reVive Light Therapy LED Ultrasonic Wrinkle System
K180445 · Led Technologies, Inc. · OHS · May 15, 2018 · General, Plastic Surgery
Device Facts
Record ID
K180445
Device Name
reVive Light Therapy LED Ultrasonic Wrinkle System
Applicant
Led Technologies, Inc.
Product Code
OHS · General, Plastic Surgery
Decision Date
May 15, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The reVive Light Therapy® LED Cleansing System is an Over-the-Counter (OTC) LED device intended for use in treating wrinkles.
Device Story
The reVive Light Therapy® LED Cleansing System is an OTC handheld device for wrinkle treatment. It utilizes four types of LEDs (605nm amber, 630nm red, 660nm red, and 880nm infrared) to deliver light energy directly to the skin. The device also features a secondary cleansing function using TPE brush-heads. The system is powered via USB and includes a power adapter. Operation is simple, with no user-adjustable settings; the user applies the device directly to the skin for a 3-minute treatment duration. The device is intended for patient self-use at home. By providing consistent light energy, the device aims to reduce the appearance of wrinkles. The healthcare provider is not involved in the routine operation of this OTC device.
Clinical Evidence
No clinical trials were conducted. Evidence consists of bench testing, including functional performance testing, software validation, and a usability study with 15 participants. 100% of participants successfully performed the light sensitivity test and operated the device. Biocompatibility testing for ABS and TPE materials was performed per ISO 10993-5 and ISO 10993-10, showing no cytotoxicity, intracutaneous reactivity, or skin sensitization.
Technological Characteristics
The device uses LED-based light therapy (605nm, 630nm, 660nm, 880nm) and a mechanical cleansing brush. Materials include ABS (housing) and TPE (brush heads). The device is powered via USB. It conforms to IEC 60601-1 (general safety) and IEC 60601-1-2 (EMC). The system is a standalone, non-networked OTC device.
Indications for Use
Indicated for the treatment of wrinkles in an Over-the-Counter (OTC) setting.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 15, 2018
LED Technologies, Inc. Jelena Barbaric Compliance Manager 6000 Greenwood Plaza Blvd., Suite 110 Greenwood Village, Colorado 80111
Re: K180445
Trade/Device Name: reVive Light Therapy LED Cleansing System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: OHS Dated: February 16, 2018 Received: February 20, 2018
Dear Jelena Barbaric:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
> Sincerely. Jennifer R. Stevenson -S3 For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices
Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known) K180445
Device Name reVive Light Therapy® LED Cleansing System
Indications for Use (Describe) marcanolis of the rapy® LED Cleansing System is an Over-the-Counter (OTC) LED device intended for use in treating wrinkles.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510 (k) Summary
This summary of 510 (k) information is being submitted in accordance with the requirements of 21 CFR § 878.4810.
Submission Date: February 16th, 2018
| 1. Submitter Information: | LED Technologies, Inc. – Jelena Barbaric<br>6000 Greenwood Plaza Blvd., Suite 110<br>Greenwood Village, CO 80111<br>Tel: 303-407-6882<br>Email: jbarbaric@ledtechnologies.com |
|---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
For Specification Developer: LED Technologies, Inc.
Attn: Lloyd Nelson
6000 Greenwood Plaza Blvd., Suite 110
Greenwood Village, CO, 80111
Tel.: 303-646-0543. Ext 117
Email: Inelson@ledtechnologies.com
### 2. General Information
- 2.1 Classification Name: Light Based Over-The-Counter Wrinkle Reduction
- 2.2 Common/usual name: reVive Light Therapy® LED Cleansing System
- 2.3 Proprietary Names: reVive Light Therapy® LED Cleansing System
- 2.4 Classification: Class II
- 2.5 Classification Number: 878.4810
- 2.6 Product Code: OHS
- 2.7 Review Panel: General & Plastic Surgery
### 3. Device Description
The reVive Light Therapy® LED Cleansing System is an over-the counter light emitting diode (LED) device, that emits energy for use in dermatology for the treatment of wrinkles. The device uses four types of LEDs: 605nm amber, 630nm red, and 880nm infrared.
{4}------------------------------------------------
In addition to the LED light therapy treatment function, reVive Light Therapy® LED Cleansing System has a separate function for cleansing. The system includes two cleansing brush-heads, constructed of TPE material.
The reVive Light Therapy® LED Cleansing System components include the device containing the LED module, USB power cord, and power adapter.
The treatment time is controlled by the user. There are no user settings or adjustments required.
The unit is applied directly to the skin to ensure consistent administration of light during each treatment. The device does not contain any user serviceable components. The device is sold as Over the Counter (OTC).
# 4. Indications/Intended Use:
The reVive Light Therapy® LED Cleansing System is an Over-the-Counter (OTC) LED device intended for use in treating wrinkles.
### 5. Predicate Device:
This device is substantially equivalent to the following predicates, which are currently in safe and effective commerce under product codes OHS:
K171390 – dpl® II Panel (LED Technologies, Inc.) K141181 - dpl® Nüve wrinkle (LED Technologiees, Inc.) K162489 – Red Light Device K152332 - Perfectio LED Infrared Device
# Predicate Chart
| Device | reVive Light<br>Therapy®<br>LED<br>Cleansing<br>System<br>K180445 | dpl® II Panel<br>LED<br>Technologies,<br>Inc.<br>K171390 | dpl® Nüve<br>wrinkle<br>LED<br>Technologies,<br>Inc.<br>K141181 | Red Light<br>Device<br>K162489 | Perfectio LED<br>Infrared<br>Device<br>K152332 |
|----------------------|-------------------------------------------------------------------|----------------------------------------------------------|-----------------------------------------------------------------|--------------------------------|------------------------------------------------|
| Wavelengths | 605nm,<br>630nm,<br>660nm,<br>880nm | 605nm,<br>630nm,<br>660nm,<br>880nm | 605nm,<br>630nm,<br>660nm, 880nm | 633nm<br>830nm | 633nm<br>830nm |
| Irradiance<br>source | LED | LED | LED | LED | LED |
| Treatment<br>Area | 18.86 cm² | 415 cm² | 27 cm² | 17 cm² | 17 cm² |
{5}------------------------------------------------
| Treatment<br>Time | 3 minutes<br>per<br>treatment | 3 minutes per<br>treatment | 3 minutes per<br>treatment | 5-7 minutes<br>per treatment<br>area | 5-7 minutes<br>per treatment<br>area |
|-------------------|---------------------------------|------------------------------|------------------------------|---------------------------------------------|---------------------------------------------|
| Type/Class | OTC | OTC | OTC | OTC | OTC |
| IFU | For<br>treatment of<br>wrinkles | For treatment<br>of wrinkles | For treatment<br>of wrinkles | For treatment<br>of periorbital<br>wrinkles | For treatment<br>of periorbital<br>wrinkles |
### Summary of the technological characteristics of the device compared to predicate device:
- 1. Has the same/similar intended use as the predicate device (l.e., treatment of wrinkles);
- 2. Has the similar output (mW/cm2) as predicate devices;
- 3. Utilizes the same/similar treatment duration) as the predicate device;
- 4. Utilizes the similar recommended treatment regimen.
The reVive Light Therapy® LED Cleansing system and the above referenced predicate devices are Over the Counter Devices used to treat wrinkles as defined in 21 CFR § 878.4810. These devices utilize red and IR diodes between 605 nm to 880 nm to provide narrow bands of light energy to treat wrinkles. The performance achieved by these devices is same/similar with similar power output. The devices are intended to be placed directly on the skin. They are manufactured out of similar materials. Based upon comparison to the predicate devices, the reVive Light Therapy® LED Cleansing System has the same/similar intended uses, with similar technological characteristics as predicate devices. The system performs as intended and does not raise any new safety or effectiveness issues.
### 5. Performance Testing and Standards:
Testing of the reVive Light Therapy® LED Cleansing System, included functional performance testing, software validation, testing, and user safety testing.
Safety and functionality testing demonstrates that the reVive Light Therapy® LED Cleansing system conforms to various international consensus standards.
IEC 60601-1: (2012): medical Electrical Equipment part 1: General Requirements for Basic Safety and Essential Performance.
IEC 60601-1-2 (2014): Medical Electrical Equipment Part 1: General Requirements for Basic Safety and Essential Performance: Collateral Standard: Electromagnetic Compatibility.
{6}------------------------------------------------
### Biocompatibility
The material that comes in contact with patient's skin during use of reVive Light Therapy® LED Cleansing system is the ABS material, the identical material as used in construction of predicate devices dpl® Nüve and dpl® II Panel.
The biocompatibility testing was conducted on the dpl® Nüve handle that is constructed of ABS material, which demonstrates the safety of the ABS material used in construction of dpl® Nüve and equivalently in the construction of the reVive Light Therapy® LED Cleansing system.
In addition, the brush head will also come into contact with patient skin. The brush head is constructed form TPE material. The biocompatibility testing was conducted for TPE material and provided as a part of biocompatibility report.
ISO 10993-5:2009 – Cytotoxicity Test
Purpose of the Test:
The purpose of the test is to determine the biological reactivity of a test cell culture in response to the test article.
### Test Summary:
Under the conditions of the study, the test article extract did not show potential toxicity to test cell culture.
ISO 10993-10:2010 – Intracutaneous reactivity test
Purpose of the Test:
The test was designed to determine the potential intracutaneous reactivity caused by the extract.
Test Summary:
The test result showed that the applied sample extract did not induce intracutaneous reactivity in the test subject under the test conditions.
ISO 10993-10:2010 – Skin Sensitization Test
Purpose of the Test:
The test was designed to evaluate the potential of a test article to cause skin sensitization.
Test Summary:
According to the results, the test article extract showed no significant evidence of causing skin sensitization in the test subjects under the conditions of the study.
{7}------------------------------------------------
The reVive Light Therapy® LED Cleansing system software was tested and validated in accordance with FDA's "Guidance for the content of Premarket Submissions for Software Contained in Medical Devices".
A Usability Study was conducted with 15 participants. The results of the study found that:
100% of the participants were able to demonstrate the light sensitivity test.
100% of the participants were able to use the device successfully.
The conclusions drawn from nonclinical tests demonstrate that the device is safe, as effective, and performs as well as the legally marketed devices.
# 6. Statement of Safety Effectiveness:
The information in this 510 (k) submission was used to support the safety and effectiveness of this device with respect to its cited properties.
### 7. Substantial Equivalence Discussion
After analysis of safety, indications, intended uses, dose rates, performance, features, design materials, power output, technological properties, treatment regiment regimens and methods of operation, the manufacturer asserts that no significant differences exist between the subject device and predicate, and no different questions of safety and effectiveness arise. Therefore, the subject device is substantial equivalence to the predicate.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.