← Product Code [HEW](/productcode/HEW) · K180429

# Advincula Delineator Uterine Manipulator (K180429)

_CooperSurgical, Inc. · HEW · Mar 14, 2018 · Obstetrics/Gynecology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K180429

## Device Facts

- **Applicant:** CooperSurgical, Inc.
- **Product Code:** [HEW](/productcode/HEW.md)
- **Decision Date:** Mar 14, 2018
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 884.1640
- **Device Class:** Class 2
- **Review Panel:** Obstetrics/Gynecology
- **Attributes:** Therapeutic

## Indications for Use

The CooperSurgical Advincula Delineator Uterine Manipulator is indicated to provide delineation of the vaginal fornices and maintain pneumoperitoneum as a uterine manipulator during Total Laparoscopic Hysterectomy (TLH), Laparoscopic Assisted Vaginal Hysterectomy (LAVH) and/or Laparoscopic Supracervical Hysterectomy (LSH).

## Device Story

Single-use, sterile uterine manipulator for laparoscopic hysterectomy (TLH, LAVH, LSH). Device features arched shaft with distal balloon and sliding colpotomy cup (Koh-Cup) with position lock. Proximal handle allows physician to manipulate uterus; Koh-Cup rim delineates vaginal fornices as anatomical landmark for resection. Occluder balloon maintains pneumoperitoneum. Used in operating room by surgeons. Benefits include improved visualization of vaginal fornices and maintenance of pneumoperitoneum during surgery.

## Clinical Evidence

Bench testing only. Design verification included pull-off and compression testing. Design validation performed in a cadaveric model. Biocompatibility testing conducted per ISO 10993-5 and ISO 10993-10 (cytotoxicity, sensitization, irritation). Shelf life testing performed per ASTM F1980-07.

## Technological Characteristics

Materials: stainless steel, silicone, various plastics, adhesives, inks. Distal balloon and sliding colpotomy cup (Koh-Cup) with position lock. Occluder balloon inflation capacity: 60-120cc. Sterilization: Ethylene Oxide. Shelf life: 1 year. Biocompatibility per ISO 10993-1:2009.

## Regulatory Identification

A culdoscope is a device designed to permit direct viewing of the organs within the peritoneum by a telescopic system introduced into the pelvic cavity through the posterior vaginal fornix. It is used to perform diagnostic and surgical procedures on the female genital organs. This generic type of device may include trocar and cannula, instruments used through an operating channel, scope preheaters, light source and cables, and component parts.

## Predicate Devices

- CooperSurgical Advincula Delineator Uterine Manipulator ([K143650](/device/K143650.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

March 14, 2018

CooperSurgical, Inc. Roaida Johnson Director, RA New Product Development 95 Corporate Drive Trumbull, CT 06611

Re: K180429

Trade/Device Name: Advincula Delineator™ Uterine Manipulator Regulation Number: 21 CFR§ 884.1640 Regulation Name: Culdoscope and Accessories Regulatory Class: II Product Code: HEW Dated: February 14, 2018 Received: February 16, 2018

Dear Roaida Johnson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Image /page/1/Picture/6 description: The image shows the name "Charles Viviano -S" in a large, sans-serif font. The text is black against a white background. The name appears to be a title or heading, possibly indicating the subject of a document or presentation. The letters are evenly spaced and clearly legible.

For Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K180429

#### Device Name

Advincula Delineator™ Uterine Manipulator

#### Indications for Use (Describe)

The CooperSurgical Advincula Delineator Uterine Manipulator is indicated to provide delineation of the vaginal fornices and maintain pneumoperitoneum as a uterine manipulator during Total Laparoscopic Hysterectomy (TLH), Laparoscopic Assisted Vaginal Hysterectomy (LAVH) and/or Laparoscopic Supracervical Hysterectomy (LSH).

| Type of Use (Select one or both, as applicable) |                                             |
|-------------------------------------------------|---------------------------------------------|
| ×  Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# K180429 Advincula Delineator™ Uterine Manipulator Special 510(k) Summary

#### Submitter Information

| Company Name:    | CooperSurgical Inc.                                                                                |
|------------------|----------------------------------------------------------------------------------------------------|
| Company Address: | 95 Corporate Drive<br>Trumbull, CT 06611<br>Telephone: 203-601-5200 Ext. 3325<br>Fax: 203-601-9870 |
| Contact Person:  | Roaida Johnson                                                                                     |

March 13, 2018 Date Prepared:

#### Device Information

| Trade Names:         | Advincula Delineator™ Uterine Manipulator |
|----------------------|-------------------------------------------|
| Common Name:         | Uterine Manipulator                       |
| Classification:      | Class II per 21 CFR 884.1640              |
| Classification Name: | Culdoscope and Accessories                |
| Product Code:        | HEW - Culdoscope (And Accessories)        |

#### Predicate Device Information

The Advincula Delineator uterine manipulator is substantially equivalent to the following predicates:

Predicate: CooperSurgical Advincula Delineator Uterine Manipulator (K143650)

The predicate device has not been subject to a design-related recall.

## Device Description

The Advincula Delineator is a uterine manipulator that is intended for use during total laparoscopic hysterectomy (TLH), laparoscopic assisted vaginal hysterectomy (LAVH), and/or laparoscopic supracervical hysterectomy (LSH) procedures. It is single-use, disposable, and provided sterile. The device has a distal balloon built into the arched shaft, and a sliding colpotomy cup, called a Koh-Cup, that locks into place with a position lock. The handle at the proximal end allows the user to hold the device and to manipulate the uterus. When properly positioned, the outer rim of the Koh-Cup delineates the vaginal fornices and provides an anatomical landmark to facilitate uterine resection. An occluder balloon is used to maintain pneumoperitoneum during the procedure.

The Advincula Delineator is available with Koh-Cups sized 2.5cm, 3.5cm or 4.0cm in diameter. There are flexible plastic and Ultem plastic Koh-Cups available for use in electrocautery procedures, and a version with a stainless-steel outer rim for use in harmonic scalpel or laser procedures.

{4}------------------------------------------------

## Indications for Use

The CooperSurgical Advincula Delineator Uterine Manipulator is indicated to provide delineation of the vaginal fornices and maintain pneumoperitoneum as a uterine manipulator during Total Laparoscopic Hysterectomy (TLH), Laparoscopic Assisted Vaginal Hysterectomy (LAVH) and/or Laparoscopic Supracervical Hysterectomy (LSH).

## Substantial Equivalence Discussion

Table 1 provides a comparison of the subject and predicate device. Table 1: Comparison of the Subject Advincula Delineator to the Predicate Devices

| Attribute                               | Subject Advincula Delineator                                                                                                                                                                                                                                                                                                                                  | Predicate Advincula<br>Delineator                                |
|-----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------|
| 510(k) Number                           | K180429                                                                                                                                                                                                                                                                                                                                                       | K143650                                                          |
| Manufacturer                            | CooperSurgical                                                                                                                                                                                                                                                                                                                                                | Same                                                             |
| Indications for Use                     | The CooperSurgical Advincula Delineator Uterine<br>Manipulator is indicated to provide delineation of<br>the vaginal fornices and maintain<br>pneumoperitoneum as a uterine manipulator during<br>Total Laparoscopic Hysterectomy (TLH),<br>Laparoscopic Assisted Vaginal Hysterectomy<br>(LAVH) and/or Laparoscopic Supracervical<br>Hysterectomy (LSH).     | Same                                                             |
| Fundamental<br>Scientific<br>Technology | The device has a distal balloon built into the arched<br>shaft, and a sliding colpotomy cup, called a Koh-<br>Cup, that locks into place. The handle at the<br>proximal end allows the user to hold the device and<br>manipulate the uterus, and the Koh-Cup delineates<br>the vaginal fornices. An occluder balloon is used to<br>maintain pneumoperitoneum. | Same                                                             |
| Manipulator<br>Material(s)              | Stainless steel, silicone, various plastics, adhesives<br>and inks                                                                                                                                                                                                                                                                                            | Same                                                             |
| Koh-Cup Materials                       | Stainless steel, plastic (soft and hard)                                                                                                                                                                                                                                                                                                                      | Different; this device<br>does not include a soft<br>plastic cup |
| Vaginal Occlusion<br>(pneumoperitoneum) | Occluder balloon with inflation capacity of 60-120cc                                                                                                                                                                                                                                                                                                          | Same                                                             |
| Environment of use                      | Operating room                                                                                                                                                                                                                                                                                                                                                | Same                                                             |
| Patient Contact                         | External communicating device, limited (<24)                                                                                                                                                                                                                                                                                                                  | Same                                                             |
| Sterilization Method                    | Ethylene Oxide                                                                                                                                                                                                                                                                                                                                                | Same                                                             |
| Number of Uses                          | Single-Use, Disposable                                                                                                                                                                                                                                                                                                                                        | Same                                                             |
| Packaging                               | Individually packaged in a flexible blister pouch<br>with a Tyvek lid, and three (3) pouches in an SBS<br>box                                                                                                                                                                                                                                                 | Same                                                             |
| Shelf Life                              | 1 Year                                                                                                                                                                                                                                                                                                                                                        | 3 Years                                                          |

The subject and predicate devices have indications for use and the same fundamental scientific technology. The only difference between the predicate and subject device is the material from which the 

{5}------------------------------------------------

Koh-Cup component is made. This difference does not raise different questions of safety and effectiveness as compared to the predicate.

## Non-Clinical Performance Testing

As part of demonstrating substantial equivalence to the predicate, the following tests were performed to evaluate the new Koh-cup for safety and performance:

#### ● Design Verification Testing

- Pull-Off Test: To verify that the modified Koh-Cup does not disassemble from the device о when used as intended. The test results show that the subject Koh-Cup met the predetermined acceptance criterion.
- Compression Test: To verify that the modified Koh-Cup can withstand adequate o compression force. The test results show that the subject Koh-Cup met the predetermined acceptance criterion.

#### Design Validation Testing .

- This test, which evaluated the use of the new Koh-Cup in a cadaveric model, validated that o the Advincula Delineator with the modified Koh-Cup performs as intended and meets user needs.
- . Biocompatibility Testing per ISO 10993-1:2009
	- Cytotoxicity per ISO 10993-5:2009; results demonstrated the new Koh-Cup was non-O cytotoxic
	- Sensitization per ISO 10993-10:2010; results demonstrated the new Koh-Cup was non o sensitizing
	- Irritation per ISO 10993-10:2010; results demonstrated the new Koh-Cup was non-O irritating

#### Shelf Life Testing per ASTM F1980-07 (Reapproved 2011) ●

- Design verification testing was performed after 1 year of accelerated aging. Devices met O predetermined acceptance criteria.
## Conclusion

The Advincula Delineator Uterine Manipulator has the same intended use as the predicate device. The addition of the new Koh-Cup material does not raise different questions of safety and effectiveness, and the results of the testing described above demonstrate that the subject device with the modified Koh-cup is as safe and effective as the predicate. Therefore, the subject device is substantially equivalent to the predicate.

---

**Source:** [https://fda.innolitics.com/device/K180429](https://fda.innolitics.com/device/K180429)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
