← Product Code [DQY](/productcode/DQY) · K173799

# NaviCross 0.018 (K173799)

_Terumo Corporation · DQY · Mar 29, 2018 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K173799

## Device Facts

- **Applicant:** Terumo Corporation
- **Product Code:** [DQY](/productcode/DQY.md)
- **Decision Date:** Mar 29, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1250
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The NaviCross 0.018" is intended to guide and support a guide wire during access of the peripheral vasculature, allow for wire exchanges and provide a conduit for the delivery of saline or diagnostic contrast agents.

## Device Story

NaviCross 0.018" is a single-use support catheter for peripheral vascular procedures. Device consists of a three-layer shaft (polyamide outer, stainless steel mesh braid, PTFE inner), a polyamide distal tip, and a hub. Features a hydrophilic coating on the distal 40 cm to facilitate navigation. Operated manually by physicians in clinical settings to guide/support guidewires, perform wire exchanges, and deliver saline or contrast media. Radiopaque tungsten markers aid visualization under fluoroscopy. Device provides a conduit for interventional procedures; benefits include improved trackability and support within peripheral vasculature. Performance verified through bench testing including tensile force, leakage, power injection, and simulated use.

## Clinical Evidence

No clinical data. Bench testing only. Performance testing included radio-detectability, tensile force, leakage, power injection, torque strength, flow rate, bending stiffness, and simulated use (reachability/usability). Biocompatibility testing conducted per ISO 10993-1 for externally communicating, circulating blood, limited contact devices.

## Technological Characteristics

Three-layer catheter shaft: polyamide outer, stainless steel mesh braid, PTFE inner. Distal tip: polyamide. Radiopaque markers: tungsten. Hydrophilic coating: dimethyl acrylamide-glycidyl methacrylate copolymer. Hub: polyamide. Anti-kink protector: polyester elastomer. Adhesive: cyanoacrylate. Dimensions: 2.6 Fr, 0.55 mm ID, 0.85 mm OD. Lengths: 65-150 cm. Sterilization: Ethylene oxide (ISO 11135:2014). Manual operation.

## Regulatory Identification

A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.

## Predicate Devices

- TERUMO SUPPORT CATHETER ([K110540](/device/K110540.md))

## Reference Devices

- CXI 0.018" support catheter ([K160884](/device/K160884.md))
- Quick-Cross 0.018" support catheter ([K033678](/device/K033678.md))
- Rubicon 0.018" support catheter ([K122394](/device/K122394.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.

March 29, 2018

Terumo Medical Corporation Ms. Yuko Watanabe Sr. Regulatory Affairs Specialist 265 Davidson Ave., Suite 320 Somerset, New Jersey 08873

#### Re: K173799

Trade/Device Name: NaviCross 0.018" Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DOY Dated: March 16, 2018 Received: March 16, 2018

#### Dear Ms. Watanabe:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);

{1}------------------------------------------------

#### Page 2 - Ms. Yuko Watanabe

and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

M.A. Willhemen

for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K173799

Device Name NaviCross 0.018"

Indications for Use (Describe)

The NaviCross 0.018" is intended to guide wire during access of the peripheral vasculature, allow for wire exchanges and provide a conduit for the delivery of saline or diagnostic contrast agents.

Type of Use (Select one or both, as applicable)

| <span></span> <span></span> Prescription Use (Part 21 CFR 801 Subpart D) |
|--------------------------------------------------------------------------|
| <span></span> Over-The-Counter Use (21 CFR 801 Subpart C)                |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(K) SUMMARY

#### A. SUBMITTER INFORMATION (807.92(a)(1))

| <b>Prepared by:</b> | Yuko Watanabe                        |
|---------------------|--------------------------------------|
|                     | Senior Regulatory Affairs Specialist |
|                     | Terumo Medical Corporation           |
|                     | Tel. (732) 302-4900                  |
|                     | Fax (410) 398-6079                   |

#### Prepared for: Owner/Operator

Terumo Corporation 44-1, 2-Chome, Hatagaya Shibuya-Ku, Tokyo Japan 151-0072 Owner/Operator Number: 801 002 6

#### Manufacturer and Sterilization Facility (Applicant)

Ashitaka Factory of Terumo Corporation 150 Maimaigi-cho Fujinomiya, Shizuoka 418-0015, Japan Registration Number: 968 183 4

Contact Person: Yuko Watanabe Senior Regulatory Affairs Specialist Terumo Medical Corporation 265 Davidson Avenue, Suite 320 Somerset, NJ 08873 Tel. (732) 302-4900 Fax (410) 398-6079 E-mail: yuko.watanabe@terumomedical.com

December 13th, 2017 Date prepared:

{4}------------------------------------------------

#### B. DEVICE NAME (807.92(a)(2))

| Proprietary Name:     | NaviCross 0.018"      |
|-----------------------|-----------------------|
| Common Name:          | Support Catheter      |
| Classification Name:  | Percutaneous Catheter |
| Classification Panel: | Cardiovascular        |
| Regulation:           | 21 CFR 870.1250       |
| Product Code:         | DQY                   |
| Classification:       | Class II              |

## C. PREDICATE DEVICE (807.92(a)(3))

The legally marketed device to which substantial equivalence is claimed is:

Predicate Device: K110540 – TERUMO SUPPORT CATHETER, manufactured by Ashitaka Factory of Terumo Corporation.

Reference Devices:

- 1. K160884: CXI 0.018" support catheter, Cook, Inc.
- 2. K033678: Quick-Cross 0.018" support catheter, Spectranetics, Inc.
- 3. K122394: Rubicon 0.018" support catheter, Boston Scientific Corp.

# D. REASON FOR 510(k) SUBMISSION

This traditional 510(k) is being submitted for the NaviCross 0.018", which will be a line extension of the currently marketed NaviCross 0.035" (K110540, cleared under the name "Terumo Support Catheter"), for the purposes of establishing substantial equivalence to a legally marketed predicate device.

## E. DEVICE DESCRIPTION (807.92(a)(4))

## Principle of Operation Technology

The NaviCross 0.018" submitted in this 510(k) and its predicate (K110540) are operated by a manual process.

{5}------------------------------------------------

## Design/Construction

The NaviCross 0.018" is a single use, ethylene oxide sterilized device that is intended to guide and support a guide wire during access of the peripheral vasculature, allow for wire exchanges and provide a conduit for the delivery of saline or diagnostic contrast agents.

NaviCross 0.018" features a three-layer construction, which consists of a stainless steel mesh braid sandwiched between an outer layer of polyamide and an inner layer of polytetrafluoroethylene. It has a distal tip that is comprised of a polyamide. It also has a hydrophilic coating over the distal 40 cm of the catheter.

The device is offered in effective lengths of 65, 90, 135 and 150 cm. French size and shaft inner diameter are as follows:

| French Size | Shaft Inner Diameter (mm) |
|-------------|---------------------------|
| 2.6Fr       | $0.55 \pm 0.02$           |

## Materials

The materials for the NaviCross 0.018" are provided in the table below.

| No. | Name of Component           |             | Raw material                                             |
|-----|-----------------------------|-------------|----------------------------------------------------------|
| 1*  | Shaft                       | Outer layer | Polyamide                                                |
|     |                             | Pigment     |                                                          |
| 2   | Braid†                      |             | Stainless steel                                          |
| 3*  | Catheter                    | Inner layer | Polytetrafluoroethylene                                  |
| 4*  |                             | Distal tip† | Polyamide                                                |
|     |                             |             | Pigment                                                  |
| 5   | Radiopaque marker           |             | Tungsten                                                 |
| 6*  | Hydrophilic polymer coating |             | Dimethyl acrylamide –<br>glycidyl methacrylate copolymer |
| 7   | Quick-drying glue           |             | Cyanoacrylate                                            |
| 8*  | Hub                         |             | Polyamide                                                |
| 9   | Anti-kink protector         |             | Polyester elastomer                                      |
|     |                             |             | Pigment                                                  |

#### Table 5.1: List of Materials

*Blood contacting material.

*Distal-tip is not braided.

{6}------------------------------------------------

#### Specifications

The specifications for the NaviCross 0.018" are provided in the table below.

| Part                             | Specification                                                           |
|----------------------------------|-------------------------------------------------------------------------|
| Catheter Size                    | 2.6 Fr.                                                                 |
| Catheter ID/OD                   | $0.55 \pm 0.02$ mm/ $0.85 \pm 0.03$ mm                                  |
| Catheter Effective Lengths*      | 65 $\pm$ 1.5 cm, 90 $\pm$ 1.5 cm,<br>135 $\pm$ 1.5 cm, 150 $\pm$ 1.5 cm |
| Hydrophilic Coating Length       | Distal 40 $\pm$ 3cm                                                     |
| Maximum guidewire outer diameter | 0.018"                                                                  |

Table 5.2: NaviCross 0.018" Specifications

*The length from the proximal anti-kink protector to the catheter distal tip.

# F. INDICATIONS FOR USE (807.92(a)(5))

The NaviCross 0.018" is intended to guide and support a guide wire during access of the peripheral vasculature, allow for wire exchanges and provide a conduit for the delivery of saline or diagnostic contrast agents.

The indications for use are equivalent to the predicate (K110540).

# G. SUBSTANTIAL EQUIVALENCE COMPARISON (807.92(a)(6))

The NaviCross 0.018", the subject of this Traditional 510(k), is substantially equivalent in its intended use, technology/principle of operation, materials, and performance to the predicate, K110540 - TERUMO SUPPORT CATHETER, manufactured by Ashitaka Factory of Terumo Corporation.

In addition to the above-listed primary predicate, Terumo has identified the following reference devices. These are market leading devices with the same intended use and basic design as the subject device. Because these devices are frequently used in clinical practice, Terumo felt it was appropriate to use them as references when setting the acceptance criteria for NaviCross 0.018" performance testing.

- 1. Cook Inc. CXI 0.018" support catheter (K160884)
- 2. Spectranetics, Inc. Quick-Cross 0.018" support catheter (K033678)
- 3. Boston Scientific Corp. Rubicon 0.018" support catheter (K122394)

{7}------------------------------------------------

A comparison of the technological characteristics is summarized in the table below.

{8}------------------------------------------------

Terumo Corporation
Traditional 510(k) – NaviCross 0.01
Section 5 510(k) Summary

| Device<br>Characteristic               | Subject Device:<br>NaviCross 0.018"                                                                                                                                                                                              | Predicate Device:<br>TERUMO SUPPORT<br>CATHETER (NaviCross<br>0.035") (K110540)                                                                                                                                               | Reference Device #1:<br>CXI 0.018" support<br>catheter<br>(K160884)                                                                                        | Reference Device #2:<br>Quick-Cross 0.018"<br>support catheter<br>(K033678)                                                                                                                                                                                                                     | Reference Device #3:<br>Rubicon 0.018"<br>support catheter<br>(K122394)                                                                                                                                                                                                            |
|----------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer                           | Ashitaka Factory of<br>Terumo Corporation                                                                                                                                                                                        | Same                                                                                                                                                                                                                          | Cook Inc.                                                                                                                                                  | Spectranetics, Inc.                                                                                                                                                                                                                                                                             | Boston Scientific Corp.                                                                                                                                                                                                                                                            |
| Intended Use<br>Indications for<br>Use | It is intended to guide and<br>support a guide wire<br>during access of the<br>peripheral vasculature,<br>allow for wire exchanges<br>and provide a conduit for<br>the delivery of saline or<br>diagnostic contrast agents.      | It is intended to guide and<br>support a guide wire<br>during access of the<br>vasculature, allow for<br>wire exchanges and<br>provide a conduit for the<br>delivery of saline or<br>diagnostic contrast agents.              | It is intended for use in<br>small vessel or<br>superselective<br>anatomy for diagnostic<br>and interventional<br>procedures, including<br>peripheral use. | It is designed for<br>use in the vascular<br>system. The catheters<br>are intended to support a<br>guidewire during access<br>of the vasculature, allow<br>for exchange of<br>guidewires, and provide<br>a conduit for the<br>delivery of saline<br>solutions or diagnostic<br>contrast agents. | It is intended to<br>facilitate placement and<br>support of guidewires<br>and other Interventional<br>devices within the<br>peripheral vasculature<br>and to allow for<br>exchange of guidewires,<br>and provide a conduit<br>for the delivery of saline<br>or contrast solutions. |
| Operation<br>Principle                 | Manual                                                                                                                                                                                                                           | Same                                                                                                                                                                                                                          | Same                                                                                                                                                       | Same                                                                                                                                                                                                                                                                                            | Same                                                                                                                                                                                                                                                                               |
| Design/<br>Construction                | Three-layer construction<br>catheter shaft with<br>hydrophilic coating, distal<br>tip and hub                                                                                                                                    | Same                                                                                                                                                                                                                          | Multi layers with<br>hydrophilic coating,<br>distal tip and hub                                                                                            | One layer with<br>hydrophilic coating,<br>distal tip and hub                                                                                                                                                                                                                                    | Multi layers with<br>hydrophilic coating,<br>distal tip and hub                                                                                                                                                                                                                    |
| Device<br>Characteristic               | Subject Device:<br>NaviCross 0.018"                                                                                                                                                                                              | Predicate Device:<br>TERUMO SUPPORT<br>CATHETER (NaviCross<br>0.035") (K110540)                                                                                                                                               | Reference Device #1:<br>CXI 0.018" support<br>catheter<br>(K160884)                                                                                        | Reference Device #2:<br>Quick-Cross 0.018"<br>support catheter<br>(K033678)                                                                                                                                                                                                                     | Reference Device #3:<br>Rubicon 0.018"<br>support catheter<br>(K122394)                                                                                                                                                                                                            |
|                                        | Catheter shaft<br>Outer layer*:<br>Polyamide w/<br>pigment                                                                                                                                                                       | Catheter shaft<br>Outer layer*:<br>Polyester                                                                                                                                                                                  | Information not<br>publicly available.                                                                                                                     | Information not publicly<br>available.                                                                                                                                                                                                                                                          | Information not publicly<br>available.                                                                                                                                                                                                                                             |
|                                        | Mesh braid: Stainless<br>steel                                                                                                                                                                                                   | Mesh braid: Stainless<br>steel                                                                                                                                                                                                |                                                                                                                                                            |                                                                                                                                                                                                                                                                                                 |                                                                                                                                                                                                                                                                                    |
|                                        | Inner layer*:<br>Polytetrafluoroethyle<br>ne                                                                                                                                                                                     | Inner layer*:<br>Polyester<br>elastomer/tungsten                                                                                                                                                                              |                                                                                                                                                            |                                                                                                                                                                                                                                                                                                 |                                                                                                                                                                                                                                                                                    |
| Materials                              | Hydrophilic coating*:<br>Dimethyl acrylamide-<br>glycidyl methacrylate<br>copolymer<br>Three radiopaque<br>markers: Tungsten                                                                                                     | Hydrophilic coating*:<br>Dimethyl acrylamide-<br>glycidyl methacrylate<br>copolymer<br>Three radiopaque<br>markers*: Platinum<br>alloy                                                                                        |                                                                                                                                                            |                                                                                                                                                                                                                                                                                                 |                                                                                                                                                                                                                                                                                    |
|                                        | Distal Tip*:<br>Polyamide w/<br>pigment<br>Hub*: Polyamide<br>Anti-kink protector:<br>Polyester elastomer<br>w/ pigment<br>Adhesive:<br>Cyanoacrylate<br>*: blood contacting<br>material                                         | Hub*: Polyamide<br>Anti-kink protector:<br>Polyester elastomer<br>w/ pigment<br>Adhesive:<br>Cyanoacrylate<br>*: blood contacting<br>material                                                                                 |                                                                                                                                                            |                                                                                                                                                                                                                                                                                                 |                                                                                                                                                                                                                                                                                    |
| Device<br>Characteristic               | Subject Device:<br>NaviCross 0.018"                                                                                                                                                                                              | Predicate Device:<br>TERUMO SUPPORT<br>CATHETER (NaviCross<br>0.035") (K110540)                                                                                                                                               | Reference Device #1:<br>CXI 0.018" support<br>catheter<br>(K160884)                                                                                        | Reference Device #2:<br>Quick-Cross 0.018"<br>support catheter<br>(K033678)                                                                                                                                                                                                                     | Reference Device #3:<br>Rubicon 0.018"<br>support catheter<br>(K122394)                                                                                                                                                                                                            |
| Package                                | Individual package on<br>which the product<br>label and the peel-off<br>labels are attached 1 unit per package                                                                                                                   | Same                                                                                                                                                                                                                          | Same                                                                                                                                                       | Same                                                                                                                                                                                                                                                                                            | Same                                                                                                                                                                                                                                                                               |
| Specifications                         | Effective lengths: 65,<br>90, 135 and 150 cm French size: 2.6Fr<br>(0.85 mm) O.D.: 0.85 mm I.D.: 0.55 mm Maximum guidewire<br>outer diameter: 0.018" Distal tip shape:<br>straight/angled Maximum injection<br>pressure: 300 psi | Effective lengths: 65,<br>90, 135 and 150 cm French size: 4Fr<br>(1.39 mm) O.D.: 1.39 mm I.D.:1.05 mm Maximum guidewire<br>outer diameter: 0.035" Distal tip shape:<br>straight/angled Maximum injection<br>pressure: 750 psi | 90, 150 cm 2.6 Fr O.D.: 0.87 mm I.D.: unknown Same (0.018") Same<br>(straight/angled) 1200 psi                                                             | 90, 135 and 150 cm 3.4 Fr O.D.: 1.12 mm<br>(Proximal shaft) I.D.: unknown Same (0.018") Straight only Same (300 psi)                                                                                                                                                                            | 90, 135 and 150 cm 4.0 Fr O.D.: 1.35 mm<br>(Proximal shaft) I.D.: unknown Same (0.018") Straight only Same (300 psi)                                                                                                                                                               |
| Sterilization                          | Ethylene oxide                                                                                                                                                                                                                   | Same                                                                                                                                                                                                                          | Same                                                                                                                                                       | Radiation                                                                                                                                                                                                                                                                                       | Same                                                                                                                                                                                                                                                                               |
| Shelf life                             | 36 months                                                                                                                                                                                                                        | Same                                                                                                                                                                                                                          | Unknown                                                                                                                                                    | Unknown                                                                                                                                                                                                                                                                                         | Unknown                                                                                                                                                                                                                                                                            |

{9}------------------------------------------------

Terumo Corporation
Traditional 510(k) — NaviCross 0.01
Section 5 510(k) Summary

{10}------------------------------------------------

Terumo Corporation
Traditional 510(k) — NaviCross 0.01
Section 5 510(k) Summary

Page 8 of 11

{11}------------------------------------------------

# H. NON CLINICAL TESTS (807.92(b)(1))

#### Performance Testing

Performance testing was conducted to ensure the safety and effectiveness of the NaviCross 0.018" throughout the shelf life, verify conformity to the applicable external and internal standards, and demonstrate substantial equivalence to the predicate device. With the exception of the Radio-detectability test, the following performance tests were performed on non-aged and accelerated aged samples. The following table provides a list of performance tests that were performed on the NaviCross 0.018".

| Test Item                                      |
|------------------------------------------------|
| Radio-detectability                            |
| Surface                                        |
| Peak tensile force                             |
| Freedom from leakage                           |
| Power injection                                |
| Distal tip                                     |
| Fluid leakage (Hub)                            |
| Sub-atmospheric pressure air leakage (Hub)     |
| Stress cracking (Hub)                          |
| Resistance to separation from axial load (Hub) |
| Resistance to separation from unscrewing (Hub) |
| Resistance to overriding (Hub)                 |
| Particulate evaluation                         |
| Coating integrity                              |
| Torque strength                                |
| Distal tip strength                            |
| Flow rate                                      |
| Product dimension                              |
| Bending stiffness of catheter shaft            |
| Flexibility and kink test                      |
| Torque transmission property                   |
| Interior sliding characteristics               |
| Exterior sliding characteristics               |
| Wire support characteristics                   |
| Simulated Use - Reachability                   |
| Simulated Use - Usability Test                 |

|  |  | Table 5.4: Summary of Performance Testing |  |
|--|--|-------------------------------------------|--|
|--|--|-------------------------------------------|--|

<sup>1</sup> Only non-aged sample was tested since the amount of contrast media contained in the product would not change over time.

{12}------------------------------------------------

Performance testing met the predetermined acceptance criteria and is acceptable for clinical use throughout its shelf life.

## Biocompatibility

In accordance with ISO 10993-1, the NaviCross 0.018" is classified as: Externally Communicating Device, Circulating Blood, Limited Contact (<24 hours). The finished device's patient contacting parts were tested in accordance with the tests recommended in the FDA Guidance for Industry and Food and Drug Administration Staff - Use of International Standard ISO-10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process." Screening tests were performed on accelerated aged devices to show that the biocompatibility is maintained throughout the shelf life of the product. The table below provides a list of biocompatibility tests conducted on the NaviCross 0.018".

| Non-aged, sterile, whole device                     |
|-----------------------------------------------------|
| Cytotoxicity                                        |
| Sensitization                                       |
| Intracutaneous Reactivity                           |
| Acute Systemic Toxicity                             |
| Pyrogenicity                                        |
| Hemolysis                                           |
| Thrombogenicity                                     |
| Complement Activation (Immunology)                  |
| Physicochemical Profile (Physicochemical and FT-IR) |
| Accelerated-aged (3 years), sterile, whole device   |
| Cytotoxicity                                        |
| Hemolysis                                           |
| Physicochemical Profile (Physicochemical and FT-IR) |

Table 5.5: Summary of ISO 10993 Biocompatibility Testing

Results of the testing demonstrate that the device is biocompatible throughout the shelf life of the product.

{13}------------------------------------------------

## Sterilization

The sterility of the device is assured using a sterilization method validated in accordance with ISO 11135:2014, Sterilization of Health Care Products - Ethylene Oxide – Requirements for the development, validation and routine control of a sterilization process for medical devices, to provide a Sterility Assurance Level (SAL) of 10-6.

# I. CLINICAL TESTS (807.92(b)(2))

This 510(k) does not include data from clinical tests.

# J. CONCLUSION (807.92(b)(3))

In summary, the NaviCross 0.018", the subject of this 510(k), is substantially equivalent in its intended use, technology/principle of operation, materials, and performance to the predicate, K110540 - TERUMO SUPPORT CATHETER, manufactured by Ashitaka Factory of Terumo Corporation.

---

**Source:** [https://fda.innolitics.com/device/K173799](https://fda.innolitics.com/device/K173799)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
