← Product Code [LZA](/productcode/LZA) · K180467

# SBG Blue Nitrile Powder Free Medical Examination Glove (K180467)

_Showa Best Glove, Inc. · LZA · Oct 10, 2018 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K180467

## Device Facts

- **Applicant:** Showa Best Glove, Inc.
- **Product Code:** [LZA](/productcode/LZA.md)
- **Decision Date:** Oct 10, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.6250
- **Device Class:** Class 1
- **Review Panel:** General Hospital

## Indications for Use

A powder free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hands or fingers to prevent contamination between patient and examiner.

## Device Story

SBG Blue Nitrile Powder Free Medical Examination Glove is a single-use, disposable, non-sterile glove made from a nitrile latex compound. Designed for use by healthcare examiners to provide a barrier against contamination between the examiner and the patient. The device is worn on the hands or fingers during medical examinations. It is intended for over-the-counter use. The glove provides a physical barrier to reduce the risk of cross-contamination, potentially benefiting both the patient and the healthcare provider by maintaining hygiene standards during clinical interactions.

## Clinical Evidence

Bench testing only. Testing included physical properties (tensile strength, ultimate elongation per ASTM D6319-10/D573), freedom from holes (ASTM D5151-06), residual powder (ASTM D6124), and biocompatibility (ISO 10993-10:2010 for primary skin irritation and dermal sensitization). Results confirmed the device meets all acceptance criteria and is not a primary skin irritant or contact sensitizer.

## Technological Characteristics

Material: Nitrile latex compound. Form factor: Disposable, non-sterile, blue, powder-free glove. Dimensions: Length 234 mm, Palm Width 97 mm, Finger Thickness 0.11 mm, Palm Thickness 0.10 mm. Standards: ASTM D6319-10 (physical properties), ASTM D5151-06 (freedom from holes), ASTM D6124 (residual powder), ISO 10993-10:2010 (biocompatibility).

## Regulatory Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

## Predicate Devices

- EMG Blue Nitrile Medical Examination Glove Powder Free ([K141590](/device/K141590.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains two logos. The logo on the left is the Department of Health & Human Services - USA logo. The logo on the right is the FDA U.S. Food & Drug Administration logo. The FDA logo is in blue.

October 10, 2018

Showa Best Glove, Inc Jeffrey Richardson Director of Operations 579 Edison Street Menlo, California 30731-6335

Re: K180467

Trade/Device Name: SBG Blue Nitrile Powder Free Medical Examination Glove Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: Class I Product Code: LZA Dated: September 4, 2018 Received: September 6, 2018

### Dear Jeffrey Richardson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice

(https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Clarence W. Murray lii III -S

For Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital. Respiratory. Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K180467

### Device Name

SBG Blue Nitrile Powder Free Medical Examination Glove

Indications for Use (Describe)

A powder free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hands or fingers to prevent contamination between patient and examiner.

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

## K180467

This summary of 510(k) is being submitted in accordance with 21 CFR 807.92.

#### 1. SUBMITTER

Showa Best Glove, Inc. 579 Edison Street Menlo, GA 30731 Tel: 706-862-6740 Fax: 706-862-2660 Date: October 1, 2018

#### 2. OFFICIAL CORRESPONDENCE/CONTACT PERSON

Jeffrey Richardson Showa Best Glove, Inc. 579 Edison Street Menlo, GA 30731 Tel: 706-862-6740 Fax: 706-862-2660 Email: cmitchell@showagroup.com

#### 3. 510(K) PREPARER

Carol Mitchell, Executive Assistant to President/COO Showa Best Glove, Inc. 579 Edison Street Menlo, GA 30731 Tel: 706-862-6724 Fax: 706-862-2660 Email: cmitchell@showagroup.com

Margaret Savage, RA/QA Supervisor Showa Best Glove, Inc. 931 Second Ave. SE Fayette, AL 35555 Tel: 205-932-3202 134 Fax: 205-932-5904 Email: tsavage@showagroup.com

{4}------------------------------------------------

#### 4. DEVICE

Brand Name of Device: SBG Blue Nitrile Powder Free Medical Examination Glove Common or Usual Name: Patient Examination Glove Classification Name: Non-powdered Patient Examination Glove Regulation: 21 CFR 880.6250 Device Classification: Class: I Product Code: LZA

### 5. PREDICATE DEVICE

Trade Name: EMG Blue Nitrile Medical Examination Glove Powder Free Common Name: Patient Examination Glove 510(k) Number: K141590 Manufacturer: ECO MEDI GLOVE SDN. BHD

### 6. INDICATIONS FOR USE

A powder free patient examination gloves is a disposable device intended for medical purposes that is worn on the examiner's hands or fingers to prevent contamination between patient and examiner.

### 7. DEVICE DESCRIPTION

The SBG Blue Nitrile Powder Free Medical Examination Glove is a single use, disposable device made from a nitrile latex compound. Blue in color, powder free and non-sterile. The proposed device was tested according to the following standards listed in the table below:

{5}------------------------------------------------

#### SUMMARY OF COMPARING TECHNOLOGICAL CHARACTERISTICS WITH PREDICATE DEVICE 8.

There is no difference in the technology characteristics when compared to the predicate device. Both the proposed and predicate devices are made from a nitrile latex compound, Blue in color, powder free and non-sterile.

|                                                                                      |                                                                                                                                                                                                                                                                           | Proposed Device                                                                                                                                                                                                                                                         | Predicate Device                                                                                                                                                                                                                                                        | Comparison                   |
|--------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------|
| Characteristics                                                                      | Acceptance<br>Criteria                                                                                                                                                                                                                                                    | SBG Blue Nitrile<br>Powder Free Medical<br>Examination Glove<br>510(k) Number<br>K180467                                                                                                                                                                                | EMG Blue Nitrile<br>Medical Examination<br>Glove Powder Free<br>510(k) Number<br>K141590                                                                                                                                                                                |                              |
| Product Code                                                                         | LZA                                                                                                                                                                                                                                                                       | LZA                                                                                                                                                                                                                                                                     | LZA                                                                                                                                                                                                                                                                     | Same                         |
| Intended Use                                                                         | A patient examination<br>glove is a disposable<br>device intended for<br>medical purposes that<br>is worn on the<br>examiner's hands or<br>fingers to prevent<br>contamination<br>between patient and<br>examiner.  The device<br>is for over-the-counter<br>use.         | A powder free patient<br>examination gloves is a<br>disposable device intended<br>for medical purposes that is<br>worn on the examiner's<br>hand or finger to prevent<br>contamination between<br>patient and examiner.  The<br>device is for over-the-<br>counter use. | A powder free patient<br>examination gloves is a<br>disposable device intended<br>for medical purposes that is<br>worn on the examiner's<br>hand or finger to prevent<br>contamination between<br>patient and examiner.  The<br>device is for over-the-<br>counter use. | Same                         |
| Material Use                                                                         | Not made from<br>Natural Rubber<br>Latex                                                                                                                                                                                                                                  | Nitrile latex compound                                                                                                                                                                                                                                                  | Nitrile latex compound                                                                                                                                                                                                                                                  | Same                         |
| Color                                                                                | Blue                                                                                                                                                                                                                                                                      | Blue                                                                                                                                                                                                                                                                    | Blue                                                                                                                                                                                                                                                                    | Same                         |
| Sterility                                                                            | Non-Sterile                                                                                                                                                                                                                                                               | Non-Sterile                                                                                                                                                                                                                                                             | Non-Sterile                                                                                                                                                                                                                                                             | Same                         |
| Dimensions                                                                           | Overall Length (mm)<br>220 mm = (sizes XS-S)<br>230 mm = (sizes M-<br>XL)<br>Width (± 10 mm)<br>Size S = 80 mm<br>Size M = 95 mm<br>Size L = 110 mm<br>Size XL = 120 mm<br>Thickness at Finger<br>(mm)<br>All Sizes = 0.05 mm<br>Thickness at Palm<br>All Sizes = 0.05 mm | Size Medium<br>Palm Width: 97 mm<br>Length: 234 mm<br>Finger Thickness: 0.11 mm<br>Palm Thickness: 0.10 mm                                                                                                                                                              | Size Medium<br>Palm Width: 94 mm<br>Length: 245 mm<br>Finger Thickness: 0.17 mm<br>Palm Thickness: 0.10 mm                                                                                                                                                              | Same                         |
| Characteristics                                                                      | Acceptance<br>Criteria                                                                                                                                                                                                                                                    | Proposed Device                                                                                                                                                                                                                                                         | Predicate Device                                                                                                                                                                                                                                                        | Comparison                   |
|                                                                                      |                                                                                                                                                                                                                                                                           | SBG Blue Nitrile<br>Medical Powder<br>Free<br>Examination<br>Glove<br>K180467                                                                                                                                                                                           | EMG Blue Nitrile<br>Medical<br>Examination Glove<br>Powder Free<br>510(k) Number<br>K141590                                                                                                                                                                             |                              |
| Physical Properties                                                                  | Tensile Strength<br>(MPa) = 14 min.                                                                                                                                                                                                                                       | Tensile Strength (MPa)<br>37                                                                                                                                                                                                                                            | Tensile Strength (MPa)<br>34                                                                                                                                                                                                                                            | Same                         |
|                                                                                      | Ultimate Elongation<br>(%) = 500 min.                                                                                                                                                                                                                                     | Ultimate Elongation<br>(%)<br>590                                                                                                                                                                                                                                       | Ultimate Elongation (%)<br>600                                                                                                                                                                                                                                          | Same                         |
|                                                                                      | After Aging ASTM 6319-10 (D573)<br>(70°C ± 2 °C for 166 hrs ± 2 hrs.)                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                         |                                                                                                                                                                                                                                                                         |                              |
|                                                                                      |                                                                                                                                                                                                                                                                           | Tensile Strength<br>(MPa) = 14 min                                                                                                                                                                                                                                      | Tensile Strength (MPa)<br>38                                                                                                                                                                                                                                            | Tensile Strength (MPa)<br>30 |
|                                                                                      | Ultimate Elongation<br>(%) = 500 min.                                                                                                                                                                                                                                     | Ultimate Elongation<br>(%)<br>533                                                                                                                                                                                                                                       | Ultimate Elongation (%)<br>560                                                                                                                                                                                                                                          | Same                         |
| Freedom from Holes<br>(D 5151)                                                       | AQL 2.5 Inspection<br>Level G-1<br>Accept at 5 failures<br>Reject at 6 failures                                                                                                                                                                                           | AQL 2.5 Inspection<br>Meets ASTM D5151-<br>06                                                                                                                                                                                                                           | AQL 2.5 Inspection<br>Meets ASTM D5151-06                                                                                                                                                                                                                               | Same                         |
| Residual Powder<br>(D 6124)                                                          | ≤ 2.0 mg/pc                                                                                                                                                                                                                                                               | 0.18 mg                                                                                                                                                                                                                                                                 | 0.20 mg                                                                                                                                                                                                                                                                 | Same                         |
| Biocompatibility test -<br>Primary Skin Irritation<br>Test<br>(ISO 10993-10:2010 (E) |                                                                                                                                                                                                                                                                           | Under the conditions<br>of the study, not a<br>primary skin irritant.                                                                                                                                                                                                   | Under the conditions of<br>the study, not a primary<br>skin irritant.                                                                                                                                                                                                   | Same                         |
| Biocompatibility test<br>Dermal Sensitization<br>Assay (ISO 10993-<br>10:2010 (E)    |                                                                                                                                                                                                                                                                           | Under conditions of<br>the study, not a<br>contact sensitizer                                                                                                                                                                                                           | Under conditions of the<br>study, not a contact<br>sensitizer                                                                                                                                                                                                           | Same                         |
| Safety Assessment<br>(Systemic Toxicity)                                             |                                                                                                                                                                                                                                                                           | The results from the<br>safety assessment<br>demonstrated the<br>device presents a low<br>potential health risk                                                                                                                                                         | /                                                                                                                                                                                                                                                                       | Different                    |

{6}------------------------------------------------

{7}------------------------------------------------

#### 9. CONCLUSIONS

The conclusions drawn from the nonclinical tests (discussed above) that demonstrate that the device is as safe, as effective, and performs as well as or better than the legally marketed device.

---

**Source:** [https://fda.innolitics.com/device/K180467](https://fda.innolitics.com/device/K180467)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
