K180394 · Synaptive Medical, Inc. · LLZ · Mar 9, 2018 · Radiology
Device Facts
Record ID
K180394
Device Name
BrightMatter Plan 1.6.0
Applicant
Synaptive Medical, Inc.
Product Code
LLZ · Radiology
Decision Date
Mar 9, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
BrightMatter Plan is indicated for: · Viewing, presentation and documentation of medical imaging, including different modules for image processing, image fusion, and image segmentation where the output can be used for image guided surgery. · Planning and simulation of cranial surgical procedures. · Reviewing of existing treatment plans. Typical users of the software are medical professionals, including but not limited to surgeons and radiologists.
Device Story
BrightMatter Plan is a standalone software workstation for pre-operative neurosurgical planning. Input: medical imaging data (including DTI). Processing: software performs automated skull segmentation, diffusion tract generation from DTI, and manual annotation of regions of interest; allows image fusion and 3D visualization. Output: surgical plan (volumetric dataset) exported to PACS for use in image-guided surgery. Used by surgeons/radiologists in office or home settings. Benefits: enables visualization of anatomical structures and tracts to assist in surgical trajectory definition and procedure simulation; improves workflow efficiency via enhanced segmentation and planning tools.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing (software verification) and design validation. Bench testing included algorithm pipeline verification, functional/unit/integration testing, and system requirements verification. Design validation confirmed conformance to clinical expectations using representative preoperative images by the intended user population.
Technological Characteristics
Standalone software workstation. Image processing, fusion, and segmentation modules. Features automated skull segmentation and DTI-based tractography generation. Supports 2D/3D visualization. Connectivity: exports data to PACS. Software-based; no specific hardware materials or energy sources defined.
Indications for Use
Indicated for medical professionals, including surgeons and radiologists, for viewing, processing, and documenting medical images, planning and simulating cranial surgical procedures, and reviewing existing treatment plans for image-guided surgery.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
March 9, 2018
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Synaptive Medical Inc. % Maham Ansari Director of Regulatory Affairs 555 Richmond Street West, Suite 800 Toronto, Ontario M5V 3B1 CANADA
Re: K180394
Trade/Device Name: BrightMatter Plan 1.6.0 Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: February 12, 2018 Received: February 13, 2018
Dear Maham Ansari:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and
Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Michael D. O'Hara For
Robert Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K180394
Device Name BrightMatter Plan 1.6.0
Indications for Use (Describe) BrightMatter Plan is indicated for:
· Viewing, presentation and documentation of medical imaging, including different modules for image processing, image fusion, and image segmentation where the output can be used for image guided surgery.
· Planning and simulation of cranial surgical procedures.
· Reviewing of existing treatment plans.
Typical users of the software are medical professionals, including but not limited to surgeons and radiologists.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 5. 510(k) Summary
[As required by 21 CFR 807.92]
| Manufacturer:<br>Address: | Synaptive Medical Inc.<br>555 Richmond Street West, Suite 800<br>Toronto, ON M5V 3B1<br>Canada |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Establishment Registration: | 3012075008 |
| Contact Name:<br>Title:<br>Phone Number:<br>Fax Number:<br>Date Prepared: | Ms. Maham Ansari, MS, RAC<br>Director, Regulatory Affairs<br>+1 647 925 3435<br>1 888 650 5230<br>12 February 2018 |
| Device Proprietary Name:<br>Device Common or Usual Name:<br>Classification Name:<br>Product Code:<br>Regulation Number:<br>Regulation Class:<br>Regulation Description: | BrightMatter Plan<br>BrightMatter Plan<br>System, Image Processing, Radiological<br>LLZ<br>21 CFR 892.2050<br>II<br>Picture Archiving and Communications<br>System |
#### Predicate Devices
Substantial equivalence is claimed to the following device:
| Trade name | Manufacturer | 510(k) Number | Date Cleared |
|---------------------------------------|---------------------------|---------------|---------------|
| BrightMatter Planning<br>Software 1.0 | Synaptive Medical<br>Inc. | K140337 | June 02, 2014 |
### Device Description
BrightMatter Plan is a treatment planning software that enables the user to view and process medical image data. The software is intended for pre-operative planning of neuro-surgical treatments based on image guided surgical systems. The planning software system provides the ability to visualize diagnostic images in 2D and 3D formats and fusion of image datasets. The software automatically segments the skull from the acquired image and generates diffusion tracts from Diffusion Tensor Imaging (DTI) data. The user can also manually annotate regions of interest, resulting in structures which can subsequently be visualized in 3D. A trained person can use the software to segment structures, define regions of interest and establish one or more trajectories.
{4}------------------------------------------------
The software, operated on a stand-alone computer workstation, is expected to be used by a Clinician in an office or home setting, in preparation for one of several possible surgical procedures. The end of the processing is a surgical plan which can be exported to a Picture Archiving and Communication Systems (PACS) for subsequent use in image guided surgery.
### Indications for Use
BrightMatter Plan is indicated for:
- . Viewing, presentation and documentation of medical imaging, including different modules for image processing, image fusion, and image segmentation where the output can be used for image guided surgery.
- . Planning and simulation of cranial surgical procedures.
- Reviewing of existing treatment plans.
Typical users of the software are medical professionals, including but not limited to surgeons and radiologists.
### Summary of Technological Comparisons
BrightMatter Plan is an updated version of the predicate: its previously cleared version 1.0 (K140337). The subject device is substantially unchanged from the predicate with respect to indications and performance claims. The primary difference between the proposed and predicate devices is additional or altered functionality which improve the planning workflow or the user experience of the device.
Key differences and advantages compared to the predicate are:
- 1. BrightMatter Plan incorporates numerous improvements to usability.
- 2. The new software enables manual segmentation of tractography, in a manner similar to segmenting anatomical structures. This enables the planner to deliberately reduce the amount of information displayed during planning to the most useful elements.
- 3. The software exports a volumetric dataset, incorporating the planned target and engagement points. This enables more efficient translation of the plan to a PACS.
- 4. Removal of sulcal paths workflow stage since it cannot be utilized in commercially available navigation systems and was consequently not used in practice.
There are no known disadvantages with the proposed device in comparison to the predicate device.
The essential underlying technology used to process images is the same. The intended uses are equivalent. Direct comparison of major functionality showed largely identical results. Any changes introduced in the subject device do not raise new concerns of safety and effectiveness.
{5}------------------------------------------------
### Non-Clinical Testing
The following bench (software verification) testing was conducted on BrightMatter Plan:
- Performance testing – algorithm pipeline verification; functional verification
- Unit level verification ●
- Integration verification ●
- . Implementation requirements verification and system integration - release testing, compatibility testing, testing of resolved anomalies, and platform testing
- . System requirements verification – review of traceability from systems requirements to implementation; labeling reviews.
### Design Validation
Device (software) was tested by the intended user population using representative preoperative images and evaluation of resulting plan for conformance to clinical expectations with production equivalent device. This was conducted to validate that the customer needs have been satisfied.
Effectiveness of risk control measures related to usability/human factors were also evaluated and validated via this activity.
### Clinical Testing
This technology is not new; therefore, a clinical study was not considered necessary prior to release. Additionally, there was no clinical testing required to support the medical device as the indications for use are equivalent to the predicate device. The substantial equivalence of the device is supported by the non-clinical testing.
#### Conclusion
BrightMatter Plan has been shown through comparison and testing to be substantially equivalent to the identified predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.