K180411 · Sorin Group Italia S.R.L. · KRH · Mar 16, 2018 · Cardiovascular
Device Facts
Record ID
K180411
Device Name
MEMO 4D
Applicant
Sorin Group Italia S.R.L.
Product Code
KRH · Cardiovascular
Decision Date
Mar 16, 2018
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.3800
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
MEMO 4D device is intended for correction of mitral insufficiencies or steno-insufficiencies. The use of the Memo 4D device is indicated for correction of congenital insufficiencies with dilatation and deformation of the mitral annulus. Type I insufficiencies, with no manifest lesions in the subvalvular apparatus, can be treated with the implant of the annuloplasty ring on its own. For type II insufficiencies, characterised by valve prolapse sustained by elongation/ breakage of the chordae tendineae and papillary muscles, and type III insufficiencies, characterised by partially immobilised leaflets due to the fusion/hypertrophy of the chordae tendineae, the device implantation must be accompanied by corrective valvuloplasty.
Device Story
MEMO 4D is a sterile, semirigid annuloplasty ring used to restore mitral valve function by mechanically reshaping the annulus. Device consists of a superelastic metallic alloy core embedded in medical-grade silicone, encased in tubular knitted fabric coated with turbostratic carbon (Carbofilm). Features a removable 'ReChord System' (surgical thread loops) to assist in sizing artificial chordae during valvuloplasty. Implanted by surgeons in an OR setting during mitral valve repair. Ring provides structural support to the annulus; helps restore leaflet coaptation. Benefits patients by correcting mitral insufficiency and preventing further annular dilatation. Design includes a 'saddle' shape and varied sizes (34-42mm) to accommodate degenerative mitral disease and excess leaflet tissue.
Clinical Evidence
No clinical testing was conducted. Substantial equivalence is supported by non-clinical performance data, including computational stress analysis (FEA), tensile strength, fatigue/durability testing, biocompatibility, MRI compatibility, and simulated use testing.
Indicated for patients with congenital or acquired mitral insufficiencies or steno-insufficiencies with dilatation and deformation of the mitral annulus. Applicable to Type I (no subvalvular lesions), Type II (valve prolapse with chordae/papillary muscle elongation/breakage), and Type III (partially immobilized leaflets due to chordae fusion/hypertrophy) insufficiencies.
Regulatory Classification
Identification
An annuloplasty ring is a rigid or flexible ring implanted around the mitral or tricuspid heart valve for reconstructive treatment of valvular insufficiency.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Guidance for Annuloplasty Rings 510(k) Submissions.”
Predicate Devices
MEMO 3D ReChord Semirigid Annuloplasty Ring (K142221)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The FDA logo is composed of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The Department of Health & Human Services logo is a stylized representation of a human figure, while the FDA part includes the acronym "FDA" in a blue square and the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 16, 2018
Sorin Group Italia S.r.1. Mr. Mauro Ercolani Director, Regulatory Affairs Via Crescentino sn 13040 Saluggia, VC, Italy
Re: K180411
Trade/Device Name: MEMO 4D Semirigid Annuloplasty Ring Regulation Number: 21 CFR 870.3800 Regulation Name: Annuloplasty Ring Regulatory Class: Class II Product Code: KRH Dated: February 13, 2018 Received: February 15, 2018
Dear Mr. Ercolani:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820);
{2}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Nicole G. Ibrahim -S
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K180411
Device Name MEMO 4D Semirigid Annuloplasty Ring
### Indications for Use (Describe)
Memo 4D device is intended for correction of mitral insufficiencies or steno-insufficiencies.
The use of the Memo 4D device is indicated for correction of congenital insufficiencies with dilatation and deformation of the mitral annulus. Type I insufficiencies, with no manifest lesions in the subvalvular apparatus, can be treated with the implant of the annuloplasty ring on its own. For type II insufficiencies, characterised by valve prolapse sustained by elongation/ breakage of the chordae tendineae and papillary muscles, and type III insufficiencies, characterised by partially immobilised leaflets due to the fusion/hypertrophy of the chordae tendineae, the device implantation must be accompanied by corrective valvuloplasty.
Type of Use (Select one or both, as applicable)
| <span> <svg class="bi bi-check-square" fill="currentColor" height="1em" viewbox="0 0 16 16" width="1em" xmlns="http://www.w3.org/2000/svg"> <path d="M14 1H2a1 1 0 0 0-1 1v12a1 1 0 0 0 1 1h12a1 1 0 0 0 1-1V2a1 1 0 0 0-1-1zM2 0a2 2 0 0 0-2 2v12a2 2 0 0 0 2 2h12a2 2 0 0 0 2-2V2a2 2 0 0 0-2-2H2z" fill-rule="evenodd"></path> <path d="M10.97 4.97a.75.75 0 0 1 1.071 1.05l-3.992 4.99a.75.75 0 0 1-1.08.02L4.324 8.384a.75.75 0 1 1 1.06-1.06l2.094 2.093 3.473-4.425a.236.236 0 0 1 .02-.022z" fill-rule="evenodd"></path> </svg> </span> Prescription Use (Part 21 CFR 801 Subpart D) |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span> <svg class="bi bi-square" fill="currentColor" height="1em" viewbox="0 0 16 16" width="1em" xmlns="http://www.w3.org/2000/svg"> <path d="M14 1H2a1 1 0 0 0-1 1v12a1 1 0 0 0 1 1h12a1 1 0 0 0 1-1V2a1 1 0 0 0-1-1zM2 0a2 2 0 0 0-2 2v12a2 2 0 0 0 2 2h12a2 2 0 0 0 2-2V2a2 2 0 0 0-2-2H2z" fill-rule="evenodd"></path> </svg> </span> Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510(k) Summary
(in accordance with 21 CFR 807.92)
510(k) Number: K180411
#### l. Submitter Information
| Submitter: | Sorin Group Italia S.r.l.<br>Via Crescentino, sn<br>13040 Saluggia (VC)<br>ITALY |
|-----------------|-----------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Mauro Ercolani<br>Director, Regulatory Affairs<br>Tel: +39 0161 487 412<br>Fax: +39 0161 487 048<br>e-mail: mauro.ercolani@livanova.com |
| Date Prepared: | February 12, 2018 |
#### II. Device Name and Classification
| Proprietary Name: | MEMO 4D Semirigid Annuloplasty Ring |
|-----------------------|-------------------------------------|
| Common/Usual Name: | Ring, Annuloplasty |
| Classification Name: | Annuloplasty Ring |
| Regulation Number: | 21 CFR 870.3800 |
| Product Code: | KRH |
| Classification: | Class II |
| Classification Panel: | Cardiovascular |
#### III. Predicate Device
The MEMO 4D Semirigid Annuloplasty Ring is substantially equivalent to its cleared predicate device identified below. Both devices have the same fundamental scientific technology and intended use:
| 510(k) Number: | K142221 |
|-----------------------|---------------------------------------------|
| Proprietary Name: | MEMO 3D ReChord Semirigid Annuloplasty Ring |
| Common/Usual Name: | Ring, Annuloplasty |
| Classification Name: | Annuloplasty Ring |
| Regulation Number: | 21 CFR 870.3800 |
| Product Code: | KRH |
| Classification: | Class II |
| Classification Panel: | Cardiovascular |
{5}------------------------------------------------
#### IV. Device Description
Like its predicate device (the MEMO 3D ReChord Semirigid Annuloplasty Ring, K142221), the MEMO 4D Semirigid Annuloplasty Ring is supplied as a sterile, non-pyrogenic ring pre-mounted on a disposable holder.
Both the predicate and subject device are manufactured by embedding a superelastic metallic alloy inner core with medical grade silicone.
The resulting silicone sheath around the inner core is then encased within a tubular knitted fabric coated with a thin layer of turbostratic carbon (Carbofilm™). The fabric is then sewn along its length with a Carbofilm™ coated polyester thread.
Both the predicate and subject device feature a fully removable system (i.e., the ReChord System) in the posterior curve of the annuloplasty ring, composed by a series of loops made by a single piece of yellow surgical thread retained in place by a single piece of blue surgical thread.
The ReChord System is designed to provide a temporary reference element to facilitate the sizing of the artificial chord length at the annular plane level when performing replacement of mitral chordae tendineae in concomitance with the implant of the annuloplasty ring.
The following changes were implemented:
The MEMO 4D is available in two new sizes: 40mm and 42mm.
The MEMO 4D features a larger silicone filler layer in the outer portion of the ring, intended to further facilitate ring suturing to the annulus.
A re-shaping of the MEMO 4D annuloplasty ring was implemented for sizes 34 to 42 mm. These sizes maintain the same core cross-sectional diameter and commissure-to-commissure dimension as the predicate MEMO 3D ReChord, but feature a gradual increase of the ring antero-posterior dimension and an out-of-plane "saddle" shape in the anterior ring portion. The ring re-shaping creates a differentiated design for the MEMO 4D with respect to the MEMO 3D ReChord. The design includes a rounder shaped implant to accommodate cases of excess leaflet tissue which may be indicative of degenerative mitral disease.
Minor design changes to the disposable holder to adapt to the modified ring shape and a new injection molding supplier for the holder were also introduced. The MEMO 4D packaging system was redesigned, introducing a double nested blister to replace the polysulfone containers used for the predicate MEMO 3D ReChord.
A complete set of accessories is available separately to properly size the annulus and implant the MEMO 4D annuloplasty ring.
{6}------------------------------------------------
#### V. Indications for Use
MEMO 4D device is intended for correction of mitral insufficiencies or stenoinsufficiencies.
The use of the MEMO 4D device is indicated for correction of congenital or acquired mitral insufficiencies with dilatation and deformation of the mitral annulus. Type I insufficiencies, with no manifest lesions in the subvalvular apparatus, can be treated with the implant of the annuloplasty ring on its own. For type II insufficiencies, characterised by valve prolapse sustained by elongation/ breakage of the chordae tendineae and papillary muscles, and type III insufficiencies, characterised by partially immobilised leaflets due to the fusion/hypertrophy of the chordae tendineae, the device implantation must be accompanied by corrective valvuloplasty.
#### VI. Substantial Equivalence Discussion
Table 1 presented in the following page compares the characteristics and features of the MEMO 4D to its predicate device.
{7}------------------------------------------------
# 510(k) Summary (continued)
| MEMO 4D<br>characteristic /<br>feature | Predicate<br>Device<br>510(k) # | Predicate<br>Device<br>Proprietary<br>Name | Device Equivalence |
|-------------------------------------------------------------|---------------------------------|---------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| - Intended Use / Indications for use | | | The intended use and indications for use reported<br>in the device labeling are the same between the<br>MEMO 4D and the predicate device. |
| - Anatomical site for implantation | | | The anatomical site for implantation and surgical<br>access of the MEMO 4D are the same with<br>respect to the predicate device. |
| - Target population | | | The differences between the predicate and the<br>subject devices do not affect the disease /<br>condition that the device is intended to treat or<br>mitigate (i.e., mitral insufficiencies or steno-<br>insufficiencies, with or without repair/replacement<br>of chordae tendineae) or the target patient<br>population (i.e., patients with repairable mitral<br>valve via surgical approaches). |
| - Physical characteristics<br>- Device design and materials | K142221 | MEMO 3D<br>ReChord<br>Semirigid<br>Annuloplasty<br>Ring | Both the MEMO 3D ReChord and the MEMO 4D<br>are closed semirigid rings with variable flexibility<br>(maximum flexibility in the posterior portion,<br>gradually decreasing towards the anterior ring<br>portion) used to restore proper leaflet coaptation<br>and valve function by mechanically reshaping the<br>mitral heart valve annulus.<br><br>The MEMO 4D annuloplasty ring is identical to<br>the predicate MEMO 3D ReChord device both in<br>terms of implanted materials, device operating<br>principle and mechanism of action. |
| - Product packaging and sterilization | | | The MEMO 4D sterile barrier consists of double<br>nested polycarbonate blisters, whereas the sterile<br>barrier of the predicate device consists of double<br>nested polysulfone containers.<br><br>The materials used for the MEMO 4D packaging<br>are widely used for similar applications, and the<br>sterile barrier has been qualified to maintain the<br>device sterile over its stated shelf-life.<br><br>The MEMO 4D is steam-sterilized using the<br>same process in the same facility with respect<br>to the predicate device. |
## Table 1 – Comparison with the predicate device
{8}------------------------------------------------
#### VII. Non-Clinical Performance Data
Sorin Group Italia S.r.l. has conducted verification and validation testing of the MEMO 4D based on the risk analysis for the device.
Non-clinical testing included:
- computational stress analysis (FEA) and tensile strength testing;
- fatique and durability testing;
- extraction tests for the ReChord System threads and disposable holder; ।
- biocompatibility testing;
- MRI compatibility testing;
- simulated distribution testing;
- package integrity and microbial barrier test after accelerated ageing; ।
- simulated use testing.
Suture pull-out, corrosion resistance, and long-term biocompatibility testing were previously conducted on the predicate device MEMO 3D ReChord and remain valid for the MEMO 4D due to similar materials, design and manufacturing technologies.
#### Clinical Performance Data VIII.
No clinical testing was conducted in support of the MEMO 4D Semirigid Annuloplasty Ring, as the indications for use are equivalent to those of its predicate device MEMO 3D ReChord. These types of devices have been on the market for many years with proven safety and efficacy of use. The nonclinical testing referred to in this submission supports the substantial equivalence of this device.
#### IX. Statement of Substantial Equivalence
The MEMO 4D Semirigid Annuloplasty Ring has been demonstrated as safe and effective for its intended use as the legally marketed predicate device. With respect to intended use and technological characteristics, the MEMO 4D is substantially equivalent to the legally marketed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.