K180123 · Noras Mri Products GmbH · MOS · Feb 16, 2018 · Radiology
Device Facts
Record ID
K180123
Device Name
Breast BI 7 MR Coil 1.5T Mammavention
Applicant
Noras Mri Products GmbH
Product Code
MOS · Radiology
Decision Date
Feb 16, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1000
Device Class
Class 2
Indications for Use
The intended use of Breast BI 7 MR Coil 1.5T Mammavention is, in conjunction with a Magnetic Resonance Scanner, the MR examination of the female breast. It is indicated for use as a diagnostic imaging device to produce transverse, sagittal, coronal and oblique images of the internal structures of the female breast. When interpreted by a trained physician, these images provide information that can be useful in determining diagnosis. The included Breast Biopsy Unit permits MR guided breast biopsy and wire localization of lesions can be performed by a trained physician.
Device Story
Device is a dedicated breast receive-only MRI coil system; used with Siemens 1.5T MRI scanners (Avanto fit, Aera, Amira). Coil detects high-frequency signals from hydrogen nuclei induced by the MRI scanner's transmitter; signals are digitized and processed by the host MRI system to generate diagnostic images. Includes a Breast Biopsy Unit for MR-guided biopsy and wire localization. Operated by trained physicians in clinical imaging environments. Output consists of high-resolution anatomical images of breast structures; assists physicians in lesion diagnosis and interventional guidance. Benefits include improved visualization of internal breast structures for diagnostic and biopsy procedures.
Clinical Evidence
Bench testing only. Device performance and safety parameters verified against FDA-recognized NEMA standards and IEC 60601-2-33:2011. Power tests conducted for the integrated system.
Technological Characteristics
Receive-only RF coil for 1.5T MRI; compatible with Siemens Avanto fit, Aera, and Amira systems. Complies with IEC 60601-2-33:2011. No internal software algorithms; functions as a passive signal receiver for the host MRI system.
Indications for Use
Indicated for MR examination of female breast to produce transverse, sagittal, coronal, and oblique images for diagnostic purposes; includes MR-guided breast biopsy and wire localization of lesions by trained physicians.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text.
February 16, 2018
Noras MRI products GmbH Jens Heidenmann Official Correspondent Leibnizstrasse 4 Hoechberg, 97204 De
Re: K180123
Trade/Device Name: Breast BI 7 MR Coil 1.5T Mammavention Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: Class II Product Code: MOS Dated: December 28, 2017 Received: January 16, 2018
Dear Jens Heidenmann:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
## Page 2 - Jens Heidenmann
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Michael D. O'Hara
For
Robert A. Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K 180123
Device Name
Breast BI 7 MR Coil Mammavention 1.5T (PN:118879)
### Indications for Use (Describe)
The intended use of Breast BI 7 MR Coil 1.5T Mammavention with a Magnetic Resonance Scanner, the MR examination of the female breast. It is indicated for use as a diagnostic imaging device to produce transverse, sagittal, coronal and oblique images of the internal structures of the female breast. When interpreted by a trained physican, these images provide information that can be useful in determining diagnosis. The included Breast Biopsy Unit permits MR guided breast biopsy and wire localization of lesions chan be performed by a trained physician.
The coil system Breast BI 7 MR Coil Mammavention 1.5T can be used with the following MRI systems: 1.5T: Siemens 1.5T: Avanto fit, Aera, Amira
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for NORAS MRI products. The word "NORAS" is in large, bold, blue letters. To the right of the word "NORAS" is a blue line that resembles a heartbeat, and the words "MRI products" are to the right of the line.
# 510(k) SUMMARY
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for NORAS MRI products. The word "NORAS" is in large, bold, blue letters. To the right of the word "NORAS" is a blue line that resembles a heartbeat. To the right of the heartbeat line are the words "MRI products" in smaller, blue letters. The logo is set against a white background with a blue shape on the left side.
## 510(k) Summary
Breast BI 7 MR Coil 1.5T Mammavention
Date of Summary Preparation: 2018 - 02 - 14
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR § 807.92.
### 1. General Information
Importer/Distributor Name and Address NORAS MRI products GmbH Leibnizstr.4 97204 Hoechberg / Germany
ERN: 3004929307 Owner/Operator Number: 9071737
Manufacturing Site Name and Address NORAS MRI products GmbH Leibnizstr.4 97204 Hoechberg / Germany
ERN: 3004929307 Owner/Operator Number: 9071737
### 2. Contact Person
Jens Heidenmann QA & RA Management NORAS MRI products GmbH Leibnizstr.4 97204 Hoechberg Germany Tel: (+49) 931 / 29927-17 Fax: (+49) 931 / 29927-20 jens.heidenmann@noras.de
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for NORAS MRI products. The logo is in blue and white, and it features the word "NORAS" in large, bold letters. To the right of the word "NORAS" is a line that resembles a heart rate monitor. To the right of the line is the text "MRI products" in smaller letters.
### 3. Device Name and Classification
| Trade Name: | Breast BI 7 MR Coil 1.5T Mammavention (PN:118879) |
|-----------------------|---------------------------------------------------|
| Common Name: | Breast BI 7 MR Coil 1.5T Mammavention |
| Classification Name: | Magnetic Resonance Diagnostic Device |
| Classification Panel: | Radiology |
| CFR Number: | 21 CFR § 892.1000 |
| Device Class: | II |
| Product Code: | 90MOS |
| Submission Number: | K180123 |
### 4. Device Description
The Breast BI 7 MR Coil 1.5T Mammavention described in this document has been designed, depending upon model type, for use with a SIEMENS MRI system with field strength of 3 T. The coil system serves solely as a receiving coil for the reception of high frequency signals from the hydrogen -(1-H) nuclei. The hydrogen nuclei are induced into precession by the transmitting coil of the MRT device. The processing magnetization induces potential differences in the Breast Bl 7 MR Coil 1.5T Mammavention which are digitized and further processed in the MRT system
#### 5. Intended Use / Indications for Use
The intended use of Breast BI 7 MR Coil 1.5T Mammavention is, in conjunction with a Magnetic Resonance Scanner, the MR examination of the female breast. It is indicated for use as a diagnostic imaging device to produce transverse, sagittal, coronal and oblique images of the internal structures of the female breast. When interpreted by a trained physician, these images provide information that can be useful in determining diagnosis.
The included Breast Biopsy Unit permits MR guided breast biopsy and wire localization of lesions can be performed by a trained physician.
The coil system Breast BI 7 MR Coil 1.5T Mammavention can be used with the following MRI systems:
1.5T: Siemens 1.5T:
Avanto fit, Aera, Amira
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for NORAS MRI products. The logo is in blue and white. The word "NORAS" is in large, bold letters. To the right of the word "NORAS" is a line that resembles a heartbeat, and the words "MRI products" are below the line. The logo is simple and professional.
### Substantial Equivalence 6.
NORAS MRI product GmbH believes that, within the meaning of the Safe Medical Devices Act of 1990, the Breast BI 7 MR Coil 1.5T Mammavention is substantially equivalent to the following multipurpose coil:
| Predicate<br>Device Name<br>and<br>Manufacturer | 510(k)<br>Number | Clearance<br>Date | Product Code | Comparable<br>Properties |
|-------------------------------------------------|------------------|--------------------|--------------|--------------------------------------------------------------------|
| Breast BI 7 MR<br>Coil 3T<br>Mammavention | K162651 | January 18<br>2017 | 90MOS | Proton imaging<br>High resolution of<br>breast anatomic<br>regions |
### Summary of Technological Characteristics of the Principal Device as 7. Compared with the predicate Device
Summary of technological characteristics of the Breast Bl 7 MR Coil 1.5T Mammavention are the same as for the predicate device Breast BI 7 MR Coil 3T Mammavention
### General Safety and Effectiveness Concerns 8.
The Breast BI 7 MR Coil 1.5T Mammavention is conform with the FDA recognized NEMA Standards for the measurement of performance and safety parameters and the IEC standards for safety issues with the Magnetic Resonance Imaging Devices, IEC 60601-2-33:2011. All device testing have been completed successfully before device clearance. This assures that the performance of this device can be considered safe and effective when used with the currently available Siemens MAGNETOM 1.5T The power tests which have been done by MRI manufacturer for the whole system can be found in 076 Third Party Power Test Breast Bl 7 MR Coil 1.5T.pdf.
### 9. Conclusion as to Substantial Equivalence
NORAS MRI products GmbH believes that, within the definition of the Safe Medical Devices Act of 1990, the Breast Bl 7 MR Coil 1.5T Mammavention is substantially equivalent to the predicate device listed above.
Jens Kude
Jens Heidenmann QM & RA Manager
February 14. 2018
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.