1. Pulpotomy 2. Temporary Filling Material 3. Repair of Root Perforations 4. Repair of Furcation Perforations 5. Repair of Perforating Internal and External Resorptions 6. Root-End Filling in Endodontic Surgery 7. Pulp Exposures (Direct Pulp Capping) 8. Protective Liner (Indirect Pulp Capping) and Base for Use Under a Variety of Substrates
Device Story
TheraCal DC is a biocompatible, dual-cured, resin-modified calcium silicate material used by dentists for treating damaged dentin in crowns and roots. It is applied during endodontic procedures and deep cavity preparations. The material features thixotropic properties for ease of placement and a command set via light-curing, allowing for immediate condensation of restorative materials. It releases calcium and hydroxide ions to support dental health. The device is used in clinical dental settings by licensed practitioners to protect the pulp and repair structural defects, facilitating restorative clinical decision-making.
Clinical Evidence
Bench testing only. Evaluated flexural strength (ISO 4049:2009), compressive strength (ANSI/ADA Spec. No. 27:1993), diametral tensile strength, working time, bond strength, radiopacity (ISO 4049:2009), water sorption/solubility (ISO 4049:2009, ISO 6876:2012), calcium release, and pH. Biocompatibility assessed per ISO 7405:2008 and ISO 10993-1:2009. Results demonstrate equivalence to predicate.
Technological Characteristics
Dual-cured, resin-modified calcium silicate containing Portland cement and hydrophilic resin. Radiopaque. Basic pH. Applied without bonding agent. Tested per ISO 4049:2009, ISO 6876:2012, and ANSI/ADA Spec. No. 27:1993.
Indications for Use
Indicated for patients requiring pulpotomy, temporary filling, repair of root/furcation perforations, repair of internal/external resorptions, root-end filling, direct pulp capping, or indirect pulp capping/base under restorative substrates.
Regulatory Classification
Identification
A calcium hydroxide cavity liner is a device material intended to be applied to the interior of a prepared cavity before insertion of restorative material, such as amalgam, to protect the pulp of a tooth.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The FDA logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. The Department of Health & Human Services logo is a stylized image of a human figure.
April 5, 2018
Bisco, Inc. Ryan Hobson RA Product Registration Manager 1100 West Irving Park Road Schaumburg, Illinois 60193
Re: K180344
Trade/Device Name: TheraCal DC Regulation Number: 21 CFR 872.3250 Regulation Name: Calcium Hydroxide Cavity Liner Regulatory Class: Class II Product Code: EJK Dated: March 5, 2018 Received: March 7, 2018
Dear Ryan Hobson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Mary S. Runner -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K180344
Device Name TheraCal DC
Indications for Use (Describe)
1. Pulpotomy
2. Temporary Filling Material
3. Repair of Root Perforations
4. Repair of Furcation Perforations
5. Repair of Perforating Internal and External Resorptions
6. Root-End Filling in Endodontic Surgery
7. Pulp Exposures (Direct Pulp Capping)
8. Protective Liner (Indirect Pulp Capping) and Base for Use Under a Variety of Substrates
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for BISCO. Above the name are five connected hexagons. The first and last hexagons are red, while the middle three are blue. The name BISCO is written in a sans-serif font, with the top half of the letters in gray and the bottom half in black.
## Special 510(k) SUMMARY
Bisco, Inc. 1100 W. Irving Park Road Schaumburg IL, 60193
Contact Person:
Ryan Hobson Tel: 847-534-6143 Fax: 847-534-6143
5 March 2018
Date Prepared:
Trade Name: Common Name: Product Code: Classification/Name:
TheraCal DC Dual-Cured Resin-Modified Calcium Silicate Pulp Protectant EJK Liner, Cavity, Calcium Hydroxide Class II per 21 CFR 872.3250
### Predicate Devices:
TheraCal DC is substantially equivalent to: TheraCal DC by Bisco Inc. K143292 Primary Predicate:
# TheraCal DC is indicated for use for:
- 1. Pulpotomy
- 2. Temporary Filling Material
- 3. Repair of Root Perforations
- 4. Repair of Furcation Perforations
- 5. Repair of Perforating Internal and External Resorptions
- 6. Root-End Filling in Endodontic Surgery
- 7. Pulp Exposures (Direct Pulp Capping)
- 8. Protective Liner (Indirect Pulp Capping) and Base for Use Under a Variety of Substrates
800-247-3368 or 847-534-6000 Fax: 847-891-5049 www.bisco.com
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the BISCO company logo. The logo consists of five connected hexagons, with the first and last hexagons being red, the second and fourth being blue, and the middle hexagon being light blue. Below the hexagons is the word "BISCO" in a gray, sans-serif font, with the bottom half of the letters being black.
## Special 510(k) SUMMARY (continued)
The indications for use of TheraCal DC are identical to the predicate device.
| TheraCal DC | Modified TheraCal DC |
|------------------------------------------------------------|------------------------------------------------------------|
| (K143292) | |
| TheraCal DC is indicated for use for: | TheraCal DC is indicated for use for: |
| 1. Pulpotomy | 1. Pulpotomy |
| 2. Temporary filling material | 2. Temporary filling material |
| 3. Repair of root perforations | 3. Repair of root perforations |
| 4. repair of furcation perforations | 4. repair of furcation perforations |
| 5. repair of perforating internal and external resorptions | 5. repair of perforating internal and external resorptions |
| 6. apexification | 6. apexification |
| 7. root-end filling in endodontic surgery | 7. root-end filling in endodontic surgery |
| 8. pulp exposures (direct pulp capping) | 8. pulp exposures (direct pulp capping) |
### Description of Applicant Device:
TheraCal DC is a biocompatible, dual-cured, resin-modified calcium silicate that is used to treat damaged dentin in both the crown and the root. TheraCal DC's precise placement allows its use in all deep cavity preparations and endodontic repairs. The dual-cure set permits immediate placement and condensation of the restorative material. Its proprietary formulation allows for a command set with a light curing unit while maintaining ease of placement due to its thixotropic properties.
#### Technological Characteristics:
All components of TheraCal DC are based upon industry standard chemistry. Comparisons of the chemical composition of TheraCal DC to the predicate is provided in the following table:
| Chemical Composition | TheraCal DC<br>(K143292) | Modified TheraCal DC |
|--------------------------|----------------------------------------------------|----------------------------------------------------|
| Filler | Portland Cement | Portland Cement |
| Resin Composition | Hydrophilic resin to facilitate calcium<br>release | Hydrophilic resin to facilitate calcium<br>release |
| Method of polymerization | Dual-Cure | Dual-Cure |
| Method of Application | bonding agent not required | bonding agent not required |
| Ions Released | Calcium and hydroxide | Calcium and hydroxide |
| Physical / Mechanical Property Comparison | TheraCal DC<br>(K143292) | Modified TheraCal DC |
|-------------------------------------------------|--------------------------|----------------------|
| Radiographic Appearance<br>(ISO 4049:2009 7.14) | Radiopaque | Radiopaque |
| Ions Released | Calcium releasing | Calcium releasing |
| pH | basic | basic |
| Shelf-Life | 12 months | 18 months |
800-247-3368 or 847-534-6000 Fax: 847-891-5049 www.bisco.com
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for BISCO. The logo consists of five connected hexagons in a row. The first, third, and fifth hexagons are red, while the second and fourth hexagons are blue. Below the hexagons is the word "BISCO" in a sans-serif font, with the top half of the letters in gray and the bottom half in black.
The primary difference between the subject device and the predicate is the stated shelf life.
#### Performance Data:
The following physical/mechanical properties of TheraCal DC were tested: flexural strength (ISO 4049:2009), compressive strength (ANSI/ADA Spec. No. 27:1993 Appendix A2), dimetral tensile strength (in-house test method), working time (in-house test method), bond strength (in-house test method), radiopacity (ISO 4049:2009), water sorption and solubility (ISO 4049:2009 and ISO 6876:2012), calcium release (in-house test method), and pH (in-house test method). An evaluation of biocompatibility of TheraCal DC was conducted using ISO 7405:2008 and ISO 10993-1:2009. Direct comparison testing demonstrates that TheraCal DC is equivalent to the predicate device.
It is concluded from the evaluation and the results of the cytotoxicity study that TheraCal DC meets the requirements of the testing.
#### Conclusion:
It is concluded from review of the predicate device indications, chemical composition, biocompatibility, and physical properties that TheraCal DC is substantially equivalent in safety and effectiveness to the predicate device.
800-247-3368 or 847-534-6000 Fax: 847-891-5049 www.bisco.com
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.