← Product Code [DQA](/productcode/DQA) · K180380

# vital signs monitor (K180380)

_Edan Instruments, Inc. · DQA · Dec 21, 2018 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K180380

## Device Facts

- **Applicant:** Edan Instruments, Inc.
- **Product Code:** [DQA](/productcode/DQA.md)
- **Decision Date:** Dec 21, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.2700
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Pediatric

## Indications for Use

The monitor is intended to be used for monitoring, storing, recording, and reviewing of, and to generate alarms for, multiple physiological parameters of adults, pediatrics and neonates. The monitor is intended for use by trained healthcare professionals in hospital environments. Monitored parameters include: NIBP, SpO2, PR (pulse rate), Quick TEMP/Infrared TEMP. The monitor is not intended for MRI environments. TEMP module is not intended for neonates.

## Device Story

iM3 vital signs monitor; measures NIBP, SpO2, PR, and temperature (Quick/Infrared). Operates in Monitor, Spot, and Round modes; Monitor mode provides continuous monitoring/alarming; Spot/Round modes facilitate quick checks. Used in hospital environments by trained healthcare professionals. Connects to EDAN MFM-CMS for remote display. Data storage includes trends, NIBP reviews, and waveforms. Provides visual/audible alarms to assist clinical decision-making; supports patient care management. Benefits include efficient vital sign tracking and centralized monitoring.

## Clinical Evidence

Bench testing only. No clinical data presented. Performance verified via electrical safety (ANSI/AAMI ES60601-1), EMC (IEC 60601-1-2), and adherence to specific parameter standards (ISO 80601-2-61 for SpO2, ISO 80601-2-56 for TEMP, IEC 80601-2-30 for NIBP). Software verification and validation conducted per FDA guidance.

## Technological Characteristics

Class II device; 8-inch color TFT touch screen. Parameters: NIBP (oscillation), SpO2, PR, TEMP (Quick/Infrared). Connectivity: USB, Network, Nurse call, Bluetooth LE 4.0, Wi-Fi (802.11b/g/n). Power: 100-240V AC or rechargeable battery. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, ISO 10993-1, IEC 62304. Sterilization: Surface cleaning/disinfection with neutral detergent, ethanol, or isopropanol.

## Regulatory Identification

An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.

## Predicate Devices

- Edan Instruments, Inc, Vital Signs Monitor Model M3 ([K131818](/device/K131818.md))

## Reference Devices

- Edan Instruments, Inc, Patient Monitor Model V5 V6 V8 ([K160981](/device/K160981.md))
- MFM-CMS Central Monitoring System ([K120727](/device/K120727.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the U.S. Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and "ADMINISTRATION" in a smaller font size below.

December 21, 2018

Edan Instruments, Inc. % Alice Yang Regulatory Engineer Edan Medical 1200 Crossman Ave. Suite 200 Sunnyvale, California 94089

Re: K180380

Trade/Device Name: iM3 vital signs monitor Regulation Number: 21 CFR 870.1130 Regulation Name: Cardiac monitor (including cardiotachometer and rate alarm) Regulatory Class: Class II Product Code: MWI, DQA, DXN, FLL Dated: November 21, 2018 Received: November 27, 2018

Dear Alice Yang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Stephen C. Browning -S5

for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

## Indications for Use

510(k) Number (if known)

#### K180380

Device Name iM3 vital signs monitor

Indications for Use (Describe)

The monitor is intended to be used for monitoring, storing, and reviewing of, and to generate alarms for, multiple physiological parameters of adults, pediatrics and neonates. The monitor is intended for use by trained healthcare professionals in hospital environments.

Monitored parameters include: NIBP, SpO2, PR (pulse rate), Quick TEMP/Infrared TEMP.

The monitor is not intended for MRI environments. TEMP module is not intended for neonates.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response. including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

{3}------------------------------------------------

# 510(K) Summary

# Prepared in accordance with the content and format regulatory requirements of 21 CFR Part 807.92

| 1. Submitter:                      | Edan Instruments, Inc.                                                                                                                                                                                                                 |
|------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                                    | #15 Jinhui Road, Jinsha Community,                                                                                                                                                                                                     |
|                                    | Kengzi Sub-District, Pingshan District,                                                                                                                                                                                                |
|                                    | Shenzhen, 518122 P.R.China.                                                                                                                                                                                                            |
|                                    | Tel: +86(0755) 26858736                                                                                                                                                                                                                |
|                                    | Fax: +1 (408) 418-4059                                                                                                                                                                                                                 |
| Contact person:                    | Alice Yang                                                                                                                                                                                                                             |
| Preparing date:                    | December 21, 2018                                                                                                                                                                                                                      |
| 2. Device name and classification: | Device Name: Vital Signs Monitor                                                                                                                                                                                                       |
|                                    | Model: iM3                                                                                                                                                                                                                             |
|                                    | Classification Name/ Product code:                                                                                                                                                                                                     |
|                                    | 21 CFR 870.2300 Cardiac monitor (including cardiotachometer and rate alarm)/MWI                                                                                                                                                        |
|                                    | 21 CFR 870.1130 Noninvasive blood pressure measurement system/DXN                                                                                                                                                                      |
|                                    | 21 CFR 870.2700, Oximeter/DQA                                                                                                                                                                                                          |
|                                    | 21 CFR 880.2910 Clinical electronic thermometer/FLL                                                                                                                                                                                    |
|                                    | Regulatory Class: Class II                                                                                                                                                                                                             |
| 3. Predicate Device(s):            | 1) Edan Instruments, Inc, Vital Signs Monitor Model M3, K131818(Primary)                                                                                                                                                               |
|                                    | 2) Edan Instruments, Inc, Patient Monitor Model V5 V6 V8, K160981 (Reference)                                                                                                                                                          |
| 4. Device Description:             | The iM3 has three work modes, Monitor, Spot and Round, to measure physiological parameters, including non-invasive blood pressure (NIBP), oxygen saturation of the blood (SpO2), pulse rate (PR), and Quick TEMP/Infrared TEMP (TEMP). |
|                                    | • For the Monitor mode, the user can do continuous measurement, monitoring, alarming and manager patient data.                                                                                                                         |
|                                    | • For the Spot mode, also call spot check mode, the user can measure parameters quickly without creating a patient within the instrument workflow.                                                                                     |

- The Round mode is designed to support hospital staff when doing ward .

{4}------------------------------------------------

rounds. The Round mode is similar as Spot mode. The difference between Round mode and Spot mode is that for one ward round, the iM3 can additionally store one ward round record automatically (the latest measurement) or manually (user can choose the measure results).

iM3 capabilities include storing, displaying measuring data. When necessary, alarms will be produced so that doctors and nurses can manage patient care appropriately.

The iM3can connect with EDAN's MFM-CMS Central Monitoring System to display the iM3 data on a remote work station. The MFM-CMS system received FDA 510(k) clearance on June 21, 2013(K120727)

Not all sensors are intended for neonatal patients. The intended Note: population for each sensor is detailed in the section 6 below.

- 5. Indication for Use The monitor is intended to be used for monitoring, storing, recording, and reviewing of, and to generate alarms for, multiple physiological parameters of adults, pediatrics and neonates. The monitor is intended for use by trained healthcare professionals in hospital environments. Monitored parameters include: NIBP, SpO2, PR (pulse rate), Quick TEMP/Infrared TEMP. The monitor is not intended for MRI environments. TEMP module is not intended for neonates.
### 6. Predicate Device Comparison

The subject device shares the same characteristics in most items with the predicate device except in the following aspects:

| Item         | <Subject Device>                                                                                                                                                                            | <Predicate Device>                                                                                                                                                                                  | <Predicate Device>                                                                                                                                                                                                                |
|--------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|              | iM3                                                                                                                                                                                         | M3                                                                                                                                                                                                  | V5/V6/V8                                                                                                                                                                                                                          |
| K#           | Current Submission                                                                                                                                                                          | K131818                                                                                                                                                                                             | K160981                                                                                                                                                                                                                           |
| Intended Use |                                                                                                                                                                                             |                                                                                                                                                                                                     |                                                                                                                                                                                                                                   |
| Description  | The monitor is intended to be<br>used for monitoring, storing,<br>recording, and reviewing of,<br>and to generate alarms for,<br>multiple physiological<br>parameters of adults, pediatrics | The monitor is intended to be<br>used by qualified physicians or<br>personnel professionally<br>trained<br>and it is for monitoring adults,<br>pediatrics and neonates in<br>hospital environments. | The monitors are intended to be<br>used for monitoring, storing,<br>and reviewing of, and to<br>generate alarms for, multiple<br>physiological parameters of<br>adults, pediatrics and neonates.<br>The monitors are intended for |

Table 1Comparison to predicates

{5}------------------------------------------------

K180380 Page 3 of 11

|                                        | and neonates. The monitor is<br>intended for use by trained<br>healthcare professionals in<br>hospital environments.<br>Monitored parameters include:<br>NIBP, SpO2, PR (pulse rate),<br>Quick TEMP/Infrared TEMP.<br>The monitor is not intended<br>for MRI environments. TEMP<br>module is not intended for<br>neonates. | This monitor is used to<br>monitor vital signals for<br>patients and is suitable for use<br>in hospital<br>environments including<br>out-patient department, wards<br>and NICU.<br>Monitored parameters include:<br>NIBP, SpO2, pulse rate, Quick<br>TEMP/Infrared TEMP. | use by trained healthcare<br>professionals in hospital<br>environments.<br>The monitored physiological<br>parameters include: ECG,<br>respiration (RESP), temperature<br>(TEMP), oxygen saturation of<br>arterial blood (SpO2), pulse rate<br>(PR), non-invasive blood<br>pressure (NIBP), invasive blood<br>pressure (IBP), carbon dioxide<br>(CO2), cardiac output (C.O.),<br>anesthetic gas (AG), bispectral<br>index (BIS), respiration<br>mechanics (RM) and impedance<br>cardiography (ICG). |
|----------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended patient<br>population         | adult, pediatric and neonatal<br>patients                                                                                                                                                                                                                                                                                  | adult, pediatric and neonatal<br>patients                                                                                                                                                                                                                                | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Intended<br>application<br>environment | Hospital environment.                                                                                                                                                                                                                                                                                                      | Hospital environment.                                                                                                                                                                                                                                                    | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |

# NIBP monitor (EDAN Module)

| Principle of<br>Operation                                                                | Oscillation                                                                                                                                                                                                                                              | Oscillation                                                                                                                                                                                                                                              |                                                                                 |
|------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------|
| Measurement<br>Range                                                                     | Adult<br>Systolic 40 to 270 mmHg<br>Diastolic 10 to 215 mmHg<br>Mean 20 to 235 mmHg<br><br>Pediatric<br>Systolic 40 to 230 mmHg<br>Diastolic 10 to 180 mmHg<br>Mean 20 to 195 mmHg<br><br>Neonate<br>Systolic 40 to 135 mmHg<br>Diastolic 10 to 100 mmHg | Adult<br>Systolic 40 to 270 mmHg<br>Diastolic 10 to 215 mmHg<br>Mean 20 to 235 mmHg<br><br>Pediatric<br>Systolic 40 to 200 mmHg<br>Diastolic 10 to 150 mmHg<br>Mean 20 to 165 mmHg<br><br>Neonate<br>Systolic 40 to 135 mmHg<br>Diastolic 10 to 100 mmHg |                                                                                 |
|                                                                                          |                                                                                                                                                                                                                                                          |                                                                                                                                                                                                                                                          |                                                                                 |
|                                                                                          | Mean 20 to 110 mmHg                                                                                                                                                                                                                                      | Mean 20 to 110 mmHg                                                                                                                                                                                                                                      |                                                                                 |
| Accuracy                                                                                 | Maximum average error: ±5 mmHg                                                                                                                                                                                                                           | Maximum average error: ±5 mmHg                                                                                                                                                                                                                           |                                                                                 |
|                                                                                          | Maximum standard deviation: 8 mmHg                                                                                                                                                                                                                       | Maximum standard deviation: 8 mmHg                                                                                                                                                                                                                       |                                                                                 |
|                                                                                          |                                                                                                                                                                                                                                                          |                                                                                                                                                                                                                                                          |                                                                                 |
| Resolution                                                                               | 1 mmHg                                                                                                                                                                                                                                                   | 1 mmHg                                                                                                                                                                                                                                                   |                                                                                 |
| Maximum<br>measuring period                                                              | Adult/Pediatric 120s<br>Neonate 90s                                                                                                                                                                                                                      | Adult/Pediatric 120s<br>Neonate 90s                                                                                                                                                                                                                      |                                                                                 |
|                                                                                          | Adult 297±3 mmHg                                                                                                                                                                                                                                         | Adult 297±3 mmHg                                                                                                                                                                                                                                         |                                                                                 |
| Overpressure<br>protection                                                               | Pediatric 245±3 mmHg                                                                                                                                                                                                                                     | Pediatric 240±3 mmHg                                                                                                                                                                                                                                     |                                                                                 |
|                                                                                          | Neonate 147±3 mmHg                                                                                                                                                                                                                                       | Neonate 147±3 mmHg                                                                                                                                                                                                                                       |                                                                                 |
| Mode                                                                                     | Manual, Auto, Continuous                                                                                                                                                                                                                                 | Manual, Auto, Continuous                                                                                                                                                                                                                                 |                                                                                 |
| Measuring interval<br>in AUTO Mode                                                       | 1/2/3/4/5/10/15/30/60/90/120/<br>180/240/360/480 mins                                                                                                                                                                                                    | 1/2/3/4/5/10/15/30/60/90/120/<br>240/480 mins                                                                                                                                                                                                            |                                                                                 |
| Continuous                                                                               | 5min                                                                                                                                                                                                                                                     | 5min                                                                                                                                                                                                                                                     |                                                                                 |
| PR from NIBP                                                                             |                                                                                                                                                                                                                                                          |                                                                                                                                                                                                                                                          |                                                                                 |
| Measurement range                                                                        | 40 to 240 bpm                                                                                                                                                                                                                                            | 40 bpm to 240 bpm                                                                                                                                                                                                                                        |                                                                                 |
| Accuracy                                                                                 | ±3 bpm or ±3.5%,<br>whichever is greater                                                                                                                                                                                                                 | ±3 bpm or 3.5%,<br>whichever is greater                                                                                                                                                                                                                  |                                                                                 |
|                                                                                          |                                                                                                                                                                                                                                                          |                                                                                                                                                                                                                                                          |                                                                                 |
| Resolution                                                                               | 1 bpm                                                                                                                                                                                                                                                    | 1 bpm                                                                                                                                                                                                                                                    |                                                                                 |
| NIBP monitor(Suntech Module)                                                             |                                                                                                                                                                                                                                                          |                                                                                                                                                                                                                                                          |                                                                                 |
| Principle of<br>Operation                                                                | oscillation                                                                                                                                                                                                                                              |                                                                                                                                                                                                                                                          | Same                                                                            |
|                                                                                          |                                                                                                                                                                                                                                                          |                                                                                                                                                                                                                                                          |                                                                                 |
|                                                                                          | Adult<br>SYS: 40 mmHg ~ 260 mmHg<br>DIA: 20 mmHg ~ 200 mmHg<br>MAP: 26 mmHg- 220 mmHg                                                                                                                                                                    |                                                                                                                                                                                                                                                          | Same                                                                            |
|                                                                                          | Pediatric<br>SYS: 40 mmHg- 230 mmHg<br>DIA: 20 mmHg- 160 mmHg<br>MAP: 26 mmHg– 183 mmHg                                                                                                                                                                  |                                                                                                                                                                                                                                                          |                                                                                 |
| Measurement<br>Range                                                                     | Neonate<br>SYS: 40 mmHg – 130 mmHg<br>DIA: 20 mmHg– 100 mmHg<br>MAP: 26 mmHg – 110 mmHg                                                                                                                                                                  |                                                                                                                                                                                                                                                          |                                                                                 |
|                                                                                          |                                                                                                                                                                                                                                                          |                                                                                                                                                                                                                                                          |                                                                                 |
| Accuracy                                                                                 | Maximum average error:                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                          | Same                                                                            |
|                                                                                          |                                                                                                                                                                                                                                                          |                                                                                                                                                                                                                                                          |                                                                                 |
|                                                                                          | ±5mmHg                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                          |                                                                                 |
|                                                                                          | Maximum standard deviation:                                                                                                                                                                                                                              |                                                                                                                                                                                                                                                          |                                                                                 |
|                                                                                          | 8mmHg                                                                                                                                                                                                                                                    |                                                                                                                                                                                                                                                          |                                                                                 |
| Resolution                                                                               | 1mmHg                                                                                                                                                                                                                                                    |                                                                                                                                                                                                                                                          | Same                                                                            |
|                                                                                          | Adult 130s                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                          | Same                                                                            |
| Maximum<br>measuring period                                                              | Pediatric 120s                                                                                                                                                                                                                                           |                                                                                                                                                                                                                                                          |                                                                                 |
|                                                                                          | Neonate 75s                                                                                                                                                                                                                                              |                                                                                                                                                                                                                                                          |                                                                                 |
| Overpressure<br>protection                                                               | Adult /Pediatric <300mmHg                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                          | Same                                                                            |
|                                                                                          | Neonate 150mmHg                                                                                                                                                                                                                                          |                                                                                                                                                                                                                                                          |                                                                                 |
| Mode                                                                                     | Manual, Auto, Continuous,<br>Average                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                          | Manual, Auto, Continuous                                                        |
| Measuring interval<br>in AUTO Mode                                                       | 1/2/3/4/5/10/15/30/60/90/120/<br>180/240/360/480 mins                                                                                                                                                                                                    |                                                                                                                                                                                                                                                          | 1/2/3/4/5/10/15/30/60/90/120/<br>240/480 mins                                   |
| Continuous                                                                               | 5min                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                          | Same                                                                            |
| PR from NIBP                                                                             |                                                                                                                                                                                                                                                          |                                                                                                                                                                                                                                                          | Same                                                                            |
| Measurement range                                                                        | 30 bpm to 220 bpm                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                          | Same                                                                            |
| Accuracy                                                                                 | ±3 bpm or 2%,<br>whichever is greater                                                                                                                                                                                                                    |                                                                                                                                                                                                                                                          | Same                                                                            |
| SpO2 monitor (EDAN Module)                                                               |                                                                                                                                                                                                                                                          |                                                                                                                                                                                                                                                          |                                                                                 |
| Measurement<br>Range                                                                     | SpO2 0-100%<br>Pulse Rate 25 to 300 bpm                                                                                                                                                                                                                  | SpO2 0-100%<br>Pulse Rate 25 to 300 bpm                                                                                                                                                                                                                  |                                                                                 |
| Accuracy                                                                                 | <b>Saturation</b><br>Adult/pediatric, non-motion<br>conditions<br>70 to 100% ±2 %<br>0-69% unspecified<br>Neonate<br>70 to 100% ±3%<br>0-69% unspecified<br><b>Pulse Rate</b><br>Adult and Neonate<br>25 to 300 bpm ±2bpm<br>(non-motion conditions)     | <b>Saturation</b><br>Adult/pediatric, non-motion<br>conditions<br>70 to 100% ±2 %<br>0-69% unspecified<br>Neonate<br>70 to 100% ±3%<br>0-69% unspecified<br><b>Pulse Rate</b><br>Adult and Neonate<br>25 to 300 bpm ±2bpm<br>(non-motion conditions)     |                                                                                 |
| Resolution                                                                               | SpO2 1%<br>Pulse Rate 1 bpm                                                                                                                                                                                                                              | SpO2 1%<br>Pulse Rate 1 bpm                                                                                                                                                                                                                              |                                                                                 |
| Sensor                                                                                   | Emitted Light Energy: ≤15 mW                                                                                                                                                                                                                             | Emitted Light Energy: ≤15 mW                                                                                                                                                                                                                             |                                                                                 |
| SpO2 monitor (NELLCOR Module)                                                            |                                                                                                                                                                                                                                                          |                                                                                                                                                                                                                                                          |                                                                                 |
| Optional Spo2                                                                            | NELLCOR SpO2 module                                                                                                                                                                                                                                      | NELLCOR SpO2 module                                                                                                                                                                                                                                      |                                                                                 |
| module                                                                                   |                                                                                                                                                                                                                                                          |                                                                                                                                                                                                                                                          |                                                                                 |
| TEMP<br>monitor(EDAN<br>quick temp<br>module)                                            |                                                                                                                                                                                                                                                          |                                                                                                                                                                                                                                                          |                                                                                 |
| Measuring range                                                                          | Monitor mode: 25 °C ~45 °C<br>Predict mode: 35.5 °C ~42 °C                                                                                                                                                                                               | Monitor mode: 25 °C ~45 °C<br>Predict mode: 35.5 °C ~42 °C                                                                                                                                                                                               |                                                                                 |
| Working<br>temperature                                                                   | 10°C ~ 40°C                                                                                                                                                                                                                                              | 10°C ~ 40°C                                                                                                                                                                                                                                              |                                                                                 |
| Sensor type                                                                              | Oral /axillary /rectal                                                                                                                                                                                                                                   | Oral /axillary /rectal                                                                                                                                                                                                                                   |                                                                                 |
| Adjustable range of<br>alarm limits                                                      | 35.5 °C ~42 °C                                                                                                                                                                                                                                           | 35.5 °C ~42 °C                                                                                                                                                                                                                                           |                                                                                 |
| Resolution                                                                               | 0.1 °C                                                                                                                                                                                                                                                   | 0.1 °C                                                                                                                                                                                                                                                   |                                                                                 |
| Accuracy                                                                                 | Monitor mode: ±0.1 °C<br>(25 °C ~ 45 °C )                                                                                                                                                                                                                | Monitor mode: ±0.1 °C<br>(25 °C ~ 45 °C )                                                                                                                                                                                                                |                                                                                 |
| Response time                                                                            | < 60 s                                                                                                                                                                                                                                                   | < 60 s                                                                                                                                                                                                                                                   |                                                                                 |
| Update time                                                                              | 1 s~ 2 s                                                                                                                                                                                                                                                 | 1 s~ 2 s                                                                                                                                                                                                                                                 |                                                                                 |
| Warm-up time                                                                             | <10 s                                                                                                                                                                                                                                                    | <10 s                                                                                                                                                                                                                                                    |                                                                                 |
| Time for predicting                                                                      | <30 s                                                                                                                                                                                                                                                    | <30 s                                                                                                                                                                                                                                                    |                                                                                 |
| Calibration                                                                              | Self-test interval: ≤5 minutes                                                                                                                                                                                                                           | Self-test interval: ≤5 minutes                                                                                                                                                                                                                           |                                                                                 |
| Clinical Bias                                                                            | (-0.2 to -0.4 )°C                                                                                                                                                                                                                                        | (-0.2 to -0.4 )°C                                                                                                                                                                                                                                        |                                                                                 |
| Limits of<br>Agreement                                                                   | 0.49                                                                                                                                                                                                                                                     | 0.49                                                                                                                                                                                                                                                     |                                                                                 |
| Stability                                                                                | 0.14 °C                                                                                                                                                                                                                                                  | 0.14 °C                                                                                                                                                                                                                                                  |                                                                                 |
| TEMP monitor(infrared ear temp module)                                                   |                                                                                                                                                                                                                                                          |                                                                                                                                                                                                                                                          |                                                                                 |
| Measuring range                                                                          | 34 °C~ 42.2 °C                                                                                                                                                                                                                                           | 34 °C~ 42.2 °C                                                                                                                                                                                                                                           |                                                                                 |
| Working<br>temperature                                                                   | 10 °C~ 40 °C                                                                                                                                                                                                                                             | 10 °C~ 40 °C                                                                                                                                                                                                                                             |                                                                                 |
| Adjustable range of<br>alarm limits                                                      | 35.5 °C~ 42 °C                                                                                                                                                                                                                                           | 35.5 °C~ 42 °C                                                                                                                                                                                                                                           |                                                                                 |
| Resolution                                                                               | 0.1 °C                                                                                                                                                                                                                                                   | 0.1 °C                                                                                                                                                                                                                                                   |                                                                                 |
| Response time                                                                            | 1 s                                                                                                                                                                                                                                                      | 1 s                                                                                                                                                                                                                                                      |                                                                                 |
| Measuring Mode                                                                           | Adjusted Mode                                                                                                                                                                                                                                            | Adjusted Mode                                                                                                                                                                                                                                            |                                                                                 |
| Clinical Accuracy                                                                        | ±0.2 C (0.4 °F) (35.5 °C<br>~42 °C ) (95°F ~107.6°F)<br><br>±0.3 C (0.5°F) (out of the<br>range mentioned above)                                                                                                                                         | ±0.2 C (0.4 °F) (35.5 °C<br>~42 °C ) (95°F ~107.6°F)<br><br>±0.3 C (0.5°F) (out of the<br>range mentioned above)                                                                                                                                         |                                                                                 |
| Response time                                                                            | 1 s                                                                                                                                                                                                                                                      | 1 s                                                                                                                                                                                                                                                      |                                                                                 |
|                                                                                          |                                                                                                                                                                                                                                                          |                                                                                                                                                                                                                                                          |                                                                                 |
| Accuracy                                                                                 |                                                                                                                                                                                                                                                          |                                                                                                                                                                                                                                                          |                                                                                 |
|                                                                                          | TEMP monitor(Covidien F3000 quick temp module)                                                                                                                                                                                                           |                                                                                                                                                                                                                                                          |                                                                                 |
| Measuring range                                                                          | 30°C~43°C                                                                                                                                                                                                                                                | 30°C~43°C                                                                                                                                                                                                                                                |                                                                                 |
| Prediction<br>measurement range                                                          | 35°C~43°C                                                                                                                                                                                                                                                | 35°C~43°C                                                                                                                                                                                                                                                |                                                                                 |
| Cold mode                                                                                |                                                                                                                                                                                                                                                          |                                                                                                                                                                                                                                                          |                                                                                 |
| prediction                                                                               | 33°C~43°C                                                                                                                                                                                                                                                | 33°C~43°C                                                                                                                                                                                                                                                |                                                                                 |
| measurement range                                                                        |                                                                                                                                                                                                                                                          |…

---

**Source:** [https://fda.innolitics.com/device/K180380](https://fda.innolitics.com/device/K180380)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
