Silverlon Island Wound Dressing, Silverlon Wound Pad Dressing (also known as Silverlon Burn Pad Dressing)

K180570 · Argentum Medical, LLC · FRO · Jan 14, 2019 · SU

Device Facts

Record IDK180570
Device NameSilverlon Island Wound Dressing, Silverlon Wound Pad Dressing (also known as Silverlon Burn Pad Dressing)
ApplicantArgentum Medical, LLC
Product CodeFRO · SU
Decision DateJan 14, 2019
DecisionSESE
Submission TypeTraditional
Device ClassClass U
AttributesTherapeutic

Indications for Use

Silverlon® Island Wound Dressing and Silverlon® Wound Pad Dressing are multi-layer, sterile, non-adherent, antimicrobial barrier wound dressings. The Over-The-Counter Indications: Local management of superficial wounds, minor burns, abrasions and lacerations. Prescription Indications: Under the supervision of a healthcare professional Silverlon® Island Wound Dressings and Silverlon® Wound Pad Dressings are intended for up to 7 day use for wounds such as vascular access or peripheral IV sites, orthopedic external pin sites, wound drain sites, surgical wounds (donor and graft sites, incisions), and partial to full thickness dermal ulcers (stage I-IV pressure sores, venous stasis ulcers, arterial ulcers, diabetic ulcers). Silverlon® Island Wound Dressings and Silverlon® Wound Pad Dressings are indicated for the management of infected wounds, as the silver in the dressing provides an antimicrobial barrier that may be helpful in managing these wounds. In addition, the moist wound healing environment and control of wound bacteria within the Silverlor® Island Wound Dressing and Silverlon® Wound Pad Dressing may help reduce the risk of wound infection and support the body's healing process. Silverlon® Island Wound Dressings and Silverlon® Wound Pad Dressings may be used for the management of painful wounds, Silverlon® Island Wound Dressings and Silverlon® Wound Pad Dressing's non-adherent wound contact layer reduces pain during dressing changes and evaporation of moisture in the dressing may soothe the wound.

Device Story

Multi-layer, sterile, non-adherent, antimicrobial wound dressing; consists of knitted nylon fiber coated with metallic silver (1% silver oxide). Moisture activates silver ions to provide antimicrobial barrier; kills bacteria within dressing; absorbs wound exudate. Used in clinical settings or home; applied by healthcare professionals or patients. Provides moist healing environment; reduces pain during dressing changes via non-adherent contact layer. MR Conditional labeling added.

Clinical Evidence

No clinical testing was necessary for this submission. Bench testing included biocompatibility (ISO 10993), antimicrobial effectiveness (Kirby-Bauer, ASTM E2315), and MR compatibility (ASTM F2052, F2213, F2182, F2119).

Technological Characteristics

5-layer dressing: (1) silver-coated knitted nylon; (2) polyethylene film; (3) absorptive laminate pad; (4) apertured high-density polyethylene film; (5) polyester fabric with acrylic adhesive (Island version only). Silver coating: 5.05 mg/cm, 0.8-1.0 µm thick. Biocompatible materials. MR Conditional.

Indications for Use

Indicated for local management of superficial wounds, minor burns, abrasions, lacerations (OTC); and for management of vascular access/IV sites, orthopedic pin sites, drain sites, surgical wounds, and partial to full-thickness dermal ulcers (stage I-IV pressure, venous stasis, arterial, diabetic) (Rx). Indicated for management of infected and painful wounds.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. January 14, 2019 Argentum Medical, LLC Kartik Patel Associate Director, Quality and Regulatory Affairs 2571 Kaneville Court Geneva, IL 60134 Re: K180570 Trade/Device Name: Silverlon Island Wound Dressing, Silverlon Wound Pad Dressing (also known as Silverlon Burn Pad Dressing) Regulatory Class: Unclassified Product Code: FRO Dated: November 29, 2018 Received: December 3, 2018 Dear Kartik Patel: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination. product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see {1}------------------------------------------------ https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # Kimberly Ferlin -S for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use ### 510(k) Number (if known) K180570 #### Device Name Silverlon® Island Wound Dressing, Silverlon® Wound Pad Dressing (also known as Silverlon® Burn Pad Dressing) #### Indications for Use (Describe) Silverlon® Island Wound Dressing and Silverlon® Wound Pad Dressing are multi-layer, sterile, non-adherent, antimicrobial barrier wound dressings. The Over-The-Counter Indications: Local management of superficial wounds, minor burns, abrasions and lacerations. Prescription Indications: Under the supervision of a healthcare professional Silverlon® Island Wound Dressings and Silverlon® Wound Pad Dressings are intended for up to 7 day use for wounds such as vascular access or peripheral IV sites, orthopedic external pin sites, wound drain sites, surgical wounds (donor and graft sites, incisions), and partial to full thickness dermal ulcers (stage I-IV pressure sores, venous stasis ulcers, arterial ulcers, diabetic ulcers). Silverlon® Island Wound Dressings and Silverlon® Wound Pad Dressings are indicated for the management of infected wounds, as the silver in the dressing provides an antimicrobial barrier that may be helpful in managing these wounds. In addition, the moist wound healing environment and control of wound bacteria within the Silverlor® Island Wound Dressing and Silverlon® Wound Pad Dressing may help reduce the risk of wound infection and support the body's healing process. Silverlon® Island Wound Dressings and Silverlon® Wound Pad Dressings may be used for the management of painful wounds, Silverlon® Island Wound Dressings and Silverlon® Wound Pad Dressing's non-adherent wound contact layer reduces pain during dressing changes and evaporation of moisture in the dressing may soothe the wound. | Type of Use (Select one or both, as applicable) | |-------------------------------------------------| | ☑ Prescription Use (Part 21 CFR 801 Subpart D) | | ☑ Over-The-Counter Use (21 CFR 801 Subpart C) | ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ## 510(k) Summary Image /page/3/Picture/2 description: The image features the logo for Argentum Medical. The logo consists of a blue square with a white "A" inside, and the words "Argentum Medical" are written in a light gray font to the right of the square. The logo is simple and modern, and the colors are clean and professional. 2571 Kaneville Court Geneva, IL 60134 PH: (630) 232.2507 TF: (888) 551-0188 510(k) Summary [As required by 21 CFR 807.92] ## 1.0 Submitted by Argentum Medical, LLC 2571 Kaneville Court Geneva, IL 60134 Phone 630.232.2507 Fax 630.232.8005 #### 2.0 Contact Person Kartik Patel Argentum Medical, LLC 2571 Kaneville CT. Geneva, IL 60134 Phone 630.232.2510 Fax 630.232.8005 kpatel@silverlon.com ### 3.0 Date Prepared November 29, 2018 #### 4.0 Device Trade Name Silverlon® Island Wound Dressing, Silverlon® Wound PadDressing (also known as Silverlon® Burn Pad Dressing) ## 5.0 Common Name Dressing, Wound, Drug {4}------------------------------------------------ ## 6.0 Classification Name Unclassified: Pre-Amendment 21 CFR Number: None Product code: FRO ## 7.0 Substantially Equivalent Devices (Predicate Devices): Silverlon Island Wound Dressing, Silverlon Wound Pad Dressing (K143001) ## 7.1. Reference Device: Silverlon Island Wound Dressing (K122817) ## 8.0 Device Description Silverlon Island Wound Dressing is a self-adhesive, elastomeric wound dressing composed of 5 distinct layers (in innermost order): The Silverlon Wound Pad Dressing is comprised of Layers 1 - 4, they are not supplied with Layer 5 (adhesive tape layer). - . Layer 1 is a non-adherent wound contact layer that consists of a single layer of knitted continuous nylon fiber coated with metallic silver at the rate of 5.05 mg/cm (50.5 g/m²); the silver surface coating is approximately 0.8 - 1.0 µm thick and contains approximately 1% silver oxide). - . Layer 2 is a polyethylene film used to bond the nylon substrate to the pad layer described below; - Layer 3 is a laminate pad with an absorptive capacity of 65 oz/yd2 to absorb wound . exudate; - . Layer 4 is an apertured high density polyethylene film that bonds the pad to the outer polyester fabric described below; and, - Layer 5 is a non-woven, medical grade polyester fabric coated on the skin-contacting . side with a self-curing acrylic, pressure-sensitive, medical grade adhesive covered with a silicone-coated (one side) paper liner. The Silverlon Wound Pad dressings do not have a tape layer. All materials are biocompatible per irritation, sensitization and cytotoxicity testing results. Layer 1 delivers antimicrobial silver ions in the dressing when activated by moisture. The silver ions in the dressing kill wound bacteria held in the dressing and provide an antimicrobial barrier to protect the wound bed. This dressing absorbs high amounts of wound fluid and the dressing intimately conforms to the wound surface. {5}------------------------------------------------ K180570 Under the direction of a healthcare professional, Silverlon Island Wound Dressing and Silverlon Wound Pad Dressing may be used for wounds such as vascular access or peripheral IV sites, orthopedic external pin sites, wound drain sites, surgical wounds (donor and graft sites, incisions), and partial to full thickness dermal ulcers (stage I-IV pressure sores, venous stasis ulcers, arterial ulcers, diabetic ulcers) or traumatic wounds left to heal by secondary intent, and management of painful wounds. Silverlon Island Wound Dressings and Silverlon Wound Pad Dressings are indicated for the management of infected wounds, as the silver in the dressing provides an antimicrobial barrier that may be helpful in managing these wounds. In addition, the moist wound healing environment and control of wound bacteria within the Silverlon Island Wound Dressings and Silverlon Wound Pad Dressings are indicated for the management of infected wounds, as the silver in the dressing provides an antimicrobial barrier that may be helpful in managing these wounds. A majority of postoperative surgical incisions are sutured, stapled or glued and are covered with some form of dressing. This first dressing, frequently referred to as the primary dressing, acts to absorb drainage, maintain a clean environment and serve as a barrier against further trauma to the delicate incision surface. Careful selection of nonadherent, absorptive dressings that do not become incorporated within the incision serves greatly to reduce the potential of surgical incision injury, and reduce pain upon dressing change when the time comes for the initial dressing to be removed. Silverlon Island Wound Dressings and Silverlon Wound Pad Dressings can be used safely and effectively as primary dressings on surgical incisions which heal by primary intent. ## 9.0 Technological Characteristics Layer 1 is a non-adherent wound contact layer that consists of a single layer of knitted continuous nylon fiber coated with metallic silver at the rate of 5.05 mg/cm (50.5 g/m²); the silver surface coating is approximately 0.8 – 1.0 µm thick and contains approximately 1% silver oxide. Layer 1 delivers antimicrobial silver ions in the dressing when activated by moisture. The silver ions in the dressing kill wound bacteria held in the dressing and provides an antimicrobial barrier to protect the wound bed. This dressing absorbs high amounts of wound fluid and the dressing intimately conforms to the wound surface. 3 {6}------------------------------------------------ #### 10.0 Indications for Use Silverlon® Island Wound Dressings and Silverlon Wound Pad Dressings are multi-layer, sterile, non-adherent, antimicrobial barrier wound dressings. ### 10.1. The Over-The-Counter Indications: Local management of superficial wounds, minor burns, abrasions and lacerations. #### 10.2. Prescription Indications: Under the supervision of a healthcare professional Silverlon Island Wound Dressings and Silverlon Wound Pad Dressings are intended for up to 7 day use for wounds such as vascular access or peripheral IV sites, orthopedic external pin sites, wound drain sites, surgical wounds (donor and graft sites, incisions), and partial to full thickness dermal ulcers (stage I-IV pressure sores, venous stasis ulcers, arterial ulcers, diabetic ulcers). Silverlon Island Wound Dressings and Silverlon Wound Pad Dressings are indicated for the management of infected wounds, as the silver in the dressing provides an antimicrobial barrier that may be helpful in managing these wounds. In addition, the moist wound healing environment and control of wound bacteria within the Silverlon Island Wound Dressing and Silverlon Wound Pad Dressing may help reduce the risk of wound infection and support the body's healing process. Silverlon Island Wound Dressings and Silverlon Wound Pad Dressings may be used for the management of painful wounds, Silverlon Island Wound Dressings and Silverlon Wound Pad Dressing's non-adherent wound contact layer reduces pain during dressing changes and evaporation of moisture in the dressing may soothe the wound. #### 11.0 Summary of Technological Characteristics of Device Compared to Predicate Device The technological characteristics have not changed from the previously cleared Silverlon Island Wound Dressing and Silverlon Wound Pad Dressing (K143001) cleared November 7, 2014. Antimicrobial activity due to the Silverlon pad is not affected. The Silverlon Island Wound Dressing has a tape layer supplied with the dressing and Silverlon Wound Pad Dressing is supplied without the tape layer. The labeling change to MR Conditional does not impact the product design or technological characteristics. There are no new questions of safety and effectiveness. {7}------------------------------------------------ ## 12.0 Summary of Nonclinical Testing Silverlon Island Wound Dressings (K122817) have been subjected to independent standard in vitro and in vivo biocompatibility tests, including cytotoxicity, sensitization and intracutaneous reactivity. All tests were performed in accordance with the International Standard Organization (ISO) 10993 Standard Series for Biological Evaluation of Medical Devices, including: - . In vitro accelerated stability testing for antimicrobial effectiveness, equivalent to 5 Years; - . Kirby-Bauer Standard Antimicrobial Susceptibility Test; and - . Microbiological Consultants - ASTM E2315 Time Kill Assay for Antimicrobial Agents Report. The expanded labeling for use of the Silverlon Island Wound Dressings and Silverlon Wound Pad Dressings in the Magnetic Resonance environment is supported by completed testing of: - . Magnetically Induced Displacement Force in accordance with ASTM F2052; - . Magnetically Induced Torque in accordance with ASTM F2213; - . Heating by RF Fields in accordance with ASTM F2182; and - . Image Artifacts in accordance with ASTM F2119. These tests were completed in accordance with FDA quidance, "Establishing Safety and Compatibility of Passive Implants in the Magnetic Resonance (MR) Environment" (December 11, 2014). ## 13.0 Summary of Clinical Testing No clinical testing was necessary for this submission. #### 14.0 Substantial Equivalence The subject device, Silverlon Island Wound Dressings (with a tape layer) and the Silverlon Wound Pad Dressings (pad without a tape layer), and the predicate device, Silverlon Island Wound Dressing and Silverlon Wound Pad Dressing (K143001), have the 1) same intended use and 2) same technological characteristics, and 3) the expanded labeling for use in an MR environment does not raise different questions of safety or effectiveness or significantly affect the safety and effectiveness in comparison to the predicate device.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...