← Product Code [BWF](/productcode/BWF) · K173918

# VersaPAP Positive Airway Pressure (PAP) Device (K173918)

_Trudell Medical International · BWF · Aug 17, 2018 · Anesthesiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K173918

## Device Facts

- **Applicant:** Trudell Medical International
- **Product Code:** [BWF](/productcode/BWF.md)
- **Decision Date:** Aug 17, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 868.5690
- **Device Class:** Class 2
- **Review Panel:** Anesthesiology
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

VersaPAP* is indicated for the treatment and prevention of atelectasis. VersaPAP* also has the ability to provide supplemental oxygen when used with compressed oxygen. The VersaPAP* device is for patients (ages 5 years and above) who are capable of following directions for Positive Airway in a hospital environment. The VersaPAP* device is a single patient, multiple use device intended to be used under the supervision of a healthcare professional.

## Device Story

Hand-held respiratory therapy device; creates positive airway pressure via Venturi-vacuum principle. Compressed air/oxygen accelerates through orifice; high velocity creates vacuum; draws atmospheric air into cylindrical channel; combined flow exits mouthpiece. Used in hospital environment under healthcare professional supervision; single-patient, multiple-use (up to 7 days). Includes manometer gauge for real-time airway pressure monitoring. Output is positive airway pressure during breathing cycle; assists lung expansion in spontaneously breathing patients. Benefits include prevention/treatment of atelectasis and supplemental oxygen delivery.

## Clinical Evidence

No clinical data. Substantial equivalence based on bench testing, including comparative performance testing against the predicate using representative patient-generated breathing patterns, and biocompatibility assessment per ISO 10993-1.

## Technological Characteristics

Pneumatic device; Venturi-vacuum principle. Components: mouthpiece, backpiece (baffle, body, end cap), manometer adapter, manometer gauge, oxygen tubing. Materials identical to reference devices (K173825, K150173). Flow rate: 5-15 L/min. Gas source: compressed air or oxygen. Non-sterile, single-patient use. No software or electronic components.

## Regulatory Identification

An incentive spirometer is a device that indicates a patient's breathing volume or flow and that provides an incentive to the patient to improve his or her ventilation.

## Predicate Devices

- EzPAP ([K991300](/device/K991300.md))

## Reference Devices

- MC 300* R Nebulizer ([K173825](/device/K173825.md))
- Aerobika* OPEP with Manometer Accessory ([K150173](/device/K150173.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

August 17, 2018

Trudell Medical International Marianne Tanton Director, Quality and Regulatory Affairs 725 Third Street London, n5V 5G4 Ca

## Re: K173918

Trade/Device Name: VersaPAP Positive Airway Pressure (PAP) Device Regulation Number: 21 CFR 868,5690 Regulation Name: Incentive Spirometer Regulatory Class: Class II Product Code: BWF Dated: July 18, 2018 Received: July 19, 2018

Dear Ms. Tanton:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrl/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Amy K. Levelle -S

for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K173918

#### Device Name

VersaPAP* Positive Airway Pressure (PAP) Device

#### Indications for Use (Describe)

VersaPAP* is indicated for the treatment and prevention of atelectasis. VersaPAP* also has the ability to provide supplemental oxygen when used with compressed oxygen. The VersaPAP* device is for patients (ages 5 years and above) who are capable of following directions for Positive Airway in a hospital environment. The VersaPAP* device is a single patient, multiple use device intended to be used under the supervision of a healthcare professional.

Type of Use (Select one or both, as applicable)

| <span style="text-decoration: overline;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|------------------------------------------------------------------------------------------------|
| ☐ Over-The-Counter Use (21 CFR 801 Subpart C)                                                  |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Traditional 510(k) - K173918/S002 VersaPAP* Positive Airway Pressure (PAP) Device Interactive Review - dated 17 Aug 2018

# Section 5 - 510(k) Summary

Prepared: 17-Aug-2018

#### 1. Submitter

Trudell Medical International 725 Third Street London, Ontario N5V 5G4, Canada

Contact: Marianne Tanton Director, Quality and Regulatory Affairs Tel: 1-519-455-7060 Email: mtanton@trudellmed.com

### 2. Device

| Trade Name:          | VersaPAP* Positive Airway Pressure (PAP) Device |
|----------------------|-------------------------------------------------|
| Common Name:         | Positive Airway Pressure (PAP) Device           |
| Classification Name: | Incentive Spirometer 21 CFR 868.5690            |
| Regulatory Class     | II                                              |
| Product Code:        | BWF                                             |

#### 3. Predicate Device

Boeing, Positive Airway Pressure (PAP) Therapy Device, K991300 DHD Healthcare Corp

DHD Healthcare Corp was purchased by Smith Medical has changed the device name to EzPAP. The predicate device is referred to as EzPAP throughout the submission.

The predicate device has not been subject to a recall.

### 4. Reference Device

MC 300* R Nebulizer - K173825 Trudell Medical International

Aerobika* OPEP with Manometer Accessory - K150173 Trudell Medical International

The reference devices have not been subject to a recall.

### 5. Device Description

The VersaPAP* device is a hand held respiratory therapy device that creates a positive airway pressure. The VersaPAP* device is intended to be used by pediatric (ages 5 years and above) and adult patients in the hospital environment, under the supervision of a healthcare professional. The VersaPAP* device is a single patient use device and may be used for two treatments per day

{4}------------------------------------------------

Traditional 510(k) – K173918/S002 VersaPAP* Positive Airway Pressure (PAP) Device Interactive Review - dated 17 Aug 2018

## Section 5 - 510(k) Summary

for 7 days. A manometer gauge is included to monitor airway pressure while operating the VersaPAP* device.

The VersaPAP* device consists of seven components: a mouthpiece, the VersaPAP* backpiece (includes 3 components: the baffle, body and end cap), a manometer adapter, a manometer gauge, and oxygen tubing.

## 6. Principle of Operation

The VersaPAP* device creates positive airway pressure using the Venturi-vacuum principle. Compressed air or compressed oxygen is driven through an orifice where it accelerates and emerges at a high velocity creating a vacuum (Venturi effect). The vacuum draws additional air from the atmosphere through a cylindrical channel where it combines with the orifice airstream and flows out the mouthpiece. This flow creates a net positive airway pressure during the breathing cycle.

## 7. Indications for Use

VersaPAP™ is indicated for the treatment and prevention of atelectasis. VersaPAP™ also has the ability to provide supplemental oxygen when used with compressed oxygen. The VersaPAP™ device is for patients (ages 5 years and above) who are capable of following directions for Positive Airway Pressure Therapy in a hospital environment. The VersaPAP™ device is a single patient, multiple use device intended to be used under the supervision of a healthcare professional.

### 8. Comparison to predicate device

The VersaPAP* device and predicate EzPAP (K991300), are identical in purpose, function, core technology and method of operation. Only minor differences exist between the VersaPAP* device and predicate, which do not affect the safety or effectiveness of the subject device. Table 1 provides a comparison of the subject and predicate devices.

{5}------------------------------------------------

# Section 5 - 510(k) Summary

#### Table 1: Comparison to Predicate Device

| Element of Comparison  | VersaPAP* device<br>(Subject Device)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | EzPAP device<br>(Predicate - K991300)                                                                                                                                                                                                                                                   |
|------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use    | VersaPAPTM is indicated for the<br>treatment and prevention of<br>atelectasis. VersaPAPTM also has<br>the ability to provide supplemental<br>oxygen when used with<br>compressed oxygen. The<br>VersaPAPTM device is for patients<br>(ages 5 years and above) who are<br>capable of following directions for<br>Positive Airway Pressure Therapy<br>in a hospital environment. The<br>VersaPAPTM device is a single<br>patient, multiple use device<br>intended to be used under the<br>supervision of a healthcare<br>professional. | EzPAP is indicated for the treatment<br>and prevention of atelectasis. EzPAP<br>facilitates opening of airways in<br>patients requiring prevention or<br>treatment of atelectasis. EzPAP also<br>has the ability to provide<br>supplemental oxygen when used with<br>compressed oxygen. |
| Technology             | Pneumatic                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |                                                                                                                                                                                                                                                                                         |
| Environment of Use     | Hospital                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |                                                                                                                                                                                                                                                                                         |
| Patient population     | Pediatric (ages 5 years and above)<br>and adult patients                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Patient population not specified                                                                                                                                                                                                                                                        |
| Single Patient Use     | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                                                                                                                                                         |
| Therapy Type           | Lung expansion therapy for patients with spontaneous breathing                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                         |
| Type of Device         | Single patient use, prescription only, non-sterile                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                         |
| Manufacturing Process  | Plastic Molding                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                                                                                                                                                                                                                                                                                         |
| Patient Interface      | Mouthpiece                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Mouthpiece, Face Mask                                                                                                                                                                                                                                                                   |
| Flow rate (L/min)      | 5-15                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |                                                                                                                                                                                                                                                                                         |
| Type of gas source     | Compressed air or compressed oxygen                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                                                                                                                                                         |
| Manometer Gauge        | Included with device to monitor airway pressure                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                                                                                                                                                                                                                                                                                         |
| Principle of Operation | Venturi-vacuum principle                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |                                                                                                                                                                                                                                                                                         |

#### 9. Performance Data

### 9.1 Summary of Performance Testing

Performance testing included comparative testing against the predicate device using representative patient generated breathing patterns. Testing confirmed that the VersaPAP* device is substantially equivalent to the predicate.

<sup>*</sup>Trade marks and registered trademarks of Trudell Medical International

{6}------------------------------------------------

Traditional 510(k) - K173918/S002 VersaPAP* Positive Airway Pressure (PAP) Device Interactive Review - dated 17 Aug 2018

## Section 5 - 510(k) Summary

### 9.2 Biocompatibility Testing Summary

To evaluate the biological safety of the VersaPAP* device, consideration was given to the following: type of patient contact, potential hazards of the materials of construction, the history of clinical use and testing of the materials of construction, biocompatibility and chemical characterization testing on a similar device, and other information available in the literature.

When used as intended, the VersaPAP* device has direct contact with the mucosal membrane and indirect contact with tissue in the patient respiratory pathway. It is considered externally communicating by way of gas pathway. In accordance with ISO 10993-1, the VersaPAP* device is considered to be an externally communicating, with limited (≤ 24 hours) indirect contact with tissue.

The materials of the VersaPAP* device components are identical to materials used in the following reference devices: MC300*R Nebulizer (K173825) and Aerobika* OPEP device (K150173)

| VersaPAP*                     | Reference Device (K Number)          |
|-------------------------------|--------------------------------------|
| Component                     |                                      |
| Mouthpiece                    | Aerobika* OPEP Device with Manometer |
| Manometer Adapter<br>and Plug | Aerobika* OPEP Device with Manometer |
| Manometer Gauge               | Aerobika* OPEP Device with Manometer |
| Oxygen Tubing                 | MC300*R Nebulizer                    |
| Backpiece - End cap           | MC300*R Nebulizer                    |
| Backpiece - Baffle            | MC300*R Nebulizer                    |
| Backpiece - Body              | MC300*R Nebulizer                    |

#### Comparison to Reference Devices:

### 9.3 Dry Gas Pathway Testing

Testing pertaining to the dry gas pathway and associated risk assessments/conclusions were conducted by an independent source. Testing included the following assessments:

- . Emissions of volatile organic compounds (VOCs)
- . Fine particles (particulate matter PM2.5)

{7}------------------------------------------------

Traditional 510(k) – K173918/S002 VersaPAP* Positive Airway Pressure (PAP) Device Interactive Review - dated 17 Aug 2018

# Section 5 - 510(k) Summary

#### 10. Clinical Performance Summary

Not applicable, the determination of substantial equivalence is not based on Clinical Performance data.

#### 11. Conclusion

The non-clinical data demonstrate that the VersaPAP* device is as safe and effective as the predicate and therefore substantially equivalent to the cleared predicate device (K991300) and does not raise any new issues of safety and efficacy.

---

**Source:** [https://fda.innolitics.com/device/K173918](https://fda.innolitics.com/device/K173918)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
