← Product Code [HSB](/productcode/HSB) · K173811

# TRIWAY Tibiotalocalcaneal (TTC) Arthrodesis System (K173811)

_In2bones Sas · HSB · Mar 15, 2018 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K173811

## Device Facts

- **Applicant:** In2bones Sas
- **Product Code:** [HSB](/productcode/HSB.md)
- **Decision Date:** Mar 15, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3020
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The TRIWAY® TTC Arthrodesis System is intended for use in tibiotalocalcaneal arthrodesis and treatment of trauma to the hindfoot and distal tibia. Examples include: - Post-traumatic and degenerative arthritis involving both ankle and subtalar joints - Rheumatoid arthritis with severe deformity - Revision of failed ankle arthrodesis with subtalar involvement or with insufficient talar body - Revision of failed total ankle arthroplasty with subtalar intrusion - Talar deficiency conditions, including avascular necrosis (requiring a tibiocalcaneal arthrodesis) - Neuroarthropathy or neuromuscular deformity or other neuromuscular disease with severe deformity of the ankle, including Charcot foot - Severe pilon fractures with trauma to the subtalar joint - Malunited tibial pilon fractures The addition of an IBS™ 6.5mm compression screw through the subtalar joint and through the nail is required.

## Device Story

TRIWAY TTC Arthrodesis System provides rigid fixation for hindfoot/distal tibia arthrodesis and trauma. System comprises angulated intramedullary nail and various screws; includes instrument set for bone preparation, screw sizing, and insertion. Targeting device facilitates positioning and insertion of cotter screws and IBS 6.5mm compression screw. Compression screw inserted obliquely through nail, calcaneus, and talus to provide compression. Used in surgical settings by orthopedic surgeons. Device stabilizes bone segments to promote fusion; benefits patients by correcting deformities and treating degenerative or traumatic conditions.

## Clinical Evidence

Bench testing only. Static and dynamic bending tests performed by independent laboratory; results met acceptance criteria. Pyrogen testing conducted via Limulus Amebocyte Lysate (LAL) test per ANSI/AAMI ST72:2011.

## Technological Characteristics

Implants (nail and screws) manufactured from Ti6Al4V (ISO 5832-1, ASTM F136). Intramedullary fixation rod design. Sizes: 10-12mm diameter, 160-250mm length. Single-use. Sterilized via gamma irradiation.

## Regulatory Identification

An intramedullary fixation rod is a device intended to be implanted that consists of a rod made of alloys such as cobalt-chromium-molybdenum and stainless steel. It is inserted into the medullary (bone marrow) canal of long bones for the fixation of fractures.

## Predicate Devices

- Panta Nail ([K091788](/device/K091788.md))
- T2 Ankle Arthrodesis Nail ([K051590](/device/K051590.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration. The Department of Health & Human Services logo is on the left, and the U.S. Food & Drug Administration logo is on the right. The FDA logo is blue and white and includes the words "FDA U.S. FOOD & DRUG ADMINISTRATION".

March 15, 2018

In2Bones SAS % Christine Scifert Executive Vice President MRC-X, LLC 6075 Poplar Avenue, suite 500 N Memphis, Tennessee 38119

Re: K173811

Trade/Device Name: TRIWAY Tibiotalocalcaneal (TTC) Arthrodesis System Regulation Number: 21 CFR 888.3020 Regulation Name: Intramedullary Fixation Rod Regulatory Class: Class II Product Code: HSB, HWC Dated: December 14, 2017 Received: December 15, 2017

Dear Christine Scifert:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);

{1}------------------------------------------------

and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

## Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

510(k) Number (if known) K173811

Device Name

TRIWAY Tibiotalocalcaneal (TTC) Arthrodesis System

Indications for Use (Describe)

The TRIWAY® TTC Arthrodesis System is intended for use in tibiotalocalcaneal arthrodesis and treatment of trauma to the hindfoot and distal tibia. Examples include:

- · Post-traumatic and degenerative arthritis involving both ankle and subtalar joints
- · Rheumatoid arthritis with severe deformity
- · Revision of failed ankle arthrodesis with subtalar involvement or with insufficient talar body
- · Revision of failed total ankle arthroplasty with subtalar intrusion
- · Talar deficiency conditions, including avascular necrosis (requiring a tibiocalcaneal arthrodesis)
- · Neuroarthropathy or neuromuscular deformity or other neuromuscular disease with severe deformity of the ankle, including Charcot foot
- · Severe pilon fractures with trauma to the subtalar joint
- Malunited tibial pilon fractures

The addition of an IBS™ 6.5mm compression screw through the subtalar joint and through the nail is required.

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D) |
|----------------------------------------------|
| Over-The-Counter Use (21 CFR 801 Subpart C)  |

| X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for IN2BONES. The logo is set on a gray oval shape. The text "IN2BONES" is in white, with the "2" in orange.

# For TRIWAY® Tibiotalocalcaneal (TTC) Arthrodesis System

| Sponsor identification                                                                                                                                                                                                                                                                                                                                                                                                                                         | In2Bones SAS<br>28 chemin du Petit Bois<br>69130 Ecully – France<br>Phone: +33.4.72.29.26.26<br>Fax: +33.4.72.29.26.29                                                                                                                                                                                                                                                                                                             |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Establishment registration number                                                                                                                                                                                                                                                                                                                                                                                                                              | 3010470577                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Date of preparation                                                                                                                                                                                                                                                                                                                                                                                                                                            | December 13th, 2017                                                                                                                                                                                                                                                                                                                                                                                                                |
| Contact person                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Christine Scifert, MS, MEM<br>MRC-X, LLC<br>6075 Poplar Avenue, Suite 500<br>Memphis, TN 38119<br>Cell: 901-831-8053<br>Email: Christine.Scifert@mrc-x.com                                                                                                                                                                                                                                                                         |
| Proprietary Name                                                                                                                                                                                                                                                                                                                                                                                                                                               | TRIWAY Tibiotalocalcaneal (TTC) Arthrodesis System                                                                                                                                                                                                                                                                                                                                                                                 |
| Common name                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Tibiotalocalcaneal (TTC) Arthrodesis System                                                                                                                                                                                                                                                                                                                                                                                        |
| Device classification regulation                                                                                                                                                                                                                                                                                                                                                                                                                               | 21 CFR 888.3020: Intramedullary fixation rod<br>Class II                                                                                                                                                                                                                                                                                                                                                                           |
| Device Product Code and Panel                                                                                                                                                                                                                                                                                                                                                                                                                                  | HSB: nail<br>HWC: screw, fixation, bone<br>87 orthopedics                                                                                                                                                                                                                                                                                                                                                                          |
| Device<br>Description                                                                                                                                                                                                                                                                                                                                                                                                                                          | The TRIWAY® Tibiotalocalcaneal (TTC) Arthrodesis System is<br>composed of an angulated nail available in several sizes and<br>different screws for bony fixation. Rigid fixation is achieved by<br>screws.                                                                                                                                                                                                                         |
|                                                                                                                                                                                                                                                                                                                                                                                                                                                                | An instrument set is available for the bone preparation, screw sizing<br>and insertion of the TRIWAY® TTC Arthrodesis System. A<br>specific targeting device is available for the positioning and the<br>insertion of the different cotter screws and the I.B.S® 6.5mm<br>compression screw, which is inserted in an oblique way to provide<br>additional compression throughout the nail, between the calcaneus<br>and the talus. |
|                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Sizes:<br>The TRIWAY® Arthrodesis Nail is available in several sizes, with<br>left and right versions (diameter 10, 11, 12mm, lengths from 160mm<br>to 250mm).<br>Cotter screws are available in 5.0mm diameter, with length ranging<br>from 25 to 110mm by 5mm increment.                                                                                                                                                         |
|                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Material: The TRIWAY® TTC Arthrodesis System implants (nail<br>and screws) are manufactured from Ti6Al4V, as per ISO 5832-1 and<br>ASTM F136.                                                                                                                                                                                                                                                                                      |
|                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Single use: The TRIWAY® TTC Arthrodesis System is designed<br>for single use only.                                                                                                                                                                                                                                                                                                                                                 |
|                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Sterilization: The TRIWAY® TTC Arthrodesis System is supplied<br>sterile, using gamma irradiation.                                                                                                                                                                                                                                                                                                                                 |
| Predicate<br>Devices                                                                                                                                                                                                                                                                                                                                                                                                                                           | Primary Predicate:<br>Panta Nail (K091788), Newdeal, Acquired by Integra LifeSciences<br>Additional Predicate:                                                                                                                                                                                                                                                                                                                     |
|                                                                                                                                                                                                                                                                                                                                                                                                                                                                | T2 Ankle Arthrodesis Nail (K051590), Stryker                                                                                                                                                                                                                                                                                                                                                                                       |
| Indications for<br>use                                                                                                                                                                                                                                                                                                                                                                                                                                         | The TRIWAY® TTC Arthrodesis System is intended for use in<br>tibiotalocalcaneal arthrodesis and treatment of trauma to the<br>hindfoot and distal tibia. Examples include:<br>• Post-traumatic and degenerative arthritis involving both ankle and<br>subtalar joints<br>• Rheumatoid arthritis with severe deformity<br>• Revision of failed ankle arthrodesis with subtalar involvement or<br>with insufficient talar body       |
|                                                                                                                                                                                                                                                                                                                                                                                                                                                                | • Revision of failed total ankle arthroplasty with subtalar intrusion                                                                                                                                                                                                                                                                                                                                                              |
| Talar deficiency conditions, including avascular necrosis (requiring a tibiocalcaneal arthrodesis) Neuroarthropathy or neuromuscular deformity or other neuromuscular disease with severe deformity or instability of the ankle, including Charcot foot Severe pilon fractures with trauma to the subtalar joint Malunited tibial pilon fractures The addition of an IBS™ 6.5mm compression screw through the subtalar joint and through the nail is required. |                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Comparison of Indications and Technological characteristics and Substantial Equivalence Summary                                                                                                                                                                                                                                                                                                                                                                | The In2Bones TRIWAY® TTC Arthrodesis System is similar to the predicate devices Newdeal Panta Nail (K091788), and Stryker T2 Ankle Arthrodesis Nail (K051590) in indications for us, intended use, design, size ranges, principle of operation and materials.                                                                                                                                                                      |
| Summary Performance Data                                                                                                                                                                                                                                                                                                                                                                                                                                       | Performance testing of the TRIWAY® TTC Arthrodesis System was assessed through mechanical bench testing performed by an independent test laboratory.<br>Testing performed included static and dynamic bending tests on TRIWAY® TTC Arthrodesis System. The results indicate that the TRIWAY® TTC Arthrodesis System met the acceptance criteria.                                                                                   |
| Pyrogen testing                                                                                                                                                                                                                                                                                                                                                                                                                                                | The method used to make the determination that the device meets pyrogen limit specification is the Limulus Amebocyte Lysate (LAL) test in accordance with ANSI/AAMI ST72:2011: Bacterial endotoxins - Test methods, routine monitoring and alternative to batch testing.                                                                                                                                                           |
| Conclusion                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Based on the comparison of indications for use and technological characteristics and the results of the testing performed, the TRIWAY® TTC Arthrodesis System is substantially equivalent to the predicate devices identified in the 510(k) submission.                                                                                                                                                                            |

{4}------------------------------------------------

{5}------------------------------------------------

---

**Source:** [https://fda.innolitics.com/device/K173811](https://fda.innolitics.com/device/K173811)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
