K180627 · Elitechgroup, Inc. · KTB · Jul 27, 2018 · Physical Medicine
Device Facts
Record ID
K180627
Device Name
Macroduct Advanced Model 3710
Applicant
Elitechgroup, Inc.
Product Code
KTB · Physical Medicine
Decision Date
Jul 27, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5525
Device Class
Class 2
Indications for Use
The Macroduct Advanced Model 3710 Sweat Collection System is intended only for clinical laboratory use by qualified medical personnel for stimulation and collection of sweat from humans for analysis for the diagnosis of cystic fibrosis.
Device Story
Macroduct Advanced Model 3710 is a microprocessor-controlled, battery-powered iontophoresis device used in clinical laboratories. It stimulates sweat production via pilocarpine iontophoresis using Pilogel discs (0.5% pilocarpine nitrate). The device delivers a controlled current (ramp to 1.5 mA, hold 5 min, ramp down) to the patient's skin. A specialized collector captures the stimulated sweat; markings on the collector indicate if sufficient sweat rate is achieved. The collected sample is subsequently analyzed using external devices (Sweat-Chek conductivity analyzer or ChloroChek chloridometer) to diagnose cystic fibrosis. Operated by qualified medical personnel, the device features a touch screen interface providing step-by-step procedural guidance, status indicators, and audio alerts. It improves upon the predicate by using an elliptical collection area to better fit limb surfaces and providing a graphical user interface to standardize the collection process.
Clinical Evidence
Bench testing only. Verification and validation activities were performed to confirm the performance of the modified device, including electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), and compliance with CLSI C34-A3 standards for sweat collection. No clinical data was required to establish substantial equivalence.
Technological Characteristics
Microprocessor-controlled iontophoresis device. Materials: High Impact Polystyrene (STYRON 478) for collector; PC-ABS blend for instrument shell. Power: 3.6V Lithium-ion batteries. Iontophoresis: 0.5% pilocarpine nitrate in agar base. Connectivity: Standalone. Standards: IEC 60601-1, IEC 60601-1-2, ANSI/AAMI ES60601-1, CLSI C34-A3.
Indications for Use
Indicated for clinical laboratory use by qualified medical personnel for stimulation and collection of sweat from humans for analysis for the diagnosis of cystic fibrosis.
Regulatory Classification
Identification
An iontophoresis device is a device that is intended to use a direct current to introduce ions of soluble salts or other drugs into the body and induce sweating for use in the diagnosis of cystic fibrosis or for other uses if the labeling of the drug intended for use with the device bears adequate directions for the device's use with that drug. When used in the diagnosis of cystic fibrosis, the sweat is collected and its composition and weight are determined. Additionally, an iontophoresis device intended for any other purposes is a prescription device that is intended to use a current to introduce ions of drugs or non-drug solutions into the body for medical purposes other than those specified in paragraph (a) of this section, meaning that the device is not intended for use in diagnosis of cystic fibrosis, or a specific drug is not specified in the labeling of the iontophoresis device.
Special Controls
*Classification.* Class II (special controls). The device is classified as class II. The special controls for this device are:(i) The following performance testing must be conducted:
(A) Testing using a drug approved for iontophoretic delivery, or a solution if identified in the labeling, to demonstrate safe use of the device as intended;
(B) Testing of the ability of the device to maintain a safe pH level; and
(C) If used in the ear, testing of the device to demonstrate mechanical safety.
(ii) Labeling must include adequate instructions for use, including sufficient information for the health care provider to determine the device characteristics that affect delivery of the drug or solution and to select appropriate drug or solution dosing information for administration by iontophoresis. This includes the following:
(A) A description and/or graphical representation of the electrical output;
(B) A description of the electrode materials and pH buffer;
(C) When intended for general drug delivery, language referring the user to drug labeling approved for iontophoretic delivery to determine if the drug they intend to deliver is specifically approved for use with that type of device and to obtain relevant dosing information; and
(D) A detailed summary of the device-related and procedure-related complications pertinent to use of the device, and appropriate warnings and contraindications, including the following warning:
*Warning:* Potential systemic adverse effects may result from use of this device. Drugs or solutions delivered with this device have the potential to reach the blood stream and cause systemic effects. Carefully read all labeling of the drug or solution used with this device to understand all potential adverse effects and to ensure appropriate dosing information. If systemic manifestations occur, refer to the drug or solution labeling for appropriate action.(iii) Appropriate analysis/testing must demonstrate electromagnetic compatibility, electrical safety, thermal safety, and mechanical safety.
(iv) Appropriate software verification, validation, and hazard analysis must be performed.
(v) The elements of the device that may contact the patient must be demonstrated to be biocompatible.
(vi) The elements of the device that may contact the patient must be assessed for sterility, for devices labeled as sterile.
(vii) Performance data must support the shelf life of the elements of the device that may be affected by aging by demonstrating continued package integrity and device functionality over the stated shelf life.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
July 27, 2018
ELITechGroup Inc. Bryce McEuen Managing Director and Business Unit Manager 370 W 1700 S Logan, Utah 84321
Re: K180627
Trade/Device Name: Macroduct Advanced Model 3710 Regulation Number: 21 CFR 890.5525 Regulation Name: Iontophoresis Device Regulatory Class: Class II Product Code: KTB Dated: May 3, 2018 Received: May 7, 2018
Dear Bryce McEuen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Michael J. Hoffmann -S
Carlos L. Peña, PhD, MS for Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K180627
Device Name Macroduct Advanced Model 3710
Indications for Use (Describe)
The Macroduct Advanced Model 3710 Sweat Collection System is intended only for clinical laboratory use by qualified medical personnel for stimulation of sweat from humans for analysis for the diagnosis of cystic fibrosis.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### Section 6.0 - 510(k) Summary
Date: July 27, 2018
| Submitter: | ELITechGroup Inc., dba Wescor Inc. |
|---------------|-------------------------------------------------|
| Address: | 370 W 1700 S, Logan, UT 84321 USA |
| Phone number: | 435-752-6011 |
| Fax number: | 435-752-4127 |
| Contact: | Bryce McEuen (Email: b.mceuen@elitechgroup.com) |
#### Device name:
Trade/proprietary Name: Macroduct® Advanced Model 3710
Common or Usual Name: Macroduct Advanced Sweat Collection System
| Code | Name | Class | Regulation | Regulation Name | Panel |
|------|-----------------------------------------|-------|--------------------|----------------------|----------------------|
| KTB | Device, Iontophoresis,<br>Specific Uses | II | 21 CFR<br>890.5525 | Iontophoresis Device | Physical<br>Medicine |
### Establishment Information:
The establishment registration number for ELITechGroup Inc. USA is 1717966. The owner operator number for ELITechGroup Inc. (Logan, UT, USA) is 1717966.
### Predicate device:
| Predicate Device | 510(k)<br>Number | Product code(s) |
|---------------------------|------------------|-----------------|
| Macroduct® Model 3700-SYS | K853973 | KTB |
{4}------------------------------------------------
### Device description:
The Macroduct Advanced Model 3710 Sweat Collection System is intended only for clinical laboratory use by qualified medical personnel for stimulation and collection of sweat from humans for analysis for the diagnosis of cystic fibrosis.
The Macroduct Advanced Sweat Collection System consists of the Macroduct Advanced Model 3710, which is a microprocessor-controlled device powered from a rechargeable lithium-ion battery, battery charging power supply and cord for charging the battery, electrode cable assembly, and a kit of single-use supplies (Pilogel discs, collectors, and straps). The system safely and efficiently accomplishes the stimulation of human sweat through pilocarpine iontophoresis using the Macroduct Advanced Model 3710. The Macroduct Advanced Sweat Collector collects a sample of the stimulated sweat. Markings on the collection tube of the collector indicate if a sufficient sweat rate is achieved during the collection of sweat. The sample can then be analyzed for indications of cystic fibrosis with the Sweat-Chek™ Sweat Conductivity Analyzer using the principle of total electrolyte concentration in the sweat sample; or with the ChloroChek® Chloridometer® using the principle of coulometric titration.
The Macroduct Advanced cannot be used in combination with other medical devices.
### Substantial Equivalence:
| Similarities | | |
|--------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------|
| | Current Device<br>Macroduct Advanced Model 3710 | Previous Device<br>Macroduct Model 3700-SYS |
| Indications for Use | The Macroduct Advanced Model 3710 Sweat Collection System is intended only for clinical laboratory use by qualified medical personnel for stimulation and collection of sweat from humans for analysis for the diagnosis of cystic fibrosis. | Same |
| Iontophoresis current and time | Ramp (approximately 25 seconds) from 0 mA to 1.5 mA; hold at 1.5 mA for 5 minutes; ramp (approximately 5 seconds) from 1.5 mA to 0 mA. | Same |
| Pilogel Disc composition | 0.5% pilocarpine nitrate in agar base. | Same |
The Macroduct Advanced Model 3710 has demonstrated substantial equivalence to the predicate device, Macroduct Model 3700-SYS.
{5}------------------------------------------------
| Similarities | | |
|----------------------------------|-------------------------------------------------------------------------------|---------------------------------------------|
| | Current Device<br>Macroduct Advanced Model 3710 | Previous Device<br>Macroduct Model 3700-SYS |
| Pilocarpine transferred per test | Up to 1 mg | Same |
| Current Safety limit | Controlled by software/hardware to 1.5 mA, limited by hardware alone to <4 mA | Same |
| Iontophoresis power supply | Microprocessor controlled | Same |
| Sweat Collector material | High Impact Polystyrene (STYRON 478) | Same |
| Differences | | |
|---------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------|
| | Macroduct Advanced Model 3710 | Predicate device<br>Macroduct Model 3700-SYS |
| Sweat Collection area | .994 in² elliptical collection surface | .994 in² circular collection surface |
| Pilogel shape | Elliptical shape to match collector | Circular shape to match collector |
| Instrument Shell | Handheld, rounded rectangular<br>in custom molded clamshell<br>case in PC-ABS blend plastic | Handheld, rectangular, stock<br>clamshell case in ABS plastic. |
| Instrument User interface | Touch screen with graphics<br>showing step-by-step<br>procedure, on-screen touch<br>switch to start and stop iontophoresis, graphic indicator for<br>current/low battery, audio signal<br>for completion/error indication | Toggle switch to start and stop<br>iontophoresis, LED indicators<br>for current/low battery, audio<br>signal for completion/error indication |
| Power Source | Three rechargeable 3.6 V Lithium ion batteries | Two 9V alkaline batteries |
| Modifications | | |
|-----------------------|----------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|
| Characteristic | Modification | Rationale for Modification |
| Sweat Collection area | Changed the shape of the collection area from 0.994 in2 round to 0.994 in2 elliptical. | The collection area shape changed to elliptical to better fit limb skin surface, including small radius arms. |
{6}------------------------------------------------
| Modifications | | |
|----------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Characteristic | Modification | Rationale for Modification |
| Pilogel shape | Changed the Pilogel shape from round to elliptical. | The Pilogel shape was changed to match the elliptical shape of the collector. |
| Instrument Shell | The size, shape and material of the instrument shell changed. | To accommodate the touch screen |
| Instrument User inter-face | Touch screen with graphics showing step-by-step procedure, on-screen touch switch to start and stop iontophoresis, graphic indicator for current/low battery, audio signal for completion/error indication | Provide users with detailed vi-sual instructions and information to further enhance and standardize the process of pilocarpine iontophoresis and sweat collection. |
# Performance Standards:
| Regulation / Standard | Organization | Regulation / Standard Description |
|-----------------------------------------------------------------------------------|-----------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 21 CFR 890.5525 | US Code of Federal<br>Regulations - FDA | Iontophoresis device intended for certain specified uses |
| C34-A3 | CLSI | Sweat Testing: Sample Collection and<br>Quantitative Chloride Analysis |
| IEC 60601-1:2005<br>(Third Edition) + CORR.<br>1:2006 + CORR. 2:2007<br>+ A1:2012 | IEC | Medical Electrical Equipment – Part 1:<br>General requirements for basic safety<br>and essential performance |
| IEC 60601-1-2:2014 | IEC | Medical Electrical Equipment – Part 1-<br>2: General requirements for basic<br>safety and essential performance –<br>Collateral Standard: Electromagnetic<br>disturbances – Requirements and tests |
| ANSI/AAMI ES60601-<br>1:2005 / A2:2010 | ANSI | US National Differences to IEC 60601-<br>1:2005 |
| EN 55022:2010 | European Commis-<br>sion | Information technology equipment –<br>Radio disturbance characteristics –<br>Limits and methods of measurement |
| 47 CFR Part 15, Sub-<br>part B | US Code of Federal<br>Regulations - FCC | Unintentional Radiators |
| ICES-003, Issue 6 | Industry Canada | Information technology equipment –<br>Radio disturbance characteristics –<br>Limits and methods of measurement |
| ISTA 3A (2008) | ISTA | Packaged products for parcel delivery<br>system shipments 70kg (150 lb) or less |
| Directive 2011/65/EU | European Parlia-<br>ment and the | Restriction of the use of certain haz-<br>ardous substances in electrical and |
{7}------------------------------------------------
| | Council of the Eu-<br>ropean Union | electronic equipment |
|--|------------------------------------|----------------------|
| | | |
#### Indications for Use:
The Macroduct Advanced Model 3710 Sweat Collection System is intended only for clinical laboratory use by qualified medical personnel for stimulation and collection of sweat from humans for analysis for the diagnosis of cystic fibrosis.
## Conclusion:
The modifications to the predicate device Macroduct Model 3700-SYS did not affect the use of the device or alter the fundamental scientific technology of the device. The design control process results for the Macroduct Advanced Model 3710 demonstrate that the Macroduct Advanced Model 3710 is substantially equivalent to the predicate device, Macroduct Model 3700-SYS, cleared under K853973.
The data demonstrates that the system is appropriate for its use and does not raise new issues of safety or effectiveness. A detailed summary of the verification, validation and risk analysis of these modifications are provided in Section 11 and complete risk analysis reports are provided in Appendix 1.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.