The IntelliVue GuardianSoftware is intended for use by healthcare providers whenever there is a need for the generation of a patient record. The IntelliVue GuardianSoftware is indicated for use in the collection, storage and management of data from Philips specified Measurements and Philips Patient Monitors that are connected through networks.
Device Story
Clinical Information Management System (CIMS) collecting/managing vital signs data from networked Philips patient monitors and measurement devices (e.g., IntelliVue MX/MP series, SureSigns, Wearable Biosensor, EarlySense InSight). Software-only product installed on customer-supplied off-the-shelf IT hardware; utilizes client-server architecture. Provides review, reporting, clinical documentation, remote viewing, and predictive trend analytics (trending, notifications, calculations, clinical advisories, EWS deterioration status). Used by healthcare providers in clinical settings to facilitate patient monitoring and clinical decision-making. Subject modification adds optional support for mobile device software clients (Android OS). Output allows clinicians to view patient data and deterioration status, supporting timely clinical intervention.
Clinical Evidence
Bench testing only. No clinical data provided. Verification and validation activities included integration-level functional/regression testing and system-level hazard analysis/performance testing. Testing conducted per ANSI/AAMI/IEC 62304:2006 standards. Results confirmed all pass/fail criteria met and design risk mitigations effective.
Technological Characteristics
Software-only product; client-server architecture. Installed on customer-provided off-the-shelf IT equipment (Microsoft OS/Databases). Supports optional mobile clients (Android OS). Connectivity via network. Software life cycle processes compliant with ANSI/AAMI/IEC 62304:2006.
Indications for Use
Indicated for healthcare providers requiring patient record generation, data collection, storage, and management from networked Philips patient monitors and specified measurement devices.
Regulatory Classification
Identification
A medical cathode-ray tube display is a device designed primarily to display selected biological signals. This device often incorporates special display features unique to a specific biological signal.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Philips Medizin Systeme Boeblingen GmbH Michael Asmalsky Senior Regulatory Affairs Engineer Hewlett-Packard-Str. 2 Boeblingen, Baden Wuerttemberg D-71034 GERMANY
September 1, 2023
Re: K180534
Trade/Device Name: Philips IntelliVue GuardianSoftware with Software Revision D.0 Regulation Number: 21 CFR 870.2450 Regulation Name: Medical Cathode-Ray Tube Display Regulatory Class: Class II Product Code: DXJ, DQK, OUG
Dear Michael Asmalsky:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated July 16, 2018. Specifically, FDA is updating this SE Letter to remove the secondary product code NSX as an administrative correction.
Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter, please contact Aneesh Deoras, OHT2: Office of Cardiovascular Devices, 240-402-4363, Aneesh.Deoras@fda.hhs.gov.
Sincerely,
Aneesh S. Deoras -S
for
Jennifer Shih Kozen Assistant Director Division of Cardiac Electrophysiology, Diagnostics, and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image contains two logos. On the left is the logo for the Department of Health & Human Services - USA, which features a stylized human figure. To the right is the logo for the U.S. Food & Drug Administration (FDA), with the FDA acronym in a blue square and the full name "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
July 16, 2018
Philips Medizin Systeme Boeblingen GmbH Michael Asmalsky Senior Regulatory Affairs Engineer Hewlett-Packard Str. 2 Boeblingen, 71034 DE Baden Wuerttemberg
Re: K180534
Trade/Device Name: Philips IntelliVue GuardianSoftware with software Revision D.0 Regulation Number: 21 CFR 870.2450 Regulation Name: Medical Cathode-Ray Tube Display Regulatory Class: Class II Product Code: DXJ, NSX, DQK, OUG Dated: May 16, 2018 Received: May 21, 2018
Dear Michael Asmalsky:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
K180534
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice
(https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Image /page/2/Picture/6 description: The image contains the name "Arielle Drummond -S" in a large, bold font. Below the name, the word "For" is written in a smaller font size. The text is centered and appears to be part of a title or heading. The background is plain and does not contain any other elements.
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K180534
Device Name
Philips IntelliVue GuardianSoftware with software Revision D.0
Indications for Use (Describe)
The IntelliVue GuardianSoftware is intended for use by healthcare providers whenever there is a need for generation of a patient record.
The IntelliVue GuardianSoftware is indicated for use in the collection, storage and management of data from Philips specified measurements and Philips Patient Monitors that are connected through networks.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------|--------------------------------------------------------------|
| <div> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "PHILIPS" in large, blue, bold letters. Below the word "PHILIPS" is the text "K180534" and "Page 1 of 3". The text is in a smaller font size than the word "PHILIPS".
## 510(k) Summary
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of the Safe Medical Devices Act of 1990 and 21 C.F.R. §807.92.
1.) The submitter of this premarket notification is:
Michael Asmalsky Philips Medizin Systeme Boeblingen GmbH Hewlett-Packard-Str. 2 D-71034 Boeblingen, Germany Tel: ++49 7031 463-1277 Fax: ++49 7031 463-2442 e-mail: michael.asmalsky@philips.com
This summary was prepared on May 15th 2018.
2.) The trade names/proprietary names of the device is:
the Philips IntelliVue GuardianSoftware with software Revision D.0.
The common/usual name is:
for the IntelliVue GuardianSoftware: Clinical Information Management System
The Classification names for the IntelliVue GuardianSoftware are as follows:
| Device Panel | Classification | ProCode | Description |
|---------------------------|----------------|---------|-----------------------------------------------------|
| Cardiovascular<br>Devices | §870.2450, II | DXJ | Display, cathode-ray tube, medical |
| General Hospital | not classified | NSX | Software, transmission and storage,<br>patient data |
| Cardiovascular<br>Devices | §870.1425, II | DQK | Programmable diagnostic computer |
| General Hospital | §880.6310, I | OUG | Medical Device Data System |
3.) The modified Philips IntelliVue GuardianSoftware (SW Rev. D.0) is substantially equivalent to the previously cleared Philips IntelliVue GuardianSoftware marketed pursuant to K161767.
4.) Description of the Devices:
The IntelliVue GuardianSoftware (866009) is a Clinical Information Management System. It collects and manages vital signs data acquired from the Philips specified measurements and Philips Patient Monitors. The IntelliVue GuardianSoftware provides review, reporting, clinical documentation, remote viewing, operating, interfacing, storage, printing and predictive trend analytics, meaning trending, notification, calculations and clinical advisories including EWS deterioration status. The IntelliVue GuardianSoftware is a software only product. It is intended to be installed on customer supplied compatible off-the shelf information technology equipment that meet the technical requirements as specified by Philips.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "PHILIPS" in blue font at the top. Below the word "PHILIPS" is the text "K180354" followed by "Page 2 of 3". The text is in a smaller, serif font.
The IntelliVue GuardianSoftware can currently acquire physiological data from the following compatible measuring devices:
- Philips IntelliVue Measurements CL SpO2 Pod, CL NBP Pod, CL Resp Pod, -
- Philips IntelliVue Patient Monitors MX400/XG50, MP5 and MP5SC, and -
- -Philips SureSigns Patient Monitors VS3/VS4
- -Philips Wearable Biosensor, and the
- -EarlySense InSight Device.
The subject modification adds the optional support for SW clients for off the shelf, customer provided mobile devices to the currently legally marketed client-server architecture of the IntelliVue GuardianSoftware.
To support the before described purposes the IntelliVue GuardianSoftware was modified to maintain a consistent numbering scheme. The modified common software revision is Rev.D.0.
## Intended Use:
The Intended Use and Indications for Use of the subject Philips IntelliVue GuardianSoftware (866009) has not changed as a result of the device modifications. The device has the following detailed Intended Use and Indications for Use Statements in its Instructions for Use:
## Philips IntelliVue GuardianSoftware:
The IntelliVue GuardianSoftware is intended for use by healthcare providers whenever there is a need for the generation of a patient record.
The IntelliVue GuardianSoftware is indicated for use in the collection, storage and management of data from Philips specified Measurements and Philips Patient Monitors that are connected through networks.
## 5.) Technological Characteristics:
The modified device has the same technological characteristics as the legally marketed predicate device. It is a software only product intended to be installed on Philips specified, customer-provided standard (of-the-shelf) IT devices. It uses a client server architecture and it is suitable for use with the specified Microsoft® Operating System and Databases. SW clients are optionally suitable for use with customer provided mobile off the shelf devices using Android OS and meeting the specifications.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. Below the word "PHILIPS" is the text "K180354" in a smaller font size. The text "K180354" is black and appears to be a product or model number.
Page 3 of 3
6.) Summary of Verification, Validation and Testing Activities and Conclusion:
Verification, validation, and testing activities establish the performance, functionality, and reliability characteristics of the subject modified IntelliVue GuardianSoftware with respect to the predicate. Testing involved software testing on integration level (functional testing and regression testing) and software testing on system level (hazard analysis testing and dedicated software performance testing).
Software verification and validation testing was conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices."
Testing as required by the hazard analysis was conducted and all specified pass/fail criteria have been met. The test results confirmed the effectiveness of the implemented design risk mitigation measures.
Verification according to the applicable safety and performance standards was conducted as described below:
| Standard | Type |
|--------------------------|----------------------------------------------------|
| ANSI/AAMI/IEC 62304:2006 | General Standard:<br>Software life cycle processes |
Pass/Fail criteria were based on the specifications cleared for the predicate device and all test results showed substantial equivalence. The results demonstrate that the Philips IntelliVue GuardianSoftware (SW Rev.D.0) meets all safety and reliability requirements and performance claims.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.