br>examiner. These gloves were tested for use with the following chemotherapy<br>drugs:<br>• Cyclophosphamide (20.0 mg/ml) No breakthrough up to 240 minutes<br>• Doxorubicin HCI (2.0 mg/ml) No breakthrough up to 240 minutes<br>• Etoposide (20.0 mg/ml) No breakthrough up to 240 minutes<br>• 5-Fluorouracil (50.0 mg/ml) No breakthrough up to 240 minutes<br>• Paclitaxel (Taxol) (6.0 mg/ml) No breakthrough up to 240 minutes<br>• Cisplatin (1.0 mg/ml) No breakthrough up to 240 minutes<br>• Dacarbazine (10.0 mg/ml) No breakthrough up to 240 minutes<br>• Ifosfamide (50.0 mg/ml) No breakthrough up to 240 minutes<br>• Mitoxantrone (2.0 mg/ml) No breakthrough up to 240 minutes<br>• Vincristine sulfate (1.0 mg/ml) No breakthrough up to 240 minutes<br>• Carmustine (3.3 mg/ml) No breakthrough up to 0.3 minutes<br>• ThioTEPA (10.0 mg.ml) No breakthrough up to 30.9 minutes<br>Warning: Not for Use With: Carmustine, ThioTEPA | | | | |
|------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|---------------------------------------------------------------------------------------------------------------------------------|---------|--|
| Comparison of Technological<br>Characteristics | Subject K180646<br>Lavender-colored, 9.5 inch, nitrile,<br>powder-free, textured fingertip,<br>ambidextrous, non-sterile patient<br>examination glove. | | Predicate K172525:<br>Blue-colored, 9.5 inch, nitrile, powder-<br>free, ambidextrous, non-sterile patient<br>examination glove. | | |
| Summary of comparison of<br>technological characteristics | There are no different technological characteristics of the subject device<br>compared to the predicate device. They are powder-free non-sterile nitrile exam<br>gloves tested for resistance to permeation by chemotherapydrugs. The<br>predicate device is a different color. This difference does not raise new issues of<br>safety or effectiveness of the subject device. | | | | |
| Comparison of the Subject Device with the Predicate Device | | | | | |
| | Subject Device | | Predicate Device K172525 | Remarks | |
| FDA Product<br>Code | LZC | | LZC | Same | |
| FDA<br>Classification | Class I | | Class I | Same | |
| Common Name | Medical Exam Glove | | Medical Exam Glove | Same | |
Blue Non-Sterile Powder Free
Nitrile Examination Gloves Tested
for Use with ChemotherapyDrugs
Device Trade
Name
Halyard Lavender Nitrile Powder-
Free Exam Gloves Tested for Use
with ChemotherapyDrugs
Similar but different
color
{5}------------------------------------------------
| Indications for<br>Use | The Halyard Lavender Powder-<br>Free Nitrile Exam Glove is a<br>disposable device intended for<br>medical purposes that is worn on<br>the examiner's hand to prevent<br>contamination between patient and<br>examiner. These gloves were<br>tested for use with chemotherapy<br>drugs listed on the label. | Non-Sterile Powder Free Nitrile<br>Examination Gloves Tested for<br>Use with Chemotherapy Drugs is<br>a disposable device intended for<br>medical purposes that is worn on<br>the examiner's hand to prevent<br>contamination between patient<br>and examiner. | Same |
|-----------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Technological<br>Characteristics | Lavender-colored, 9.5 inch, nitrile,<br>powder-free, textured fingertip,<br>ambidextrous, non-sterile patient<br>examination glove. | Technological characteristics and<br>design meet ASTM D6319-10 and<br>follow the FDA's Medical Glove<br>Guidance Manual. | Similar |
| | | Performance Data | |
| Standard | Results<br>Subject Device | Results<br>K172525 | Remarks |
| ASTM D6978-05<br>Standard<br>Practice for<br>Assessment of<br>Resistance of<br>Medical Gloves<br>to Permeation<br>by<br>Chemotherapy<br>Drugs | No signs of breakthrough after 4<br>hours for 10 drugs.<br>Carmustine showed no signs of<br>breakthrough until 0.3 minutes and<br>Thiotepa showed no signs of<br>breakthrough until 30.9 minutes.<br>Result: Meets acceptance criteria.<br><br>• Cyclophosphamide (20.0 mg/ml) No<br>breakthrough up to 240 minutes<br>• Doxorubicin HCl (2.0 mg/ml) No<br>breakthrough up to 240 minutes<br>• Etoposide (20.0 mg/ml) No<br>breakthrough up to 240 minutes<br>• 5-Fluorouracil (50.0 mg/ml) No<br>breakthrough up to 240 minutes<br>• Paclitaxel (Taxol) (6.0 mg/ml) No<br>breakthrough up to 240 minutes<br>• Cisplatin (1.0 mg/ml) No<br>breakthrough up to 240 minutes<br>• Dacarbazine (10.0 mg/ml) No<br>breakthrough up to 240 minutes<br>• Ifosfamide (50.0 mg/ml) No<br>breakthrough up to 240 minutes<br>• Mitoxantrone (2.0 mg/ml) No<br>breakthrough up to 240 minutes<br>• Vincristine sulfate (1.0 mg/ml) No<br>breakthrough up to 240 minutes<br><span style="color:red">• Carmustine (3.3 mg/ml) No<br/>breakthrough up to 0.3 minutes</span><br><span style="color:red">• ThioTEPA (10.0 mg.ml) No<br/>breakthrough up to 30.9 minutes</span> | Results from the Permeation data<br>for the predicate device.<br><br>• 5-Fluorouracil (50.0 mg/ml) No<br>breakthrough up to 240 minutes<br>• Etoposide (20.0 mg/ml) No<br>breakthrough up to 240 minutes<br>• Cyclophosphamide (20.0 mg/ml)<br>No breakthrough up to 240 minutes<br>• Carmustine (3.3 mg/ml) No<br>breakthrough up to 12.4 minutes<br>• ThioTEPA (10.0 mg.ml) No<br>breakthrough up to 24.4 minutes<br>• Paclitaxel (Taxol) (6.0 mg/ml) No<br>breakthrough up to 240 minutes<br>• Doxorubicin HCl (2.0 mg/ml) No<br>breakthrough up to 240 minutes<br>• Dacarbazine (10.0 mg/ml) No<br>breakthrough up to 240 minutes<br>• Cisplatin (1.0 mg/ml) No<br>breakthrough up to 240 minutes<br>• Ifosfamide (50.0 mg/ml) No<br>breakthrough up to 240 minutes<br>• Mitoxantrone (2.0 mg/ml) No<br>breakthrough up to 240 minutes<br>• Vincristine sulfate (1.0 mg/ml) No<br>breakthrough up to 240 minutes | Similar |
{6}------------------------------------------------
| ASTM D5151-06<br>Standard Test<br>Method for<br>Detection of<br>Holes in Medical<br>Gloves | Testing of the subject device<br>shows it meets the 2.5% AQL<br>requirement in the standards for<br>leakage with an actual AQL of<br>1.0%. | Testing results from the<br>Predicate device:<br>In accordance with ASTM D5151-<br>06, following ASTM D6319 AQL<br>2.5/Inspection Level G-I | Similar |
|--------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| ASTM D6124-06<br>Standard Test<br>Method for<br>Residual<br>Powder on<br>Medical Gloves | Residual powder on the subject<br>device is an average of 0.04<br>mg/glove within the powder-free<br>limit of < 2 mg maximum powder<br>per glove and meets the<br>acceptance criteria for powder-<br>free. | Residual Powder on the predicate<br>device is a max of 0.52 mg per<br>glove within the powder-free limit<br>of < 2 mg maximum powder per<br>glove and meets the acceptance<br>criteria for powder-free | Similar |
{7}------------------------------------------------
| ASTM D6319-10<br>Standard<br>Specification for<br>Nitrile<br>Examination<br>Gloves for<br>Medical<br>Applications | The physical dimensions of the<br>subject device is within the limits of<br>the standard and the physical<br>properties of the subject device<br>meet the requirements for tensile<br>strength with an average before<br>aging of 30.56 MPa and after<br>aging of 37.53 MPa and elongation<br>of 593% before aging and 533%<br>after aging. | The physical dimensions of the<br>subject device is within the limits<br>of the standard and the physical<br>properties of the subject device<br>meet the requirements for tensile<br>strength with an average before<br>aging of 15 MPa and after aging<br>of 14 MPa and elongation of<br>500% before aging and 400%<br>after aging. | Similar |
|-------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|
| Biological<br>evaluation of<br>medical devices –<br>Part 10: Test for<br>irritation and skin<br>sensitization | Under the conditions of the study<br>the subject device extracts (polar<br>and non-polar) were not found to<br>cause a sensitizing response or an<br>irritation response in the animal<br>model. | Under the conditions of the study<br>the predicate device extracts<br>(polar and non-polar) were not<br>found to cause a sensitizing<br>response or an irritation response<br>in the animal model. | Same |
| Biological<br>evaluation of<br>medical devices –<br>Part 11: Tests for<br>Systemic Toxicity | Under the condition of the study the<br>subject device extracts (polar and<br>non-polar) were not found to caus e<br>a systemic response in the animal<br>model | Data not available from<br>manufacturer | Different |
| Conclusion: | Performance data support the conclusion that the subject device is as safe, as<br>effective, and performs as well as the legally marketed device that was<br>submitted and cleared under K172525. | | |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.