Shockwave Medical Intravascular Lithotripsy (IVL) System
K180454 · Shockwave Medical, Inc. · PPN · Jun 27, 2018 · Cardiovascular
Device Facts
Record ID
K180454
Device Name
Shockwave Medical Intravascular Lithotripsy (IVL) System
Applicant
Shockwave Medical, Inc.
Product Code
PPN · Cardiovascular
Decision Date
Jun 27, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1250
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Shockwave Medical Intravascular Lithotripsy (IVL) System is intended for lithotripsy-enhanced balloon dilatation of lesions, including calcified lesions, in the peripheral vasculature, including the iliac, femoral, popliteal, infrapopliteal, and renal arteries. Not for use in the coronary or cerebral vasculature.
Device Story
System comprises proprietary balloon catheter, generator, and connector cable. Catheter delivered percutaneously to calcified peripheral arterial lesion. Balloon partially inflated; generator delivers energy through cable to integrated lithotripsy emitters within balloon. Emitters generate pulsatile mechanical energy to disrupt vascular calcium; facilitates subsequent low-pressure balloon dilatation of stenosis. Used in clinical setting by physicians. Benefits include improved tissue flexibility and vessel patency in difficult-to-treat calcified lesions.
Clinical Evidence
Prospective, non-randomized, multi-center study (DISRUPT BTK) of 20 patients with calcified infrapopliteal lesions. Mean age 79; 75% Rutherford Category 5. Primary safety endpoint (MAE at 30 days) was 0.0%. Primary effectiveness endpoint (acute reduction in percent diameter stenosis) was 46.5%; residual stenosis 26.2%. Procedural success (residual stenosis ≤50%) achieved in 100% of subjects. No device-related adverse events reported.
Technological Characteristics
System includes balloon catheter with integrated lithotripsy emitters, generator, and connector cable. Catheter features hydrophilic coating. Energy source: electrical energy converted to pulsatile mechanical energy. Sterilization: sterile (catheter), non-sterile (generator/cable). Design includes varying diameters (2.5mm-7.0mm) and lengths (40mm-60mm).
Indications for Use
Indicated for lithotripsy-enhanced balloon dilatation of calcified lesions in peripheral vasculature (iliac, femoral, popliteal, infrapopliteal, renal arteries). Not for coronary or cerebral use.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 27, 2018
Shockwave Medical, Inc. Ms. Plessy Paul Sr. Regulatory Affairs Specialist 48501 Warm Springs Blvd. Suite 108 Fremont, California 94539
Re: K180454
Trade/Device Name: Shockwave Medical Intravascular Lithotripsy (IVL) System Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous catheter Regulatory Class: Class II Product Code: PPN Dated: June 1, 2018 Received: June 4, 2018
Dear Ms. Paul:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{2}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
> Sincerely, 2018.06.27 Eleni Whatley 15:10:26 -04'00' For Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K180454
#### Device Name
Shockwave Medical Intravascular Lithotripsy (IVL) System
Indications for Use (Describe)
The Shockwave Medical Intravascular Lithotripsy (IVL) System is intended for lithotripsy-enhanced balloon dilatation of lesions, including calcified lesions, in the peripheral vasculature, including the iliac, femoral, popliteal, infrapopliteal, and renal arteries. Not for use in the coronary or cerebral vasculature.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|
| <span style="font-size:100%;">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size:100%;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
### K180454
### 510(k) Summary
This 510(k) Summary is submitted in accordance with the requirements of the Safe Medical Devices Act (SMDA) of 1990 and 21 CFR 807.92.
### Name, Address, Phone, and Fax Number of Applicant
Shockwave Medical, Inc. 48501 Warm Springs Blvd., Suite 108 Fremont, CA 94539 Phone: (510) 279-4262 Fax: (510) 279-5934
### Contact Person
Plessy Paul Principal Regulatory Affairs Specialist
#### Date Prepared
June 1, 2018
| Trade Name: | Shockwave Medical Intravascular Lithotripsy (IVL) System |
|----------------------|----------------------------------------------------------|
| Common Name: | Catheter, angioplasty, peripheral, transluminal |
| CFR Classification: | 21 CFR 870.1250 |
| Classification Name: | Percutaneous catheter |
| Product Code: | PPN |
#### Device Name and Classification
#### Predicate Device
The predicate device is the Shockwave Medical Lithoplasty System, K163306, cleared by FDA on December 22, 2016.
### Indications for Use / Intended Use
The Shockwave Medical Intravascular Lithotripsy (IVL) System is intended for lithotripsy-enhanced balloon dilatation of lesions, including calcified lesions, in the peripheral vasculature. including the iliac. femoral, popliteal. infra-popliteal. and renal arteries. Not for use in the coronary or cerebral vasculature.
### Device Description
The Shockwave Medical IVL System has three components: a proprietary balloon Catheter, a Generator, and a Connector Cable. The IVL Catheter has integrated lithotripsy emitters and is designed to enhance percutaneous transluminal angioplasty by enabling delivery of the calcium disrupting capability of lithotripsy prior to full balloon dilatation at low pressures. The application of lithotripsy mechanical pulse waves alters
{5}------------------------------------------------
the structure of an occlusive vascular deposit (stenosis) prior to low-pressure balloon dilation of the stenosis and facilitates the passage of blood.
The IVL Catheter is delivered through the peripheral arterial system of the lower extremities to the site of an otherwise difficult to treat lesion. The balloon is partially inflated and the lithotripsy emitters are energized thereby generating pulsatile mechanical energy within the balloon at the target treatment site and allowing subsequent dilation of a peripheral artery stenosis using low balloon pressure. The IVL Generator delivers energy through the IVL Connector Cable to the pulse emitters located inside the balloon in the IVL Catheter. The IVL Catheter is a single-use device supplied sterile to the customer. The IVL Generator and Connector Cable are non-sterile reusable devices.
### Technological Comparison
This Traditional 510(k) Premarket Notification is being submitted for a line extension to the Shockwave Medical Inc. Peripheral Intravascular Lithotripsy (IVL) System. An additional four (4) Catheter sizes (2.5mm, 3.0mm, 3.5mm and 4.0mm in diameter and 40mm length) are being added to the existing product offering of 3.5mm, 4.5mm, 5.0mm, 5.5mm, 6.0mm, 6.5 mm and 7.0mm diameter and 60mm length. These additional 4 catheter sizes are referred to as Shockwave S4 Peripheral Intravascular Lithotripsy (IVL) Catheter. The additional sizes are modifications of the 510(k) cleared Shockwave Medical IVL Catheter design which utilizes lithotripsy and balloon technology to disrupt calcification. The Shockwave S4 IVL Catheter sizes feature shorter and smaller balloon configurations with smaller crossing profile and the presence of a hydrophilic coating.
The Shockwave S" IVL Catheter has the same intended use, principles of operation and has substantially equivalent technological characteristics including same fundamental scientific technology, design, energy source, shelf life, and sterilization as the already 510(k) cleared Shockwave Medical Peripheral IVL Catheter. The Shockwave Sª IVL Catheter models use the same basic design of the predicate IVL Catheter balloon technology (an IVL catheter with integrated lithotripsy emitters to enable the localized delivery of pulsatile mechanical energy) and utilize lithotripsy to disrupt and restructure calcified lesions to improve tissue flexibility.
Updates were made to the Shockwave St IVL Catheter materials such as balloon material, Inner Lumen Tubing outer layer and colorant as well as adding hydrophilic coating. To accommodate the shorter, smaller balloon configurations with smaller crossing profile updates were made to the numbers of emitters, emitter design and total maximum pulses per catheter. In addition, a minor update to the IVL Catheter Connector was made to reduce system component interdependency. Lastly, the IVL Catheter packaging was updated.
### Summary of Performance Data
Objective evidence demonstrating that the Shockwave S4 IVL Catheter design output meets the product design input requirements as well as that device performance characteristics conform to user needs and intended uses as defined in the product specification was provided. Testing was conducted in accordance with Shockwave
{6}------------------------------------------------
Medical's Risk Analysis procedures, applicable FDA guidance documents and relevant international standards. Results demonstrate that the performance of the Shockwave St IVL Catheter meets its design specifications and is substantially equivalent to the predicate device.
## Performance Testing - Bench
Objective evidence demonstrating that the Shockwave S* IVL Catheter design output meets the product design input requirements as well as that device performance characteristics conform to user needs and intended uses as defined in the product specification was provided. The following tests were conducted for the Shockwave S4 IVL Catheter:
- Catheter diameter and balloon profile ●
- Tensile strength
- Kink resistance / flexibility ●
- Catheter torsional strength ●
- Balloon Inflation / deflation time
- Minimum Burst Strength (RBP)
- Balloon compliance ●
- Fatigue (multiple inflations)
- Pushability and trackability
- Fluoroscopic visibility
- Particulate evaluation
- Pulsing cycles and output
- Temperature Rise
- Guidewire compatibility
- Catheter Sheath Introducer Compatibility
- Catheter Working Length ●
- Distal Tip Profile ●
- Distal Tip Durability ●
- Catheter lubricity
- Coating Length ●
- Coating uniformity / integrity
- Marker Band Spacing and Alignment ●
- Emitter Spacing ●
- Emitter and Marker Band Bond Strength Integrity ●
The IVL Catheter packaging configuration was evaluated for the impact of product handling, distribution, and the storage environment through to the end-user in accordance with applicable standards.
Bench testing results demonstrate that the Shockwave St IVL Catheter meets its design specifications, and is substantially equivalent to the identified predicate device.
### Summary of Biocompatibility Testing
{7}------------------------------------------------
The Shockwave Sd IVL Catheter is categorized as "Externally communicating, Circulating Blood, (A) Limited exposure (<24hrs)". The biocompatibility of the Shockwave S4 IVL Catheter was assessed in accordance with ISO 10993-1:2009 -Biological evaluation of medical devices, Part 1 - Evaluation and tests within a risk management process and in accordance with the provisions of 21 CFR 58 Good Laboratory Practices. The following testing was performed: cytotoxicity, sensitization, irritation/intracutaneous reactivity, acute system toxicity, material -mediated pyrogenicity, chemical characterization, and hemocompatibility. The biocompatibility test results confirm that the Shockwave S 1VL Catheter is non-cytotoxic, nonsensitizing, non-irritating, not systemically toxic, non-hemolytic, and hemocompatible when evaluated under the respective test conditions.
Based on the successful completion of testing, the Shockwave S4 IVL System meets the product specification and user needs requirements specified in the test protocols and satisfies the design verification testing for dimensions, function, simulated use, system compatibility, and IVL treatment. Testing and results meet the requirements of applicable international standards and FDA guidance documents, demonstrating that the Shockwave S4 IVL System is substantially equivalent to the identified predicate device.
### Clinical Study
The 510(k) submission included the 30 day clinical study results of the DISRUPT BTK Study conducted on a configuration of the predicate Shockwave Medical Peripheral Lithoplasty Catheters in Europe, and included device sizes in 2.5mm, 2.75mm, 3.0mm, 3.25mm and 3.5mm x 60 mm.
The DISRUPT BTK study was a prospective, non-randomized, multi-center 20 patient study including the treatment of stenotic, infrapopliteal arteries with the Shockwave Medical IVL System. A total of twenty subjects were enrolled across three (3) study sites in three (3) countries (Austria, Germany and New Zealand). Baseline characteristics were consistent with a calcified patient population with infrapopliteal lesions. The mean age was 79 years, 40.0% of patients presented with renal insufficiency and the majority of subjects (75.0%) presented with Rutherford Category 5 critical limb ischemia. The mean RVD was 3.2 mm. the mean diameter stenosis 72.6%, and mean lesion length was 52.2 mm.
The IVL Catheter was delivered to the target lesion in 19 subjects (95.0%). In one subject the IVL Catheter could not be delivered, and the procedure was aborted.
The primary safety endpoint, Major Adverse Events (MAE) at 30 days, was 0.0%. There were no deaths, myocardial infarctions, emergency surgical revascularizations or amputations at 30 days post-procedure. No procedure and/or device related adverse events were reported though the 30 day follow-up.
The primary effectiveness endpoint, acute reduction in percent diameter stenosis of target lesion, was 46.5%. Acute effectiveness results showed residual stenosis of 26.2% postlithotripsy. Vascular complications included one Type B dissection. None of the
{8}------------------------------------------------
subjects experienced a thrombus, abrupt closure, no-reflow, distal embolization or perforation event.
The secondary endpoint was procedural success defined as the ability of the Shockwave Medical IVL System to achieve a post-IVL residual diameter stenosis of ≤50%. All subjects achieved residual diameter stenosis of ≤50%.
The clinical outcomes of this study demonstrated that the Peripheral IVL System can safely and effectively deliver localized pulsatile mechanical energy for balloon dilatation of calcified, stenotic infrapopliteal arteries.
### Basis for Substantial Equivalence
The IVL System with the updated Shockwave Sd IVL Catheter shares the same intended use, principles of operation, overall technical and functional capabilities, and similar design and materials as the identified predicate device. Any differences between the Shockwave S IVL Catheter and predicate catheter were evaluated through design verification and validation testing which demonstrated device performance and confirmed that there are no new questions of safety or effectiveness compared to the predicate device. Based on the nonclinical and clinical tests, the Shockwave Sª IVL Catheter has been determined to be substantially equivalent to the legally marketed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.