Ensizor® Monopolar Endoscopic Scissors are designed to cut and dissect tissue, cauterize tissue, cut sutures and cut non-metallic materials during all flexible endoscopic procedures.
Device Story
Single-use, sterile, monopolar electrosurgical scissors for flexible endoscopy. Device consists of an actuation handle, flexible shaft, and distal scissor blades. Operated by physician during endoscopic procedures; compatible with flexible endoscopes (min 2.8 mm channel). Nitinol actuator wire enables navigation through tortuous paths; rotation feature allows radial positioning. Blades feature blunt tips for atraumatic dissection and rounded surfaces for cauterization when closed. Connects to external monopolar HF generator for electrosurgical cutting and cauterization. Provides mechanical cutting and electrosurgical tissue management; enhances procedural efficiency in tissue dissection and suture cutting.
Clinical Evidence
Bench testing only. Performance evaluated via comparative testing against predicate devices, including cutting efficacy on various materials (rubber, polyethylene, multiple suture types) and electrosurgical functionality (cauterization on raw chicken). Insulation testing performed. All units passed all tests.
Technological Characteristics
Single-use, ethylene oxide sterilized. Materials include nitinol (actuator wire). Dimensions: 2.6 mm diameter, 165 cm or 235 cm length. Energy source: external monopolar HF generator. Features: Edge Flex™ blade technology, radial rotation, blunt-tip design for atraumatic dissection. No software or electronic components.
Indications for Use
Indicated for patients undergoing flexible endoscopic procedures requiring tissue dissection, cutting, cauterization, or cutting of sutures and non-metallic materials.
Regulatory Classification
Identification
An endoscopic electrosurgical unit and accessories is a device used to perform electrosurgical procedures through an endoscope. This generic type of device includes the electrosurgical generator, patient plate, electric biopsy forceps, electrode, flexible snare, electrosurgical alarm system, electrosurgical power supply unit, electrical clamp, self-opening rigid snare, flexible suction coagulator electrode, patient return wristlet, contact jelly, adaptor to the cord for transurethral surgical instruments, the electric cord for transurethral surgical instruments, and the transurethral desiccator.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 29, 2018
Slater Endoscopy, LLC % Craig Pagan Consultant C2C Development, LLC 1135 W NASA Blvd. Suite 500 Melbourne, Florida 32901
Re: K180134
Trade/Device Name: Ensizor Monopolar Endoscopic Scissors Regulation Number: 21 CFR 876.4300 Regulation Name: Endoscopic electrosurgical unit and accessories Regulatory Class: Class II Product Code: KGE, OCZ Dated: Mav 30, 2018 Received: May 30, 2018
Dear Craig Pagan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
## K180134
Device Name Ensizor® Monopolar Endoscopic Scissors
Indications for Use (Describe)
Ensizor® Monopolar Endoscopic Scissors are designed to cut and dissect tissue, cut sutures and cut nonmetallic materials during all flexible endoscopic procedures ..
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
## Submitter's name, address, telephone number, a contact person and date (1) summary was prepared:
Slater Endoscopy, LLC 14000 NW 58 Court Miami Lakes, FL 33014 Telephone: (305) 889-3350 Contact: John Starkey, COO / VP Quality Assurance Jan 10, 2018
#### Name(s) of device: (2)
| Proprietary/Trade Name: | Ensizor® Monopolar Endoscopic Scissors |
|-------------------------|-------------------------------------------------|
| Common Name: | Forceps, Biopsy, Electric |
| Regulation Number: | 876.4300 |
| Regulation Name: | Endoscopic electrosurgical unit and accessories |
| Classification Panel: | Gastroenterology/Urology |
| Product Code: | KGE, OCZ |
| Regulatory Class: | II |
## Legally Marked Predicate Device to which the submitter claims substantial (3) equivalence:
The Ensizor® Monopolar Endoscopic Scissors is substantially equivalent to the Ensizor® Endoscopic Scissors (K150939) and the Apollo Endosurgery Endoscopic Scissors (K090583).
#### Description of device(s): (4)
Ensizor® Monopolar Endoscopic Scissors
- Cat # MES-26165 2.6 mm Monopolar Endoscopic Scissors x 165cm insertion length
- Cat # MES-26235 2.6 mm Monopolar Endoscopic Scissors x 235cm insertion ● length
The Ensizor Monopolar Endoscopic Scissors is a single-use ethylene oxide sterile electrosurgical scissors. This device is designed to be used with an electrosurgical monopolar HF generator to cut and dissect tissue, cauterize tissue, cut sutures and cut non-metallic materials during all flexible endoscopic procedures.
The device is compatible with flexible endoscopes with a minimum channel diameter of 2.8 mm. The maximum insertion portion diameter is 2.6 mm, and insertion lengths are 165 cm and 235 cm. The patented Edge Flex™ technology used for the Scissor Blades ("Blades") enhances the cutting performance of this flexible endoscopic
{4}------------------------------------------------
scissors. The Ensizor's rotation feature enables accurate-radial positioning, and the nitinol actuator wire allows for tortuous endoscopic paths while minimizing loss of distal control. The Blades are designed with blunt tips to enable the atraumatic dissection of tissue planes and when closed, make a non-sharp rounded cauterizing surface.
### Statement of intended use: (5)
Ensizor® Monopolar Endoscopic Scissors are designed to cut and dissect tissue, cauterize tissue, cut sutures and cut non-metallic materials during all flexible endoscopic procedures.
### Comparison of Technological Characteristics to Predicate Device: (6)
The Ensizor® Monopolar Endoscopic Scissors have the same technological characteristics as the predicate devices as described below:
- Devices are designed to dissect, cut and cauterize tissue during endoscopic procedures.
- . Devices consist principally of an actuation handle and a flexible shaft body terminating in a pair of cutting Scissors.
- . Scissors function just like standard scissors for mechanical cutting. By operating the handle, the Scissors open and close.
### (7) Performance Data:
The following nonclinical testing was performed in order to evaluate the substantial equivalence of the Ensizor® Monopolar Endoscopic Scissors to the predicate devices.
| | Test | Ensizor<br>Monopolar<br>Endoscopic<br>Scissors | Apollo<br>Endosurgery<br>Endoscopic<br>Monopolar<br>Scissors |
|-----------------------------------------------|------------------|------------------------------------------------|--------------------------------------------------------------|
| Product Attribute Testing | All units passed | All units passed | |
| Product Performance Testing | All units passed | All units passed | |
| Sample Cutting | | | |
| .008 In. thickness rubber (Dental Dam) | All units passed | All units passed | |
| .002 In. thickness LDPE Polyethylene<br>Sheet | All units passed | All units passed | |
| 2-0 Polydioxanone Suture | All units passed | All units passed | |
| 3-0 Polydioxanone Suture | All units passed | All units passed | |
| 2-0 Polypropylene Suture | All units passed | All units passed | |
Table 2 510(k) Summary - Testing Summary
{5}------------------------------------------------
| Test | Ensizor<br>Monopolar<br>Endoscopic<br>Scissors | Apollo<br>Endosurgery<br>Endoscopic<br>Monopolar<br>Scissors |
|------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------|--------------------------------------------------------------|
| 3-0 Polypropylene Suture | All units passed | All units passed |
| 2-0 Silk Sutures | All units passed | All units passed |
| 3-0 Silk Sutures | All units passed | All units passed |
| 2-0 Vicryl Sutures | All units passed | All units passed |
| 3-0 Vicryl Sutures | All units passed | All units passed |
| Raw chicken for the electrocautery<br>functionality test (cut at least 10 times each<br>sample material and 10 tissue cauterization<br>tests). | All units passed | All units passed |
| Insulation Test | All units passed | All units passed |
### (8) Conclusions:
Based on the non-clinical performance data performed comparing the Ensizor® Monopolar Endoscopic Scissors to the predicate devices, it is concluded that the device is as safe, as effective, and performs as well as or better than the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.