Focus DAILIES and Focus DAILIES Toric (nelfilcon A) One-Day soft contact lenses are indicated for the optical correction of refractive ametropia (myopia, hyperopia and astigmatism) in not-aphakic persons with nondiseased eyes. Focus DAILIES Progressives (nelfilcon A) One-Day soft contact lenses are indicated for daily wear for the optical correction of refractive ametropia (myperopia) and/or presbyopia in not-aphakic persons with non-diseased eyes who require a reading addition of +3.00 diopters (D) or less and who may have 2.00 diopters (D) or less of astigmatism that does not interfere with visual acuity. The lenses are to be prescribed for single use, daily disposable wear. The lenses are not intended to be cleaned or disinfected and should be discarded after a single use. DAILIES AquaComfort Plus (nelfilcon A) One-Day Contact Lenses are indicated for daily wear for the optical correction of refractive ametropia (myopia and hyperopia) in not-aphakic persons with non-diseased eyes with up to approximately 1.50 diopters (D) of astigmatism that does not interfere with visual acuity. DAILIES AquaComfort Plus Toric (nelfilcon A) One-Day Contact Lenses are indicated for daily wear for the optical correction of refractive ametropia (myopia and hyperopia) in not-aphakic persons with non-diseased eyes with 6.00 diopters (D) or less of astigmatism. DAILIES AquaComfort Plus Multifocal (nelfilcon A) One-Day Contact Lenses are indicated for daily wear for the optical correction of refractive ametropia (myperopia) and/or presbyopia in not-aphakic persons with non-diseased eyes who may require a reading addition of +3.00 diopters (D) or less and who may have 2.00 diopters (D) or less of astigmatism that does not interfere with visual acuity. All DAILIES AquaComfort Plus (nelfilcon A) One-Day Contact Lenses are to be prescribed for single use, daily disposable wear. The lenses are not intended to be cleaned or disinfected and should be discarded after a single use. FreshLook spherical (nelfilcon A) One Day Color Contact Lenses are indicated for daily wear for the optical correction of refractive ametropia (myopia, hyperopia and astigmatism) in not-aphakic persons with non-diseased eyes. The FreshLook spherical (nelfilcon A) One Day Color Contact Lenses also act to enhance or alter the apparent color of the eye. The lenses are to be prescribed for single-use Daily Disposable Wear. The lenses are not intended to be cleaned or disinfected and should be discarded after a single-use. DAILIES TOTAL (delefilcon A) spherical soft contact lenses are indicated for the optical correction of refractive ametropia (myopia and hyperopia) in phakic persons with non-diseased eves with up to approximately 1.50 diopters (D) of astigmatism. DAILIES TOTAL 1 (delefilcon A) multifocal soft contact lenses are indicated for the optical correction of refractive ametropia (myopia and hyperopia) and/or presbyopia in phakic persons with non-diseased eyes who may require a reading addition of +3.00 diopters (D) or less and who may have up to approximately 1.50 diopters of astigmatism . The lenses are to be prescribed for single use, daily disposable wear. The lenses are not intended to be cleaned or disinfected and should be discarded after a single use.
Device Story
Soft contact lenses for optical correction of refractive errors; available in spherical, toric, and multifocal designs. Materials include nelfilcon A (hydrophilic) or delefilcon A (silicone hydrogel). Lenses are single-use, daily disposable; not intended for cleaning or disinfection. Supplied sterile in foil-blister packs with buffered saline. Prescribed by eye care professionals for patient use. Benefits include vision correction and, for FreshLook, cosmetic color alteration. Modification involves updated lidding material for primary packaging.
Clinical Evidence
Bench testing only. Biocompatibility (cytotoxicity, ocular irritation, systemic toxicity) and stability testing performed per ISO 10993 and FDA guidance. No clinical data required due to similarity to predicate devices.
Technological Characteristics
Materials: nelfilcon A (PVA-based) or delefilcon A (silicone hydrogel). Manufacturing: Lightstream Technology (full mold cast). Sterilization: Moist heat (autoclave). Packaging: Polypropylene blister with polyester-coated aluminum foil lidding. Connectivity: None. Software: None.
Indications for Use
Indicated for daily wear optical correction of refractive ametropia (myopia, hyperopia, astigmatism, presbyopia) in not-aphakic or phakic persons with non-diseased eyes. Specific astigmatism limits apply by model (up to 1.50D, 2.00D, or 6.00D). Presbyopic models require reading addition of +3.00D or less. FreshLook lenses also provide eye color enhancement. Single-use, daily disposable wear.
Regulatory Classification
Identification
A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized representation of a caduceus, a symbol associated with medicine and healthcare. The FDA name is written in blue, with "U.S. FOOD & DRUG" on the top line and "ADMINISTRATION" on the bottom line.
April 10, 2018
Alcon Laboratories, Inc. Andreas Friese Regulatory Project Director, GRA Alcon Vision Care Alcon / CIBA Vision GmbH Industriering 1, Grosswallstadt, DE 63868 Bavaria
Re: K180398
Trade/Device Name: Focus DAILIES / Focus DAILIES Toric / Focus DAILIES Progressives, DAILIES AquaComfort Plus (DACP) / DACP Toric / DACP Multifocal, FreshLook One-Day, DAILIES Total1, DAILIES Total1 Multifocal Regulation Number: 21 CFR 886.5925 Regulation Name: Soft (Hydrophilic) Contact Lens Regulatory Class: Class II Product Code: LPL, MVN Dated: February 12, 2018 Received: February 13, 2018
Dear Andreas Friese:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely yours.
# J Angelo Green -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
510(k) Number (if known) K180398
Device Name Focus DAILIES, Focus DAILIES Toric, Focus DAILIES Progressives
### Indications for Use (Describe)
Focus DAILIES and Focus DAILIES Toric (nelfilcon A) One-Day soft contact lenses are indicated for the optical correction of refractive ametropia (myopia, hyperopia and astigmatism) in not-aphakic persons with nondiseased eyes.
Focus DAILIES Progressives (nelfilcon A) One-Day soft contact lenses are indicated for daily wear for the optical correction of refractive ametropia (myperopia) and/or presbyopia in not-aphakic persons with non-diseased eyes who require a reading addition of +3.00 diopters (D) or less and who may have 2.00 diopters (D) or less of astigmatism that does not interfere with visual acuity.
The lenses are to be prescribed for single use, daily disposable wear. The lenses are not intended to be cleaned or disinfected and should be discarded after a single use.
Type of Use *(Select one or both, as applicable)*
| <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> |
|-------------------------------------------------------------------------------------------------------|
| <span> <span style="font-size: 16px;">□</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{3}------------------------------------------------
510(k) Number (if known) K180398
Device Name
DAILIES AquaComfort Plus, DAILIES AquaComfort Plus Toric, DAILIES AquaComfort Plus Multifocal
### Indications for Use (Describe)
DAILIES AquaComfort Plus (nelfilcon A) One-Day Contact Lenses are indicated for daily wear for the optical correction of refractive ametropia (myopia and hyperopia) in not-aphakic persons with non-diseased eyes with up to approximately 1.50 diopters (D) of astigmatism that does not interfere with visual acuity.
DAILIES AquaComfort Plus Toric (nelfilcon A) One-Day Contact Lenses are indicated for daily wear for the optical correction of refractive ametropia (myopia and hyperopia) in not-aphakic persons with non-diseased eyes with 6.00 diopters (D) or less of astigmatism.
DAILIES AquaComfort Plus Multifocal (nelfilcon A) One-Day Contact Lenses are indicated for daily wear for the optical correction of refractive ametropia (myperopia) and/or presbyopia in not-aphakic persons with non-diseased eyes who may require a reading addition of +3.00 diopters (D) or less and who may have 2.00 diopters (D) or less of astigmatism that does not interfere with visual acuity.
All DAILIES AquaComfort Plus (nelfilcon A) One-Day Contact Lenses are to be prescribed for single use, daily disposable wear. The lenses are not intended to be cleaned or disinfected and should be discarded after a single use.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------|
| <div> <span> <span style="text-decoration: overline;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{4}------------------------------------------------
510(k) Number (if known) K180398
Device Name FreshLook One-Day
### Indications for Use (Describe)
FreshLook spherical (nelfilcon A) One Day Color Contact Lenses are indicated for daily wear for the optical correction of refractive ametropia (myopia, hyperopia and astigmatism) in not-aphakic persons with non-diseased eyes. The FreshLook spherical (nelfilcon A) One Day Color Contact Lenses also act to enhance or alter the apparent color of the eye.
The lenses are to be prescribed for single-use Daily Disposable Wear. The lenses are not intended to be cleaned or disinfected and should be discarded after a single-use.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------|------------------------------------------------------------------------------|
| <div> <span> </span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <span> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{5}------------------------------------------------
510(k) Number (if known) K180398
Device Name DAILIES TOTAL1, DAILIES TOTAL1 Multifocal
### Indications for Use (Describe)
DAILIES TOTAL (delefilcon A) spherical soft contact lenses are indicated for the optical correction of refractive ametropia (myopia and hyperopia) in phakic persons with non-diseased eves with up to approximately 1.50 diopters (D) of astigmatism.
DAILIES TOTAL 1 (delefilcon A) multifocal soft contact lenses are indicated for the optical correction of refractive ametropia (myopia and hyperopia) and/or presbyopia in phakic persons with non-diseased eyes who may require a reading addition of +3.00 diopters (D) or less and who may have up to approximately 1.50 diopters of astigmatism .
The lenses are to be prescribed for single use, daily disposable wear. The lenses are not intended to be cleaned or disinfected and should be discarded after a single use.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------|----------------------------------------------------------------------------------|
| <div> <span> ☑ Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{6}------------------------------------------------
#### K180398 510(k) SUMMARY
This 510(k) summary document has been prepared in accordance with section 21 CFR 807.92.
#### Submitter of 510(k) I.
| Company: | Alcon Laboratories, Inc.<br>6201 South Freeway<br>Fort Worth, TX 76134-2099, USA |
|-------------------------|----------------------------------------------------------------------------------|
| Primary Contact Person: | Dr. Andreas Friese<br>Regulatory Project Director, GRA Alcon Vision Care |
| Phone: | +49 6022-240-514 |
| Fax: | +49 6022-240-512 |
| Email: | andreas.friese@alcon.com |
| Back up Contact Person: | Sherri Lakota,<br>Head, GRA Alcon Vision Care |
| Phone: | 817-615-5472 |
| Fax: | 817-551-4630 |
| Email: | sherri.lakota@alcon.com |
| Date Prepared: | February 12, 2018 |
# II. Devices Subject to this 510(k)
| Trade Names: | DAILIES AquaComfort Plus soft contact lenses:<br>DAILIES AquaComfort Plus, DAILIES AquaComfort Plus<br>Toric, DAILIES AquaComfort Plus Multifocal |
|--------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|
| | Focus DAILIES soft contact lenses:<br>Focus DAILIES, Focus DAILIES Toric, Focus DAILIES<br>Progressives |
| | DAILIES Total1 soft contact lenses:<br>DAILIES Total1, DAILIES Total1 Multifocal |
| | FreshLook One-Day soft contact lenses:<br>FreshLook One-Day |
| Common Name:<br>Classification Name: | Soft Contact Lenses<br>Soft (Hydrophilic) Contact Lens |
{7}------------------------------------------------
| Device Classification: | Class II [21 CFR 886.5925] |
|------------------------|----------------------------|
| Product Code: | LPL, MVN |
# III. Predicate Device
The 510(k) devices are modifications of the same predicate devices, i.e. the currently legally commercialized devices in the US per the following most recent US FDA 510(k) clearances: K172066 (DAILIES AquaComfort Plus products), K153643 (Focus DAILIES products), K050213 (FreshLook One-Day product) and K113168 (DAILIES Total1 products).
# IV. Device Description
Focus DAILIES, DAILIES AquaComfort Plus, FreshLook One-Day and DAILIES Total1 are soft contact lenses intended for the optical correction of refractive error. They are available in spherical, toric and multifocal designs as applicable:
Spherical lenses: Focus DAILIES, DAILIES AquaComfort Plus, FreshLook One-Day and DAILIES Total1 Toric lenses: Focus DAILIES Toric and DAILIES AquaComfort Plus Toric Multifocal lenses: Focus DAILIES Progressives, DAILIES AquaComfort Plus Multifocal and DAILIES Total1 Multifocal
The lens material of Focus DAILIES, DAILIES AquaComfort Plus, and FreshLook One-Day products is nelfilcon A, a high water, non-ionic hydrophilic lens material consisting of approximately 31% PVA (polyvinyl alcohol partially acetalized with Nformylmethyl acrylamide) and 69% water. For Focus DAILIES and DAILIES AquaComfort Plus products the lens material further contains non-functionalized PVA (high- and/or ultra-high molecular weight PVA) and the color additive phthalocyanine blue to create a light blue edge to edge tint (Visitint®) to make the lenses easier to see when handling. Additionally, lenses may be printed with inks containing one or more of the following color additives: phthalocyanine blue (CFR 74.3045) and phthalocyanine green (CFR 73.3124). FreshLook One-Day color contact lenses are printed with an intermittent coating containing the following pigments (either alone or in combination): iron oxides, titanium dioxide, Cu-phthalocyanine blue, chromium oxide and phthalocyanine green. All pigments are approved color additives for use in contact lenses.
DAILIES Total1 products are made from a silicone hydrogel material containing approximately 33% water and 67% delefilcon A. Delefilcon A soft contact lenses are surface coated with hydrophilic components. The lenses contain a color additive (copper phthalocyanine blue) to assist handling tint). In addition, lenses
{8}------------------------------------------------
contain 1,2-Dimyristoyl-sn-glycero-3-phosphocholine which is considered a 'comfort agent' to improve lens wearing comfort by physical mode of action (water binding and lens lubricating effect).
Focus DAILIES, DAILIES AquaComfort Plus, FreshLook One-Day and DAILIES Total1 products are supplied sterile. The lenses immersed in buffered saline solution are packaged in individual foil-blister-packs primary packaging system and are terminally sterilized in a validated autoclave (moist heat, steam under pressure).
The blister pack primary packaging system consists of an injection molded polypropylene blister shell sealed with a polyester coated aluminum foil lidding material top. The lenses are supplied in strips of five foil sealed blister packs each containing approximately 0.65 ml (for DAILIES AquaComfort Plus and DAILIES Total1) and/or approximately 0.85 ml (for Focus DAILIES and FreshLook One-Day) of phosphate-acetate buffered saline solution. The packaging saline may contain up to 0.05% Poloxamer 108. The package saline of DAILIES AquaComfort Plus family lenses additionally contains the comfort additives hydroxypropylmethyl cellulose (HPMC) and polyethylene glycol 400 (PEG 400). Sealed blister strips are provided in secondary packaging carton boxes containing 5, 30 or 90 lenses each (Focus DAILIES, DAILIES AquaComfort Plus, DAILIES Total1) and/or 10 lenses each (FreshLook One-Day), respectively.
# V. Indications for Use
The 510(k) devices are a modification of the same predicate devices. The Indications for Use remain the same, as follows:
# DAILIES AquaComfort Plus:
DAILIES AquaComfort Plus (nelfilcon A) One-Day Contact Lenses are indicated for daily wear for the optical correction of refractive ametropia (myopia and hyperopia) in not-aphakic persons with non-diseased eyes with up to approximately 1.50 diopters (D) of astigmatism that does not interfere with visual acuity.
DAILIES AquaComfort Plus Toric (nelfilcon A) One-Day Contact Lenses are indicated for daily wear for the optical correction of refractive ametropia (myopia and hyperopia) in not-aphakic persons with non-diseased eyes with 6.00 diopters (D) or less of astigmatism.
DAILIES AquaComfort Plus Multifocal (nelfilcon A) One-Day Contact Lenses are indicated for daily wear for the optical correction of refractive ametropia (myopia or hyperopia) and/or presbyopia in not-aphakic persons with non-diseased eyes who may require a reading addition of +3.00 diopters (D) or less and who may have 2.00 diopters (D) or less of astigmatism that does not interfere with visual acuity.
{9}------------------------------------------------
All DAILIES AquaComfort Plus (nelfilcon A) One-Day Contact Lenses are to be prescribed for single use, daily disposable wear. The lenses are not intended to be cleaned or disinfected and should be discarded after a single use.
# Focus DAILIES:
Focus DAILIES and Focus DAILIES Toric (nelfilcon A) One-Day soft contact lenses are indicated for daily wear for the optical correction of refractive ametropia (myopia, hyperopia and astigmatism) in not-aphakic persons with nondiseased eyes.
Focus DAILIES Progressives (nelfilcon A) One-Day soft contact lenses are indicated for daily wear for the optical correction of refractive ametropia (myopia or hyperopia) and/or presbyopia in not-aphakic persons with non-diseased eyes who require a reading addition of +3.00 diopters (D) or less and who may have 2.00 diopters (D) or less of astigmatism that does not interfere with visual acuity.
The lenses are to be prescribed for single use, daily disposable wear. The lenses are not intended to be cleaned or disinfected and should be discarded after a single use.
## FreshLook One-Day:
FreshLook spherical (nelfilcon A) One Day Color Contact Lenses are indicated for daily wear for the optical correction of refractive ametropia (myopia, hyperopia and astigmatism) in not-aphakic persons with non-diseased eyes. The FreshLook spherical (nelfilcon A) One Day Color Contact Lenses also act to enhance or alter the apparent color of the eye.
The lenses are to be prescribed for single-use Daily Disposable Wear. The lenses are not intended to be cleaned or disinfected and should be discarded after a single-use.
## DAILIES Total1:
DAILIES TOTAL 1 (delefilcon A) spherical soft contact lenses are indicated for the optical correction of refractive ametropia (myopia and hyperopia) in phakic or aphakic persons with non-diseased eyes with up to approximately 1.50 diopters (D) of astigmatism.
DAILIES TOTAL 1 (delefilcon A) multifocal soft contact lenses are indicated for the optical correction of refractive ametropia (myopia and hyperopia) and/or presbyopia in phakic or aphakic persons with non-diseased eyes who may require a reading addition of +3.00 diopters (D) or less and who may have up to approximately 1.50 diopters of astigmatism .
The lenses are to be prescribed for single use, daily disposable wear. The lenses are not intended to be cleaned or disinfected and should be discarded after a single use.
{10}------------------------------------------------
# VI. Comparison of Technological Characteristics with the Predicate Device
The following matrix (Table 1) summarizes the characteristics of the modified devices as compared to the predicate devices:
| Element of<br>Comparison | Predicate Devices | Modified Devices |
|--------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------|
| Administrative / Regulatory Information | | |
| <b>510(k) Number</b> | K172066 (DAILIES<br>AquaComfort Plus)<br>K153643 (Focus DAILIES)<br>K050213 (FreshLook One-<br>Day)<br>K113168 (DAILIES<br>Total1) | To be assigned |
| <b>Product Name</b> | DAILIES AquaComfort<br>Plus, DAILIES<br>AquaComfort Plus Toric,<br>DAILIES AquaComfort<br>Plus Multifocal<br>Focus DAILIES, Focus<br>DAILIES Toric, Focus<br>DAILIES Progressives<br>FreshLook One-Day<br>DAILIES Total1, DAILIES<br>Total1 Multifocal | Same |
| <b>Device<br/>Classification<br/>Information</b> | Class II, Daily Wear Soft<br>(Hydrophilic) Contact Lenses,<br>21 CFR 886.5925 (b) (1) | Same |
| Indications For Use Information | | |
| <b>Intended Use</b> | One-day contact lenses for the<br>optical correction of vision.<br>Single use, daily disposable<br>wear. | Same |
| Element of<br>Comparison | Predicate Devices | Modified Devices |
| <b>Technology Information</b> | | |
| Lens Material | nelfilcon A (Focus<br>DAILIES, DAILIES<br>AquaComfort Plus and<br>FreshLook One-Day)<br>delefilcon A (DAILIES<br>Total1) | Same |
| Water Content | 69% (Focus DAILIES,<br>DAILIES AquaComfort<br>Plus and FreshLook One-<br>Day)<br>33% (DAILIES Total1) | Same |
| Visibility Tint | Focus DAILIES, DAILIES<br>AquaComfort Plus and<br>DAILIES Total1: Light<br>blue (copper phthalo-<br>cyanine blue)<br>FreshLook One-Day: No<br>tint | Same |
| Manufacturing<br>Method | Lightstream Technology: Full<br>mold cast | Same |
| Lens Designs | Spherical, toric, multifocal<br>(Focus DAILIES and<br>DAILIES AquaComfort<br>Plus)<br>Spherical (FreshLook One-<br>Day)<br>Spherical, multifocal<br>(DAILIES Total1) | Same |
| Sterilization | Steam sterilization, validated<br>autoclave | Same |
| Primary<br>Packaging<br>System in<br>General | Foil blister pack container<br>system: Polypropylene blister<br>shell sealed with a polyester<br>coated aluminum foil lidding | Same |
Table 1: Substantial Equivalence Comparison
{11}------------------------------------------------
| | Table 1: Substantial Equivalence Comparison | | | |
|--|---------------------------------------------|--|--|--|
| | | | | |
{12}------------------------------------------------
| Element of<br>Comparison | Predicate Devices | Modified Devices |
|---------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Primary<br>Packaging<br>Blister Shell | Injection-molded<br>polypropylene blister shell<br>made from Flint Hills<br>Resources P4C5N-046<br>polypropylene or Formosa<br>4142T polypropylene | Same |
| Primary<br>Packaging Foil<br>Lidding | Multi-layer laminate<br>structure polyester-coated<br>aluminum foil lidding<br>supplied by Constantia-<br>Hueck Folien<br>Current 2017 grade as<br>supplied by Constantia-<br>Hueck Folien | Multi-layer laminate<br>structure polyester-coated<br>aluminum foil lidding<br>supplied by Constantia-<br>Hueck Folien<br>Modified 2018 grades as<br>supplied by Constantia-<br>Hueck Folien |
| Package Storage<br>/ Saline Solution | Nelfilcon A products:<br>Phosphate-acetate buffered<br>saline with up to 0.05%<br>Poloxamer 108.<br>Additionally contains PEG<br>and HPMC (for DAILIES<br>AquaComfort Plus)<br>Delefilcon A products:<br>Phosphate buffered saline<br>solution with<br>approximately 0.3% of<br>polymeric wetting agents<br>consisting of copolymers of<br>polyamidoamine and poly<br>(acrylamide-acrylic acid) | Same |
{13}------------------------------------------------
| Element of<br>Comparison | Predicate Devices | Modified Devices |
|--------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------|
| Performance Specifications including any Testing | | |
| Refractive<br>Index | Nelfilcon A products: 1.38<br>Delefilcon A products: 1.42 | Same |
| Light<br>Transmittance | Nelfilcon A products:<br>92%T ( 88%T for<br>FreshLook One-Day)<br>Delefilcon A products:<br>93%T | Same |
| Oxygen<br>Permeability<br>(Dk) | Nelfilcon A products: 26<br>Delefilcon A products: 140 | Same |
| Elastic Modulus | Nelfilcon A products: 0.9<br>MPa<br>Delefilcon A products: 0.7<br>MPa | Same |
| Biocompatibility | Biocompatible as confirmed by<br>appropriate biocompatibility<br>testing | Same |
| Shelf-life | Up to 60 months as<br>confirmed by shelf-life<br>stability testing (Focus<br>DAILIES and DAILIES<br>AquaComfort Plus)<br>Up to 48 months as<br>confirmed by shelf-life<br>stability testing (FreshLook<br>One-Day, DAILIES Total1) | Same |
### Table 1: Substantial Equivalence Comparison
In accordance with the criteria for claims of substantial equivalence in the FDA guidance Premarket Notification 510(k) Guidance Document for Daily Wear Contact Lenses, May 1994, the information provided supports the claim of substantial equivalence to a lens with an existing USAN and the same manufacturing process.
{14}------------------------------------------------
# VII. Performance Data
Performance testing was conducted in consideration of the May 1994 FDA guideline titled Premarket Notification 510(k) Guidance Document for Class II Contact Lenses and applicable ISO standards for contact lenses. The following performance data are provided in support of the substantial equivalence determination.
### Biocompatibility Testing
Test results from a series of in vivo biocompatibility evaluations, including cytotoxicity, ocular irritation and systemic toxicity testing, confirm that the minor device modification does not negatively impact the safety of the devices and that the modified devices are non-toxic and biocompatible.
All biocompatibility testing was conducted in accordance with the GLP regulation (21 CFR Part 58) and relevant ISO 10993 series biocompatibility standards.
### Stability Testing
Successful stability testing supports the labeled expiration date for the modified devices.
### Clinical Testing
Based on the great similarity of the modified devices to the predicate devices and successful results from nonclinical testing, clinical testing was not required to establish substantial equivalence.
# VIII. Conclusions
The cumulative results of all performance testing demonstrate the safety, efficacy and performance of the modified devices and, thus, substantial equivalence to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.