TRIWAY Tibiotalocalcaneal (TTC) Arthrodesis System
K173811 · In2bones Sas · HSB · Mar 15, 2018 · Orthopedic
Device Facts
Record ID
K173811
Device Name
TRIWAY Tibiotalocalcaneal (TTC) Arthrodesis System
Applicant
In2bones Sas
Product Code
HSB · Orthopedic
Decision Date
Mar 15, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3020
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The TRIWAY® TTC Arthrodesis System is intended for use in tibiotalocalcaneal arthrodesis and treatment of trauma to the hindfoot and distal tibia. Examples include: - Post-traumatic and degenerative arthritis involving both ankle and subtalar joints - Rheumatoid arthritis with severe deformity - Revision of failed ankle arthrodesis with subtalar involvement or with insufficient talar body - Revision of failed total ankle arthroplasty with subtalar intrusion - Talar deficiency conditions, including avascular necrosis (requiring a tibiocalcaneal arthrodesis) - Neuroarthropathy or neuromuscular deformity or other neuromuscular disease with severe deformity of the ankle, including Charcot foot - Severe pilon fractures with trauma to the subtalar joint - Malunited tibial pilon fractures The addition of an IBS™ 6.5mm compression screw through the subtalar joint and through the nail is required.
Device Story
TRIWAY TTC Arthrodesis System provides rigid fixation for hindfoot/distal tibia arthrodesis and trauma. System comprises angulated intramedullary nail and various screws; includes instrument set for bone preparation, screw sizing, and insertion. Targeting device facilitates positioning and insertion of cotter screws and IBS 6.5mm compression screw. Compression screw inserted obliquely through nail, calcaneus, and talus to provide compression. Used in surgical settings by orthopedic surgeons. Device stabilizes bone segments to promote fusion; benefits patients by correcting deformities and treating degenerative or traumatic conditions.
Clinical Evidence
Bench testing only. Static and dynamic bending tests performed by independent laboratory; results met acceptance criteria. Pyrogen testing conducted via Limulus Amebocyte Lysate (LAL) test per ANSI/AAMI ST72:2011.
Technological Characteristics
Implants (nail and screws) manufactured from Ti6Al4V (ISO 5832-1, ASTM F136). Intramedullary fixation rod design. Sizes: 10-12mm diameter, 160-250mm length. Single-use. Sterilized via gamma irradiation.
Indications for Use
Indicated for patients requiring tibiotalocalcaneal arthrodesis or treatment of hindfoot/distal tibia trauma, including post-traumatic/degenerative arthritis, rheumatoid arthritis, failed ankle arthrodesis or total ankle arthroplasty, talar deficiency (e.g., avascular necrosis), neuroarthropathy, neuromuscular deformity (e.g., Charcot foot), and pilon fractures.
Regulatory Classification
Identification
An intramedullary fixation rod is a device intended to be implanted that consists of a rod made of alloys such as cobalt-chromium-molybdenum and stainless steel. It is inserted into the medullary (bone marrow) canal of long bones for the fixation of fractures.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration. The Department of Health & Human Services logo is on the left, and the U.S. Food & Drug Administration logo is on the right. The FDA logo is blue and white and includes the words "FDA U.S. FOOD & DRUG ADMINISTRATION".
March 15, 2018
In2Bones SAS % Christine Scifert Executive Vice President MRC-X, LLC 6075 Poplar Avenue, suite 500 N Memphis, Tennessee 38119
Re: K173811
Trade/Device Name: TRIWAY Tibiotalocalcaneal (TTC) Arthrodesis System Regulation Number: 21 CFR 888.3020 Regulation Name: Intramedullary Fixation Rod Regulatory Class: Class II Product Code: HSB, HWC Dated: December 14, 2017 Received: December 15, 2017
Dear Christine Scifert:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
## Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K173811
Device Name
TRIWAY Tibiotalocalcaneal (TTC) Arthrodesis System
Indications for Use (Describe)
The TRIWAY® TTC Arthrodesis System is intended for use in tibiotalocalcaneal arthrodesis and treatment of trauma to the hindfoot and distal tibia. Examples include:
- · Post-traumatic and degenerative arthritis involving both ankle and subtalar joints
- · Rheumatoid arthritis with severe deformity
- · Revision of failed ankle arthrodesis with subtalar involvement or with insufficient talar body
- · Revision of failed total ankle arthroplasty with subtalar intrusion
- · Talar deficiency conditions, including avascular necrosis (requiring a tibiocalcaneal arthrodesis)
- · Neuroarthropathy or neuromuscular deformity or other neuromuscular disease with severe deformity of the ankle, including Charcot foot
- · Severe pilon fractures with trauma to the subtalar joint
- Malunited tibial pilon fractures
The addition of an IBS™ 6.5mm compression screw through the subtalar joint and through the nail is required.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) |
|----------------------------------------------|
| Over-The-Counter Use (21 CFR 801 Subpart C) |
| X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for IN2BONES. The logo is set on a gray oval shape. The text "IN2BONES" is in white, with the "2" in orange.
# For TRIWAY® Tibiotalocalcaneal (TTC) Arthrodesis System
| Sponsor identification | In2Bones SAS<br>28 chemin du Petit Bois<br>69130 Ecully – France<br>Phone: +33.4.72.29.26.26<br>Fax: +33.4.72.29.26.29 |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Establishment registration number | 3010470577 |
| Date of preparation | December 13th, 2017 |
| Contact person | Christine Scifert, MS, MEM<br>MRC-X, LLC<br>6075 Poplar Avenue, Suite 500<br>Memphis, TN 38119<br>Cell: 901-831-8053<br>Email: Christine.Scifert@mrc-x.com |
| Proprietary Name | TRIWAY Tibiotalocalcaneal (TTC) Arthrodesis System |
| Common name | Tibiotalocalcaneal (TTC) Arthrodesis System |
| Device classification regulation | 21 CFR 888.3020: Intramedullary fixation rod<br>Class II |
| Device Product Code and Panel | HSB: nail<br>HWC: screw, fixation, bone<br>87 orthopedics |
| Device<br>Description | The TRIWAY® Tibiotalocalcaneal (TTC) Arthrodesis System is<br>composed of an angulated nail available in several sizes and<br>different screws for bony fixation. Rigid fixation is achieved by<br>screws. |
| | An instrument set is available for the bone preparation, screw sizing<br>and insertion of the TRIWAY® TTC Arthrodesis System. A<br>specific targeting device is available for the positioning and the<br>insertion of the different cotter screws and the I.B.S® 6.5mm<br>compression screw, which is inserted in an oblique way to provide<br>additional compression throughout the nail, between the calcaneus<br>and the talus. |
| | Sizes:<br>The TRIWAY® Arthrodesis Nail is available in several sizes, with<br>left and right versions (diameter 10, 11, 12mm, lengths from 160mm<br>to 250mm).<br>Cotter screws are available in 5.0mm diameter, with length ranging<br>from 25 to 110mm by 5mm increment. |
| | Material: The TRIWAY® TTC Arthrodesis System implants (nail<br>and screws) are manufactured from Ti6Al4V, as per ISO 5832-1 and<br>ASTM F136. |
| | Single use: The TRIWAY® TTC Arthrodesis System is designed<br>for single use only. |
| | Sterilization: The TRIWAY® TTC Arthrodesis System is supplied<br>sterile, using gamma irradiation. |
| Predicate<br>Devices | Primary Predicate:<br>Panta Nail (K091788), Newdeal, Acquired by Integra LifeSciences<br>Additional Predicate: |
| | T2 Ankle Arthrodesis Nail (K051590), Stryker |
| Indications for<br>use | The TRIWAY® TTC Arthrodesis System is intended for use in<br>tibiotalocalcaneal arthrodesis and treatment of trauma to the<br>hindfoot and distal tibia. Examples include:<br>• Post-traumatic and degenerative arthritis involving both ankle and<br>subtalar joints<br>• Rheumatoid arthritis with severe deformity<br>• Revision of failed ankle arthrodesis with subtalar involvement or<br>with insufficient talar body |
| | • Revision of failed total ankle arthroplasty with subtalar intrusion |
| Talar deficiency conditions, including avascular necrosis (requiring a tibiocalcaneal arthrodesis) Neuroarthropathy or neuromuscular deformity or other neuromuscular disease with severe deformity or instability of the ankle, including Charcot foot Severe pilon fractures with trauma to the subtalar joint Malunited tibial pilon fractures The addition of an IBS™ 6.5mm compression screw through the subtalar joint and through the nail is required. | |
| Comparison of Indications and Technological characteristics and Substantial Equivalence Summary | The In2Bones TRIWAY® TTC Arthrodesis System is similar to the predicate devices Newdeal Panta Nail (K091788), and Stryker T2 Ankle Arthrodesis Nail (K051590) in indications for us, intended use, design, size ranges, principle of operation and materials. |
| Summary Performance Data | Performance testing of the TRIWAY® TTC Arthrodesis System was assessed through mechanical bench testing performed by an independent test laboratory.<br>Testing performed included static and dynamic bending tests on TRIWAY® TTC Arthrodesis System. The results indicate that the TRIWAY® TTC Arthrodesis System met the acceptance criteria. |
| Pyrogen testing | The method used to make the determination that the device meets pyrogen limit specification is the Limulus Amebocyte Lysate (LAL) test in accordance with ANSI/AAMI ST72:2011: Bacterial endotoxins - Test methods, routine monitoring and alternative to batch testing. |
| Conclusion | Based on the comparison of indications for use and technological characteristics and the results of the testing performed, the TRIWAY® TTC Arthrodesis System is substantially equivalent to the predicate devices identified in the 510(k) submission. |
{4}------------------------------------------------
{5}------------------------------------------------
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.