← Product Code [OBJ](/productcode/OBJ) · K173820

# OptiCross 6 HD, 60 MHz Coronary Imaging Catheter (Paper DFU), OptiCross 6 HD, 60 MHz Coronary Imaging Catheter (e-DFU) (K173820)

_Boston Scientific Corporation · OBJ · Jan 17, 2018 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K173820

## Device Facts

- **Applicant:** Boston Scientific Corporation
- **Product Code:** [OBJ](/productcode/OBJ.md)
- **Decision Date:** Jan 17, 2018
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 870.1200
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

This catheter is intended for ultrasound examination of coronary intravascular pathology only. Intravascular ultrasound imaging is indicated in patients who are candidates for transluminal coronary interventional procedures.

## Device Story

OptiCross 6 HD is a sterile, short-rail intravascular ultrasound (IVUS) imaging catheter. It consists of an imaging core with a 60 MHz ultrasonic transducer at the distal tip and a catheter body with a telescoping section for 15 cm of linear movement. The device connects to a Motordrive Unit (MDU5 PLUS) via an electro-mechanical interface. Operated by physicians in a clinical setting, the catheter is flushed with heparinized saline to provide acoustic coupling. The transducer rotates to capture ultrasound signals, which are processed by the BSC IVUS system to produce cross-sectional images of coronary arteries. These images allow clinicians to visualize intravascular pathology, aiding in the assessment of coronary anatomy and guiding interventional procedures. The device benefits patients by providing high-resolution imaging to support clinical decision-making during coronary interventions.

## Clinical Evidence

No clinical data. Substantial equivalence is supported by non-clinical performance data, including bench-top testing (deliverability, crossability, image resolution, NURD, measurement accuracy), biocompatibility (ISO 10993-1), electromagnetic compatibility (IEC 60601-1-2), and a GLP pre-clinical porcine study to validate in-vivo image quality.

## Technological Characteristics

60 MHz ultrasonic transducer; short-rail catheter design with telescoping shaft; 0.014" guidewire compatible; 6F guide catheter compatible; heparinized saline acoustic coupling; electro-mechanical interface to MDU5 PLUS; biocompatible materials per ISO 10993-1; compliant with IEC 60601-1-2 (3rd Edition).

## Regulatory Identification

An intravascular diagnostic catheter is a device used to record intracardiac pressures, to sample blood, and to introduce substances into the heart and vessels. Included in this generic device are right-heart catheters, left-heart catheters, and angiographic catheters, among others.

## Predicate Devices

- OptiCross 6 40 MHz Coronary Imaging Catheter ([K153617](/device/K153617.md))

## Reference Devices

- OptiCross 18, 30 MHz Peripheral Imaging Catheter ([K160514](/device/K160514.md))
- OptiCross, 40 MHz Coronary Imaging Catheter ([K123621](/device/K123621.md))
- OptiCross X, 40 MHz Coronary Imaging Catheter ([K161125](/device/K161125.md))
- OptiCross HD, 60 MHz Coronary Imaging Catheter ([K173284](/device/K173284.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

January 17, 2018

Boston Scientific Corporation Mugdha Dongre Regulatory Affairs Specialist II 47215 Lakeview Boulevard Fremont, California 94538

Re: K173820

Trade/Device Name: OptiCross 6 HD, 60 MHz Coronary Imaging Catheter Regulation Number: 21 CFR 870.1200 Regulation Name: Diagnostic Intravascular Catheter Regulatory Class: Class II Product Code: OBJ, ITX Dated: December 15, 2017 Received: December 18, 2017

Dear Mugdha Dongre:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);

{1}------------------------------------------------

## Page 2 - Mugdha Dongre

and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

M.A. Wilhelm

for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

510(k) Number (if known) K173820

Device Name

OptiCross™ 6 HD, 60 MHz Coronary Imaging Catheter

#### Indications for Use (Describe)

This catheter is intended for ultrasound examination of coronary intravascular pathology only. Intrasound imaging is indicated in patients who are candidates for transluminal coronary interventional procedures.

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

{3}------------------------------------------------

# 510(k) Summary per 21 CFR 807.92

| Submitter's Name and<br>Address | Boston Scientific Corporation<br>47215 Lakeview Boulevard<br>Fremont, CA 94538<br>USA                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |         |               |
|---------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|---------------|
| Contact Name and<br>Information | Mugdha Dongre<br>Regulatory Affairs Specialist II<br>Tel: 510.624.1314<br>Fax: 510.440.7698<br>Email: Mugdha.Dongre@bsci.com                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |         |               |
| Date Prepared                   | December 15, 2017                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |         |               |
| Trade Name                      | OptiCross™ 6 HD, 60MHz Coronary Imaging Catheter                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |         |               |
| Common Name                     | Diagnostic Intravascular Catheter, Ultrasound Transducer                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |         |               |
| Classification Name             | Catheter, Ultrasound, Intravascular (OBJ) has been<br>classified as Class II per 21 CFR 870.1200<br>Transducer Ultrasonic (ITX) has been classified as Class II<br>per 21 CFR 892.1570                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |         |               |
| Predicate Name                  | OptiCross™ 6 40 MHz<br>Coronary Imaging Catheter                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | K153617 | 25-Feb-2016   |
| Reference Devices               | OptiCross™ 18, 30 MHz<br>Peripheral Imaging Catheter                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | K160514 | 22-June-2017  |
|                                 | OptiCross™, 40 MHz Coronary<br>Imaging Catheter                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | K123621 | 15-April-2013 |
|                                 | OptiCross™ X, 40 MHz<br>Coronary Imaging Catheter                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | K161125 | 19-May-2016   |
|                                 | OptiCross™ HD, 60 MHz<br>Coronary Imaging Catheter                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | K173284 | 16-Nov-2017   |
| Device Description              | OptiCross™ 6 HD (60 MHz Coronary Imaging Catheters) is a<br>sterile, short rail imaging catheter.<br>It consists of two main assemblies:<br>1. Imaging Core<br>2. Catheter Body<br>The imaging core is composed of a hi-torque, flexible, rotating<br>drive cable with a radial looking 60 MHz ultrasonic transducer at<br>the distal tip. An electro-mechanical connector interface at the<br>proximal end of the catheter makes the connection to the<br>Motordrive Unit (MDU5 PLUS™). The MDU5 PLUS-catheter<br>interface consists of an integrated mechanical drive socket and<br>electrical connection.<br>The catheter body is comprised of three sections: |         |               |

{4}------------------------------------------------

|                                                                               | 1. Distal Imaging Window Lumen                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
|-------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                                                                               | 2. Proximal Shaft Lumen                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
|                                                                               | 3. Telescoping Section                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
|                                                                               | The distal imaging window lumen and proximal shaft lumen<br>sections comprise the "working length" of the catheter, and the<br>telescoping section remains outside of the guiding catheter.<br>The catheter body has a distal imaging window lumen with<br>proximal exit 1.6 cm from the distal end. A radiopaque (RO)<br>marker is embedded in the catheter body at 0.5 cm from the distal<br>tip. In addition, two insertion depth markers are located on the<br>proximal shaft lumen at 90 cm and 100 cm from the distal tip to aid<br>in estimating catheter position relative to the distal guide catheter<br>tip. The proximal shaft lumen is attached to the telescoping section<br>via a strain relief connection.<br>The telescoping shaft (section) allows the imaging core to be<br>advanced and retracted for 15 cm of linear movement.<br>The corresponding movement of the transducer occurs from the<br>proximal end of the guidewire exit port to the proximal end of the<br>distal imaging window lumen. The telescope section has proximal<br>markers for lesion length assessment, consisting of a series of<br>marks spaced 1 cm apart on the telescope body.<br>A flush port with a one-way check valve is used to flush the interior<br>of the catheter body and maintain a flushed condition. The catheter<br>must be flushed with heparinized saline prior to use, as this<br>provides the acoustic coupling media required for ultrasonic<br>imaging. The one-way check valve helps retain saline in the<br>catheter during use. |
| Intended Use/Indications<br>for Use                                           | This catheter is intended for ultrasound examination of coronary<br>intravascular pathology only. Intravascular ultrasound imaging is<br>indicated in patients who are candidates for transluminal coronary<br>interventional procedures.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Device Technology<br>Characteristics and<br>Comparison to Predicate<br>Device | The OptiCross™ 6 HD, catheter is a 60 MHz ultrasound coronary<br>imaging catheter intended to operate with a BSC IVUS system for<br>diagnostic imaging. It will be used with a minimum 6F (>0.064"<br>ID) guide catheter / introducer sheath and a 0.014" guidewire.<br><br>The OptiCross™ 6 HD Coronary Imaging Catheter incorporates a<br>transducer with new design features intended to utilize a 60MHz<br>frequency to enhance image resolution while maintaining similar<br>depth of penetration as well as similar performance characteristics<br>as that of its predicate OptiCross™ 6 40Mhz Coronary Imaging<br>Catheter.<br><br>Modifications with respect to the predicate device include:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
|                                                                               | A new transducer (new design with new materials) PCA board populated with two zero Ohm resistors to generate DC coupling.              In support of a substantial equivalence determination, BSC has<br>compared and evaluated the material and design differences<br>between the subject and predicate device.<br>Non-clinical performance evaluations, as described below, indicate<br>that the subject device is substantially equivalent to, and at least as                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
|                                                                               | safe and effective as the predicate device OptiCross™ 6, 40 MHz<br>Coronary Imaging Catheter (K153617).                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Non-Clinical<br>Performance Data                                              | Determination of substantial equivalence is based on an assessment<br>of non-clinical performance data.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
|                                                                               | Non-clinical data includes bench-top performance testing,<br>packaging validation, biological safety, electromagnetic<br>compatibility, acoustic output testing, and a GLP pre-clinical<br>animal study to validate imaging performance in the porcine model.<br><br>Bench Testing:<br>Bench testing was performed to evaluate physical integrity,<br>functionality, and overall performance of the catheter.<br>Performance criteria includes deliverability, crossability,<br>guide catheter compatibility, lubricity, retraction capability,<br>image resolution, image penetration, non-uniform rotational<br>distortion (NURD), measurement accuracy, pullback reliability,<br>dimensional requirements, visibility under fluoroscopy, interface<br>with ancillary devices, environmental requirements, user interface<br>requirements, catheter robustness and simulated use structural<br>integrity.<br><br>Pre-clinical Evaluation:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
|                                                                               | The purpose of this study was to support design validation and<br>business need evaluations claims associated with in-vivo image<br>quality.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
|                                                                               | Biological Safety Testing:<br>OptiCross™ 6 HD catheter utilizes existing materials from the<br>commercially available OptiCross™ family of catheters which have<br>similar configurations and are manufactured under the same<br>processing conditions. Hence the biocompatibility test data has<br>been leveraged from the commercially available OptiCross™ 6 and<br>OptiCross™ 18 Imaging Catheter to meet the requirements of ISO<br>10993-1.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
|                                                                               | Electrical and Mechanical Safety:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
|                                                                               | Acoustic Output was evaluated in accordance with FDA<br>Guidance, Information for Manufacturers Seeking Marketing<br>Clearance of Diagnostic Ultrasound Systems and<br>Transducers (September 9, 2008). Acoustic Output test results for<br>OptiCross™ 6 HD are below the FDA Track 1 limits.<br>Electromagnetic compatibility testing was also conducted;<br>demonstrating compliance to IEC 60601-1-2 (3rd Edition).                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
|                                                                               | Packaging Validation:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
|                                                                               | Packaging of OptiCross™ 6 HD catheters is equivalent to the<br>packaging of other commercially available BSC IVUS catheters in<br>the OptiCross™ product family. Thus, the OptiCross™ HD<br>packaging will leverage the packaging DV results of the currently<br>available OptiCross catheters to demonstrate it meets the required<br>product specifications.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
|                                                                               | Conclusion:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
|                                                                               | Non-clinical performance evaluations, as described above,<br>indicate that the subject device is substantially equivalent to,<br>and at least as safe and effective as the predicate device,<br>OptiCross™ 6, 40 MHz Coronary Imaging Catheter (K153617).                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Clinical Performance<br>Data                                                  | Not applicable; determination of substantial equivalence is based on<br>an assessment of non-clinical performance data.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Conclusion                                                                    | With respect to the predicate, design modifications incorporated by<br>OptiCross™ 6 HD do not alter the fundamental scientific<br>technology or the indications for use. Based on Failure Mode<br>Effects Analysis (FMEA), comprehensive verification and<br>validation activities were successfully completed; raising no new<br>issues of safety or effectiveness.<br>Non-clinical performance data supports a determination that the<br>subject device, OptiCross™ 6 HD, is substantially equivalent to the<br>predicate device, OptiCross™ 6, 40 MHz Coronary Imaging<br>Catheter (K153617); and that it is at least as safe and effective for<br>its intended use                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |

{5}------------------------------------------------

{6}------------------------------------------------

---

**Source:** [https://fda.innolitics.com/device/K173820](https://fda.innolitics.com/device/K173820)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
