K180460 · Theradome, Inc. · OAP · May 18, 2018 · Physical Medicine
Device Facts
Record ID
K180460
Device Name
Theradome LH40
Applicant
Theradome, Inc.
Product Code
OAP · Physical Medicine
Decision Date
May 18, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5500
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Theradome LH40 (Theragrow) is an over-the-counter (OTC) therapeutic device intended to treat Androgenetic Alopecia and promote hair growth in males who have Norwood-Hamilton Classifications of lla to V patterns of hair loss and to treat Androgenetic Alopecia and promote hair growth in females who have Ludwig-Savin Scale I-1 to I-4, II-1, II-2; both with Fitzpatrick Skin Types I to IV.
Device Story
Theradome LH40 is a low-level laser therapy (LLT) helmet for home use. Device utilizes 40 red visible light diode lasers (678 nm) to deliver non-thermal energy to scalp hair follicles. Photobiostimulation promotes hair growth. User operates device via single button; audible timer automatically terminates 20-minute treatment session. Hands-free design allows patient self-use. Output is non-thermal laser energy; healthcare provider does not interact with device output. Benefit is hair growth stimulation for androgenetic alopecia.
Clinical Evidence
No clinical data required. Substantial equivalence established via comparison of technological characteristics and intended use to predicate devices. Bench testing, biocompatibility, mechanical, software, and EMC testing from predicate device remain applicable.
Technological Characteristics
Low-level laser therapy (LLT) device. 40 coherent light diodes (Class 3R), 678 nm wavelength, 5mW power. Helmet form factor (17.5x19.6x27.3 cm, 425g). Battery-operated. Single-button interface with audible timer. No software algorithm complexity; simple timer-based control.
Indications for Use
Indicated for males with Norwood-Hamilton IIa to V hair loss and females with Ludwig-Savin I-1 to I-4, II-1, II-2 hair loss; both with Fitzpatrick Skin Types I to IV.
Regulatory Classification
Identification
An infrared lamp is a device intended for medical purposes that emits energy at infrared frequencies (approximately 700 nanometers to 50,000 nanometers) to provide topical heating.
Special Controls
*Classification.* Class II (special controls). The device, when it is an infrared therapeutic heating lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 890.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, featuring a stylized human figure. To the right is the FDA logo, with the letters "FDA" in a blue square, followed by "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
May 18, 2018
Theradome, Inc. % Jacqueline Hauge Regulatory Affairs Consultant Jacqueline A. Hauge 23300 Manning Trail N Scandia, Minnesota 55073
Re: K180460
Trade/Device Name: Theradome LH40 Regulation Number: 21 CFR 890.5500 Regulation Name: Infrared Lamp Regulatory Class: Class II Product Code: OAP Dated: February 14, 2018 Received: February 20, 2018
Dear Jacqueline Hauge:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good
{1}------------------------------------------------
manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
> Sincerely, Jennifer R. Stevenson -ਟਤੋ For Binita S. Ashar, M.D., M.B.A., F.A.C.S.
Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
### K180460
Device Name Theradome LH40 (Theragrow)
### Indications for Use (Describe)
The Theradome LH40 (Theragrow) is an over-the-counter (OTC) therapeutic device intended to treat Androgenetic Alopecia and promote hair growth in males who have Norwood-Hamilton of IIa to V patterns of hair loss and to treat Androgenetic Alopecia and promote hair growth in females who have Ludwig-Savin Scale I-1 to I-4, II-1, II-2; both with Fitzpatrick Skin Types I to IV.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| ☐ Research Use (Part 21 CFR 821 Subpart B) | ☑ Same Tier Contracting (21 CFR 821 Subpart C) |
|--------------------------------------------|------------------------------------------------|
|--------------------------------------------|------------------------------------------------|
_ | Prescription Use (Part 21 CFR 801 Subpart D)
|X Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "theradome" in a stylized font. The first part of the word, "thera," is in gray, while the second part, "dome," is in light blue. Above the word, there are three curved lines in light blue, adding a design element to the logo.
#### l. SUBMITTER
| Date Prepared: | February 14, 2018 |
|------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | Jacqueline A. Hauge<br>Regulatory Consultant<br>Phone: 763.742.8227<br>Email: jacqueline.hauge@gmail.com |
| Sponsor: | Theradome, Inc.<br>4900 Hopyard Road, Suite 100<br>Pleasanton, CA 94588<br>Phone: 510-714-0525<br>Fax: 925-374-1202<br>Establishment Registration# 3008295459 |
| | |
| Sponsor Contact: | Tamim Hamid, PhD, CEO<br>Phone: 510-714-0525 |
| II. DEVICE | |
#### Trade Name: Theradome LH40 (Theragrow) Common Name: Lamp, non-heating, for promotion of hair growth Classification Name: Infrared Lamp Product Code OAP Regulatory Class: Class II Regulation Number: 21 CFR 890.5500 Panel: General and Plastic Surgery
#### lll. PREDICATE DEVICES
| Primary: | K161046 Theradome LH40 Evo (Theradome, Inc.) |
|-------------|--------------------------------------------------------------|
| Additional: | K171775 Theradome LH80 Pro (Theradome, Inc.) |
| Reference: | K152473 HairMax LaserComb 41 (Lexington International, LLC.) |
#### IV. PURPOSE OF SUBMISSION
The purpose of this Premarket Notification is to obtain FDA clearance for the expanded indications for use statement for the Theradome LH40 device to include treatment of Androgenetic Alopecia and promotion of hair growth in males who have Norwood-Hamilton Classifications of IIa to V patterns of hair loss.
#### DEVICE DESCRIPTION V.
The Theradome LH40 (Theragrow) laser helmet is a low level laser therapy (LLT) device containing red, visible light diode lasers operating at 678 nanometers, designed to deliver non-thermal energy to the hair follicles to promote hair growth via photobiostimulation of the scalp. The lasers are contained inside a lightweight, one-size-fits-all helmet. The Theradome LH40 device utilizes 40 laser diodes to deliver laser stimulation to the entire scalp for hands-free operation during treatment. The device is one-button operated, and has an audible timer that automatically turns the lasers off after the 20 minute treatment is complete.
{4}------------------------------------------------
#### INDICATIONS FOR USE VI.
The Theradome LH40 (Theragrow) is an over-the-counter (OTC) therapeutic device intended to treat Androgenetic Alopecia and promote hair growth in males who have Norwood-Hamilton Classifications of lla to V patterns of hair loss and to treat Androgenetic Alopecia and promote hair growth in females who have Ludwig-Savin Scale I-1 to I-4, II-1, II-2; both with Fitzpatrick Skin Types I to IV.
### VII. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
Low level laser therapy (LLT) is the principle technology for both the subject and predicate devices. At a high-level, the Theradome LH40 (Theragrow) device has the same technological characteristics as the predicate device, including:
- Number of Laser Diode ●
- Laser Class
- Laser Power ●
- Wavelength
- Laser Delivery Method
- Helmet Design
- Single Button Operation
- . Audible Timer
- Treatment Time and Duration ●
- Materials of Construction ●
- Risk Profile
- Battery Operation
The fundamental scientific technology of the Theradome LH40 device was not changed as part of this submission.
| Technological<br>Characteristic | Theradome<br>LH40 | Theradome<br>LH80 | HairMax<br>LaserComb 41 | Theradome<br>LH40 |
|------------------------------------------|--------------------------------|--------------------------------|------------------------------------------------------|--------------------------------|
| Regulatory Information | | | | |
| Manufacturer | Theradome, Inc. | Theradome, Inc. | Lexington Int'l, LLC | Theradome, Inc. |
| 510(k) Number | K161046 | K171775 | K142573 | K180460 |
| Product Code | OAP | OAP | OAP | OAP |
| Device Type | OTC Only | OTC Only | OTC Only | OTC Only |
| Technological Characteristics Comparison | | | | |
| # of Coherent Light Diodes | 40 | 80 | 41 | 40 |
| Coherent Light Class | Class 3R | Class 3R | Class 3R | Class 3R |
| Wavelength (nm) | 678 (+7) | 678 (+7) | 655 (+10) | 678 (+7) |
| Laser Power for Classification | 5mW | 5mW | <5mW | 5mW |
| Power Flux (mW/cm²) | 0.476 | 0.952 | Not Reported | 0.476 |
| Radiant Energy per Treatment (J) | 432 | 216 | Not Reported | 432 |
| Fluence (J/cm²) | 0.52 | 1.03 | 0.0 | 0.52 |
| Treatment Schedule | 4 times/week | 2 times/week | 3-4 times/week<br>(every other day) | 4 times/week |
| Treatment Time | 1200 sec<br>(20 min) | 1200 secs<br>(20 min) | 30 seconds at each of<br>6 positions on the<br>scalp | 1200 sec<br>(20 min) |
| Total Treatment Time | 4800 sec/week<br>(80 min/week) | 2400 sec/week<br>(40 min/week) | 720 seconds<br>(9-12 min/week) | 4800 sec/week<br>(80 min/week) |
| Device Configuration | | | | |
| Device Configuration | Helmet | Helmet | Band | Helmet |
| Device Dimensions (cm) | 17.5x19.6x27.3 | 17.5x19.5x26.8 | 20x5x13 | 17.5x19.6x27.3 |
| Device Weight (g) | 425 | 425 | 156 | 425 |
| Nominal Scalp Coverage (cm²) | 420 | 420 | Not Reported | 420 |
# Comparison of Theradome LH40 to Predicate Devices
{5}------------------------------------------------
#### PERFORMANCE DATA VIII.
Performance testing was not required to support a substantial equivalence determination for the Theradome LH40 device. The existing performance testing; including biocompatibility, mechanical, software, and electrical and electromagnetic compatibility (EMC), are unchanged from the predicate device. Animal testing was not required to support a substantial equivalence determination for the Theradome LH40 device. Clinical testing was not required to support a substantial equivalence determination for the Theradome LH40 device.
#### CONCLUSIONS IX.
The Theradome LH40 device has been demonstrated to be substantially equivalent to the predicate device based on the equivalence of technological characteristics and intended use. Because there are no new or changed technological characteristics presented in this submission, the existing risk profile for the device remains unchanged. Theradome has demonstrated that the Theradome device is safe and effective for the expanded indications for use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.