K180144 · Boston Scientific Corporation · ESW · Sep 21, 2018 · General, Plastic Surgery
Device Facts
Record ID
K180144
Device Name
Agile Esophageal Stent System
Applicant
Boston Scientific Corporation
Product Code
ESW · General, Plastic Surgery
Decision Date
Sep 21, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.3610
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Agile Esophageal Stent System is intended for maintaining esophageal luminal patency in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors, and occlusion of concurrent esophageal fistulas.
Device Story
Agile Esophageal Stent System is a self-expanding, braided nitinol stent used to maintain esophageal luminal patency in malignant strictures and fistulas; device is MR conditional. Stent is preloaded in a coaxial delivery system compatible with therapeutic gastroscope working channels (min 3.7mm). Physician performs endoscopic placement under direct visualization; delivery system includes radiopaque markers for fluoroscopic guidance and visual markers on the handle for endoscopic positioning. Suture at stent flares allows for repositioning during initial deployment. Stent is fully or partially covered with silicone. Device provides mechanical support to the esophageal wall to prevent occlusion; benefits include restored luminal patency and improved patient quality of life.
Clinical Evidence
No clinical data. Substantial equivalence is supported by extensive bench testing, including biocompatibility (cytotoxicity, sensitization, systemic toxicity, genotoxicity, implantation), and mechanical performance testing (deployment/reconstrainment, tensile strength, corrosion, fatigue, and comparative hoop expansion/compression force).
Technological Characteristics
Self-expanding braided nitinol stent; silicone covering (fully or partially covered). Coaxial delivery system with radiopaque and visual markers. MR conditional. Dimensions: 14mm-23mm diameter, 59mm-150mm length. Sterile, single-use. No software or electronic components.
Indications for Use
Indicated for patients with esophageal strictures caused by intrinsic and/or extrinsic malignant tumors and concurrent esophageal fistulas requiring maintenance of esophageal luminal patency.
Regulatory Classification
Identification
An esophageal prosthesis is a rigid, flexible, or expandable tubular device made of a plastic, metal, or polymeric material that is intended to be implanted to restore the structure and/or function of the esophagus. The metal esophageal prosthesis may be uncovered or covered with a polymeric material. This device may also include a device delivery system.
Special Controls
The special control for this device is FDA's “Guidance for the Content of Premarket Notification Submissions for Esophageal and Tracheal Prostheses.”
Predicate Devices
Wallflex Esophageal Fully & Partially Covered Stent System (K091510/ K073266)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
September 21, 2018
Boston Scientific Corporation Kayla Mackey Regulatory Affairs Specialist II 100 Boston Scientific Way Marlborough, Massachusetts 01752
Re: K180144
Trade/Device Name: Agile Esophageal Stent System Regulation Number: 21 CFR 878.3610 Regulation Name: Esophageal Prosthesis Regulatory Class: Class II Product Code: ESW Dated: August 22, 2018 Received: August 23, 2018
Dear Kayla Mackey:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice
(https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Image /page/1/Picture/7 description: The image shows the FDA logo on the left side. On the right side of the image, there is text that says "Jeffrey W" followed by "2018.09" and then "-04'00'". The FDA logo is a stylized version of the letters FDA.
Jeffrey W. Cooper -S 2018.09.21 15:47:15
for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
| Form Approved: OMB No. 0910-0120 | |
|----------------------------------|--|
| Expiration Date: 06/30/2020 | |
| See PRA Statement below. | |
| 510(k) Number (if known) | Unknown K180144 |
|--------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | Agile Esophageal Stent System |
| Indications for Use (Describe) | The Agile Esophageal Stent System is intended for maintaining esophageal luminal patency in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors, and occlusion of concurrent esophageal fistulas. |
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------|
| <span style="text-decoration: underline;">Prescription Use (Part 21 CFR 801 Subpart D)</span> | <span style="text-decoration: underline;">Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (7/17)
Page 1 of 1
19
{3}------------------------------------------------
## SECTION 5: 510(K) SUMMARY
# 1. Submitter:
Boston Scientific Corporation 100 Boston Scientific Way Marlborough, MA 01752
Kayla Mackey Regulatory Affairs Specialist II Telephone: 508-683-4534 Fax: 508-683-5939
Date Prepared: January 16, 2018
# 2. Device:
| Trade Name: | Agile™ Esophageal Stent System |
|----------------------|--------------------------------|
| Classification Name: | Esophageal Prosthesis |
| Regulation Number: | 21 CFR 878.3610 |
| Product Code: | ESW |
| Classification: | Class II |
## 3. Predicate Device(s):
| Trade Name: | Wallflex Esophageal Fully &<br>Partially Covered Stent System |
|----------------------|---------------------------------------------------------------|
| 510(k) Number: | K091510/ K073266 |
| Classification Name: | Esophageal Prosthesis |
| Regulation Number: | 21 CFR 878.3610 |
| Product Code: | ESW |
| Classification: | Class II |
| Trade Name: | Taewoong Niti-S Esophageal TTS<br>Stent |
| 510(k) Number: | K123205 |
| Classification Name: | Esophageal Prosthesis |
| Regulation Number: | 21 CFR 878.3610 |
| Product Code: | ESW |
| Classification: | Class II |
| Trade Name: | Merit Alimaxx-ES Esophageal<br>Stent |
| 510(k) Number: | K080838 |
{4}------------------------------------------------
Classification Name: Regulation Number: Product Code: Classification:
Esophageal Prosthesis 21 CFR 878.3610 ESW Class II
#### Reference Device:
Trade Name: 510(k) Number: Classification Name: Regulation Number: Product Code: Classification:
Ultraflex Esophageal NG Stent System K091816 Esophageal Prosthesis 21 CFR 878.3610 ESW Class II
## 4. Device Description:
The Agile™ Esophageal Stent System is intended for the endoscopic placement of the selfexpanding nitinol stent for maintaining luminal patency in esophageal strictures. The flexible Agile Stent and through-the scope Delivery System allow for stent placement through direct visualization. The Agile™ Esophageal Stent System is supplied sterile.
The Agile™ Stent is a flexible. MR conditional, self-expanding, braided nitinol stent. The proximal and distal ends of the stent have a flare which is larger in diameter than the stent body. The stent wires at each end of the flare are looped. A suture is threaded within the looped ends of both the proximal and distal stent flares to aid in repositioning during initial placement. The proposed Agile Stent is both fully and partially covered in a silicone covering.
The Agile™ Stent comes preloaded and constrained in the delivery system. The delivery system consists of two coaxial shafts compatible with a minimum 3.7mm working channel of a therapeutic gastroscope. The exterior tube is used to constrain the stent before deployment and to reconstrain the stent after partial deployment. The delivery system has radiopaque and visual markers to aid the user in accurate stent placement. There are three visual markers on the delivery system handle to aid in the placement of the stent under endoscopic visualization and four radiopaque markers on the interior of the delivery system to aid in placement of the stent under fluoroscopic visualization.
| Table 5-1: Agile Partially Covered Stent Sizes | | | |
|------------------------------------------------|----------------|--------------|----------------|
| UPN | Stent Size | | |
| | Stent Diameter | Stent Length | Flare Diameter |
| M00517200 | 14mm | 62 mm | 19mm |
| M00517210 | 14mm | 102 mm | 19mm |
| M00517220 | 14mm | 119 mm | 19mm |
| M00517230 | 14mm | 148 mm | 19mm |
| M00517240 | 18mm | 59 mm | 23mm |
| M00517250 | 18mm | 97 mm | 23mm |
| M00517260 | 18mm | 119 mm | 23mm |
| M00517270 | 18mm | 149 mm | 23mm |
#### Proposed Model Configurations:
{5}------------------------------------------------
| Table 5-1: Agile Partially Covered Stent Sizes | | | |
|------------------------------------------------|----------------|--------------|----------------|
| UPN | Stent Size | | |
| | Stent Diameter | Stent Length | Flare Diameter |
| M00517280 | 23mm | 62 mm | 28mm |
| M00517290 | 23mm | 101 mm | 28mm |
| M00517300 | 23mm | 120 mm | 28mm |
| M00517310 | 23mm | 150 mm | 28mm |
| Table 5-2: Agile Fully Covered Stent Sizes | | | |
|--------------------------------------------|----------------|--------------|----------------|
| UPN | Stent Size | | |
| | Stent Diameter | Stent Length | Flare Diameter |
| M00517400 | 14mm | 62 mm | 19mm |
| M00517410 | 14mm | 102 mm | 19mm |
| M00517420 | 14mm | 119 mm | 19mm |
| M00517430 | 14mm | 148 mm | 19mm |
| M00517440 | 18mm | 59 mm | 23mm |
| M00517450 | 18mm | 97 mm | 23mm |
| M00517460 | 18mm | 119 mm | 23mm |
| M00517470 | 18mm | 149 mm | 23mm |
| M00517480 | 23mm | 62 mm | 28mm |
| M00517490 | 23mm | 101 mm | 28mm |
| M00517500 | 23mm | 120mm | 28mm |
| M00517510 | 23mm | 150mm | 28mm |
### 5. Indications for Use:
The Agile™ Esophageal Stent System is intended for maintaining esophageal luminal patency in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors, and occlusion of concurrent esophageal fistulas.
### 6. Technological Characteristics
The intended use of the proposed Agile™ Esophageal Stent System is identical to the predicate Wallflex Esophageal Fully and Partially Covered Stent System. Taewoong Niti-S TTS Esophageal Stent, and Merit Alimaxx-ES Esophageal Stent.
The proposed Agile™ Esophageal Stent Delivery system is similiar to the predicate Taewoong Niti-S TTS Esophageal delivery system in overall length and working length, outer diameter, method of placement, and guidewire compatibility.
The proposed Agile™ Esophageal Stent is similar to the predicate Wallflex Esophageal Fully and Partially Covered Stents, Taewoong Niti-S TTS Esophageal Stents, and Merit Alimaxx-ES Esophageal Stents in terms of performance, stent diameter, and stent length. The
{6}------------------------------------------------
stent materials are similar between the proposed Agile Esophageal Stent System and the predicate Wallflex Esophageal Fully and Partially Covered Stents.
# 7. Biocompatibility Analysis
The following biocompatibility tests have been completed for the proposed Agile Esophageal Fully Covered Stent System.
| Table 5-3 Summary of Agile Esophageal Fully Covered Stent Biocompatibility Testing | | | |
|------------------------------------------------------------------------------------|---------|--|--|
| Test | Results | | |
| MEM Elution Cytotoxicity/Part 5 | Pass | | |
| Sub-acute (Subchronic)/Part 11/ Intravenous Toxicity | Pass | | |
| Guinea Pig Maximization Sensitization/Part 10 | Pass | | |
| Intracutaneous Reactivity/Part 10 | Pass | | |
| Sub-acute (Subchronic)/Part 11/ Intraperitoneal Toxicity | Pass | | |
| Materials Mediated Rabbit Pyrogen/Part 11 | Pass | | |
| Ames Mutagenicity/ Part 3 | Pass | | |
| Acute Systemic Injection/Part 11 | Pass | | |
| Mouse Lymphoma / Part 3 | Pass | | |
| Analytical Testing/ Part 18 (NVR; GCMS; LC-MS; ICPMS;<br>HS-GC-MS) | Pass | | |
| Implantation (Intramuscular)/ Part 6<br>(4weeks) | Pass | | |
| Implantation (Intramuscular)/ Part 6<br>(13 weeks) | Pass | | |
| Test | Results |
|-----------------------------------------------|---------|
| MEM Elution Cytotoxicity/Part 5 | Pass |
| Intracutaneous Reactivity/Part 10 | Pass |
| Guinea Pig Maximization Sensitization/Part 10 | Pass |
| Analytical testing | Pass |
{7}------------------------------------------------
Ultraflex Esophageal NG Stent System (K091816) was used as a reference device to support the biocompatibility of the Agile Esophageal Partially Covered Stent. The following biocompatibility tests have been leveraged from Ultraflex Esophgeal NG Stent System (K091816).
| Table 5-5 Summary of Ultraflex Esophageal NG Stent System Biocompatibility Testing | | |
|------------------------------------------------------------------------------------|---------|--|
| Test | Results | |
| Cytotoxicity | Pass | |
| Sensitization | Pass | |
| Intracutaneous Reactivity | Pass | |
| Systemic Toxicity- Acute Systemic Toxicity | Pass | |
| Subacute Toxicity- Intravenous | Pass | |
| Subacute Toxicity- Intraperitoneal | Pass | |
| Genotoxicity- Ames Assay | Pass | |
| Mouse Lymphoma | Pass | |
| Mouse Lymphoma / Part 3 | Pass | |
| Implantation (12 weeks) | Pass | |
| USP Physiochemical | Pass | |
### 8. Performance Data
The proposed Agile™ Esophageal Stent System successfully passed all pre-defined product specifications for the tests performed. Below is a summary of the tests performed on the Agile Esophageal Stent System:
| Table 5-6 Summary of Bench Tests and Results | | |
|----------------------------------------------|--------------------------------------------|------------------------|
| No. | Test | Results<br>(Pass/Fail) |
| 1 | Guidewire Passage | Pass |
| 2 | Trackability/Pushability | Pass |
| 3 | Deployment and Reconstrainment | Pass |
| 4 | Delivery System Removal | Pass |
| 5 | Immediate Stent Opening | Pass |
| 6 | Delivery System Withdrawal | Pass |
| 7 | Stent Hoop Expansion and Compression Force | Pass |
| 8 | Delivery System Working Length | Pass |
| 9 | Delivery System Outer Diameter | Pass |
| 10 | Visual Transition Zone Length | Pass |
| 11 | Stent Foreshortening | Pass |
| 12 | Final Unconstrained Stent Length | Pass |
| 13 | Stent Flare Length | Pass |
{8}------------------------------------------------
| 14 | Non-Covered Length (Partially Covered Codes Only) | Pass |
|----|------------------------------------------------------|------|
| 15 | Stent Covering Length (Partially Covered Codes Only) | Pass |
| 16 | Stent Diameter | Pass |
| 17 | Flare Diameter | Pass |
| 18 | Tip to Inner Member Bond Tensile | Pass |
| 19 | Inner Jacket to Handle Assembly Tensile | Pass |
| 20 | Reconstrainment Band to Inner Jacket Tensile | Pass |
| 21 | Exterior Tube to Handle Bond Tensile | Pass |
| 22 | Inner Jacket to Inner Member Bond Tensile | Pass |
| 23 | Exterior Tube Transition Zone Tensile | Pass |
| 24 | Suture Pull Strength Test | Pass |
| 25 | Corrosion Testing | Pass |
| 26 | Fatigue Testing | Pass |
| 27 | Deployment Accuracy | Pass |
| 28 | Magnetic Resonance Testing | Pass |
| Table 5-7 Summary of the Comparative Bench Tests | |
|--------------------------------------------------|------------------------------------------|
| No. | Test |
| 29 | Stent Hoop Expansion & Compression Force |
## 9. Comparative Bench Testing
Comparative bench testing was conducted on the Agile Esophageal Stent System and its substantially equivalent predicate device. The comparative testing demonstrated that the Stent Hoop Expansion and Compression Force of the Agile Esophageal Stents falls within the range of the currently marketed predicates for the propose stent diameters.
### Conclusion:
The information Boston Scientific provided in this submission demonstrates that the proposed Agile Esophageal Stent System is substantially equivalent to the currently cleared predicate devices thru the design and performance characteristics and indication for use.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.