← Product Code [DXJ](/productcode/DXJ) · K180534

# Philips IntelliVue GuardianSoftware (K180534)

_Philips Medizin Systeme Boeblingen GmbH · DXJ · Jul 16, 2018 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K180534

## Device Facts

- **Applicant:** Philips Medizin Systeme Boeblingen GmbH
- **Product Code:** [DXJ](/productcode/DXJ.md)
- **Decision Date:** Jul 16, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.2450
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Software as a Medical Device

## Indications for Use

The IntelliVue GuardianSoftware is intended for use by healthcare providers whenever there is a need for the generation of a patient record. The IntelliVue GuardianSoftware is indicated for use in the collection, storage and management of data from Philips specified Measurements and Philips Patient Monitors that are connected through networks.

## Device Story

Clinical Information Management System (CIMS) collecting/managing vital signs data from networked Philips patient monitors and measurement devices (e.g., IntelliVue MX/MP series, SureSigns, Wearable Biosensor, EarlySense InSight). Software-only product installed on customer-supplied off-the-shelf IT hardware; utilizes client-server architecture. Provides review, reporting, clinical documentation, remote viewing, and predictive trend analytics (trending, notifications, calculations, clinical advisories, EWS deterioration status). Used by healthcare providers in clinical settings to facilitate patient monitoring and clinical decision-making. Subject modification adds optional support for mobile device software clients (Android OS). Output allows clinicians to view patient data and deterioration status, supporting timely clinical intervention.

## Clinical Evidence

Bench testing only. No clinical data provided. Verification and validation activities included integration-level functional/regression testing and system-level hazard analysis/performance testing. Testing conducted per ANSI/AAMI/IEC 62304:2006 standards. Results confirmed all pass/fail criteria met and design risk mitigations effective.

## Technological Characteristics

Software-only product; client-server architecture. Installed on customer-provided off-the-shelf IT equipment (Microsoft OS/Databases). Supports optional mobile clients (Android OS). Connectivity via network. Software life cycle processes compliant with ANSI/AAMI/IEC 62304:2006.

## Regulatory Identification

A medical cathode-ray tube display is a device designed primarily to display selected biological signals. This device often incorporates special display features unique to a specific biological signal.

## Predicate Devices

- Philips IntelliVue GuardianSoftware ([K161767](/device/K161767.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Philips Medizin Systeme Boeblingen GmbH Michael Asmalsky Senior Regulatory Affairs Engineer Hewlett-Packard-Str. 2 Boeblingen, Baden Wuerttemberg D-71034 GERMANY

September 1, 2023

Re: K180534

Trade/Device Name: Philips IntelliVue GuardianSoftware with Software Revision D.0 Regulation Number: 21 CFR 870.2450 Regulation Name: Medical Cathode-Ray Tube Display Regulatory Class: Class II Product Code: DXJ, DQK, OUG

Dear Michael Asmalsky:

The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated July 16, 2018. Specifically, FDA is updating this SE Letter to remove the secondary product code NSX as an administrative correction.

Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter, please contact Aneesh Deoras, OHT2: Office of Cardiovascular Devices, 240-402-4363, Aneesh.Deoras@fda.hhs.gov.

Sincerely,

Aneesh S. Deoras -S

for

Jennifer Shih Kozen Assistant Director Division of Cardiac Electrophysiology, Diagnostics, and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image contains two logos. On the left is the logo for the Department of Health & Human Services - USA, which features a stylized human figure. To the right is the logo for the U.S. Food & Drug Administration (FDA), with the FDA acronym in a blue square and the full name "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

July 16, 2018

Philips Medizin Systeme Boeblingen GmbH Michael Asmalsky Senior Regulatory Affairs Engineer Hewlett-Packard Str. 2 Boeblingen, 71034 DE Baden Wuerttemberg

Re: K180534

Trade/Device Name: Philips IntelliVue GuardianSoftware with software Revision D.0 Regulation Number: 21 CFR 870.2450 Regulation Name: Medical Cathode-Ray Tube Display Regulatory Class: Class II Product Code: DXJ, NSX, DQK, OUG Dated: May 16, 2018 Received: May 21, 2018

Dear Michael Asmalsky:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{2}------------------------------------------------

K180534

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice

(https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Image /page/2/Picture/6 description: The image contains the name "Arielle Drummond -S" in a large, bold font. Below the name, the word "For" is written in a smaller font size. The text is centered and appears to be part of a title or heading. The background is plain and does not contain any other elements.

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

# Indications for Use

510(k) Number (if known) K180534

Device Name

Philips IntelliVue GuardianSoftware with software Revision D.0

Indications for Use (Describe)

The IntelliVue GuardianSoftware is intended for use by healthcare providers whenever there is a need for generation of a patient record.

The IntelliVue GuardianSoftware is indicated for use in the collection, storage and management of data from Philips specified measurements and Philips Patient Monitors that are connected through networks.

| Type of Use (Select one or both, as applicable)               |                                                              |
|---------------------------------------------------------------|--------------------------------------------------------------|
| <div>    Prescription Use (Part 21 CFR 801 Subpart D)  </div> | <div>    Over-The-Counter Use (21 CFR 801 Subpart C)  </div> |

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the word "PHILIPS" in large, blue, bold letters. Below the word "PHILIPS" is the text "K180534" and "Page 1 of 3". The text is in a smaller font size than the word "PHILIPS".

## 510(k) Summary

This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of the Safe Medical Devices Act of 1990 and 21 C.F.R. §807.92.

1.) The submitter of this premarket notification is:

Michael Asmalsky Philips Medizin Systeme Boeblingen GmbH Hewlett-Packard-Str. 2 D-71034 Boeblingen, Germany Tel: ++49 7031 463-1277 Fax: ++49 7031 463-2442 e-mail: michael.asmalsky@philips.com

This summary was prepared on May 15th 2018.

2.) The trade names/proprietary names of the device is:

the Philips IntelliVue GuardianSoftware with software Revision D.0.

The common/usual name is:

for the IntelliVue GuardianSoftware: Clinical Information Management System

The Classification names for the IntelliVue GuardianSoftware are as follows:

| Device Panel              | Classification | ProCode | Description                                         |
|---------------------------|----------------|---------|-----------------------------------------------------|
| Cardiovascular<br>Devices | §870.2450, II  | DXJ     | Display, cathode-ray tube, medical                  |
| General Hospital          | not classified | NSX     | Software, transmission and storage,<br>patient data |
| Cardiovascular<br>Devices | §870.1425, II  | DQK     | Programmable diagnostic computer                    |
| General Hospital          | §880.6310, I   | OUG     | Medical Device Data System                          |

3.) The modified Philips IntelliVue GuardianSoftware (SW Rev. D.0) is substantially equivalent to the previously cleared Philips IntelliVue GuardianSoftware marketed pursuant to K161767.

4.) Description of the Devices:

The IntelliVue GuardianSoftware (866009) is a Clinical Information Management System. It collects and manages vital signs data acquired from the Philips specified measurements and Philips Patient Monitors. The IntelliVue GuardianSoftware provides review, reporting, clinical documentation, remote viewing, operating, interfacing, storage, printing and predictive trend analytics, meaning trending, notification, calculations and clinical advisories including EWS deterioration status. The IntelliVue GuardianSoftware is a software only product. It is intended to be installed on customer supplied compatible off-the shelf information technology equipment that meet the technical requirements as specified by Philips.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the word "PHILIPS" in blue font at the top. Below the word "PHILIPS" is the text "K180354" followed by "Page 2 of 3". The text is in a smaller, serif font.

The IntelliVue GuardianSoftware can currently acquire physiological data from the following compatible measuring devices:

- Philips IntelliVue Measurements CL SpO2 Pod, CL NBP Pod, CL Resp Pod, -
- Philips IntelliVue Patient Monitors MX400/XG50, MP5 and MP5SC, and -
- -Philips SureSigns Patient Monitors VS3/VS4
- -Philips Wearable Biosensor, and the
- -EarlySense InSight Device.

The subject modification adds the optional support for SW clients for off the shelf, customer provided mobile devices to the currently legally marketed client-server architecture of the IntelliVue GuardianSoftware.

To support the before described purposes the IntelliVue GuardianSoftware was modified to maintain a consistent numbering scheme. The modified common software revision is Rev.D.0.

## Intended Use:

The Intended Use and Indications for Use of the subject Philips IntelliVue GuardianSoftware (866009) has not changed as a result of the device modifications. The device has the following detailed Intended Use and Indications for Use Statements in its Instructions for Use:

## Philips IntelliVue GuardianSoftware:

The IntelliVue GuardianSoftware is intended for use by healthcare providers whenever there is a need for the generation of a patient record.

The IntelliVue GuardianSoftware is indicated for use in the collection, storage and management of data from Philips specified Measurements and Philips Patient Monitors that are connected through networks.

## 5.) Technological Characteristics:

The modified device has the same technological characteristics as the legally marketed predicate device. It is a software only product intended to be installed on Philips specified, customer-provided standard (of-the-shelf) IT devices. It uses a client server architecture and it is suitable for use with the specified Microsoft® Operating System and Databases. SW clients are optionally suitable for use with customer provided mobile off the shelf devices using Android OS and meeting the specifications.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. Below the word "PHILIPS" is the text "K180354" in a smaller font size. The text "K180354" is black and appears to be a product or model number.

Page 3 of 3

6.) Summary of Verification, Validation and Testing Activities and Conclusion:

Verification, validation, and testing activities establish the performance, functionality, and reliability characteristics of the subject modified IntelliVue GuardianSoftware with respect to the predicate. Testing involved software testing on integration level (functional testing and regression testing) and software testing on system level (hazard analysis testing and dedicated software performance testing).

Software verification and validation testing was conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices."

Testing as required by the hazard analysis was conducted and all specified pass/fail criteria have been met. The test results confirmed the effectiveness of the implemented design risk mitigation measures.

Verification according to the applicable safety and performance standards was conducted as described below:

| Standard                 | Type                                               |
|--------------------------|----------------------------------------------------|
| ANSI/AAMI/IEC 62304:2006 | General Standard:<br>Software life cycle processes |

Pass/Fail criteria were based on the specifications cleared for the predicate device and all test results showed substantial equivalence. The results demonstrate that the Philips IntelliVue GuardianSoftware (SW Rev.D.0) meets all safety and reliability requirements and performance claims.

---

**Source:** [https://fda.innolitics.com/device/K180534](https://fda.innolitics.com/device/K180534)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
