K180306 · Next Sight Srl · HKI · Jun 8, 2018 · Ophthalmic
Device Facts
Record ID
K180306
Device Name
Nexy
Applicant
Next Sight Srl
Product Code
HKI · Ophthalmic
Decision Date
Jun 8, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.1120
Device Class
Class 2
Indications for Use
Nexy device is an automatic eye-fundus camera intended for taking digital images of a human retina without the use of a mydriatic agent.
Device Story
Nexy is an automatic non-mydriatic eye-fundus camera; captures 45° color retinal images. System comprises optical head, forehead/cheekbone rests, base, and tablet interface running dedicated software. User initiates automatic examination; device aligns to pupil center, performs autofocus, and captures image. Output is JPEG retinal photo saved to internal database. Used in hospitals and clinics by trained medical personnel, health professionals, or optometrists. Provides clinicians with retinal visualization to assist in screening for eye or systemic diseases; benefits patient through non-invasive, mydriatic-free imaging.
Clinical Evidence
Bench testing only. Testing included ophthalmic performance per ISO 15004-1/2 and ISO 10940, electrical safety per IEC 60601-1, electromagnetic compatibility per IEC 60601-1-2, and software verification/validation per IEC 62304.
Technological Characteristics
Non-mydriatic fundus camera; 45° field of view; CMOS image sensor (2.1 Mpixel); white LED flash (<50ms); IR LED focusing; polarized light; +15D/-15D diopter adjustment. Connectivity: DICOM standard. Standards: IEC 60601-1, IEC 60601-1-2, ISO 10940, ISO 15004-1/2, ISO 14971, ISO 62366, ISO 62304.
Indications for Use
Indicated for taking digital images of a human retina without the use of a mydriatic agent. Intended for use in hospitals, public/private health clinics, and environments where eye or systemic diseases are screened. Not for use on children or minors without adult assistance.
Regulatory Classification
Identification
An ophthalmic camera is an AC-powered device intended to take photographs of the eye and the surrounding area.
Special Controls
*Classification.* Class II (special controls). The device, when it is a photorefractor or a general-use ophthalmic camera, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services seal on the left and the FDA acronym and name on the right. The FDA acronym is in a blue square, and the full name "U.S. Food & Drug Administration" is in blue text.
June 8, 2018
Next Sight srl Arianna Dalla Bá Quality and Regulatory Affairs Specialist Via IV Novembre 6/E Limena (Padova), 35010 IT
Re: K180306
Trade/Device Name: Nexy Regulation Number: 21 CFR 886.1120 Regulation Name: Ophthalmic Camera Regulatory Class: Class II Product Code: HKI Dated: March 13, 2018 Received: March 15, 2018
Dear Arianna Dalla Bá:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing
U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov
{1}------------------------------------------------
(21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Denise L. Hampton -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose, and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K180306
Device Name Nexy
Indications for Use (Describe)
Nexy device is an automatic eye-fundus camera intended for taking digital images of a human retina without the use of a mydriatic agent.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Nextsight. The word "next" is in orange, and the word "sight" is in gray. To the right of the words is a stylized X, with one side in orange and the other in gray. Below the logo is the phrase "SEEN THE DIFFERENCE ?"
Headquarters VIA IV NOVEMBRE, 6/E 35010 LIMENA (PD) - ITALY TEL. +39 049 - 76 99 33
# 510(K) Summary
# K180306
#### GENERAL INFORMATION:
| Company Name: | Next Sight srl |
|-----------------|-------------------------------------------|
| Address: | Via IV Novembre 6/E |
| | 35010 - Limena (Padova) |
| | Italy |
| Phone: | +39049769933 |
| Contact Person: | Arianna Dalla Bà |
| Title: | Quality and Regulatory Affairs Specialist |
| Address: | Via IV Novembre 6/E |
| | 35010 - Limena (Padova) |
| | Italy |
| Email: | arianna.dallaba@nextsight.info |
| Phone: | +39049769933 |
| Date Prepared: | January 31, 2018 |
#### DEVICE IDENTIFICATION:
| Trade Name: | Nexy |
|----------------------|--------------------------------|
| Generic Device Name: | Camera, Ophthalmic, Ac-Powered |
| Review Panel: | Ophthalmic |
| Classification: | Class II |
| Regulation Number: | 21 CFR 886.1120 |
| Product Code: | HKI |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image is a logo for a company called "next sight". The word "next" is in orange, and the word "sight" is in gray. To the right of the words is a symbol that looks like two overlapping X's, one in orange and one in gray. Below the words is the phrase "SEEN THE DIFFERENCE ?" in gray.
NEXT SIGHT SRL
WWW.NEXTSIGHT.INFO
Headquarters
VIA IV NOVEMBRE, 6/E
35010 LIMENA (PD) - ITALY
TEL. +39 049-76 99 33
## Device Description:
The Nexy medical device is an automatic eye-fundus camera without the use of a mydriatic agent, designed to take color photos of the retinal fundus with a 45° field.
The device consists of two fundamental components: the eye-fundus camera and the tablet which represents the user interface where running a dedicated software application.
The instrument body includes: optical head, forehead and cheekbones rests and a base. The device is provided also with a front lens cap, a power cord and AC/DC adaptor.
Through the tablet, the user starts the automatic examination that provides as output the photo of the patient's retina saved in the internal database of the device in JPEG format.
The device can be used in a hospital environment, public and private health clinics and in public and private environments where eye or systemic diseases are screened.
The device must be used by medical personnel in some cases assisted by health professionals and/or optometry professionals. Users must be adequately trained in the use of the device and informed about the risks and side effects as well as being in possession of the qualifications required by law.
#### Indications For Use:
Nexy device is an automatic eye-fundus camera intended for takinq digital imaqes of a human retina without the use of a mydriatic agent.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Next Sight. The word "next" is in orange, and the word "sight" is in gray. To the right of the words is a symbol that looks like two interlocking links of a chain, with one link in orange and the other in gray. Below the words is the phrase "SEEN THE DIFFERENCE ?" in gray.
Headquarters
VIA IV NOVEMBRE, 6/E
35010 LIMENA (PD) - ITALY
TEL. +39 049 - 76 99 33
#### Identification of predicate devices:
The predicate devices selected for comparison with the subject device are identified as follows:
| Predicate<br>Device | 510(k)<br>Holder<br>Name | 510(K) no. | Date<br>cleared |
|---------------------------------------|-----------------------------------------------------|------------|----------------------|
| Digital<br>Retinography<br>System DRS | CenterVue | K101935 | 27 October<br>2010 |
| MiiS Horus<br>Scope DEC<br>100 | Medimaging<br>Integrated<br>Solution Inc.<br>(MiiS) | K120982 | 28 September<br>2012 |
#### Bench tests
As the predicate device DRS (K101935), Nexy device complies with the following standards:
- IEC 60601-1:2005;
- IEC 60601-1-2:2007; .
- ISO 10940-2:2009; ●
- . ISO 15004-2 2007.
Nexy complies also with the following standards:
- . ISO 15004-1:2006
- ISO 14971: 2012
- ISO 62366:2007 ●
- ISO 62304: 2006 ●
- DICOM standard .
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image is a logo for Next Sight. The word "next" is in orange, and the word "sight" is in gray. To the right of the words is an orange and gray symbol that looks like two crossing lines. Below the words is the phrase "SEEN THE DIFFERENCE ?" in gray.
Headquarters VIA IV NOVEMBRE, 6/E 35010 LIMENA (PD) - ITALY TEL. +39 049 - 76 99 33
## Predicate Device comparison
Table 1 provides a comparison between the subject device and the predicate devices.
| Table 1 - Predicate device comparison table | | | |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ATTRIBUTE /<br>CHARACTERISTICS | Nexy<br>Subject Device | Digital Retinography<br>System DRS<br>Predicate Device PD1 | MiiS Horus Scope DEC<br>100<br>Predicate Device PD2 |
| Device Name | Nexy | Digital Retinography<br>System DRS | MiiS Horus Scope DEC<br>100 |
| Device<br>Manufacturer | Next Sight srl | CentreVue | Medimaging<br>Integrated Solution<br>Inc. (MiiS) |
| FDA Product<br>Code | HKI | HKI | HKI |
| Comparison: Both predicate devices PD1 and PD2 have the same FDA product code as subject<br>device. | | | |
| 510k reference | K180306 | K101935 | K120982 |
| FDA<br>Classification<br>Name | Camera, Ophthalmic, Ac-<br>Powered | Camera, Ophthalmic,<br>Ac-Powered | Camera, Ophthalmic,<br>Ac-Powered |
| Comparison: Both predicate devices PD1 and PD2 have the same FDA Classification Name as<br>subject device. | | | |
| Classification | II | II | II |
| Comparison: Both predicate devices PD1 and PD2 have the same Classification as subject device. | | | |
| FDA Regulation<br>Number | 886.1120 | 886.1120 | 886.1120 |
| Comparison: Both predicate devices PD1 and PD2 have the same FDA Regulation Number as<br>subject device. | | | |
| Type of device | Fundus non-mydriatic<br>camera | Fundus non-mydriatic<br>camera | Fundus non-mydriatic<br>camera |
| Comparison: Both predicate devices PD1 and PD2 have the same Type of device as subject device. | | | |
| Indication for<br>Use | Nexy device is an<br>automatic eye-fundus<br>camera intended for taking<br>digital images of a human<br>retina<br>without the use of a<br>mydriatic agent. | The CenterVue Digital<br>Retinography System<br>DRS is intended for<br>taking digital images of<br>a human<br>retina without the use<br>of a mydriatic agent. | MiiS Horus Scope DEC<br>100 is a digital hand-<br>held eye-fundus<br>camera used to record<br>digital photographs<br>and video of fundus of<br>the human eye and<br>surrounding area. |
| Comparison: Predicate device PD1 has the same Indication for Use as subject device. The predicate<br>device PD2 has similar Indication for Use to the subject device. | | | |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Nextsight. The word "next" is in orange, and the word "sight" is in gray. To the right of the words is a graphic of two overlapping lines, one orange and one gray. Below the logo is the phrase "seen the difference ?"
Headquarters
VIA IV N0VEMBRE, 6/E 35010 LIMENA (PD) - ITALY
TEL. +39 049 - 76 99 33
| Table 1 - Predicate device comparison table | | | |
|---------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------|
| ATTRIBUTE / CHARACTERISTICS | Nexy Subject Device | Digital Retinography System DRS Predicate Device PD1 | MiiS Horus Scope DEC 100 Predicate Device PD2 |
| Treatable areas | Eyes | Eyes | Eyes |
| | Comparison: Both predicate devices PD1 and PD2 have the same treatable areas as subject device. | | |
| Target Population | All population with some exceptions: the device cannot be used on children and in any case on minors, without the assistance of an adult. | Not Available | Not Available |
| | Comparison: This information is not available in the competitors' documentation. | | |
| Alignment and examination | Automatic - the device automatically aligns to the center of the pupil prior to autofocusing and capture. | Automatic - the DRS device automatically aligns to the center of the pupil prior to autofocusing and capture. | Manual - MiiS Horus Scope DEC 100 is a digital hand-held eye-fundus camera. |
| | Comparison: Predicate device PD1 has the same alignment and examination as subject device. The alignment of the predicate device PD2 is different from the subject device. | | |
| Field of horizontal vision | 45° | 45° | 40° |
| Field of vertical vision | 45° | 40° | 40° |
| | Comparison: | | |
| | Field of view represents the portion of the field of vision of the retina that can be framed with the device. The predicate device PD1 as the same filed of horizontal vision as the subjected device. The Field of horizontal vision of the predicate device PD2 is different from the subject device. | | |
| Minimum pupil size | 3.8mm | 4mm | 3.5mm |
| | Comparison: | | |
| | The minimum pupil size parameter represents the size of the smallest pupil in order to take the exam. | | |
| | Both predicate devices PD1 and PD2 have similar minimum pupil size as subject device. | | |
| Image sensor | CMOS | CMOS | CMOS |
| | Comparison: Both predicate devices PD1 and PD2 have the same image sensor as subject device. | | |
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image is a logo for a company called "next sight". The word "next" is in orange, and the word "sight" is in gray. To the right of the words is a symbol that looks like two intersecting orange and gray lines. Below the words is the phrase "SEEN THE DIFFERENCE ?" in gray.
Headquarters
VIA IV N0VEMBRE, 6/E
35010 LIMENA (PD) - ITALY
TEL. +39 049 - 76 99 33
| Table 1 - Predicate device comparison table | | | |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------|
| ATTRIBUTE /<br>CHARACTERISTICS | Nexy<br>Subject Device | Digital Retinography<br>System DRS<br>Predicate Device PD1 | MiiS Horus Scope DEC<br>100<br>Predicate Device PD2 |
| Sensor size | 2.1 Mpixel | 5 Mpixel | 2 Mpixel |
| Comparison:<br>Predicate device PD2 has a similar sensor size as subject device. The sensor size of the predicate device PD1 is different from the subject device. | | | |
| Image Format | JPEG, DICOM | JPEG, DICOM | JPEG and Video (AVI) |
| Comparison: Predicate device PD1 has the same image format as subject device. The image format of the predicate device PD2 is the same for JPEG and it is different for AVI from the subject device. | | | |
| Resolution on<br>retina | 14 $ \mu $ m | 15 $ \mu $ m | This information is not<br>available for this<br>predicate device. |
| Comparison: Predicate device PD1 has similar resolution on retina as subject device.<br>This parameter indicates the level of distinguishable detail, Nexy resolution on retina differs by 1 $ \mu $ m from the predicate device PD1. | | | |
| Light source | LEDs | LEDs | LEDs |
| Comparison: Both predicate devices PD1 and PD2 have the same light source as subject device. | | | |
| Type of fixation<br>light | Green and continuous<br>(during the exam time) | Green and continuous<br>(during the exam time) | Not Applicable |
| Comparison: Both predicate devices PD1 and PD2 have the same type fixation light as subject device. | | | |
| Focus on the<br>retina — type of<br>light | IR (infrared) LEDs and<br>continuous source (during<br>the focusing time) | IR (infrared) LEDs and<br>continuous source<br>(during the focusing<br>time) | This information is not<br>available for this<br>predicate device. |
| Comparison: Both predicate devices PD1 and PD2 use the infrared LEDs to perform the focus on the retina. | | | |
| Type of flash | White LEDs in the visible<br>range and pulsed (duration<br>less than 50ms) with<br>annular shape on the<br>cornea. | White LEDs in the<br>visible range and pulsed<br>(duration less than<br>50ms) with annular<br>shape on the cornea. | This information is not<br>available for this<br>predicate device. |
| Comparison: Both predicate devices PD1 and PD2 use and comparable (color, technology, shape on<br>the cornea type of light to flash the retina. | | | |
| Exposure<br>conditions | In accordance to the flash<br>duration time. | In accordance to the<br>flash duration time. | This information is not<br>available for this<br>predicate device. |
| Comparison: Both exposure conditions of predicate devices PD1 and PD2 are in accordance to the flash duration time. | | | |
| Light feature | Polarized | Not polarized | Polarized |
| Comparison: Predicate devices PD2 has the same light source as subject device. It is different for | | | |
| Table 1 - Predicate device comparison table | | | |
| ATTRIBUTE /<br>CHARACTERISTICS | Nexy<br>Subject Device | Digital Retinography<br>System DRS<br>Predicate Device PD1 | MiiS Horus Scope DEC<br>100<br>Predicate Device PD2 |
| the predicate device PD2. | | | |
| Diopter<br>adjustment | +15D / -15D | +15D / -15D | +20D / -20D |
| Comparison:<br>Predicate devices PD1 has the same diopter adjustment as subject device. It is different for the<br>predicate device PD2. | | | |
| Number of<br>Fixings | 7 | 7 | Not Applicable |
| Comparison:<br>Predicate devices PD1 has the same Number of Fixings as subject device. | | | |
| Fixation target | Internal LEDs | Internal LEDs | Not Applicable |
| Comparison: Predicate devices PD1 has the same Fixation target as subject device. | | | |
| DICOM<br>standard<br>compliance | Yes | Yes | This information is not<br>available for this<br>predicate device. |
| Comparison: Predicate devices PD1 complies to DICOM standard as subject device. | | | |
| Performance<br>Standard<br>Compliance | Nexy device complies with<br>the following standards:<br>● IEC 60601-1:2005;<br>● IEC 60601-1-2:2007;<br>● ISO 10940-2:2009;<br>● ISO 15004-2 2007.<br><br>Nexy complies also with<br>the following standards:<br>● ISO 15004-1:2006<br>● ISO 14971: 2012<br>● ISO 62366:2007<br>● ISO 62304: 2006<br><br>Comparison:<br>Predicate devices PD1 complies to the same standards as subject device. Furthermore the subject<br>device complies with other standards. | The DRS complies with<br>the following standards:<br>IEC 60601-1:2005;<br>IEC 60601-1-2:2007;<br>ISO 10940-2:2009;<br>ISO 15004-2 2007. | This information is not<br>available for this<br>predicate device. |
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image contains the logo for Next Sight. The logo has the word "next" in orange above the word "sight" in gray. To the right of the words is a graphic of two crossing lines, one orange and one gray. Below the logo is the phrase "SEEN THE DIFFERENCE ?" in gray.
Headquarters VIA IV NOVEMBRE, 6/E 35010 LIMENA (PD) - ITALY
TEL. +39 049 - 76 99 33
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the logo for Nextsight. The word "next" is in orange, and the word "sight" is in gray. Below the words is the phrase "SEEN THE DIFFERENCE ?" in gray. To the right of the words is a stylized image of two crossing bars, one in orange and one in gray.
NEXT SIGHT SRL
WWW.NEXTSIGHT.INFO
Headquarters
VIA IV NOVEMBRE, 6/E
35010 LIMENA (PD) - ITALY
TEL. +39 049 - 76 99 33
# TESTING IN SUPPORT OF SUBSTANTIAL EQUIVALENCE DETERMINATION
All necessary bench testing was conducted on Nexy to support a determination of substantial equivalence to the predicate devices. The tests performed include:
- Ophthalmic Testing per ISO 15004-1, and ISO15004-2 ●
- Ophthalmic Testing about Fundus Cameras per ISO 10940 Ophthalmic Instruments --. Fundus Cameras
- Software Verification and Validation per AAMI/ANSI/IEC 62304 The software of this . device was considered as a "Moderate" level of concern based on the FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" issued on: May 11, 2005
- . Electrical Safety Testing per IEC60601-1 and Electromagnetic Compatibility Testing per IEC60601-1-2.
The collective performance testing demonstrates that the Nexy device does not raise any new questions of safety or effectiveness when compared to the predicate devices. The results of the performance testing demonstrate that the Nexy device performs as intended and does not raise any new questions of safety or effectiveness.
#### Conclusion
None of the differences highlighted in the Table 1 introduce new issues of safety and effectiveness compared to predicate devices.
The remaining technical aspects of the Nexy device compared with the predicate devices are identical or very similar.
Based on the information contained in this submission, it is concluded that the Nexy device is substantially equivalent to the identified predicate devices already in commerce within the USA and that any differences that do exist have no effect on the safety and effectiveness of the device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.