Tissue of Origin Test Kit-FFPE

K173839 · Cancer Genetics, Inc. · OIW · Mar 15, 2018 · Clinical Toxicology

Device Facts

Record IDK173839
Device NameTissue of Origin Test Kit-FFPE
ApplicantCancer Genetics, Inc.
Product CodeOIW · Clinical Toxicology
Decision DateMar 15, 2018
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 862.3100
Device ClassClass 2
AttributesAI/ML

AI Performance

OutputAlgorithmAcceptanceObservedDev DSDev ReadersTest DSTest Readers
Tissue of Origin Classification——90.8% concordance (95% CI: 83.6% to 95.3%)——Validation study: 142 specimens including all 15 subtypes in the database.—
Tissue of Origin Classification (Software Concordance)——100% agreement (95% CI: 95.9% to 100%)——Validation study: 20 FFPE tumor specimens.—

Indications for Use

The Pathwork Tissue of Origin Test is an in vitro diagnostic intended to measure the degree of similarity between the RNA expression patterns in a patient's formalin-fixed, paraffin-embedded (FFPE) tumor and the RNA expression patterns in a database of fifteen tumor types (poorly differentiated, undifferentiated, and metastatic cases) that were diagnosed according to then current clinical and pathological practice. This test should be evaluated by a qualified physician in the context of the patient's clinical history and other diagnostic test results. Limitations: The Pathwork® Tissue of Origin Test is not intended to establish the origin of tumors (e.g. cancer of unknown primary) that cannot be diagnosed according to current clinical and pathological practice. It is not intended to subclassify or modify the classification of tumors that can be diagnosed by current clinical and pathological practice, nor to predict disease course or survival or treatment efficacy, nor to distinguish primary from metastatic tumor. Tumor types not in the Pathwork® Tissue of Origin Test database may have RNA expression patterns that are similar to patterns in the database. Therefore, results cannot be used to distinguish tumor types in the database from tumor types not in the database.

Device Story

In vitro diagnostic test; measures RNA expression patterns in FFPE tumor samples; compares patterns against database of 15 tumor types. Input: FFPE tumor tissue; RNA extracted and amplified using GeneChip™ 3' IVT Pico Kit. Processing: Microarray analysis via Pathchip®; Genechip™ System 3000 Dx v.2 scanner. Output: Similarity score between patient sample and database tumor types. Used in clinical laboratory settings; interpreted by qualified physicians alongside clinical history. Benefits: Assists in identifying tumor origin for poorly differentiated/metastatic cases.

Clinical Evidence

No clinical data provided. Bench testing only: 142 FFPE tumor specimens compared between original RampUp amplification and new GeneChip 3' IVT Pico Kit (90.8% concordance, 95% CI: 83.6-95.3%). 20 specimens compared between original Dx software and updated Dx v.2 software (100% concordance, 95% CI: 95.9-100%).

Technological Characteristics

In vitro diagnostic microarray system. Uses Pathchip® microarray, hybridization oven, and Genechip™ System 3000 Dx v.2 scanner. RNA amplification via GeneChip™ 3' IVT Pico Kit. Internal quality controls include Percent Positive, Overall Signal, and Regional Discontinuity. Software-based analysis of probe intensities.

Indications for Use

Indicated for patients with poorly differentiated, undifferentiated, or metastatic tumors to measure similarity of RNA expression patterns to a database of 15 known tumor types. Not intended to establish tumor origin for cases undiagnosable by current practice, subclassify tumors diagnosable by current practice, predict disease course/survival/treatment efficacy, or distinguish primary from metastatic tumors.

Regulatory Classification

Identification

An amphetamine test system is a device intended to measure amphetamine, a central nervous system stimulating drug, in plasma and urine. Measurements obtained by this device are used in the diagnosis and treatment of amphetamine use or overdose and in monitoring levels of amphetamine to ensure appropriate therapy.

Special Controls

*Classification.* Class II (special controls). An amphetamine test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).

Predicate Devices

Submission Summary (Full Text)

{0} SPECIAL 510(k): Device Modification OIR Decision Summary To: THE FILE RE: K173839 This 510(k) submission contains information/data on modifications made to the applicant's own class II or class I devices requiring 510(k). The following items are present and acceptable: 1. The name and 510(k) number of the applicant's previously cleared device: K120489: Pathwork Tissue of Origin Test Kit - FFPE 2. Applicant's statement that the INDICATION/INTENDED USE of the modified device as described in its labeling HAS NOT CHANGED along with the proposed labeling which includes instructions for use, package labeling, and, if available, advertisements or promotional materials. 3. A description of the device MODIFICATION(S), including clearly labeled diagrams, engineering drawings, photographs, user's and/or service manuals in sufficient detail to demonstrate that the FUNDAMENTAL SCIENTIFIC TECHNOLOGY of the modified device has not changed. The following two changes were made: 1) the GeneChip™ 3' IVT Pico Kit is used for amplification and target labeling instead of the RampUp RNA Amplification Kit™ reagents; and 2) the Chip processing software was updated to the Genechip™ System 3000 Dx v.2 from the Genechip™ System 3000 Dx, which is no longer supported by the manufacturer. The principles of the target preparation process have not changed. 4. Comparison Information (similarities and differences) to applicant's legally marketed predicate device including, intended use, equipment, reagents and limitations, are summarized in the table below: {1} | Feature | K120489 (Predicate) | K173839 | | --- | --- | --- | | Device Name | Pathwork Tissue of Origin Test Kit-FFPE | Cancer Genetics Tissue of Origin Test Kit - FFPE | | Intended use | The Pathwork Tissue of Origin Test is an in vitro diagnostic intended to measure the degree of similarity between the RNA expression patterns in a patient's formalin-fixed, paraffin-embedded (FFPE) tumor and the RNA expression patterns in a database of fifteen tumor types (poorly differentiated, undifferentiated, and metastatic cases) that were diagnosed according to then current clinical and pathological practice. This test should be evaluated by a qualified physician in the context of the patient's clinical history and other diagnostic test results. Limitations: The Pathwork® Tissue of Origin Test is not intended to establish the origin of tumors (e.g. cancer of unknown primary) that cannot be diagnosed according to current clinical and pathological practice. It is not intended to subclassify or modify the classification of tumors that can be diagnosed by current clinical and pathological practice, nor to predict disease course or survival or treatment efficacy, nor to distinguish primary from metastatic tumor. Tumor types not in the Pathwork® Tissue of Origin Test database may have RNA expression patterns that are similar to patterns in the database. Therefore, results cannot be used to distinguish tumor types in the database from tumor types not in the database. | Same | | Microarray | Pathchip® | Same | | Equipment | Hybridization oven | Same | | | Scanner | Same | | | Genechip™ System 3000 Dx | Genechip™ System 3000 Dx v.2 | | RNA isolation | Extraction Reagents | Same | | Amplification Reagents | RampUp RNA Amplification Kit™ | GeneChip™ 3' IVT Pico Kit | | Purification of biotinylated cDNA | Ethanol precipitation | GeneChip™ 3' IVT Pico Kit | | Internal Processing Quality Control | Percent Positive Overall Signal Regional Discontinuity | Same | | Analysis | Algorithm as described in K092967 | Same | 5. A Design Control Activities Summary which includes: a) Identification of Risk Analysis method(s) used to assess the impact of the modification on the device and its components, and the results of the analysis {2} b) Based on the Risk Analysis, an identification of the verification and/or validation activities required, including methods or tests used and acceptance criteria to be applied The Risk Analyses were performed to assess the impact of the modifications on the device by identifying risks, their possible causes, and appropriate control mechanisms. The Risk Analyses took into account device hazards associated with the intended use of the device. No additional hazards, no additional causes, and no required additional controls were identified. No adverse advents or reportable incidents have been associated with the device. Based on the Risk Analyses conducted, the following studies were performed to verify and/or validate the modifications to the device: - Changing the amplification and labeling reagents can potentially alter the expression ratios of the targets if the process does not accurately reflect the endogenous levels of these targets before amplification. As a result, the gene expression patterns may be skewed toward certain subtypes that will increase the chance of misclassification. A validation study was, therefore, performed with the modified device using 142 specimens that included all 15 subtypes in the database and that were studied in the original validation study with the predicate test. The concordance in the test results between paired specimens processed with the RampUp RNA Amplification Kit™ and the GeneChip™ 3' IVT Pico Kit was 90.8% (95% confidence interval of 83.6% to 95.3%). - For the software modification, there are risks related to potential alterations to relative probe intensities on the array which could affect the classification results. Therefore, a validation study was performed to assess the concordance between the Genechip™ System 3000 Dx v.2 and the Genechip™ System 3000 Dx using 20 FFPE tumor specimens. All results were concordant (100% agreement with 95% confidence interval of 95.9% to 100%). The labeling for this modified subject device has been reviewed to verify that the indication/intended use for the device is unaffected by the modification. In addition, the applicant's description of the particular modifications and the comparative information between the modified and unmodified devices demonstrate that the fundamental scientific technology has not changed. The applicant has provided the design control information as specified in The New 510(k) Paradigm and on this basis, I recommend the device be determined substantially equivalent to the previously cleared (or their preamendment) device. 3
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...