K173936 · Natus Medical Incorporated Dba Excel-Tech Ltd. (Xltek) · GWE · Jun 15, 2018 · Neurology
Device Facts
Record ID
K173936
Device Name
Natus Photic Stimulator
Applicant
Natus Medical Incorporated Dba Excel-Tech Ltd. (Xltek)
Product Code
GWE · Neurology
Decision Date
Jun 15, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.1890
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Natus Photic Stimulator is indicated for photic activation of the EEG during an EEG study and in the generation of visual evoked potentials.
Device Story
Natus Photic Stimulator is an arm-mounted LED lamp assembly used by EEG technicians in hospital environments (fixed, mobile, or portable systems). Device receives trigger pulses from external acquisition software to generate 1ms white light photic flashes. Flash frequency (0.5 Hz to 60 Hz) and intensity (up to 75,000 lux at 30 cm) are controlled by the acquisition software. Used to test photo-sensibility related to epilepsy and to stimulate visual evoked potentials. Lamp head is manually positioned 30 cm from patient face via ergonomic arm and handgrip. Output provides visual stimulation to patient; healthcare provider observes patient response via EEG or evoked potential monitoring to aid in neurological diagnosis.
Clinical Evidence
Bench testing only. Performance verified against IEC 60601-1, IEC 60601-1-2, IEC 60601-2-40, IEC 60601-1-6, IEC 62366, IEC 62471, and ANSI Z80.36-2016. Results confirm compliance with predetermined specifications and safety standards.
Indicated for photic activation of the EEG during an EEG study and in the generation of visual evoked potentials in patients of all ages.
Regulatory Classification
Identification
An evoked response photic stimulator is a device used to generate and display a shifting pattern or to apply a brief light stimulus to a patient's eye for use in evoked response measurements or for electroencephalogram (EEG) activation.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo, which features a stylized caduceus. To the right is the FDA logo, which is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 15, 2018
Natus Medical Incorporated DBA Excel-Tech Ltd. (XLTEK) Sanjay Mehta Director, Quality Assurance & Regulatory Affairs 2568 Bristol Circle Oakville, ON L6H 5S1 Canada
Re: K173936
Trade/Device Name: Natus Photic Stimulator Regulation Number: 21 CFR 882.1890 Regulation Name: Evoked Response Photic Stimulator Regulatory Class: Class II Product Code: GWE Dated: May 14, 2018 Received: May 16, 2018
Dear Sanjay Mehta:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
## Page 2 - Sanjay Mehta
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jay R. Gupta -S
For Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K173936
Device Name Natus Photic Stimulator
Indications for Use (Describe)
The Natus Photic Stimulator is indicated for photic activation of the EEG study and in the generation of visual evoked potentials.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image contains the logo for Natus Neurology. The word "natus" is in a teal sans-serif font, with a small registered trademark symbol next to the "s". Below "natus" is the word "neurology" in a smaller, italicized, black sans-serif font. The logo is simple and clean, with a focus on the company name.
Page 1 of 4
| Submission Date: | 20 December 2017 | | |
|-----------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------|
| Submitter: | Natus Medical Incorporated DBA Excel-Tech Ltd. (XLTEK)<br>2568 Bristol Circle<br>Oakville, Ontario, L6H 5S1<br>Canada | | |
| Submitter and<br>Application<br>Correspondent | Mr. Sanjay Mehta<br>Phone: +1 (905) 287-5055<br>Fax: +1 (905) 829-5304<br>Email: sanjay.mehta@natus.com | | |
| Manufacturing Site: | Creation Technologies LP<br>6820 Creditview Road<br>Mississauga, Ontario L5N 0A9<br>Canada | | |
| Trade Name: | Natus Photic Stimulator | | |
| Common and<br>Classification<br>Name: | Stimulator, Photic, Evoked Response | | |
| Classification<br>Regulation: | 21 CFR §882.1890 | | |
| Product Code: | GWE | | |
| Substantially<br>Equivalent Devices: | New XLTEK Model | Predicate 510(k)<br>Number | Predicate<br>Manufacturer / Model |
| | XLTEK / Natus Photic<br>Stimulator | K101691 | Lifelines Ltd. / Lifeline<br>Photic Stimulator |
| | | | 510(k): NATUS PHOTIC STIMULATOR<br>510(k) SUMMARY |
| | | | PAGE 2 of 4 |
| <b>Device Description:</b> | The Natus Medical Incorporated (Natus) DBA Excel-Tech Ltd. (XLTEK) Photic Stimulator is used by trained medical staff in a medical environment to apply photic flashes to the patient during neurophysiology studies such as electroencephalographic (EEG) studies, where it is used as an activation to test photo-sensibility related to epilepsy.<br><br>The Natus Photic Stimulator is typically used by EEG technicians in the hospital environment in fixed, mobile or portable systems, and with patients of all ages. The lamp head assembly is mounted on an arm allowing easy placement at 30 cm away from the patient's face. The arm is ergonomically designed so that the lamp can be move and rotated in the direction of the patient using a handgrip on the lamp.<br><br>Trigger pulses applied to the trigger input of the Natus Photic Stimulator generate 1 millisecond photic flashes at specific frequencies, typically in the range of 0.5 Hz to 60 Hz by a white light emitting diode (LED) lamp. The basic operating mode consists of generating a single flash per trigger pulse applied. Flash intensity has a maximum range from 22,000 lux up to 75,000 lux measured at 30 cm from the LED lamp at the position of highest intensity and 3,520 lux to 12,000 lux calculated at 75cm. The frequency and intensity of the flash is controlled by the acquisition software.<br><br>It can also be used along with evoked potential devices for stimulating visual evoked potentials. | | |
| <b>Intended Use:</b> | The Natus Photic Stimulator is indicated for photic activation of the EEG during an EEG study and in the generation of visual evoked potentials. | | |
| <b>Technology Comparison:</b> | The Natus Photic Stimulator employs the same technological characteristics as the predicate device. | | |
| System Characteristic | Lifelines Ltd.<br>Lifelines Photic Stimulator<br>(K101691) | Excel-Tech Ltd. (XLTEK)<br>Natus Photic Stimulator<br>(Proposed Device) | |
| Indications for Use | The Lifelines Photic<br>Simulator is indicated for<br>photic activation of the EEG<br>during an EEG study and in<br>the generation of visual<br>evoked potentials. | The Natus Photic Stimulator is<br>indicated for photic activation<br>of the EEG during an EEG<br>study and in the generation of<br>visual evoked potentials. | |
| Mode of Operation | Arm-mounted photic<br>stimulator generates flashes<br>of white light by means of<br>light emitting diode (LED) | Arm-mounted photic<br>stimulator generates flashes of<br>white light by means of LEDs. | |
| Light Source | Single, high-intensity LED | Surface mounted, high-<br>intensity LED array | |
| System<br>Characteristic | Lifelines Ltd.<br>Lifelines Photic Stimulator<br>(K101691) | Excel-Tech Ltd. (XLTEK)<br>Natus Photic Stimulator<br>(Proposed Device) | |
| Luminous Flux | 700 lm typical, 900 lm max<br>13,000 lux at one (1) foot | Intensity measured at 30 cm (~<br>1 foot) distance, position 12:<br>● Minimum: 22,000 lux<br>● Maximum: 75,000 lux<br><br>Typical observed intensity is<br>39,900 lux. | |
| Flash Rate | 1 to 60 Hz or single manual<br>flash | 0.5 to 60 Hz or single manual<br>flash | |
{4}------------------------------------------------
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the logo for Natus Neurology. The word "natus" is in a teal, sans-serif font, with a registered trademark symbol next to it. Below "natus" is the word "neurology" in a smaller, italicized, sans-serif font, also in teal.
## Summary of Performance Testing:
| Electrical Safety | The Natus Photic Stimulator was verified for performance in accordance<br>with the following standard: |
|----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | • IEC 60601-1: 2005, Am1: 2012, Medical electrical equipment – Part<br>1: General requirements for basic safety and essential performance. |
| | Results indicate that the Natus Photic Stimulator complies with the<br>applicable standards. |
| Electromagnetic<br>Compatibility | The Natus Photic Stimulator was verified for performance in accordance<br>with the following standard: |
| | • IEC 60601-1-2: 2007, Medical electrical equipment – Part 1-2:<br>General requirements for basic safety and essential performance –<br>Collateral standard: Electromagnetic compatibility – Requirements<br>and tests. |
| | • IEC 60601-2-40: 2016, Medical electrical equipment – Part 2-40:<br>Particular requirements for the basic safety and essential<br>performance of electromyographs and evoked response equipment. |
| | Results indicate that the Natus Photic Stimulator complies with the<br>applicable standards. |
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for Natus Neurology. The word "natus" is in a teal color and in a larger font. Below that, the word "neurology" is in black and in a smaller font. The logo is simple and modern.
| Performance<br>Testing – Bench | The Natus Photic Stimulator was verified for performance in accordance<br>with internal requirements and the applicable clauses of the following<br>standards:<br>IEC 60601-1-6: 2010, Am1: 2013, Medical electrical equipment –<br>Part 1-6: General requirements for basic safety and essential<br>performance - Collateral standard: Usability. IEC 60601-2-40: 2016, Medical electrical equipment – Part 2-40:<br>Particular requirements for the basic safety and essential<br>performance of electromyographs and evoked response equipment. IEC 62366: 2007, Am1: 2014, Medical devices – Application of<br>usability engineering to medical devices. IEC 62471: 2006, Photobiological safety of lamps and lamp systems. ANSI Z80.36-2016, Light Hazard Protection for Ophthalmic<br>Instruments. |
|--------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Results indicate that the Natus Photic Stimulator complies with its<br>predetermined specifications and the applicable standards. |
| Conclusion | Verification and validation activities were conducted to establish the<br>performance and safety characteristics of the device modifications made<br>to the Natus Photic Stimulator. The results of these activities demonstrate<br>that the Natus Photic Stimulator is as safe, as effective, and performs as<br>well as or better than the predicate devices.<br><br>Therefore, the Natus Photic Stimulator is considered substantially<br>equivalent to the predicate devices. |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.