The Materialise PKA Guide System is intended to be used as a surgical instrument to assist in the positioning of Partial Knee Replacement components intra-operatively and in guiding the marking of bone before cutting and to guide cutting of the bone. The Materialise PKA Guide System is to be used with ZUK UNI, JOURNEY™ UNI, JOURNEY II UNI, JZ (Hybrid) UNI knee systems. Vanguard™ M Unicompartmental Knee System and Oxford® Partial Knee System prostheses families only. The Zimmer Biomet Patient Specific Instruments are compatible for use with the Oxford® Partial Knee System as approved in P010014. The Materialise PKA Guides are intended for single use only.
Device Story
System uses MRI images as input to generate pre-surgical plan via Materialise PKA Planner software; surgeon inspects, fine-tunes, and approves plan pre-operatively. System produces patient-specific templates (femur/tibia guides) and bone models. Used in clinical setting by orthopedic surgeons to transfer pre-operative plan to patient intra-operatively. Guides assist in positioning/aligning partial knee replacement components, marking bone, and guiding bone cuts. Benefits include improved surgical accuracy and alignment for partial knee replacement procedures.
Clinical Evidence
No new clinical data or bench testing performed. Substantial equivalence supported by previous testing for biocompatibility, sterility, cleaning, debris, and dimensional stability from predicate devices. Usability evaluations conducted for software validation.
Technological Characteristics
Patient-specific surgical templates (femur/tibia) and bone models. Software-based pre-surgical planning system. Single-use. Compliant with ISO 9000/13485 and FDA 21 CFR 820/Part 11. Planning software includes Oxford-specific and Vanguard M-specific algorithms.
Indications for Use
Indicated for patients undergoing partial knee replacement surgery requiring assistance in component positioning, bone marking, and bone cutting. Compatible with ZUK UNI, JOURNEY UNI, JOURNEY II UNI, JZ (Hybrid) UNI, Vanguard M Unicompartmental, and Oxford Partial Knee systems.
Regulatory Classification
Identification
A knee joint femorotibial metal/polymer non-constrained cemented prosthesis is a device intended to be implanted to replace part of a knee joint. The device limits minimally (less than normal anatomic constraints) translation in one or more planes. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral condylar resurfacing component or components made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components made of ultra-high molecular weight polyethylene and are intended for use with bone cement (§ 888.3027).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font below.
July 6, 2018
Materialise NV Oliver Clemens Regulatory Officer Technologielaan 15 Leuven, 3001 Be
Re: K173970
Trade/Device Name: Materialise PKA Guide System Regulation Number: 21 CFR 888.3520 Regulation Name: Knee Joint Femorotibial Metal/Polymer Non-Constrained Cemented Prosthesis Regulatory Class: Class II Product Code: HSX, OOG Dated: December 29, 2017 Received: December 29, 2017
Dear Oliver Clemens:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
K173970
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
> Sincerely, Vincent J. Devlin -S for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K173970
Device Name Materialise PKA Guide System
## Indications for Use (Describe)
The Materialise PKA Guide System is intended to be used as a surgical instrument to assist in the positioning of Partial Knee Replacement components intra-operatively and in guiding the marking of bone before cutting and to guide cutting of the bone.
The Materialise PKA Guide System is to be used with ZUK UNI, JOURNEY™ UNI, JOURNEY II UNI, JZ (Hybrid) UNI knee systems. Vanguard™ M Unicompartmental Knee System and Oxford® Partial Knee System prostheses families only.
The Zimmer Biomet Patient Specific Instruments are compatible for use with the Oxford® Partial Knee System as approved in P010014.
The Materialise PKA Guides are intended for single use only.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
The following section is included as required by the Safe Medical Devices Act (SMDA) of 1990 and 21CFR 807.92
| Company name | Materialise N.V. |
|-----------------------------------|-----------------------------------|
| Establishment registration number | 3003998208 |
| Street Address | Technologielaan 15 |
| City | Leuven |
| Postal code | 3001 |
| Country | Belgium |
| Phone number | +32 16 39 62 80 |
| Fax number | +32 16 39 66 06 |
| Principal Contact person | Oliver Clemens |
| Contact title | Regulatory Affairs Officer |
| Contact e-mail address | Regulatory.Affairs@materialise.be |
| Additional contact person | Wim Claassen |
| Contact title | Portfolio Manager |
| Contact e-mail address | Wim.Claassen@materialise.be |
#### Submission date
The date of the Traditional 510(k) submission is December 21, 2017.
# Submission information
| Device name | Materialise PKA Guide System |
|--------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name | Materialise PKA Planner<br>Materialise PKA Guide<br>SurgiCase Planner<br>Visionaire UNI cutting guides<br>Signature Planner<br>Signature Guides<br>Zimmer Patient Specific Instruments Planner<br>Zimmer Patient Specific Instruments |
| Common Name | Knee prosthesis |
| Classification Name | Knee joint patellofemorotibial polymer /metal /polymer semi-<br>constrained cemented prosthesis |
| Primary product code | HSX (21 CFR 888.3520) |
| Subsequent product codes | OOG |
{4}------------------------------------------------
# Predicate Devices
The predicate devices to which substantial equivalence is claimed:
| Trade or proprietary or model name | Materialise PKA Guide System |
|------------------------------------|------------------------------|
| 510(k) number | K172650 |
| Decision date | November 20, 2017 |
| Classification product code | HSX (21 CFR 888.3520) |
| Subsequent product codes | OOG |
| Manufacturer | Materialise N.V. |
| Trade or proprietary or model name | Signature Personalized Patient Care System |
|------------------------------------|--------------------------------------------|
| 510(k) number | K110415 |
| Decision date | May 16, 2011 |
| Classification product code | HRY (21 CFR 888.3530) |
| Subsequent product codes | JWH, MBH, OIY, OOG |
| Manufacturer | Materialise N.V. |
| Trade or proprietary or model name | Materialise TKA Guide System |
|------------------------------------|------------------------------|
| 510(k) number | K173445 |
| Decision date | February 2, 2018 |
| Classification product code | JWH (21 CFR 888.3560) |
| Subsequent product codes | MBH, OIY, OOG |
| Manufacturer | Materialise N.V. |
#### Device Description
Materialise PKA Guides are patient-specific medical devices that are designed to implant partial knee prosthesis.
The Materialise PKA Guides must only be used in conjunction with the 510(k) cleared, legally marketed prosthesis. Consult the prosthesis labeling and instructions for specific patient indications, contraindications, associated
{5}------------------------------------------------
risks, information for use, warnings and precautions. Moterialise PKA Guides is an instrument set containing a femur and/or tibia template(s) and bone models.
## Intended Use
The Materialise PKA Guide System is intended to be used as a surgical instrument to assist in the positioning of Partial Knee Replacement components intra-operatively and in guiding the marking of bone before cutting and to guide cutting of the bone.
The Materialise PKA Guide System is to be used with ZUK UNI, JOURNEY II UNI, JZ (Hybrid) UNI knee systems, Vanguard™ M Unicompartmental Knee System and Oxford® Partial Knee System prostheses families only.
The Zimmer Biomet Patient Specific Instruments are compatible for use with the Oxford® Partial Knee System as approved in P010014.
The Materialise PKA Guides are intended for single use only.
# Functioning of the Device
The Materialise PKA Guide System generates a pre-surgical plan based on MRI images using the Materialise PKA Planner. The software device then is used pre-operatively by a qualified surgeon to inspect, fine-tune and approve the pre-surgical plan. Next, Materialise PKA Guides are designed based on the approved presurgical plan. Materialise PKA Guides are patient specific templates which transfer the pre-operatively determined positioning of the chosen partial knee replacement components to the patient intra-operatively, assisting the surgeon in positioning and aligning the actual partial knee replacement components by guiding the marking of bone before cutting and to guide cutting of the bone.
#### Technological Characteristics
A detailed comparison shows the subject device is substantially equivalent in intended use, design, functionality, operating principles, materials and performance characteristics to the main difference between the subject device and previously cleared predicate device K172650 is the Oxford and Vanguard M implants in the Planner software for the surgeon to select during stage. The software algorithm was extended with an Oxford specific planning algorithm which was previously cleared as part of the software component of predicate device K110415. Furthermore, the medical coordinate system and clinical relevant landmarks for the PKA Oxford specific planning algorithm were aligned with those of the TKA procedure as cleared in predicate device K173445. The changes do not affect the safety and effectiveness of the devices.
#### Performance Data
No new testing of the Materialise PKA Guides was done in support of this pre-market notification. Previous testing for biocompatibility, sterility, cleaning, debris and dimensional stability are applicable to the subject device and demonstrate substantial equivalence with the predicate device. Testing verified that the accuracy and performance of the system is adequate to perform as intended. The stability of the device placement, surgical technique, intended use and functional elements of the subject device are the same as that of the predicate Signature Personalized Patient Care System (K110415) and therefore previous cadaver testing on predicate device K110415 is considered applicable to the subject device. New software validation/verification testing of the Materialise PKA Planner was done in support of this pre-market notification in the form of usability evaluations.
{6}------------------------------------------------
#### Summary
The characteristics that determine the functionality and performance of the Materialise PKA Guide System are substantially equivalent to the devices cleared under K172650 and K110415. The non-clinical testing indicates that the subject device is as safe, as effective, and performs as well as the predicates. The Materialise PKA Guide System will be manufactured in compliance with FDA (CFR 820 & Part 11) and ISO quality system (9000 and 13485) requirements.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.