← Product Code [FGE](/productcode/FGE) · K180086

# Gyrus ACMI – EZDilate 3-Stage Balloon Dilatation Catheter (K180086)

_Gyrus Acmi, Inc. · FGE · Mar 8, 2018 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K180086

## Device Facts

- **Applicant:** Gyrus Acmi, Inc.
- **Product Code:** [FGE](/productcode/FGE.md)
- **Decision Date:** Mar 8, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.5010
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

The Gyrus ACMI EZDilate 3-Stage Balloon Dilatation Catheters are in adult and adolescent populations to endoscopically dilate strictures of the alimentary tract. It is also indicated in adults for endoscopic dilatation of the Sphincter of Oddi with or without prior sphincterotomy.

## Device Story

The Gyrus ACMI EZDilate 3-Stage Balloon Dilatation Catheter is a wire-guided, semi-compliant nylon balloon catheter used for endoscopic dilation of alimentary tract strictures and the Sphincter of Oddi. The device consists of a catheter shaft and a distal balloon that provides progressive radial expansion through three target sizes. It is operated by a physician using a separately sold 60cc inflation device. The balloon is coated with silicone to facilitate insertion, positioning, and removal. The device is used in a clinical setting (endoscopy) and is compatible with 0.035-inch guidewires. The output is the mechanical dilation of the target stricture, which assists in treating obstructions or facilitating access. The device is provided sterile for single-patient use.

## Clinical Evidence

Bench testing only. Performance testing included visual inspection, dimensional measurements, tensile, fatigue, luer gauging, working length, tip stiffness, compliance, burst, insertion/retrieval force, friction, deflation, and endoscope compatibility testing.

## Technological Characteristics

Semi-compliant nylon blow-molded balloon; silicone coating; 0.035-inch guidewire compatible; ethylene oxide sterilization; 3-year shelf life; 240cm catheter length; 5.5cm/6.5cm balloon lengths; mechanical inflation via external 60cc device.

## Regulatory Identification

A biliary catheter and accessories is a tubular flexible device used for temporary or prolonged drainage of the biliary tract, for splinting of the bile duct during healing, or for preventing stricture of the bile duct. This generic type of device may include a bile collecting bag that is attached to the biliary catheter by a connector and fastened to the patient with a strap.

## Special Controls

*Classification.* Class II (special controls). The device, when it is a bile collecting bag or a surgical biliary catheter that does not include a balloon component, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.

## Predicate Devices

- Gyrus ACMI – EZDilate 3-Stage Balloon Dilatation Catheter ([K143609](/device/K143609.md))
- Boston Scientific CRE Dilatation Balloon ([K112994](/device/K112994.md))

## Reference Devices

- Gastroenterology and Urology Guidewire ([K935198](/device/K935198.md))
- Atrion Inflation Device ([K032840](/device/K032840.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

March 8, 2018

Gyrus ACMI, Inc. Mary Anne Patella Senior Specialist Regulatory Affairs 136 Turnpike Road Southborough, MA 01772

Re: K180086

Trade/Device Name: Gyrus ACMI EZDilate 3-Stage Balloon Dilatation Catheter Regulation Number: 21 CFR§ 876.5010 Regulation Name: Biliary catheter and accessories Regulatory Class: II Product Code: FGE, KNO Dated: January 11, 2018 Received: January 12, 2018

Dear Mary Anne Patella:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice. labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Benjamin R. Fisher -S

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K180086

Device Name

Gyrus ACMI EZDilate 3-Stage Balloon Dilatation Catheter

Indications for Use (Describe)

The Gyrus ACMI EZDilate 3-Stage Balloon Dilatation Catheters are in adult and adolescent populations to endoscopically dilate strictures of the alimentary tract. It is also indicated in adults for endoscopic dilatation of the Sphincter of Oddi with or without prior sphincterotomy.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

| <span style="font-size: 16px;"> <input checked="true" type="checkbox"/> </span> Prescription Use (Part 21 CFR 801 Subpart D) |
|------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-size: 16px;"> <input type="checkbox"/> </span> Over-The-Counter Use (21 CFR 801 Subpart C)                 |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Traditional 510(k) Notification January 11, 2018

#### 510(k) Summary Gyrus ACMI, Inc. Gyrus ACMI - EZDilate 3-Stage Balloon Dilatation Catheter

## General Information

| Contract Manufacturer:                                                                    | Vention Medical, Inc.<br>261 Cedar Hill Drive<br>Marlborough, MA 01752<br>Phone: 508-481-6233                                           |
|-------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|
| Establishment Registration Number:                                                        | 3004734318                                                                                                                              |
| 510(k) Submitter:                                                                         | Olympus Surgical Technologies America<br>Gyrus ACMI, Inc.<br>136 Turnpike Rd.<br>Southborough, MA 01772-2104                            |
| Establishment Registration Number:                                                        | 3003790304                                                                                                                              |
| Contact Person:                                                                           | Mary Anne Patella<br>Senior Specialist Regulatory Affairs<br>508-804-2771<br>Maryanne.patella@olympus-osta.com                          |
| Date Prepared:                                                                            | January 11, 2018                                                                                                                        |
| Device Description                                                                        |                                                                                                                                         |
| Classification Name:<br>CFR citation<br>Procode<br>Classification<br>Classification Panel | Biliary catheter and accessories<br>Esophageal dilator<br>21 CFR 876.5010, 876.5365<br>FGE, KNQ<br>Class II<br>Gastroenterology/Urology |
| Trade Name:                                                                               | Gyrus ACMI – EZDilate 3-Stage Balloon<br>Dilatation Catheter                                                                            |
| Generic/Common Name:                                                                      | Balloon Dilation Catheter                                                                                                               |
| Predicate Devices                                                                         |                                                                                                                                         |

| Gyrus ACMI – EZDilate 3-Stage Balloon Dilatation Catheter | K143609 |
|-----------------------------------------------------------|---------|
| Boston Scientific CRE Dilatation Balloon                  | K112994 |

{4}------------------------------------------------

K180086 Page 2 of 4

#### Product Description

Both the predicate and proposed Gyrus ACMI EZDilate 3-Stage Balloon Dilatation Catheters consist of a semi-compliant nylon blow molded balloon; and, a shaft that communicates a fluid passage for expansion and collapse of the balloon portion, operated by an inflation device.

Like the predicate, the EZDilate Wire Guided Balloon is designed to be used with a 0.035 in. (0.89mm) guidewire and has a wire lumen sized appropriately. Each balloon catheter is packaged with a soft tip 0.035 in. (0.89mm) guidewire pre-loaded into the guidewire lumen. The guidewire is manufactured by Lake Region Medical; and, is cleared under K935198, Gastroenterology and Urology Guidewire Modifications. All wire-guided balloons with have lengths of 5.5cm and a catheter length of 240cm.

The balloon is inflated with a 60cc inflation device, which will be sold separately. The suggested inflation device is manufactured by Atrion (K032840).

#### Technological Characteristics

The proposed Gyrus ACMI EZDilate 3-Stage Balloon Dilatation Catheter has the same intended use, design, and scientific technology as the predicate EZDilate 3-Stage Balloon Dilatation Catheter (K143609). Both devices are the same design and there were no new issues of safety or effectiveness with the proposed device.

The Gyrus ACMI EZDilate 3-StageWire Guided Balloon Dilatation Catheter is a reinforced catheter attached to a distal dilatation balloon. The semi-compliant balloon allows for a progressive radial expansion through three target sizes using one balloon. A silicone coating applied to the balloon increases ease of insertion, positioning of the balloon within the alimentary tract, and device removal.

#### Material

No material changes were made to the predicate EZDilate 3-Stage Balloon Dilatation Catheter cleared under K143609.

#### Intended Uses

The Gyrus ACMI EZDilate 3-Stage Balloon Dilatation Catheters are indicated for use in adult and adolescent populations to endoscopically dilate strictures of the alimentary tract. It is also indicated in adults for endoscopic dilatation of the Sphincter of Oddi with or without prior sphincterotomy.

{5}------------------------------------------------

### Summary of Sterilization and Shelf Life Discussion

Like the predicate EZDilate 3-Stage Balloon Dilatation Catheter (K143609), the proposed EZDilate 3-Stage Balloon Dilatation Catheter will be distributed in a sterile state and is intended for single patient use only. The sterilization method used is ethylene oxide and has a shelf life of three (3) years.

#### Summary of Performance Testing

The predicate EZDilate 3-Stage Balloon Dilatation Catheter (K143609) and the proposed EZDilate 3-Stage Balloon Dilatation Catheter are identical in every respect. Minor process changes to the predicate EZDilate - Wire Guided 18-19-20 balloon catheter resulted in rounders to the balloon; enhancing balloon visualization. As a result of the process changes, the following performance tests were repeated.

- Visual Inspection ●
- Dimensional Measurements ●
- Tensile Testing
- Fatigue Testing
- Luer Gauging Test ●
- Balloon Working Length
- Tip Stiffness Testing
- Compliance Testing
- Balloon Burst Testing ●
- Balloon Insertion Force Testing
- Balloon Retrieval Force Testing
- Balloon Friction Testing ●
- Balloon Deflation Testing ●
- Balloon Endoscope Compatibility Testing

#### Substantial Equivalence

The proposed Gyrus ACMI EZDilate 3-Stage Balloon Dilatation Catheter has the same design, and scientific technology as the Gyrus ACMI EZDilate 3-Stage Balloon Dilatation Catheter; and the same intended use population as the Predicate Boston Scientific CRE Dilatation Balloon (K112994). Both devices are of similar design and there were no new issues of safety or effectiveness with the proposed device. Please see the following Substantial Equivalence Comparison Table:

{6}------------------------------------------------

| Traditional 510(k) Notification |
|---------------------------------|
| January 11, 2018                |

| Design Feature                                                 | Proposed Gyrus ACMI<br>EZDilate 3-Stage<br>Balloon Dilatation<br>Catheter - Wire Guided                                                                                                                                                                                                                                              | Predicate Gyrus ACMI<br>EZDilate 3-Stage<br>Balloon Dilatation<br>Catheter - Wire Guided<br>(K143609)                                                                                                                                                                                                                 | Predicate Boston<br>Scientific CRE™<br>Dilatation Balloon<br>(K112994)                                                                                                                                                                                                                                                            | Comments                                                                                                  |                                                                                               |                                                                                |                                                                                                           |
|----------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------|
| Indication for<br>Use                                          | The EZDilate 3-Stage<br>Balloon Dilatation<br>Catheters are indicated<br>for use in adult and<br>adolescent populations to<br>endoscopically dilate<br>strictures of the<br>alimentary tract. It is also<br>indicated in adults for<br>endoscopic dilatation of<br>the Sphincter of Oddi<br>with or without prior<br>sphincterotomy. | The EZDilate 3-Stage<br>Balloon Dilatation<br>Catheters are indicated<br>for use in adult<br>populations to<br>endoscopically dilate<br>strictures of the<br>alimentary tract. It is also<br>indicated in adults for<br>endoscopic dilatation of<br>the Sphincter of Oddi<br>with or without prior<br>sphincterotomy. | The CRE™ Wireguided<br>Balloon Dilatation<br>Catheter is indicated for<br>use in adult and<br>adolescent populations to<br>endoscopically dilate<br>strictures of the<br>alimentary tract. It is also<br>indicated in adults for<br>endoscopic dilatation of<br>the Sphincter of Oddi<br>with or without prior<br>sphincterotomy. | Identical to the<br>predicate Boston<br>Scientific CRETM<br>Dilatation Balloon<br>(K112994)               |                                                                                               |                                                                                |                                                                                                           |
| Balloon Length                                                 | 5.5cm, 6.5cm                                                                                                                                                                                                                                                                                                                         | 5.5cm, 6.5cm                                                                                                                                                                                                                                                                                                          | 5.5cm                                                                                                                                                                                                                                                                                                                             | Similar (The 6-7-8 and<br>8.5-9.5-10.5 balloons<br>are 6.5cm)                                             |                                                                                               |                                                                                |                                                                                                           |
| Inflated Outer<br>Diameter<br>(OD)/Inflation<br>Pressure (ATM) | Inflated<br>OD<br>6-7-8<br>8.5-9.5-<br>10.5<br>11-12-13<br>13.5-<br>14.5-15.5<br>16-17-18<br>18-19-20                                                                                                                                                                                                                                | Pressure<br>ATM<br>2.0-5.5-9.5<br>2.0-5.5-9.5<br>2.0-3.5-5.5<br>2.5-4.0-6.0<br>2.5-4.0-5.5<br>2.5-3.5-5.0                                                                                                                                                                                                             | Inflated<br>OD<br>6-7-8<br>8.5-9.5-<br>10.5<br>11-12-13<br>13.5-<br>14.5-<br>15.5<br>16-17-18<br>18-19-20                                                                                                                                                                                                                         | Pressure<br>ATM<br>2.0-5.5-9.5<br>2.0-5.5-9.5<br>2.0-3.5-5.5<br>2.5-4.0-6.0<br>2.5-4.0-5.5<br>2.5-3.5-5.0 | Inflated<br>OD<br>6-7-8<br>8-9-10<br>10-11-12<br>12-13.5-<br>15<br>15-16.5-<br>18<br>18-19-20 | Pressure<br>ATM<br>3-6-10<br>3.5-5-9<br>3-5-8<br>3-4.5-8<br>3-4.5-7<br>3-4.5-6 | Similar (Identical<br>maximum/minimum<br>inflated outer<br>diameters.<br>Intermediate sizes may<br>vary.) |
| Device<br>Working<br>Length                                    | 240 cm                                                                                                                                                                                                                                                                                                                               | 240 cm                                                                                                                                                                                                                                                                                                                | 240 cm (Esophageal,<br>Pyloric, Colonic &<br>Biliary)<br>180 cm (Esophageal,<br>Pyloric & Biliary)                                                                                                                                                                                                                                | Similar (Boston<br>Scientific offers an<br>additional length.)                                            |                                                                                               |                                                                                |                                                                                                           |
| Compatible<br>Guidewire                                        | 0.035"                                                                                                                                                                                                                                                                                                                               | 0.035"                                                                                                                                                                                                                                                                                                                | 0.035"                                                                                                                                                                                                                                                                                                                            | Identical                                                                                                 |                                                                                               |                                                                                |                                                                                                           |
| Semi-<br>Compliant<br>Balloon                                  | Yes                                                                                                                                                                                                                                                                                                                                  | Yes                                                                                                                                                                                                                                                                                                                   | Yes                                                                                                                                                                                                                                                                                                                               | Identical                                                                                                 |                                                                                               |                                                                                |                                                                                                           |
| Radiopaque<br>Marker                                           | Yes                                                                                                                                                                                                                                                                                                                                  | Yes                                                                                                                                                                                                                                                                                                                   | Yes                                                                                                                                                                                                                                                                                                                               | Similar (Marker<br>location may vary.)                                                                    |                                                                                               |                                                                                |                                                                                                           |
| Rounded<br>balloon taper                                       | Yes                                                                                                                                                                                                                                                                                                                                  | Yes                                                                                                                                                                                                                                                                                                                   | Yes                                                                                                                                                                                                                                                                                                                               | Similar (Facilitates<br>endoscopic<br>visualization of the<br>stricture through the<br>balloon.)          |                                                                                               |                                                                                |                                                                                                           |

|  |  | Substantial Equivalence Comparison Table |  |  |
|--|--|------------------------------------------|--|--|
|--|--|------------------------------------------|--|--|

## Conclusion:

In summary, the Gyrus ACMI EZDilate 3-Stage Balloon Dilatation Catheter is substantially equivalent to the predicate devices and presents no new questions of safety or effectiveness.

---

**Source:** [https://fda.innolitics.com/device/K180086](https://fda.innolitics.com/device/K180086)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).