Indigo Aspiration System (CAT RX and Separator 4)- Penumbra Engine Pump and Canister

K180412 · Penumbra, Inc. · QEW · May 15, 2018 · Cardiovascular

Device Facts

Record IDK180412
Device NameIndigo Aspiration System (CAT RX and Separator 4)- Penumbra Engine Pump and Canister
ApplicantPenumbra, Inc.
Product CodeQEW · Cardiovascular
Decision DateMay 15, 2018
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.5150
Device ClassClass 2
AttributesTherapeutic

Intended Use

INDIGO CAT RX Aspiration Catheters and INDIGO Separator 4 As part of the INDIGO Aspiration System, the INDIGO CAT RX Aspiration Catheters and Separator 4 are indicated for the removal of fresh, soft emboli and thrombi from vessels in the coronary and peripheral vasculature. INDIGO Aspiration Tubing As part of the INDIGO Aspiration System, the INDIGO Sterile Aspiration Tubing is indicated to connect the INDIGO CAT RX Aspiration Catheters to the Penumbra Aspiration Pump. Penumbra Aspiration Pump The Penumbra Aspiration Pump is indicated as a vacuum source for Penumbra Aspiration Systems.

Device Story

Indigo Aspiration System removes thrombus from coronary and peripheral vasculature via continuous aspiration. System includes dual-lumen rapid exchange CAT RX Aspiration Catheter, Separator 4, aspiration tubing, and Penumbra Engine Pump with 1000 mL canister. Physician guides catheter over guidewire to occlusion site under fluoroscopy; aspiration applied via pump to remove thrombus into canister. Separator 4 clears catheter lumen if blocked. Engine Pump is electromechanical vacuum source; operates at 100-240 Vac, 50/60 Hz; maintains vacuum for 3 hours continuous use; 500-hour operating life. Used in OR or interventional labs. Pump and canister do not contact patient. Output allows physician to clear occlusions, potentially restoring blood flow and improving patient outcomes.

Clinical Evidence

Bench testing only. No clinical data. Performance verified via electrical safety, EMC, environmental, dimensional, and 500-hour life testing. Packaging validated per ASTM D4169.

Technological Characteristics

Electromechanical vacuum pump; 100-240 Vac, 50/60 Hz. 1000 mL reservoir canister with stop-flow filter. Catheters: dual-lumen rapid exchange, fluoroscopically visible. Compliance: IEC 60601-1, IEC 60601-1-2, EN ISO 10079-1. Packaging: corrugated cardboard with foam inserts (pump), clay-coated SBS boxes (canister).

Indications for Use

Indicated for removal of fresh, soft emboli and thrombi from coronary and peripheral vessels. Used with Penumbra Aspiration System components.

Regulatory Classification

Identification

An embolectomy catheter is a balloon-tipped catheter that is used to remove thromboemboli, i.e., blood clots which have migrated in blood vessels from one site in the vascular tree to another.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. May 15, 2018 Penumbra, Inc. Michaela Mahl Senior Manager of Regulatory Affairs One Penumbra Place Alameda, California 94502 # Re: K180412 Trade/Device Name: Indigo Aspiration System (CAT RX and Separator 4) - Penumbra Engine Pump and Canister Regulation Number: 21 CFR 870.5150 Regulation Name: Embolectomy catheter Regulatory Class: Class II Product Code: DXE Dated: March 14, 2018 Received: March 15, 2018 Dear Michaela Mahl: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. M.A. Wilhelm for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K180412 Device Name Indigo Aspiration System (CAT RX Aspiration Catheter and Separator 4) - Penumbra Engine Pump and Canister Indications for Use (Describe) INDIGO CAT RX Aspiration Catheters and INDIGO Separator 4 As part of the INDIGO Aspiration System, the INDIGO CAT RX Aspiration Catheters and Separator 4 are indicated for the removal of fresh, soft emboli and thrombi from vessels in the coronary and peripheral vasculature. INDIGO Aspiration Tubing As part of the INDIGO Aspiration System, the INDIGO Sterile Aspiration Tubing is indicated to connect the INDIGO CAT RX Aspiration Catheters to the Penumbra Aspiration Pump. Penumbra Aspiration Pump The Penumbra Aspiration Pump is indicated as a vacuum source for Penumbra Aspiration Systems. Type of Use (Select one or both, as applicable) X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the word "Penumbra" in a red serif font, followed by a red circle with a white "P" inside. The "P" is stylized with a line extending from the top of the "P" to the bottom of the circle. The logo is simple and modern, and the red color gives it a sense of energy and excitement. #### 1 510(k) Summary (K180412) (as required by 21 CFR 807.92) Pursuant to Section 12, Part (a)(i)(3A) of the Safe Medical Devices Act of 1990, Penumbra, Inc. is providing the summary of Substantial Equivalence for the Indigo Aspiration System® (CAT RX and Separator 4) - Penumbra Engine™ Pump and Canister. #### 1.1 Sponsor/Applicant Name and Address Penumbra, Inc. One Penumbra Place Alameda, CA 94502 USA Phone: (510) 748-3200 #### 1.2 Sponsor Contact Information Michaela Mahl Senior Manager of Regulatory Affairs Phone: (510) 748-3288 FAX: (510) 217-6414 Email: mmahl@penumbrainc.com #### 1.3 Date of Preparation of 510(k) Summary May 14, 2018 #### 1.4 Device Trade or Proprietary Name Indigo Aspiration System® (CAT RX and Separator 4) - Penumbra Engine™ Pump and Canister #### 1.5 Device Classification | Regulatory Class: | II | |-----------------------|-------------------------------| | Classification Panel: | Cardiovascular | | Classification Name: | Catheter, Embolectomy | | Common Name: | Embolectomy Aspiration System | | Regulation Number: | 21 CFR §870.5150 | | Product Code: | DXE | #### 1.6 Predicate and Reference Devices | 510(k) Number / Clearance Date | Name of Device | Name of Manufacturer | |---------------------------------|--------------------------------------------------------------------------|---------------------------------------------------------------| | Predicate Device | | | | K163618 cleared on May 24, 2017 | Indigo Aspiration System (CAT RX<br>Aspiration Catheter and Separator 4) | Penumbra, Inc.<br>One Penumbra Place<br>Alameda, CA 94502 USA | {4}------------------------------------------------ Image /page/4/Picture/0 description: The image contains the word "Penumbra" in a red, sans-serif font. To the right of the word is a red circle with a white "P" inside. The "P" is stylized with a horizontal line extending from the right side of the "P" to the edge of the circle. | 510(k) Number / Clearance Date | Name of Device | Name of Manufacturer | |------------------------------------|-------------------|---------------------------------------------------------------| | Reference Device | | | | K122756 cleared on October 2, 2012 | Penumbra Pump MAX | Penumbra, Inc.<br>One Penumbra Place<br>Alameda, CA 94502 USA | ### 1.7 Predicate Comparison | | Predicate Device | Subject Device | |-----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------| | Trade Name | Indigo System (CAT RX and Separator 4)<br>(with Penumbra Pump MAX and Canister) | Indigo System (CAT RX and Separator 4)<br>(with Penumbra Engine Pump and<br>Canister) | | 510(k) No. | K163618 | K180412 | | Classification | Class II, DXE | SAME | | Indication for Use | INDIGO CAT RX Aspiration Catheters and<br>INDIGO Separator 4<br>As part of the INDIGO Aspiration System,<br>the INDIGO CAT RX Aspiration Catheters<br>and Separator 4 are indicated for the<br>removal of fresh, soft emboli and thrombi<br>from vessels in the coronary and peripheral<br>vasculature.<br><br>INDIGO Aspiration Tubing<br>As part of the INDIGO Aspiration System,<br>the INDIGO Sterile Aspiration Tubing is<br>indicated to connect the INDIGO CAT RX<br>Aspiration Catheters to the Penumbra<br>Aspiration Pump.<br><br>Penumbra Aspiration Pump<br>The Penumbra Aspiration Pump is indicated<br>as a vacuum source for Penumbra<br>Aspiration Systems. | SAME | | Aspiration Pump | | | | Trade Name | Penumbra Pump MAX and Canister | Penumbra Engine Pump and Canister | | IEC 60601-1<br>Compliance | Yes | SAME | | IEC 60601-1-2<br>Compliance | Yes | SAME | | Voltage | 100-115 Vac/230 Vac | 100-240 Vac | | Frequency | 50 Hz/60 Hz | SAME | | Sterilization | Non sterile | SAME | | Shelf Life | N/A | SAME | {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the logo for Penumbra, Inc., a healthcare company. The logo consists of the word "Penumbra" in a red, sans-serif font, followed by a red circle with a white "P" inside. The "P" is stylized with a thin, light blue line running through it. The logo is simple and modern, and the red color gives it a sense of energy and urgency. | | Predicate Device | Subject Device | |---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------|---------------------------------------------------------------------------------------| | Trade Name | Indigo System (CAT RX and Separator 4)<br>(with Penumbra Pump MAX and Canister) | Indigo System (CAT RX and Separator 4)<br>(with Penumbra Engine Pump and<br>Canister) | | 510(k) No. | K163618 | K180412 | | Classification | Class II, DXE | SAME | | The Aspiration Catheters, Separators, and Aspiration Tubings are unchanged and remain identical<br>(including all labelling) to those of the currently cleared Indigo System (CAT RX and Separator 4)<br>(K163618). | | | #### Device Description 1.8 The Indigo System (CAT RX and Separator 4) is designed to remove thrombus from the coronary and peripheral vessels by aspirating the proximal side of the thrombus. The intended users for this device are physicians who have received appropriate training in interventional techniques. The Indigo System (CAT RX and Separator 4) was most recently cleared under K163618. The Indigo System (CAT RX and Separator 4) is designed to remove thrombus from the vasculature using continuous aspiration. The INDIGO CAT RX Aspiration Catheter is a dual lumen rapid exchange catheter that targets aspiration from the Pump directly to the thrombus, removing it via the INDIGO Aspiration Tubing and depositing it in the Pump Canister. The INDIGO Separator 4 may be used, if needed, to clear the lumen of the INDIGO CAT RX Aspiration Catheter should it become blocked with thrombus. The INDIGO CAT RX Aspiration Catheter is introduced through a guide catheter or long introducer sheath and into the coronary or peripheral vasculature and guided over a guidewire to the site of the primary occlusion. The INDIGO CAT RX Aspiration Catheter may be provided with a rotating hemostasis valve and a peelable sheath. The INDIGO Separator 4 is provided with an introducer and torque device. The INDIGO CAT RX Aspiration Catheter and INDIGO Separator 4 are visible under fluoroscopy. The Penumbra Engine Pump and Canister The Penumbra Engine Pump ("Engine Pump") is an electromechanical device designed to create vacuum pressure for use with the Indigo System (CAT RX and Separator 4). The {6}------------------------------------------------ Image /page/6/Picture/0 description: The image shows the Penumbra logo. The word "Penumbra" is written in a red, sans-serif font. To the right of the word is a red circle with a white "P" inside. The logo is simple and modern. Engine Pump maintains vacuum pressure for 3 hours of continuous use and has an operating life of ≥ 500 hours. The pump can be used in environments with 100 - 240 Vac and 50/60 Hz. The Engine Pump is intended for use in operating rooms or interventional catheterizations laboratories. The Engine Pump and Canister do not come into contact with the patient. The Engine Pump Canister is a 1000 mL minimum volume reservoir which is an accessory to the Engine Pump. The Canister has a lid with a stop-flow filter to prevent excess fluid from entering the pump. The patient port on the lid is sized to accept the Suction Connector on the Indigo Aspiration Tubing. #### 1.9 Indications for Use ### INDIGO CAT RX Aspiration Catheters and INDIGO Separator 4 As part of the INDIGO Aspiration System, the INDIGO CAT RX Aspiration Catheters and Separator 4 are indicated for the removal of fresh, soft emboli and thrombi from vessels in the coronary and peripheral vasculature. ## INDIGO Aspiration Tubing As part of the INDIGO Aspiration System, the INDIGO Sterile Aspiration Tubing is indicated to connect the INDIGO CAT RX Aspiration Catheters to the Penumbra Aspiration Pump. ### Penumbra Aspiration Pump The Penumbra Aspiration Pump is indicated as a vacuum source for Penumbra Aspiration Systems. ## 1.10 Summary of Non-Clinical Data As required under Section 12, Part (a)(i)(3A) of the Safe Medical Devices Act of 1990, a summary of any information regarding substantial equivalence of the device follows. Included in this section are summary descriptions of the testing (or rationale for not testing if not applicable to the Engine Pump and Canister) which substantiates the performance of the subject Indigo System (CAT RX and Separator 4) with Engine Pump and Canister as well as its substantial equivalence to the predicate device: {7}------------------------------------------------ Image /page/7/Picture/0 description: The image shows the Penumbra logo. The word "Penumbra" is written in a red serif font. To the right of the word is a red circle with a white "P" inside. The logo is simple and clean, and the red and white colors are eye-catching. - Biocompatibility ● - Design Verification (Bench-Top Testing and Electrical Safety/EMC Testing) . - Shelf Life . - Sterilization ● - Packaging ● The subject Indigo System (CAT RX and Separator 4) with Engine Pump and Canister met all established requirements. ## 1.10.1 Biocompatibility The subject and predicate Indigo System (CAT RX and Separator 4) sterile devices are identical. There are no changes to the previously provided biocompatibility data of the Indigo System (CAT RX and Separator 4) materials sterile devices, which were reviewed and cleared under K163618. No additional biocompatibility testing is required or was performed for the Indigo System (CAT RX and Separator 4) sterile devices. The Indigo System (CAT RX and Separator 4) Engine Pump is a non-sterile reusable piece of capital equipment. The pump does not contact the patient, nor is it introduced into the sterile field. As such, biocompatibility testing is not required and was not performed for the pump. The Indigo System (CAT RX and Separator 4) Engine Pump Canister also does not contact the patient, nor is it introduced into the sterile field. Furthermore, blood or body fluids collected in the canister are not re-introduced to the patient. As such, biocompatibility testing is not required and was not performed for the Canister. ### 1.10.2 Design Verification (Bench-top Testing and Electrical Safety/EMC Testing) The subject and predicate Indigo System (CAT RX and Separator 4) sterile devices are identical. There are no changes to the design specifications and performance characteristics of the Indigo System (CAT RX and Separator 4) sterile devices. Therefore, all previous bench-top testing data which were reviewed and cleared under K163618 continue to support the subject Indigo System (CAT RX and Separator 4) sterile devices. No additional bench-top testing is required or was performed for the Indigo System (CAT RX and Separator 4) sterile devices. {8}------------------------------------------------ Image /page/8/Picture/0 description: The image shows the word "Penumbra" in a red, sans-serif font. To the right of the word is a red circle with a white "P" inside. The "P" is stylized with a horizontal line extending from the right side of the "P" to the edge of the circle. Bench-top testing was conducted to evaluate the physical and mechanical properties of the subject Indigo System (CAT RX and Separator 4) Engine Pump and Canister. All bench-top studies were conducted using good scientific practices and statistical sampling methods as required by the Penumbra Design Control procedures. Performance testing was based on the design specifications, risk analysis, performance standards, and guidance documents. The Indigo System (CAT RX and Separator 4) Engine Pump and Canister also underwent electrical safety and EMC testing in accordance with the requirements of IEC 60601-1 and IEC 60601-1-2. Testing was performed by Intertek, a nationally recognized test laboratory. The Engine Pump and Canister passed all tests and met all acceptance criteria. | Attribute | Sample<br>Size | Specification | Acceptance Criteria | Results | |------------------------------------------------------------|----------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------|-----------| | Pump - Electrical<br>Safety | N=1 | The Pump conforms to IEC 60601-1<br>and IEC 60601-1-2 requirements<br>including international worldwide<br>variants (CB Scheme). The Pump is<br>compliant with EN ISO 10079-1. | 100% Pass | 100% Pass | | Pump - Environmental<br>Testing | N=5 | Pump performance specifications<br>under environmental conditions | 100% Pass | 100% Pass | | Pump - Dimensional<br>Inspection | N=5 | Dimensional specifications per<br>Product Specification | 100% Pass | 100% Pass | | Pump - Inspection of<br>Design Features | N=5 | Design specifications per Product<br>Specification | 100% Pass | 100% Pass | | Pump - Performance | N=5 | Performance specifications per<br>Product Specification | 100% Pass | 100% Pass | | Pump - Performance at<br>variable voltage and<br>frequency | N=5 | Performance specifications per<br>Product Specification | 100% Pass | 100% Pass | | Pump and Canister -<br>500 hour Use Testing | N=5 | Performance specifications after 500<br>hour use | 100% Pass | 100% Pass | {9}------------------------------------------------ Image /page/9/Picture/0 description: The image shows the Penumbra logo. The word "Penumbra" is written in a red sans-serif font. To the right of the word is a red circle with a white "P" inside. | Attribute | Sample<br>Size | Specification | Acceptance Criteria | Results | |-----------------------------------------------------|----------------|---------------------------------------------------------|---------------------|-----------| | Pump Canister -<br>Inspection of Design<br>Features | N=5 | Design specifications per Product<br>Specification | 100% Pass | 100% Pass | | Pump Canister -<br>Performance | N=5 | Performance specifications per<br>Product Specification | 100% Pass | 100% Pass | | Pump and Canister -<br>Simulated Use | N=5 | Pump and Canister use<br>specifications | 100% Pass | 100% Pass | ## 1.10.3 Shelf Life The subject and predicate Indigo System (CAT RX and Separator 4) sterile devices are identical. There are no changes to the previously provided shelf life data for the sterile devices which were reviewed and cleared under K163618. No additional shelf life testing is required or was performed for the Indigo System (CAT RX and Separator 4) sterile devices. The proposed Indigo System (CAT RX and Separator 4) Engine Pump is a reusable piece of capital equipment that is provided non-sterile. Therefore, shelf life testing is not applicable to the Engine Pump and shelf life testing was not performed. The Engine Pump is established for 500 hours of use based on completed life (reliability) testing. The Engine Pump Canister does not have an established shelf life. ### 1.10.4 Sterilization The subject and predicate Indigo System (CAT RX and Separator 4) sterile devices are identical. There are no changes to the previously provided sterilization data of the devices, which were reviewed and cleared under K163618. No additional sterilization testing is required or was performed for these devices. Sterilization testing is not applicable to the proposed Indigo System (CAT RX and Separator 4) Engine Pump and Canister. Both are supplied non-sterile and are not intended to be sterilized. {10}------------------------------------------------ Image /page/10/Picture/0 description: The image shows the logo for Penumbra, Inc., a global healthcare company focused on innovative therapies. The logo features the word "Penumbra" in a red, sans-serif font, followed by a stylized "P" enclosed in a red circle. The "P" is formed by a white line, and there is a small, light blue dot where the vertical and curved lines of the "P" meet. ### 1.10.5 Packaging The packaging materials and process of the subject and predicate Indigo System (CAT RX and Separator 4) sterile devices are identical. There are no changes to the previously provided packaging material listing or the packaging process for these devices, which were reviewed and cleared under K163618. No additional packaging testing is required or was performed for the Indigo System (CAT RX and Separator 4) sterile devices. The packaging materials for the proposed Indigo System (CAT RX and Separator 4) Engine Pump and Canister are similar to those used for the predicate Indigo System (CAT RX and Separator 4) Penumbra Pump MAX and Canister. The Engine Pump and Canister are packaged to ensure that damage does not occur during shipping. The Engine Pump is packaged in a protective, corrugated, cardboard 275 B/C Flute Double Wall RSC shipping container with custom shaped foam inserts. Pumps used for Design Verification and Packaging Validation testing first underwent transportation conditioning per ASTM D4169, Distribution Cycle 3, Assurance Level 3 to ensure packaging integrity. The Canister is packaged individual in Clay coated Solid Bleach Sulfate product boxes. The Canister box is then packaged in protective, corrugated, cardboard 275 B/C Flute Double Wall RSC shipping containers (18 canister boxes per shipper). ### 1.11 Summary of Substantial Equivalence The subject Indigo System (CAT RX and Separator 4) disposable devices are unchanged and remain identical to those of the predicate Indigo System (CAT RX and Separator 4) with regard to indications, function, design, materials, biocompatibility, packaging, and sterilization. The subject Indigo System (CAT RX and Separator 4) Engine Pump and Canister are substantially equivalent to the predicate device with regard to intended use, operating principle, design concept, materials, and packaging processes.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...