K180619 · Innomed Healthscience, Inc. · OHP · Aug 16, 2018 · Dental
Device Facts
Record ID
K180619
Device Name
Bongo
Applicant
Innomed Healthscience, Inc.
Product Code
OHP · Dental
Decision Date
Aug 16, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.5570
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Bongo is indicated for use in the treatment of mild to moderate obstructive sleep apnea (OSA) in adults >66 lbs.
Device Story
Bongo is an intranasal appliance for treating mild to moderate obstructive sleep apnea. It consists of valve assemblies and soft nasal inserts placed inside the nostrils. During inhalation, valves open to allow airflow; during exhalation, valves close, forcing air through small vent ports to create expiratory positive airway pressure (EPAP). This EPAP props the airway open until the next inhalation. Used in homes, hospitals, or sleep labs, the device is operated by the patient. It is a multi-use, cleanable device. By maintaining airway patency during sleep, it aims to reduce the apnea-hypopnea index (AHI). Healthcare providers use the device to manage OSA symptoms, with efficacy monitored via polysomnography (PSG).
Clinical Evidence
Prospective, non-randomized, open-label, single-center study (N=10). Primary endpoint was AHI reduction. Mean diagnostic AHI was 15.7, reducing to 7.1 with Bongo. Median diagnostic AHI was 15.4, reducing to 7.0. Results showed significant AHI reduction compared to baseline. No serious adverse events reported.
Technological Characteristics
Intranasal appliance with valve assemblies and soft nasal inserts. Operates via passive mechanical flap valves to create EPAP. Materials tested per ISO 10993-1 for surface/mucosal contact and permanent duration. Multi-use, cleanable design. No electronic components or software.
Indications for Use
Indicated for adults >66 lbs. with mild to moderate obstructive sleep apnea (OSA).
Regulatory Classification
Identification
Intraoral devices for snoring and intraoral devices for snoring and obstructive sleep apnea are devices that are worn during sleep to reduce the incidence of snoring and to treat obstructive sleep apnea. The devices are designed to increase the patency of the airway and to decrease air turbulence and airway obstruction. The classification includes palatal lifting devices, tongue retaining devices, and mandibular repositioning devices.
Special Controls
*Classification.* Class II (special controls). The special control for these devices is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intraoral Devices for Snoring and/or Obstructive Sleep Apnea; Guidance for Industry and FDA.”
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA acronym followed by the full name of the agency, "U.S. Food & Drug Administration," on the right. The FDA part of the logo is in blue.
August 16, 2018
InnoMed Healthscience, Inc. % Paul Dryden Consultant InnoMed Healthscience, Inc. c/o ProMedic, LLC 131 Bay Point Dr. NE Saint Petersburg, FL 33704
Re: K180619 Trade/Device Name: Bongo Regulation Number: 21 CFR 872.5570 Regulation Name: Intraoral Devices for Snoring and Intraoral Devices for Snoring and Obstructive Sleep Apnea Regulatory Class: Class II Product Code: OHP Dated: July 16, 2018 Received: July 17, 2018
Dear Paul Dryden:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device. please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
{2}------------------------------------------------
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely yours.
# Srinivas Nandkumar -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K180619
Device Name Bongo
Indications for Use (Describe)
Bongo is indicated for use in the treatment of mild to moderate obstructive sleep apnea (OSA) in adults >66 lbs.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
| Date Prepared: | August 10, 2018 |
|----------------------------|--------------------------------------------------------------------------------------------------|
| | InnoMed Healthscience, Inc.<br>1701 W. Hillsboro Blvd.<br>Suite 303<br>Deerfield Beach, FL 33442 |
| Official Contact: | Javier Collazo<br>General Manager and Vice President<br>Tel - 954.773.9656 |
| Proprietary or Trade Name: | Bongo |
| Common/Usual Name: | Expiratory resistance valve, intranasal, for obstructive sleep apnea |
| Classification Name: | 21CFR 872.5570<br>ProCode - OHP<br>Class II |
| Predicate Device: | Ventus ProVent (K102404) |
Device Description: The Bongo intranasal appliance is a simple device which is placed just inside the nostrils. The device directs expiratory flow through small vent ports which increases intranasal pressure similar to the expiratory portion of the breathing cycle during use of the predicate device. The Bongo consists of valve assemblies coupled with nasal inserts. The Bongo has soft nasal inserts to form a seal within the nasal openings. The nasal inserts are offered in various sizes.
| Indications for Use: | Bongo is indicated for use in the treatment of mild to moderate obstructive<br>sleep apnea (OSA) in adults >66 lbs. |
|----------------------|---------------------------------------------------------------------------------------------------------------------|
| Patient Population: | For adults (>66 lbs.) with mild to moderate OSA. |
| Environments of Use: | Home use, hospitals, and sleep laboratories |
{5}------------------------------------------------
# K180619
## 510(k) Summary Page 2 of 4
| Attributes | Proposed Bongo | Predicate<br>Ventus ProVent (K102404) | Comments |
|------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------|
| Indications for Use | For use in the treatment of mild to moderate<br>obstructive sleep apnea (OSA). | For use in the treatment of obstructive sleep<br>apnea (OSA). | Similar |
| Patient Population | Adults (>66 lbs. / 30 kg.) | Not specified | Similar |
| Environment of Use | Home<br>Hospitals<br>Sleep Laboratories | Not specified | Similar |
| Duration of Use | Single patient, multi-use<br>Used on a daily basis | Single patient, disposable<br>Used on a daily basis | Bongo can be cleaned<br>and reused |
| Prescriptive | Yes | Yes | Identical |
| Technology / Mode of Operation | | | |
| Creation of expiratory positive<br>airway pressure | During exhalation the valves close, directing<br>air through small ports, increasing resistance.<br>This resistance creates EPAP (Expiratory<br>Positive Airway Pressure), which props open<br>the airway until you inhale again. | During exhalation the valves close, directing<br>air through small ports, increasing resistance.<br>This resistance creates EPAP (Expiratory<br>Positive Airway Pressure), which props open<br>the airway until you inhale again. | Identical |
| Sealing method | Soft nasal inserts | Adhesive seals around the nostril openings | Similar |
| Features | | | |
| Multiple sizes of nasal inserts | Yes | Not required | Similar |
| One expiratory flow / resistance | Yes | Yes | Similar |
| Single Patient | Yes, but may be cleaned and reused | Yes, but is single use disposable | Similar |
| Technical Specifications / Performance Testing | | | |
| Inhalation Resistance | Comparative inhalation resistance testing was<br>performed evaluating the subject device and the<br>predicate device at various flow rates. | Comparative inhalation resistance testing was<br>performed evaluating the subject device and<br>the predicate device at various flow rates. | Similar inhalation<br>resistance |
| Exhalation Flow (ml/sec)<br>@ a given pressure | Comparative exhalation flow testing was<br>performed evaluating the subject device and the<br>predicate device at various pressures. | Comparative exhalation flow testing was<br>performed evaluating the subject device and the<br>predicate device at various pressures. | Similar<br>exhalation flow |
| Classification of Patient Contacting<br>Materials for Biocompatibility | ISO 10993-1<br>Surface Contact / External Communicating<br>Permanent duration of use | ISO 10993-1<br>Surface Contact / External Communicating<br>Permanent duration of use | Similar |
| Attributes | Proposed Bongo | Predicate<br>Ventus ProVent (K102404) | Comments |
| Additional Non-clinical Testing | Age / Shelf-life<br>Cleaning<br>Drop test<br>Environmental conditions<br>Valve cycling | Not specified in 510(k) summary | Subject device meets<br>its performance<br>specifications |
| Clinical Testing | Subjects were evaluated under PSG and the<br>results demonstrate that AHI was reduced with<br>Bongo. | Available clinical data showed a reduction in<br>AHI with the device. | Similar |
{6}------------------------------------------------
#### K180619
#### 510(k) Summary Page 3 of 4
The Bongo intranasal appliance is viewed as substantially equivalent to the predicate device because:
#### Indications -
- The Bongo is intended to be used in the treatment of mild to moderate OSA. ●
- The predicate ProVent (K102404) is intended to be used in the treatment OSA, therefore the indications for use are considered . substantially equivalent.
#### Patient Population -
- The Bongo and the predicate ProVent (K102404) are both intended for adults (>66 Ibs.). ●
- Therefore, the patient population is considered substantially equivalent.
#### Technology and Principle of Operation -
- The Principle of Operation is to create expiratory positive airway pressure (EPAP). ●
- The proposed Bongo and the predicate ProVent (K102404) employ similar technology, i.e., use of flap valves which open during . inhalation and close during exhalation to create the EPAP.
- The principle of operation is considered substantially equivalent. .
- The Bongo and the predicate ProVent (K102404) both have nasal into the nasal into the nasal openings. The Bongo . utilizes soft nasal inserts, which are offered in various sizes, to form a seal within the nasal openings. The predicate ProVent (K102404), which is offered in one size, utilizes adhesive to seal around the nasal openings.
#### Environment of Use -
- The devices are intended for use in the home, hospitals, and sleep laboratories. ●
- Substantially equivalent to predicate ProVent (K102404). ●
{7}------------------------------------------------
#### Non-clinical Testing -
## Biocompatibility of Patient Contacting Materials -
- The materials have been tested per ISO 10993-1 and they have been used in legally marketed . devices which have equivalent patient contact and duration of use.
- . ISO 10993-1 would consider the type of patient contact as Surface Contact / mucosal and Externally Communicating / Tissue and permanent duration of use.
#### Performance Compared to the Predicate
Comparative testing was performed evaluating the Bongo vs. the predicate ProVent (K102404) for the key performance specifications which could demonstrate substantial equivalence. These test included:
- . Inhalation Resistance
- . Expiration Flow rate
A summary of the testing was presented in the above table.
- The results of the comparative performance testing demonstrated that the Bongo is substantially ● equivalent to the predicate ProVent (K102404).
In addition to these comparative tests we also performed testing to demonstrate that the Bongo meets its performance specifications after being subjected to aging, environmental temperature changes, cleaning, and being dropped.
- . The results support that the Bongo meets it performance specifications and is not affected by aging or cleaning.
## Clinical Testing -
A prospective, non-randomized, open label, single-center clinical study was performed.
The clinical study had ten (N=10) patients who completed the study, with seven patients responding with AHI reduction using the Bongo. For the patients (N=10):
- The mean diagnostic AHI was 15.7 and the mean AHI for the final PSG with Bongo was 7.1. ●
- . The median diagnostic AHI was 15.4 and the median AHI for the final PSG with Bongo was 7.0.
The results demonstrated that AHI was significantly reduced with the Bongo compared to the baseline diagnostic. There were no serious adverse events.
#### Discussion:
In comparing available clinical data for the predicate and subject device reduced AHI with the device compared without the device. The reductions were substantially equivalent.
#### Substantial Equivalence Conclusion:
The sponsor has demonstrated through performance testing, design and features, non-clinical, and clinical testing that the proposed device does not raise different risks of safety and effectiveness compared to the predicate and are substantially equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.