← Product Code [IYO](/productcode/IYO) · K180265

# COMPACT TOUCH Ophthalmic Ultrasound System (K180265)

_Quantel Medical · IYO · Mar 2, 2018 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K180265

## Device Facts

- **Applicant:** Quantel Medical
- **Product Code:** [IYO](/productcode/IYO.md)
- **Decision Date:** Mar 2, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1560
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** 3rd-Party Reviewed

## Indications for Use

The COMPACT TOUCH Ophthalmic Ultrasound System and the probes that are used with it are indicated for diagnostic imaging and biometric measurement of the eye including: - Visualization of the interior of the eye and the orbit by A and B scans. - Axial Length measurement of the eye by ultrasonic means. - Implanted IOL power calculation, using the Axial Length measurement. - Measurement of corneal thickness by ultrasonic means.

## Device Story

The COMPACT TOUCH Ophthalmic Ultrasound System is an ultrasonic diagnostic device for ophthalmology. It utilizes a base unit connected to A-scan and B-scan probes to capture ultrasonic echoes from ocular structures. The system processes these signals to generate images of the eye and orbit and to perform biometric measurements, including axial length and corneal thickness. These measurements are used to calculate implanted IOL power. The device is operated by clinicians in a clinical setting. The updated B-scan probe features an increased transducer frequency (15 MHz) compared to the predicate, resulting in improved measurement accuracy. The system includes a Wi-Fi connection and an HDMI port for external display. Clinicians use the visual output and biometric data to inform diagnostic and surgical planning decisions, potentially improving patient outcomes through precise ocular measurements and visualization.

## Clinical Evidence

Bench testing only. Performance testing demonstrated compliance with AAMI ANSI ES60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 62366-1, and IEC 60601-2-37. Hardware and software validation activities confirmed the device performs as intended. Measurement accuracy was verified to be ± 0.115 mm.

## Technological Characteristics

Ultrasonic pulsed echo imaging system. Components: Base unit and probes (A-scan and B-scan). B-scan transducer frequency: 15 MHz. Connectivity: Wi-Fi, HDMI. Electrical safety: IEC 60601-1 (Type B protection). Standards: AAMI ANSI ES60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 62366-1, IEC 60601-2-37.

## Regulatory Identification

An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Special Controls

*Classification.* Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

## Predicate Devices

- Quantel Medical Compact Touch ([K094038](/device/K094038.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, followed by the words "U.S. FOOD & DRUG" in large blue letters. Below that is the word "ADMINISTRATION" in smaller blue letters.

March 2, 2018

Quantel Medical % Mr. Mark Job Responsible Third Party Official Regulatory Technology Services, LLC 1394 25th Street. NW BUFFALO MN 55313

Re: K180265

Trade/Device Name: COMPACT TOUCH® Ophthalmic Ultrasound System Regulation Number: 21 CFR 892.1560 Regulation Name: Ultrasonic pulsed echo imaging system Regulatory Class: II Product Code: IYO, ITX Dated: February 26, 2018 Received: February 27, 2018

Dear Mr. Job:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Michael D. O'Hara
For

Robert Ochs, Ph.D Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

### Indications for Use

510(k) Number (if known)

K180265

Device Name

COMPACT TOUCH® Ophthalmic Ultrasound System

Indications for Use (Describe)

The COMPACT TOUCH Ophthalmic Ultrasound System and the probes that are used with it are indicated for diagnostic imaging and biometric measurement of the eye including:

- · Visualization of the interior of the eye and the orbit by A and B scans.
- · Axial Length measurement of the eve by ultrasonic means.
- · Implanted IOL power calculation, using the Axial Length measurement.

· Measurement of corneal thickness by ultrasonic means.

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

{3}------------------------------------------------

# Page : 2/5

#### System: Compact Touch

Intended Use : Diagnostic ultrasound imaging of the human body as follows :

| Clinical Application         |                                    | Mode of Operation |   |     |     |                  |                           |                    |
|------------------------------|------------------------------------|-------------------|---|-----|-----|------------------|---------------------------|--------------------|
| General<br>(Track 1<br>Only) | Specific<br>(Tracks 1 & 3)         | B                 | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) B/M | Other<br>(Specify) |
| Ophthalmic                   | Ophthalmic                         | P                 |   |     |     |                  |                           | (A-mode)<br>P      |
|                              | Fetal                              |                   |   |     |     |                  |                           |                    |
|                              | Abdominal                          |                   |   |     |     |                  |                           |                    |
|                              | Intra-operative (Specify)          |                   |   |     |     |                  |                           |                    |
|                              | Intra-operative (Neuro)            |                   |   |     |     |                  |                           |                    |
|                              | Laparoscopic                       |                   |   |     |     |                  |                           |                    |
|                              | Pediatric                          |                   |   |     |     |                  |                           |                    |
|                              | Small organ (Specify)              |                   |   |     |     |                  |                           |                    |
|                              | Neonatal Cephalic                  |                   |   |     |     |                  |                           |                    |
|                              | Adult Cephalic                     |                   |   |     |     |                  |                           |                    |
|                              | Trans-rectal                       |                   |   |     |     |                  |                           |                    |
|                              | Trans-vaginal                      |                   |   |     |     |                  |                           |                    |
|                              | Trans-urethral                     |                   |   |     |     |                  |                           |                    |
|                              | Trans-esoph (non-Card.)            |                   |   |     |     |                  |                           |                    |
| Fetal<br>Imaging &<br>Other  | Musculo-skeletal<br>(Conventional) |                   |   |     |     |                  |                           |                    |
|                              | Musculo-Skeletal (Superficial)     |                   |   |     |     |                  |                           |                    |
|                              | Intravascular                      |                   |   |     |     |                  |                           |                    |
|                              | Other (Specify)                    |                   |   |     |     |                  |                           |                    |
|                              | Cardiac Adult                      |                   |   |     |     |                  |                           |                    |
|                              | Cardiac Pediatric                  |                   |   |     |     |                  |                           |                    |
|                              | Intravascular (Cardiac)            |                   |   |     |     |                  |                           |                    |
| Cardiac                      | Trans-esoph (Cardiac)              |                   |   |     |     |                  |                           |                    |
|                              | Intra-cardiac                      |                   |   |     |     |                  |                           |                    |
|                              | Other (Specify)                    |                   |   |     |     |                  |                           |                    |
| Peripheral<br>Vessel         | Peripheral vessel                  |                   |   |     |     |                  |                           |                    |
|                              | Other (Specify)                    |                   |   |     |     |                  |                           |                    |

N= new indication ; P=previously cleared by FDA ; E=added under Appendix E

P=Previously cleared in K094038

{4}------------------------------------------------

Page : 3/5

System: Compact Touch

Transducer : 15 MHz B-scan Probe (ref model number B1)

Intended Use : Diagnostic ultrasound imaging of the human body as follows :

| Clinical Application         |                                    | Mode of Operation |   |     |     |                  |                           |                    |
|------------------------------|------------------------------------|-------------------|---|-----|-----|------------------|---------------------------|--------------------|
| General<br>(Track 1<br>Only) | Specific<br>(Tracks 1 & 3)         | B                 | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) B/M | Other<br>(Specify) |
| Ophthalmic                   | Ophthalmic                         | N                 |   |     |     |                  |                           |                    |
| Fetal<br>Imaging &<br>Other  | Fetal                              |                   |   |     |     |                  |                           |                    |
|                              | Abdominal                          |                   |   |     |     |                  |                           |                    |
|                              | Intra-operative (Specify)          |                   |   |     |     |                  |                           |                    |
|                              | Intra-operative (Neuro)            |                   |   |     |     |                  |                           |                    |
|                              | Laparoscopic                       |                   |   |     |     |                  |                           |                    |
|                              | Pediatric                          |                   |   |     |     |                  |                           |                    |
|                              | Small organ (Specify)              |                   |   |     |     |                  |                           |                    |
|                              | Neonatal Cephalic                  |                   |   |     |     |                  |                           |                    |
|                              | Adult Cephalic                     |                   |   |     |     |                  |                           |                    |
|                              | Trans-rectal                       |                   |   |     |     |                  |                           |                    |
|                              | Trans-vaginal                      |                   |   |     |     |                  |                           |                    |
|                              | Trans-urethral                     |                   |   |     |     |                  |                           |                    |
|                              | Trans-esoph (non-Card.)            |                   |   |     |     |                  |                           |                    |
|                              | Musculo-skeletal<br>(Conventional) |                   |   |     |     |                  |                           |                    |
|                              | Musculo-Skeletal (Superficial)     |                   |   |     |     |                  |                           |                    |
|                              | Intravascular                      |                   |   |     |     |                  |                           |                    |
|                              | Other (Specify)                    |                   |   |     |     |                  |                           |                    |
| Cardiac                      | Cardiac Adult                      |                   |   |     |     |                  |                           |                    |
|                              | Cardiac Pediatric                  |                   |   |     |     |                  |                           |                    |
|                              | Intravascular (Cardiac)            |                   |   |     |     |                  |                           |                    |
|                              | Trans-esoph (Cardiac)              |                   |   |     |     |                  |                           |                    |
|                              | Intra-cardiac                      |                   |   |     |     |                  |                           |                    |
|                              | Other (Specify)                    |                   |   |     |     |                  |                           |                    |
| Peripheral<br>Vessel         | Peripheral vessel                  |                   |   |     |     |                  |                           |                    |
|                              | Other (Specify)                    |                   |   |     |     |                  |                           |                    |

N= new indication ; P=previously cleared by FDA ; E=added under Appendix E

{5}------------------------------------------------

Page : 4/5

System: Compact Touch

 11 MHz Biometry Probe (ref model number : TP-01-b/ TP-02-las) Transducer :

Intended Use : Diagnostic ultrasound imaging of the human body as follows :

| Clinical Application         |                                    | Mode of Operation |   |     |     |                  |                           |                    |  |
|------------------------------|------------------------------------|-------------------|---|-----|-----|------------------|---------------------------|--------------------|--|
| General<br>(Track 1<br>Only) | Specific<br>(Tracks 1 & 3)         | B                 | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) B/M | Other<br>(Specify) |  |
| Ophthalmic                   | Ophthalmic                         |                   |   |     |     |                  |                           | (A-mode)<br>P      |  |
| Fetal<br>Imaging &<br>Other  | Fetal                              |                   |   |     |     |                  |                           |                    |  |
|                              | Abdominal                          |                   |   |     |     |                  |                           |                    |  |
|                              | Intra-operative (Specify)          |                   |   |     |     |                  |                           |                    |  |
|                              | Intra-operative (Neuro)            |                   |   |     |     |                  |                           |                    |  |
|                              | Laparoscopic                       |                   |   |     |     |                  |                           |                    |  |
|                              | Pediatric                          |                   |   |     |     |                  |                           |                    |  |
|                              | Small organ (Specify)              |                   |   |     |     |                  |                           |                    |  |
|                              | Neonatal Cephalic                  |                   |   |     |     |                  |                           |                    |  |
|                              | Adult Cephalic                     |                   |   |     |     |                  |                           |                    |  |
|                              | Trans-rectal                       |                   |   |     |     |                  |                           |                    |  |
|                              | Trans-vaginal                      |                   |   |     |     |                  |                           |                    |  |
|                              | Trans-urethral                     |                   |   |     |     |                  |                           |                    |  |
|                              | Trans-esoph (non-Card.)            |                   |   |     |     |                  |                           |                    |  |
|                              | Musculo-skeletal<br>(Conventional) |                   |   |     |     |                  |                           |                    |  |
|                              | Musculo-Skeletal (Superficial)     |                   |   |     |     |                  |                           |                    |  |
|                              | Intravascular                      |                   |   |     |     |                  |                           |                    |  |
|                              | Other (Specify)                    |                   |   |     |     |                  |                           |                    |  |
| Cardiac                      | Cardiac Adult                      |                   |   |     |     |                  |                           |                    |  |
|                              | Cardiac Pediatric                  |                   |   |     |     |                  |                           |                    |  |
|                              | Intravascular (Cardiac)            |                   |   |     |     |                  |                           |                    |  |
|                              | Trans-esoph (Cardiac)              |                   |   |     |     |                  |                           |                    |  |
|                              | Intra-cardiac                      |                   |   |     |     |                  |                           |                    |  |
|                              | Other (Specify)                    |                   |   |     |     |                  |                           |                    |  |
| Peripheral<br>Vessel         | Peripheral vessel                  |                   |   |     |     |                  |                           |                    |  |
|                              | Other (Specify)                    |                   |   |     |     |                  |                           |                    |  |

N= new indication ; P=previously cleared by FDA ; E=added under Appendix E

P=Previously cleared in K094038

{6}------------------------------------------------

Page : 5/5

System: Compact Touch

Transducer : Pachymetry Probe (ref model number : P1 : (20 MHz A-scan)

Intended Use : Diagnostic ultrasound imaging of the human body as follows :

| Clinical Application         |                                    | Mode of Operation |   |     |     |                  |                           |                    |
|------------------------------|------------------------------------|-------------------|---|-----|-----|------------------|---------------------------|--------------------|
| General<br>(Track 1<br>Only) | Specific<br>(Tracks 1 & 3)         | B                 | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) B/M | Other<br>(Specify) |
| Ophthalmic                   | Ophthalmic                         |                   |   |     |     |                  |                           | (A-mode)<br>P      |
|                              | Fetal                              |                   |   |     |     |                  |                           |                    |
|                              | Abdominal                          |                   |   |     |     |                  |                           |                    |
|                              | Intra-operative (Specify)          |                   |   |     |     |                  |                           |                    |
|                              | Intra-operative (Neuro)            |                   |   |     |     |                  |                           |                    |
|                              | Laparoscopic                       |                   |   |     |     |                  |                           |                    |
|                              | Pediatric                          |                   |   |     |     |                  |                           |                    |
|                              | Small organ (Specify)              |                   |   |     |     |                  |                           |                    |
|                              | Neonatal Cephalic                  |                   |   |     |     |                  |                           |                    |
|                              | Adult Cephalic                     |                   |   |     |     |                  |                           |                    |
|                              | Trans-rectal                       |                   |   |     |     |                  |                           |                    |
|                              | Trans-vaginal                      |                   |   |     |     |                  |                           |                    |
|                              | Trans-urethral                     |                   |   |     |     |                  |                           |                    |
|                              | Trans-esoph (non-Card.)            |                   |   |     |     |                  |                           |                    |
|                              | Musculo-skeletal<br>(Conventional) |                   |   |     |     |                  |                           |                    |
| Fetal<br>Imaging &<br>Other  | Musculo-Skeletal (Superficial)     |                   |   |     |     |                  |                           |                    |
|                              | Intravascular                      |                   |   |     |     |                  |                           |                    |
|                              | Other (Specify)                    |                   |   |     |     |                  |                           |                    |
|                              | Cardiac Adult                      |                   |   |     |     |                  |                           |                    |
|                              | Cardiac Pediatric                  |                   |   |     |     |                  |                           |                    |
|                              | Intravascular (Cardiac)            |                   |   |     |     |                  |                           |                    |
|                              | Trans-esoph (Cardiac)              |                   |   |     |     |                  |                           |                    |
| Cardiac                      | Intra-cardiac                      |                   |   |     |     |                  |                           |                    |
|                              | Other (Specify)                    |                   |   |     |     |                  |                           |                    |
| Peripheral<br>Vessel         | Peripheral vessel                  |                   |   |     |     |                  |                           |                    |
|                              | Other (Specify)                    |                   |   |     |     |                  |                           |                    |

N= new indication ; P=previously cleared by FDA ; E=added under Appendix E

P=Previously cleared in K094038

{7}------------------------------------------------

### 510(k) SUMMARY

# Quantel Medical COMPACT TOUCH® Ophthalmic Ultrasound System

### 510(k) Owner

Quantel Medical 11 rue Bois Joli CS40015 63808 Cournon D'Auvergne-Cedex France 33 04 73 745 732 Phone: Contact Person: Bruno Pagès

### Submission Correspondent:

Maureen O'Connell O'Connell Regulatory Consultants, Inc. 5 Timber Lane North Reading, MA 01864 Phone: 978-207-1245

Date Prepared: February 28, 2018

### Trade Name of Device

COMPACT TOUCH OPHTHALMIC ULTRASOUND SYSTEM

# Common or Usual Names

Ultrasonic Pulsed Echo Imaging system Ultrasonic Diagnostic Transducer

### Classification Names

Ultrasonic Pulsed Echo Imaging system; 21 C.F.R. 892.1560 Class II Product Codes: IYO

Transducer, Ultrasonic, Diagnostic; 21 C.F.R. 892.1570 Class II Product Codes: ITX

# Predicate Device

Quantel Medical Compact Touch (K094038)

{8}------------------------------------------------

# Device Description

The COMPACT TOUCH Ophthalmic Ultrasound System is an ultrasonic system for ophthalmology. It is a modification to the COMPACT TOUCH® cleared in K094038. Similar to the previous version of the device, it consists of a base and probes. The Base performs the same calculations and displays as the predicate COMPACT TOUCH®. The A-scan probe for biometry and pachymeter is identical to the predicate while the B-scan probe has an increased transducer frequency of 15 Hz which allows for improved measurement accuracy of ± 0.115 mm.

# Intended Use / Indications for Use

The COMPACT TOUCH Ophthalmic Ultrasound System and the probes that are used with it are indicated for diagnostic imaging and biometric measurement of the eye including:

- . Visualization of the interior of the eye and the orbit by A and B scans.
- . Axial Length measurement of the eye by ultrasonic means.
- Implanted IOL power calculation. using the Axial Length measurement. ●
- Measurement of corneal thickness by ultrasonic means. ●

# Substantial Equivalence

The COMPACT TOUCH Ophthalmic Ultrasound System described in this notification and for use under the conditions of the proposed labeling is substantially equivalent to a legally marketed predicate device that is a Class II medical device which is the Quantel Medical COMPACT TOUCH® (K094038). The indications for use statement for the COMPACT TOUCH Ophthalmic Ultrasound System is exactly the same as the indications for use statement for the predicate device.

The COMPACT TOUCH Ophthalmic Ultrasound System and the predicate device, the previous version of the same device, are equivalent regarding technological characteristics. Both systems include diagnostic imaging with biometry and pachymetry performed with an A-Scan probe and diagnostic imaging and axial length measurements performed with a B-Scan probe. The A-scan probe used in both systems is identical with the same transducer used in both systems.

There have been minor modifications to the B-Scan probe where the transducer frequency of the B-scan probe has been increased from 10 MHz for the cleared COMPACT TOUCH® to 15 MHz for the new version. This variation of frequency does not affect the safety or performance of the device. The IEC 60601-2-37 tests have been followed in order to cover the new transducer specification (the MI value is inferior to 0.23).

All other aspects of the two systems remain identical with the exception of the following items which have been determined not to impact the safety and effectiveness:

- Physical Characteristics: revised dimensions and look of device and addition of Wi-Fi connection. Electrical safety and electromagnetic compatibility testing have been

{9}------------------------------------------------

performed which shows the new version of the device continues to comply with the electrical safety standards.

- -Electronic: the degree of protection against electric shock has been changed from BF to B due to the change of the power supply. The IEC 60601-1 tests have been followed in order to cover this new configuration.
- Measurement accuracy has been improved (from ± 0.2 mm to ± 0.115 mm) due to the increase of the frequency of the probe.
- General Software: the software has been revised to accommodate the other device changes listed here.
- -Documentation: an HDMI port has been added to the device.

In summary, the new version of the COMPACT TOUCH Ophthalmic Ultrasound System has the same indications for use and similar technological characteristics to the predicate device and is therefore, substantially equivalent.

# Performance Data

Performance testing was conducted in order to demonstrate compliance with recognized consensus standards:

- AAMI ANSI ES60601-1:2005 + Corr. 1:2006+Corr. 2:2007+A1:2012 Medical . electrical equipment-Part 1: General requirements for basic safety and essential performance
- . IEC 60601-1-2:2007: Medical electrical equipment-Part 1-2: General requirements for basic safety and essential performance-Collateral standard: Electromagnetic compatibility-Requirements and tests
- IEC 60601-1-6:2010+A1:2013 Medical electrical equipment-Part 1-6: General requirements for safety-Collateral Standard Usability
- IEC 62366-1 Medical Devices-Part 1: Application of usability engineering to medical . devices
- IEC 60601-2-37: "Ed2 + AM1 with 60601-1 (ed.3), AM1 with correc 1 & correc 2 ● Particular requirements for the safety of ultrasonic medical diagnostic and monitoring equipment

Additionally, hardware and software validation activities were performed to ensure the device performed as intended and software documentation appropriate for the Moderate level of concern was provided.

---

**Source:** [https://fda.innolitics.com/device/K180265](https://fda.innolitics.com/device/K180265)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
