← Product Code [GXH](/productcode/GXH) · K180138

# AtriCure cryoICE cryo-ablation probe (CRYO2) (K180138)

_AtriCure, Inc. · GXH · Feb 15, 2018 · Neurology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K180138

## Device Facts

- **Applicant:** AtriCure, Inc.
- **Product Code:** [GXH](/productcode/GXH.md)
- **Decision Date:** Feb 15, 2018
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 882.4250
- **Device Class:** Class 2
- **Review Panel:** Neurology
- **Attributes:** Therapeutic

## Indications for Use

AtriCure's cryolCE cryo-ablation probes are sterile, single use devices intended for use in the cryosurgical treatment of cardiac arrhythmias by freezing target tissues, creating an inflammatory response (cryonecrosis) that blocks the electrical conduction pathway. The probe is also intended for use in blocking pain by temporarily ablating peripheral nerves.

## Device Story

Handheld, single-use cryosurgical probe; used during cardiac surgery. Operates via Joule-Thompson effect using high-pressure nitrous oxide (N2O) supplied by AtriCure Cryo Module. Malleable aluminum cryotip reaches temperatures below -40°C to induce cryonecrosis in target tissues. Includes thermocouple for real-time temperature monitoring on console. Surgeon-operated; used to block electrical conduction pathways in heart or peripheral nerves. Benefits include precise tissue ablation for arrhythmia treatment and pain management.

## Clinical Evidence

Bench testing only. Testing included mechanical reliability, transit testing, shelf-life, and cryogen performance/thermal insulation.

## Technological Characteristics

Materials: Aluminum boiler (cryotip), insulated rigid shaft. Energy: High-pressure nitrous oxide (N2O) cryogen. Principle: Joule-Thompson effect for rapid cooling. Form factor: Malleable 4-mm diameter cryotip (up to 10-cm length). Connectivity: Connects to AtriCure Cryo Module (ACM) via inlet/exhaust tubes and thermocouple wire. Sterilization: Sterile, single-use.

## Regulatory Identification

A cryogenic surgical device is a device used to destroy nervous tissue or produce lesions in nervous tissue by the application of extreme cold to the selected site.

## Predicate Devices

- AtriCure cryoICE cryo-ablation probe ([K142203](/device/K142203.md))

## Reference Devices

- AtriCure cryoICE cryoablation probe ([K142441](/device/K142441.md))
- AtriCure cryoFORM cryoablation probe ([K152337](/device/K152337.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, featuring a stylized human figure. To the right is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. The logos are placed side by side.

February 15, 2018

AtriCure, Inc. Melissa Smallwood Regulatory Affairs Specialist 7555 Innovation Way Mason, Ohio 45040

Re: K180138

Trade/Device Name: AtriCure cryoICE cryo-ablation probe (CRYO2) Regulation Number: 21 CFR 882.4250 Regulation Name: Cryogenic Surgical Device Regulatory Class: Class II Product Code: GXH Dated: January 15, 2018 Received: January 17, 2018

Dear Melissa Smallwood:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn

(http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

M.A. Willemen

for
 Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known)

Subject of this Special 510(k) submission

K180138

Device Name

AtriCure cryoICE cryo-ablation probe (CRYO2)

Indications for Use (Describe)

AtriCure's cryolCE cryo-ablation probes are sterile, single use devices intended for use in the cryosurgical treatment of cardiac arrhythmias by freezing target tissues, creating an inflammatory response (cryonecrosis) that blocks the electrical conduction pathway. The probe is also intended for use in blocking pain by temporarily ablating peripheral nerves.

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D) | <span> <span style="font-size: 20px;"><b>     ×    </b> </span> </span> |
|----------------------------------------------|-------------------------------------------------------------------------|
| Over-The-Counter Use (21 CFR 801 Subpart C)  | <span> </span>                                                          |

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the AtriCure logo. The logo is in blue and orange. The word "Atri" is in blue, and the word "Cure" is in orange. There is a blue dot in the middle of the "C" in "Cure". There is a registered trademark symbol next to the "e" in "Cure".

# 510(k) Summary

#### l. Submitter

| Manufacturer:        | AtriCure, Inc.<br>7555 Innovation Way<br>Mason, Ohio 45040<br>P: 1 (513) 755-4100 |
|----------------------|-----------------------------------------------------------------------------------|
| Contact Person:      | Melissa Smallwood<br>Regulatory Affairs Specialist                                |
| Alternate Contact:   | Jonathan McElwee<br>Manager, Regulatory Affairs                                   |
| Date Prepared:       | 01/12/2018                                                                        |
| II.<br>Device        |                                                                                   |
| Name of Device:      | AtriCure cryoICE cryo-ablation probe                                              |
| Common Name:         | Cryosurgical Probe                                                                |
| Classification Name: | Unit, Cryosurgical, Accessories<br>Surgical, General and Plastic Surgery          |
| Regulatory Class:    | Class II                                                                          |
| Product Code:        | GXH                                                                               |

#### lll. Predicate Device

The device proposed for modification in this submission is the AtriCure cryolCE cryo-ablation probe cleared via K142203 on November 25, 2014.

The following reference devices were also used in this submission:

- K142441 AtriCure cryoICE cryoablation probe ●
- K152337 AtriCure cryoFORM cryoablation probe o

#### IV. Device Description

The cryosurgical handpiece, or cryo-ablation probe, utilized in the AtriCure Cryosurgical System is a hand held, single use, cryosurgical instrument intended for the cryosurgical treatment of cardiac arrhythmias during cardiac surgery; it is also intended for use in blocking pain by temporarily ablating peripheral nerves. The cryosurgical handpiece utilizes a high-pressure cryogen (nitrous oxide, N2O) to freeze target tissues, creating an inflammatory response, and ultimately, cryonecrosis. The cryosurgical handpiece provides probe temperatures below -40°C, the temperature at which intracellular ice formation occurs which is considered lethal to cells. When high pressure nitrous oxide is supplied to the cryoprobe via the AtriCure Cryo Module, rapid cooling is achieved via the Joule-Thompson effect. The end effector, or cryotip, of the probe is malleable to allow access to varying and anatomical sites.

{4}------------------------------------------------

The cryo-ablation probes are comprised of a cryotip end effector, shaft, handle, thermocouple, inlet tube, and exhaust tube. The cryotip consists of an aluminum boiler and three internal inlet orifices distributed throughout the cryotip internally to provide uniform cooling. The 4-mm diameter, cryotip is malleable throughout its 10-cm length. A supplied forming tool can be used to bend the desired form. The cryotip is attached to an insulated rigid shaft that allows the surgeon to adjust the length of the exposed cryotip up to 10 cm in therapeutic length. A thermocouple is affixed to the proximal external surface of the shaft to display real time temperatures on the console. The handle is attached to the shaft. Inlet and exhaust tubes and thermocouple wire pass through the handle and connect to the AtriCure Cryo Module (ACM).

### V. Indications For Use

AtriCure's cryolCE cryo-ablation probes are sterile, single use devices intended for use in the cryosurgical treatment of cardiac arrhythmias by freezing target tissues, creating an inflammatory response (cryonecrosis) that blocks the electrical conduction pathway. The probe is also intended for use in blocking pain by temporarily ablating peripheral nerves.

### VI. Comparison Of Technological Characteristics With The Predicate Device

Minor modifications were made to the non-patient contacting tube-sets of the cryo-ablation probes. The packaging was modified to align with AtriCure's current product lines.

- . The devices have the same intended use, and;
- No changes were made in operating principle, or specifications of performance, and;
- Both the previously cleared and proposed CRYO2 probe use the same patient contacting materials, ● and:
- . The results of the verification and validation testing:
	- o Demonstrated equivalency in performance
	- Did not raise any new issues of safety o

The modifications to the proposed AtriCure cryolCE cryoablation probe are intended to 1.) align components with the previously cleared cryolCE cryoFORM cryoablation probe, improve manufacturability, and 2.) align the device packaging with AtriCure's commercial product lines.

#### VII. Performance Data

Non-clinical Bench Testing

- Mechanical Reliability
- Transit ●
- Shelf-life ●
- Cryogen Performance/Thermal Insulation ●

#### VIII. Conclusions

The modified AtriCure cryolCE cryoablation probe (CRYO2) is equivalent to the previously cleared AtriCure cryolCE cryoablation probe (CRYO2) as there is no change to intended use, operating principals, or function of the device.

---

**Source:** [https://fda.innolitics.com/device/K180138](https://fda.innolitics.com/device/K180138)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
