K173813 · Shen Zhen Cosbeauty Co., Ltd. · OHT · Sep 7, 2018 · General, Plastic Surgery
Device Facts
Record ID
K173813
Device Name
IPL Hair Removal Device Joy Version
Applicant
Shen Zhen Cosbeauty Co., Ltd.
Product Code
OHT · General, Plastic Surgery
Decision Date
Sep 7, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The IPL Hair Removal Device Joy Version is indicated for the removal of unwanted hair. The device is also indicated for the permanent reduction in hair regrowth, defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime. The device is used for adults with Fitzpatrick skin types I – IV.
Device Story
Home-use IPL hair removal device; utilizes Xenon lamp to emit light pulses (510-1200nm) to heat hair follicles; reduces hair growth without cutting/pulling. Operated by user via finger switch; includes treatment window head, facial adaptor, and external power supply. Features skin proximity sensor to prevent pulse emission unless in full contact with skin. Three lamp cartridges (body, face, bikini) with varying spot sizes (4.2cm² or 2cm²) and energy levels. User selects intensity level; device provides visual feedback on power/intensity. Benefits include permanent hair reduction with minimal pain. Used in home environment; intended for adults.
Clinical Evidence
Bench testing only. Electrical safety (IEC 60601-1, IEC 60601-2-57) and electromagnetic compatibility (IEC 60601-1-2) testing performed. Usability study (OTC study) conducted with 78 subjects demonstrated that lay users can correctly self-select, operate the device, and understand labeling, contraindications, and instructions for use.
Technological Characteristics
IPL device; Xenon arc flashlamp; 510-1200nm spectrum; 5 intensity levels; max fluence 5.1 J/cm²; pulse duration 9.2-11.2ms. Housing/lamp cartridge material: ABS (biocompatible per ISO 10993). Power: 100-240V AC external adapter. Connectivity: None. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-57. Microprocessor-controlled.
Indications for Use
Indicated for removal of unwanted hair and permanent reduction in hair regrowth in adults with Fitzpatrick skin types I-IV.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
September 7, 2018
Shen Zhen CosBeauty Co., Ltd Ms. Iris Fung Regulation manager SGS-CSTC Standards Technical Services Co., Ltd. 198 KEZHU Road, SCIENTECH Park Guangzhou Guangzhou. Cn
Re: K173813
Trade/Device Name: IPL Hair Removal Device Joy Version Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: OHT Dated: August 28, 2018 Received: August 31, 2018
Dear Ms. Iris Fung:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmr/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
> Sincerely, Jennifer Stevensor
For Binita S. Ashar, M.D., M.B.A., F.A.C.S Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K173813
#### Device Name
IPL Hair Removal Device Joy Version, Model: CB-027
Indications for Use (Describe)
The IPL Hair Removal Device Joy Version is indicated for the removal of unwanted hair. The device is also indicated for the permanent reduction in hair regrowth, defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime. The device is used for adults with Fitzpatrick skin types I - IV.
Type of Use (Select one or both, as applicable)
] Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
# Date of the summary prepared: September 7, 2018
# 510(k) Summary
This summary of 510(K) safety and effectiveness information is being submitted in accordance with the requirement of 21 CFR 807.92.
# 1. Submitter's Information
# Sponsor
- � Company Name: Shenzhen CosBeauty Technology Co., Ltd.
- Address: UnitA-3F, Qiao De Industrial Park, Tian Liao, Guang Ming District, Shenzhen, � China
- Phone: +86 1831 685 8036 �
- � Fax: +86 755 8629 0505
- � Contact Person (including title): Nicole Hu
- E-mail: lizan@cos-beauty.com �
# Application Correspondent:
- � SGS-CSTC Standards Technical Services Co., Ltd.
- Address: 198 KEZHU Road, SCIENTECH Park Guangzhou Economic & Technology �
Development
District, Guangzhou, Guangdong, CHINA
- � Contact Person: Ms. Iris Fung
- Tel: +86-20-32136908 �
- Email: Iris.Fung@sgs.com �
#### 2. Subject Device Information
- � Trade Name: IPL Hair Removal Device Joy Version, Model: CB-027
- Light Based Over-The-Counter Hair Removal � Common Name:
- Laser surgical instrument for use in general and plastic � Classification name: surgery and in dermatology
- General & Plastic Surgery � Review Panel:
- ОНТ Product Code: �
- 2 � Regulation Class:
- 878.4810 � Regulation Number:
{4}------------------------------------------------
# 3. Predicate Device Information
| Sponsor | Cyden Limited | Shenzhen CosBeauty<br>Technology Co., Ltd. |
|----------------------|-----------------------------------------------------|--------------------------------------------|
| Device Name | Ipulse<br>Smoothskin Gold<br>Hair Removal<br>Device | PerfectSmooth |
| 510(k) Number | K160968 | K161428 |
| Product Code | ONF | OHT |
| Regulation<br>Number | 878.4810 | 878.4810 |
| Regulation Class | 2 | 2 |
### 2. Device Description
IPL Hair Removal Device Model: CB-027, a small over-the-counter, is a home-use device for the permanent reduction of hair growth based on Intense Pulsed Light (IPL). It works below the skin's surface and does not involve any cutting or pulling, reducing hair growth with minimal pain. A personal Light-Based Hair Removal System. Emission is by finger switch. Device includes a treatment window head, a facial adaptor and battery charger/AC cord.It is used AC Powered(100-240 V AC; 50/60 Hz).The weight of the device is 402g, and the size is 126 x 78 x 200mm (H x W x D).The device incorporates Intense Pulse Light (IPL) technology. The purpose of the light is to heat the root where the hair grows.
# 5. Intended Use / Indications for Use
The IPL Hair Removal Device Joy Version is indicated for the removal of unwanted hair. The device is also indicated for the permanent reduction in hair regrowth, defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime. The device is used for adults with Fitzpatrick skin types I – IV.
#### 6. Design
The device contains a Xenon Lamp and a skin proximity sensor to detect appropriate skin contact. If the IPL Hair Removal Device is not properly applied to the treatment area (in full contact with the skin), the device cannot be triggerred a pulse emitting. Body lamp cartridge can be used for body hair below the neck. It can cover an area of 4.2cm2, and it is specially designed for large areas on legs, arms, back and abdomen. It can remove hair in a large scope rapidly. Do not use Body lamp cartridge on face. A filter has been installed in the face lamp cartridge. It can cover an area of 2cm2, and it can be used to remove hair on upper lip, chin and sideburns. Do not use this lamp cartridge on the areas around eyes or near eyelid. The product can flash for 300,000 times, and it shall not be used when the flash counter is displaying "000000".
{5}------------------------------------------------
# 7. Materials
There are two patient- directly contracting components in the subject device as the following list.
| Component of<br>Device Requiring<br>Biocompatibility | Material of<br>Component | Body Contact<br>Category<br>(ISO 10993-1) | Contact Duration<br>(ISO 10993-1) |
|------------------------------------------------------|--------------------------|-------------------------------------------|-----------------------------------|
| Housing | ABS | Surface-contacting<br>device: skin | Maximum 30<br>minutes(< 24hours) |
| Plastic plate of lamp<br>cartridge | ABS | Surface-contacting<br>device: skin | Maximum 30<br>minutes(< 24hours) |
Both materials is identical to the material of the device PerfectSmooth, which is cleared by FDA with K161428, produced by us. So these parts' biocompatibility is safe.
# 8. Physical characteristics
| Basic Unit Characteristics | |
|-------------------------------------------|--------------------------------------------------------------|
| Compliance* with 21 CFR 898 | No |
| Main Unit Weight | 402g |
| Main Unit Dimension | 126*78*200mm |
| Housing Materials of main unit | PC2805 |
| Indicator | Indicates power information, intensity level information. |
| Environment for operation | Temperature: 5°C~35°C |
| Storage and Transport Conditions | Temperature: -10°C~55°C |
| Compliance with Voluntary Standards | Yes, Comply with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-57, |
| Patient leakage current | Comply with IEC 60601-1 |
| Power Source | Supplied by external adapter |
| Software/Firmware/Microprocessor Control? | Yes |
| Specification | |
| Output Intensity Level | 5 |
| Output energy with body lamp cartridge | Level 1: 11.77J |
| | Level 2: 14.12J |
| | Level 3: 17.12J |
| | Level 4: 19.95J |
| | Level 5: 22,21J |
| Output energy with facial lamp cartridge | Level 1: 3.65J |
| | Level 2: 4.59J |
| | Level 3: 5.33J |
| | Level 4: 6.16J |
| | Level 5: 7.04J |
| Output energy with bikini lamp cartridge | Level 1: 3.84J |
| | Level 2: 4.92J |
{6}------------------------------------------------
| | Level 3: 5.43J |
|---------------------------------|------------------------------------------------|
| | Level 4: 6.23J |
| | Level 5: 7.22J |
| Emitted Light Spectrum | 510nm~1200nm Max |
| Max Energy density: | 5.1 J/cm², |
| Pulse duration | 9.20~11.20milliseconds |
| Treatment Area (regular window) | Body: 4.2 [cm²], Bikini area and Face: 2 [cm²] |
| Flashes: | 300,000 times |
| Power Supply | 100-240 VAC, 50/60Hz |
| Technology | IPL |
### 10. Test Summary
IPL Hair Removal Device Joy Version, Model: CB-027 has been evaluated the safety and performance by lab bench testing as following:
- � Electrical safety test according to IEC 60601-1 and IEC 60601-2-57 standards
- � Electromagnetic compatibility test according to IEC 60601-1-2 standard
- � Software verification and validation test according to the requirements of the FDA "Guidance for PreMarket Submissions and for Software Contained in Medical Devices"
- � Usability study:
Usability testing (OTC Study) was completed in 78 subjects to evaluate device human factors and label comprehension.
These usability study tests demonstrate that the device and its labeling can meet with the requirement:
1) the lay user can self-select themselves as being appropriate users of this device by the external box labeling, 2) the lay user can apply the treatment safely and correctly according to the instructions for use, and 3) the lay user can understand all indications, contraindications, warnings and precautions, and be able to identify whether they are within any contraindicated group; and be able to understand the user manual.
# 11. Comparison to predicate device and conclusion
The technological characteristics, features, specifications, materials, and intended use of IPL Hair Removal Device Joy Version, model: CB-027 is substantially equivalent to the predicate devices quoted above.
The differences between the subject device and predicate devices do not raise new issues of safety or effectiveness.
{7}------------------------------------------------
| Elements of<br>Comparison | Subject Device | Predicate Device I | Predicate Device II | Remark | |
|------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|
| Device Name and<br>Model | IPL Hair Removal Device<br>Joy Version, Model: CB-<br>027 | iPulse SmoothSkin<br>Gold | PerfectSmooth (CB-<br>014) | -- | |
| 510(k) Number | Applying | K160968 | K161428 | -- | |
| Manufacturer | Shenzhen CosBeauty<br>Technology Co., Ltd. | CyDen Ltd | Shenzhen CosBeauty<br>Technology Co., Ltd. | -- | |
| Intended Use | The IPL Hair Removal<br>Device Joy Version is<br>indicated for the removal<br>of unwanted hair. The<br>device is also indicated for<br>the permanent reduction<br>in hair regrowth, defined<br>as the long-term, stable<br>reduction in the number of<br>hairs regrowing when<br>measured at 6, 9 and 12<br>months after the<br>completion of a treatment<br>regime. The device is<br>used for adults with<br>Fitzpatrick skin types I –<br>IV. | The iPulse SmoothSkin<br>Gold Hair Removal<br>System is indicated for<br>the removal of<br>unwanted hair. The<br>iPulse SmoothSkin Gold<br>is also indicated for the<br>permanent reduction in<br>hair regrowth, defined<br>as the long-term,<br>stable reduction in the<br>number of hairs<br>regrowing when<br>measured at 6, 9 and 12<br>months after the<br>completion of a<br>treatment regime. | The IPL Hair Removal<br>Device Joy Version is<br>an over-the-counter<br>device intended for<br>removal of unwanted<br>hair such<br>as but not limited to<br>small areas such as<br>underarm and facial hair<br>below the chin line and<br>large areas<br>such as legs. | SE | |
| Source Energy | Supplied by external<br>adapter | AC Mains | an external power<br>supply | SE<br>Note 1 | |
| 'Use'<br>Classification | OTC | OTC | OTC | SE | |
| Device<br>Classification | Class II | Class II | Class II | SE | |
| Device Type | Intense Pulsed Light | Intense Pulsed Light | Intense Pulsed Light | SE | |
| Wavelength (nm) | 510nm~1200nm | 510nm~1100nm | 475nm~1200nm | SE<br>Note 2 | |
| Max. Fluence<br>(J/cm2) | 5.1 [Joules/cm²] | 3-6 [Joules/cm²] | 4.7[Joules/cm²] | SE<br>Note 3 | |
| Spot Size (cm2) | Body: 4.2 [cm²], Bikini and<br>Face: 2 [cm²] | 3[cm²] | 4.5[cm²] | SE<br>Note 4 | |
| Pulse duration | 9.20-11.20milliseconds | 2 ms- 10 ms | 11-13milliseconds | SE | |
| Elements of<br>Comparison | Subject Device | Predicate Device I | Predicate Device II | Remark | |
| Energy medium | Xenon Arc Flashlamp | Xenon Arc Flashlamp | Xenon Arc Flashlamp | SE | |
| Pulsing Control | Finger switch | Finger switch | Finger switch | SE | |
| Number of Output<br>Channels | One channel | One channel | One channel | SE | |
| Output energy<br>AND Energy<br>spectrum | body<br>lamp<br>cartridge<br> | Level 1: 11.77J@<br>510nm~1200nm<br>Level 2: 14.12J@<br>510nm~1200nm<br>Level 3: 17.12J@<br>510nm~1200nm<br>Level 4: 19.95J@<br>510nm~1200nm<br>Level 5: 22,21J@<br>510nm~1200nm | 9- 18J @<br>510nm~1100nm | Level 1: 11.5J @<br>475nm~1200nm<br>Level 2: 15.0J @<br>475nm~1200nm<br>Level 3: 17.0J@<br>475nm~1200nm<br>Level 4: 20.0J@<br>475nm~1200nm<br>Level 5: 22,0J@<br>475nm~1200nm | SE<br>Note 5 |
| | facial<br>lamp<br>cartridge | Level 1: 3.65J@<br>512nm~1197nm<br>Level 2: 4.59J@<br>512nm~1197nm<br>Level 3: 5.33J@<br>512nm~1197nm<br>Level 4: 6.16J@<br>512nm~1197nm<br>Level 5: 7.04J@<br>512nm~1197nm | | | |
| | bikini<br>lamp<br>cartridge | Level 1: 3.84J@<br>511nm~1200nm<br>Level 2: 4.92J@<br>511nm~1200nm<br>Level 3: 5.43J@<br>511nm~1200nm<br>Level 4: 6.23J@<br>511nm~1200nm<br>Level 5: 7.22J @<br>511nm~1200nm | | | |
| Software/Firmware/Microprocessor<br>Control? | Yes | Yes | Yes | SE | |
| 60601Compliance<br>with Voluntary<br>Standards | Yes<br>Comply with IEC 60601-1<br>and IEC 60601-1-2,<br>IEC60601-2-57 | Yes<br>Comply with IEC 60601-<br>1 and IEC 60601-1-2,<br>IEC60601-2-57, | Yes<br>Comply with IEC<br>60601-1 and IEC<br>60601-1-2,IEC60601-2-<br>57 | SE | |
| Elements of<br>Comparison | Subject Device | Predicate Device I | Predicate Device II | Remark | |
| Compliance* with<br>21 CFR 898 | No | No | No | SE | |
| Weight | 402g | 1Kg | -- | SE<br>Note 1 | |
| Dimensions | 126*78*200mm | 132*35*65mm (H*W*D) | -- | SE<br>Note 1 | |
| Standards | | | | | |
| Biocompatibility | All user directly contacting<br>materials are compliance<br>with ISO10993-5 and<br>ISO10993-10<br>requirements. | All user directly<br>contacting materials are<br>compliance with<br>ISO10993-5 and<br>ISO10993-10<br>requirements. | All user directly<br>contacting materials<br>are compliance with<br>ISO10993-5 and<br>ISO10993-10<br>requirements. | SE | |
| Electrical Safety | Comply with<br>IEC60601-1 and<br>IEC60601-2-57 | Comply with<br>IEC 60601-1 and<br>IEC 60601-2-57 | Comply with<br>IEC 60601-1 and<br>IEC 60601-2-57 | SE | |
{8}------------------------------------------------
{9}------------------------------------------------
# Comparison in Detail(s):
# Note 1:
"Power Source(s)", "Weight", "Dimensions" is belong to basic characteristics. Although it is a little different from the predicate devices, it will not affect the main function and the intended use of the device. They all also comply with IEC 60601-1 requirements. So the differences will not raise any safety or effectiveness issue.
# Note 2:
Although the Max. Fluence of subject device is a little larger than the one of the predicate device II, but comparing to Predicate device II, the energy density of subject device is less than 10 Joules/cm². And they all comply with IEC 60601-1, IEC 60601-2-57 requirement.
So the differences of function specification will not raise any safety or effectiveness issue.
# Note 3:
Although the wavelength of subject device is a little different from the predicate devices, but they all comply with IEC 60601-1, IEC 60601-2-57 requirement.
And the wavelength of subject device is in the range of the one of predicted devices "PerfectSmooth"
So the differences of function specification will not raise any safety or effectiveness issue.
{10}------------------------------------------------
# Note 4:
There are three types of "Spot Size (cm2)" of subject device, and there is minor difference between the subject device and the predicate device I, but in predicate device II, there is similar spot size to the size of bikini area /face lamp cartridge in subject device. And they all comply with IEC 60601-1.IEC60601-2-57 requirement.
So the differences of Spot size will not raise any safety or effectiveness issue.
# Note 5:
For IPL light, the energy spectrum is Continuous wavelength range. So even there is minor value difference on the energy spectrum between subject device and predicate devices, but the total energy output density (Max. Fluence value) is similar; and they comply with IEC60601-2-57 Requirement. So the differences of energy spectrum will not raise any safety or effectiveness issue.
And for facial lamp cartridge and bikini lamp cartridge, even the max output energy is lower than the data of predicate devices of K161428/K160968 due to the difference of spot size, but the total energy output density (Max. Fluence value) is similar; and the maximum output energy density of subject device for facial lamp cartridge is about 3.52 J/cm2; and 3.61 J/cm2 for bikini lamp cartridge. Comparing with the predicate device of K 160968, the value range of energy density of predicate device of K160968 is 3-6 J/cm2, which can cover the maximum output energy density of subject device for facial lamp cartridge and bikini lamp cartridge. The phototherapy effectiveness is crucially related to the output energy density and wavelength parameters for IPL, and the output energy of one pulse is minor different to predicate devices because of its spot window size. The spot size of device only affect the treatment target skin area per each pulse flashing, it do not raise any safety or effectiveness issue.
So the differences of output energy for facial lamp cartridge and bikini lamp cartridge will not raise any safety or effectiveness issue.
# Finial Conclusion:
The subject device " IPL Hair Removal Device Joy Version, Model: CB-027" is substantial Equivalence to all predicate devices.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.