XactTrace Single Use Respiratory Effort Belt System
K173793 · Embla Systems · MNR · Mar 1, 2019 · Anesthesiology
Device Facts
Record ID
K173793
Device Name
XactTrace Single Use Respiratory Effort Belt System
Applicant
Embla Systems
Product Code
MNR · Anesthesiology
Decision Date
Mar 1, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.2375
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The XactTrace® Single Use Respiratory Effort Belt System is intended to measure respiratory effort to assist in the diagnosis of sleep disorders or sleep related respiratory effort signals measured are processed to provide electrical signals suitable for connection to the inputs of physiological recording equipment. The intended environments are hospitals, institutions, sleep centers, sleep clinics. The XactTrace® Single Use Respiratory Effort Belt System is intended for diagnostics purposes only and is not intended to be used as an apnea monitor.
Device Story
Device uses Respiratory Inductive Plethysmography (RIP) to measure changes in abdominal and thoracic circumference; converts inductance changes into electrical signals for physiological recording equipment. Available in 'Cut-to-Fit' (disposable belt with locks) and 'Pre-Sized' (disposable belt with snap sensors) variations. Used in hospitals, sleep centers, and clinics by medical professionals. Interfaces with compatible Embla PSG system amplifiers or an optional external Inductive Interface Cable (RIP processor). Provides qualitative/quantitative respiratory effort data to assist clinicians in diagnosing sleep disorders. Benefits include disposable, single-patient use design reducing cross-contamination risks while maintaining signal equivalence to reusable predicate systems.
Clinical Evidence
Bench testing only. Evaluations confirmed dimensional requirements, component compatibility, resistance measurements, cleaning durability of reusable connectors, environmental robustness, and signal output equivalence compared to the primary predicate (Universal XactTrace) across both Embla amplifier and Inductive Interface Cable configurations. EMC compliance verified per IEC 60601-1-2, 4th edition.
Technological Characteristics
Respiratory Inductive Plethysmography (RIP) sensor. Materials: Texturized polyester (warp/weft) and Dorlastan filament (elastane). Connectivity: 1.5mm female touchproof connectors; compatible with Embla PSG systems or external Inductive Interface Cable. Power: Passive sensor; signal interface powered by external cable or PSG system. Form factor: Disposable belts (Cut-to-Fit or Pre-Sized). Complies with 21 CFR 898 performance standards for electrode lead wires/cables.
Indications for Use
Indicated for adult and pediatric patients requiring respiratory effort measurement during sleep disorder studies to assist in diagnosis of sleep-related respiratory disorders. Not for use as an apnea monitor.
Regulatory Classification
Identification
A breathing (ventilatory) frequency monitor is a device intended to measure or monitor a patient's respiratory rate. The device may provide an audible or visible alarm when the respiratory rate, averaged over time, is outside operator settable alarm limits. This device does not include the apnea monitor classified in § 868.2377.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" and "ADMINISTRATION" in blue text.
March 1, 2019
Embla Systems Sanjay Mehta Director, Quality Assurance & Regulatory Affairs 1 Hines Road. Suite 202 Kanata, k2k 3C7 Ca
Re: K173793
Trade/Device Name: XactTrace Single Use Respiratory Effort Belt System Regulation Number: 21 CFR 868.2375 Regulation Name: Breathing Frequency Monitor Regulatory Class: Class II Product Code: MNR Dated: January 29, 2019 Received: January 30, 2019
Dear Sanjay Mehta:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# James J. Lee -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital. Respiratory. Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K173793
Device Name
The XactTrace® Single Use Respiratory Effort Belt System
Indications for Use (Describe)
The XactTrace® Single Use Respiratory Effort Belt System is intended to measure respiratory effort to assist in the diagnosis of sleep disorders or sleep related respiratory effort signals measured are processed to provide electrical signals suitable for connection to the inputs of physiological recording equipment.
The intended environments are hospitals, institutions, sleep centers, sleep clinics.
The XactTrace® Single Use Respiratory Effort Belt System is intended for diagnostics purposes only and is not intended to be used as an apnea monitor.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image contains the logo for Natus Neurology. The word "natus" is in a teal sans-serif font, with a small registered trademark symbol next to it. Below "natus" is the word "neurology" in a smaller, italicized, black sans-serif font.
# K173793 510(k) Summary
#### Submitter Information
| Submitter's Name: | Embla Systems |
|---------------------------|----------------------------------------------------------------------|
| Submitter's Address: | 1 Hines Road, Suite 202<br>Kanata, ON, CA, K2K 3C7 |
| Submitter's Phone: | (905) 287-5055 |
| Contact Person: | Mr. Sanjay Mehta<br>Director, Quality Assurance & Regulatory Affairs |
| Date Summary<br>Prepared: | February 27, 2019 |
### Subject Device
| Trade Name: | XactTrace® Single Use Respiratory Effort Belt System |
|----------------------|----------------------------------------------------------|
| Common Name: | Respiratory Effort Sensor |
| Classification Name: | 21 CFR 868.2375, Ventilatory Effort Recorder<br>Class II |
| Product Code: | MNR |
Product Code:
#### Predicate Device(s)
| 510(k) Number: | K043132 (Primary Predicate) |
|----------------------|----------------------------------------------|
| Trade Name: | Universal XactTrace |
| Submitter Name: | Medcare Flaga |
| Classification Name: | 21 CFR 868.2375, Breathing frequency monitor |
| Product Code: | MNR |
K923402 Crystal Trace Piezo Respiratory Effort Sensor Pro-Tech, Inc. 21 CFR 868.2375 Breathing frequency monitor BZQ
510(k) Number: Trade Name: Submitter Name: Classification Name: Product Code:
K042253 SleepSense Sleep Sensors SLP Ltd. 21 CFR 868.2375, Breathing frequency monitor MNR
#### Reference Device(s)
| 510(k) Number: | K024322 | K0417 |
|----------------------|-------------------------------------------------|-------------------------------------------------|
| Trade Name: | Embla N7000 | Compass |
| Submitter Name: | Medcare Flaga | Medcare |
| Classification Name: | 21 CFR 868.2375, Breathing frequency<br>monitor | 21 CFR 868.2375, Breathing frequency<br>monitor |
| Product Code: | MNR | MNR |
24 ss M10 System re Flaga R 868.2375, Breathing frequency MNR
#### Device Description
The XactTrace Single Use Respiratory Effort Belt System subject device applies Respiratory Inductive Plethysmography (RIP) Sensor Belts to support respiratory effort signals used in the diagnosis of sleep disorders or sleep related respiratory disorders. The XactTrace Single Use Respiratory Effort Belt System subject device supports a measure of changes in inductance as a result of changes in abdominal, thorax circumference.
There are two variations of the XactTrace® Single Use Respiratory Effort Belt System:
1. XactTrace® Single Use Cut-to-Fit Respiratory Effort Belt System, and
2. XactTrace® Single Use Pre-Sized Respiratory Effort Belt System.
The XactTrace Single Use Cut-to-Fit Respiratory Effort Belt variation consists of a single use disposable Sensor Belt and Locks (belt connectors) with a cable that connects to a recording device. The system supports conversion
{4}------------------------------------------------
of changes in inductance to a digital signal that provides both qualitative information of respiratory effort. The Sensor Belt of the XactTrace Single Use Cut-To-Fit Belt variation is cut by the medical professional to appropriate lengths to fit patient abdomen, thorax circumference.
The XactTrace Single Use Pre-Sized Respiratory Effort Belt variation consists of a single use disposable Sensor Belt and Snap Sensors (Sensor Belt connectors) with a cable that connects to a recording device. The XactTrace Single Use Pre-Sized Respiratory Effort Belt variation supports conversion of Sensor Belt changes in inductance to a signal that provides both qualitative and quantitative information of respiratory effort. XactTrace Single Use Pre-Sized Respiratory Effort belts are available in four sizes: Pediatric, Small, Medium and Large.
An optional Inductive Interface Cable (RIP Processor) is available to interface with PSG systems that accept the 1.5mm female touchproof connectors, and do not have internal RIP Sensor Belt technology.
Both variations of XactTrace Respiratory Effort Belts are interfaced with either an optional external Inductive Interface Cable or a RIP technology compatible Embla PSG system amplifier. Compatible Embla amplifiers are identified by keyhole connectors for Sensor Belt input.
### Statement of Intended Use
The XactTrace® Single Use Respiratory Effort Belt System is intended to measure respiratory effort to assist in the diagnosis of sleep disorders or sleep related respiratory effort signatory effort signals measured are processed to provide electrical signals suitable for connection to the inputs of physiological recording equipment. The intended environments are hospitals, institutions, sleep clinics. The XactTrace® Single Use Respiratory Effort Belt System is intended for diagnostics purposes only and is not intended to be used as an apnea monitor.
#### Summary of Technological Characteristics and Comparison to Predicate and Reference Devices
The fundamental RIP technology of the XactTrace® Single Use Respiratory Effort Belt System is the primary predicate Universal XactTrace.
The comparison to the primary predicate Universal XactTrace Belt to create the XactTrace® Single Use Respiration Belt System under review are summarized as:
- 1. Creation of two single use variations:
- XactTrace Single Use Pre-Sized Respiratory Effort Belt a. Equivalent to the predicate Universal XactTrace reusable belt sizes, the single use disposable belt is available in pediatric, small, medium, and large sizes.
- b. XactTrace Single Use Cut-to-Fit Respiratory Effort Belt A single use disposable belt the medical professional cuts to fit the patient size. The Cut-to-Fit RIP sensor belt with associated Locks (connectors) is the same as described in K024322 and K041724.
- 2. The XactTrace Single Use Respiration Belts under review are either interfaced with RIP technology compatible Embla PSG systems or an external Inductive Interface Cable, which is an internal battery powered signal interface. The primary predicate Universal XactTrace is powered by a battery for interface within the Sensor Belt connector.
The following table provides a side-by-side comparison of the XactTrace Single Use Respiration Effort Belt System under review to the predicate devices applied to support this pre-market notification.
| Feature | XactTrace Single<br>Use Respiratory<br>Effort Belt<br>System<br>(under review) | Universal<br>XactTrace<br>(Primary<br>Predicate,<br>K043132) | Crystal Trace<br>Piezo Respiratory<br>Effort Sensor<br>(K923402) | SLP Ltd.<br>RIP Sensor, RIP<br>Processor<br>(K042253) | Equivalence<br>Comments,<br>Discussion |
|----------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>Use | The XactTrace®<br>Single Use<br>Respiratory Effort<br>Belt System is | Universal<br>XactTrace is<br>intended to<br>measure | The Crystal Trace<br>Piezo Respiratory<br>Effort Sensor is<br>intended for use | SleepSense sleep<br>sensors provide a<br>qualitative<br>measure of some | Identical to the<br>primary predicate<br>Universal XactTrace |
| | | | | | |
| Feature | XactTrace Single<br>Use Respiratory<br>Effort Belt<br>System<br>(under review) | Universal<br>XactTrace<br>(Primary<br>Predicate,<br>K043132) | Crystal Trace<br>Piezo Respiratory<br>Effort Sensor<br>(K923402) | SLP Ltd.<br>RIP Sensor, RIP<br>Processor<br>(K042253) | Equivalence<br>Comments,<br>Discussion |
| | intended to<br>measure<br>respiratory effort<br>to assist in the<br>diagnosis of sleep<br>disorders or sleep<br>related respiratory<br>disorders. The<br>respiratory effort<br>signals measured<br>are processed to<br>provide electrical<br>signals suitable for<br>connection to the<br>inputs of<br>physiological<br>recording<br>equipment. The<br>intended<br>environments are<br>hospitals,<br>institutions, sleep<br>centers, sleep<br>clinics. The<br>XactTrace Single<br>Use Respiratory<br>Effort Belt System<br>is intended for<br>diagnostics<br>purposes only and<br>is not intended to<br>be used as an<br>apnea monitor. | respiratory effort<br>to assist in the<br>diagnosis of sleep<br>disorders or sleep<br>related respiratory<br>disorders. The<br>respiratory effort<br>signals measured<br>are processed to<br>provide electrical<br>signals suitable for<br>connection to the<br>inputs of<br>physiological<br>recording<br>equipment. The<br>intended<br>environments are<br>hospitals,<br>institutions, sleep<br>centers, sleep<br>clinics, or other<br>test environments.<br>Universal<br>XactTrace is<br>intended for<br>diagnostics<br>purposes only and<br>is not intended to<br>be used as an<br>apnea monitor. | during sleep<br>disorder studies to<br>detect respiratory<br>effort for recording<br>onto a<br>physiological<br>recorder. | physiological<br>parameters for<br>recording onto an<br>FDA-cleared data<br>acquisition system.<br>Their target<br>population:<br>Children and adult<br>patients who are<br>screened during<br>sleep disorder<br>studies. Their<br>environment of use<br>is usually at a<br>sleep laboratory or<br>sometimes at the<br>patient's home. | with the exception of<br>product name change.<br>Equivalent to the<br>predicate Crystal<br>Trace Piezo<br>Respiratory Effort<br>Sensor and SLP Ltd.<br>RIP Sensor and RIP<br>Processor.<br>Note:<br>The SLP Ltd 510(k)<br>addresses various<br>sleep sensors and<br>interface cables,<br>including a RIP<br>Sensor and RIP<br>Processor. |
| Population of<br>use | Adult and pediatric | Adult and pediatric | Adult and pediatric | Children and<br>adults | Same |
| Environment of<br>Use | Hospitals,<br>institutions, sleep<br>centers, sleep<br>clinics. | Hospitals,<br>institutions, sleep<br>centers, sleep<br>clinics, or other<br>test environments. | Hospitals,<br>institutions, sleep<br>centers, sleep<br>clinics, or other<br>test environments. | Hospitals,<br>institutions, sleep<br>centers, sleep<br>clinics, or other<br>test environments. | Same |
| Single patient<br>use | Yes | No, reusable | No, reusable | No, reusable | The XactTrace®<br>Single Use<br>Respiratory Effort<br>Belt System, sensor<br>belts under review is<br>single patient use,<br>disposable |
| Duration of use | Duration of | Duration of | Duration of | Duration of | Same |
| | | | | | |
| Feature | XactTrace Single<br>Use Respiratory<br>Effort Belt<br>System<br>(under review) | Universal<br>XactTrace<br>(Primary<br>Predicate,<br>K043132) | Crystal Trace<br>Piezo Respiratory<br>Effort Sensor<br>(K923402) | SLP Ltd.<br>RIP Sensor, RIP<br>Processor<br>(K042253) | Equivalence<br>Comments,<br>Discussion |
| | anticipated to be ≤<br>12 hours. | anticipated to be ≤<br>12 hours. | anticipated to be ≤<br>12 hours. | anticipated to be ≤<br>12 hours. | |
| Fundamental<br>technology | Respiratory<br>Inductive<br>Plethysmography<br>(RIP) | RIP | Piezo | RIP | Both RIP and Piezo<br>technologies apply a<br>change in<br>circumference of a<br>flexible Sensor Belt<br>to provide a low level<br>signal that is related<br>to the change in<br>circumference. |
| Variations | Cut-to-Fit<br>Pre-sized | Pre-sized | Pre-sized | Pre-sized | Equivalent Pre-sized<br>variations. The cut-to-<br>fit variation was<br>included in the<br>Reference device<br>510(k) review. |
| Universal Rip<br>variations<br>(Use with non-<br>Embla System) | Yes (two)<br>Abdomen and<br>Thorax | Yes (two)<br>Abdomen and<br>Thorax | Not applicable | Yes (two)<br>Abdomen and<br>Thorax | Same |
| Patient applied<br>location | Abdomen and<br>Thorax | Abdomen and<br>Thorax | Abdomen and<br>Thorax | Abdomen and<br>Thorax | Same |
| User Adjustable<br>Sensor Belt<br>Lengths | Yes | Yes | Yes | Yes | Same |
| Pre-Sized Sensor<br>Belt<br>(Circumference) | XactTrace Single<br>Use Respiratory<br>Effort Belt<br>System: Pre-sized<br>variations of large,<br>medium, small and<br>pediatric (inches).<br>Large: 40 to 75<br>Medium: 30 to 55<br>Small: 16 to 35<br>Pediatric: 11 to 22 | XactTrace<br>Reusable Embla<br>RIP Belts: Pre-size<br>variations of large,<br>medium, small and<br>pediatric (inches).<br>Large: 47 to 66<br>Medium: 33 to 56<br>Small: 22 to 41<br>Pediatric: 19 to 28 | Yes<br>Adult Size:<br>28 to 80 inches<br>Pediatric (Ped)<br>Size: 10.5 to<br>34 inches | Disposable:<br>120 cm and 150<br>cm | Both the primary<br>predicate Universal<br>XactTrace and<br>XactTrace Single Use<br>Respiratory Effort<br>Belt System under<br>review may be<br>purchased in Pre-<br>Sized belt lengths.<br>The primary predicate<br>Universal XactTrace<br>510(k) is applied to<br>adult and pediatric<br>patient populations,<br>the same as the device<br>under review.<br>Variations are<br>medical professional<br>adjustable with regard<br>to Sensor Belt length. |
| Patient contact | Not in direct<br>contact with bare<br>skin. | Not in direct<br>contact with bare<br>skin. | Yes | Unknown | Same as the primary<br>predicate Universal<br>XactTrace. |
| Feature | XactTrace Single<br>Use Respiratory<br>Effort Belt<br>System<br>(under review) | Universal<br>XactTrace<br>(Primary<br>Predicate,<br>K043132) | Crystal Trace<br>Piezo Respiratory<br>Effort Sensor<br>(K923402) | SLP Ltd.<br>RIP Sensor, RIP<br>Processor<br>(K042253) | Equivalence<br>Comments,<br>Discussion |
| Belt connectors<br>and cables to<br>recorder | Touchproof safety<br>connection,<br>complying with<br>the 21 CFR 898<br>Performance<br>Standard for<br>Electrode Lead<br>Wires and Patient<br>Cables. | Touchproof safety<br>connection,<br>complying with<br>the 21 CFR 898<br>Performance<br>Standard for<br>Electrode Lead<br>Wires and Patient<br>Cables. | Recessed female<br>safety connection. | Recessed female<br>safety connection,<br>complying with<br>the 21 CFR 898<br>Performance<br>Standard for<br>Electrode Lead<br>Wires and Patient<br>Cables. | Same. |
| Sensor Belt<br>Material | Warp: Texturized<br>polyester raw<br>white<br>Weft: Texturized<br>polyester raw<br>white<br>Elasthane:<br>Dorlastan Filament<br>Type | Warped:<br>Texturized<br>polyester colored;<br>Weft: Cotton;<br>Elastic: Dorlastan<br>Filament Type | Unknown | Unknown | The Sensor Belt<br>material does not<br>contact the patient. |
| Power source | XactTrace Single<br>Use Respiratory<br>Effort Belt System,<br>Single Use Cut-to-<br>Fit and Pre-Sized<br>Universal RIP<br>Belts: Interface<br>with optional<br>Inductive Interface<br>Cable (RIP<br>processor)<br>(Non-replaceable 3<br>volt battery)<br>XactTrace Single<br>Use Respiratory<br>Effort Belt System,<br>Single Use Cut-to-<br>Fit and Pre-Sized<br>variations:<br>Interface with<br>Embla PSG system | XactTrace<br>Reusable<br>Universal RIP<br>Belts:<br>Non-replaceable 3<br>volt battery in<br>connector (Snap<br>Sensor)<br>XactTrace<br>Reusable Embla<br>RIP Belts:<br>Powered by Embla<br>PSG system | None<br>The piezo sensor<br>belt self generates<br>a small voltage in<br>response to<br>circumference<br>change. | Non-replaceable 3<br>volt battery in RIP<br>Processor | Same.<br>The RIP technology<br>power source is either<br>a non-replaceable 3<br>volt battery or the<br>Embla PSG system. |
| Maximum signal<br>amplitude<br>(peak to peak) | 10 mV<br>(Inductive<br>Interface Cable<br>(RIP Processor )) | 11.2 mV | Unknown | 10 mV | The medical<br>professional's<br>recording equipment<br>accommodates signal<br>amplitudes with<br>amplifier settings. |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the logo for Natus Neurology. The word "natus" is written in a teal, sans-serif font. Below it, the word "neurology" is written in a smaller, italicized, dark-colored font. The logo is simple and modern.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Natus Neurology. The word "natus" is in a teal sans-serif font, with a small registered trademark symbol next to it. Below "natus" is the word "neurology" in a smaller, black, italicized sans-serif font.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image contains the logo for Natus Neurology. The word "natus" is in a teal sans-serif font, with a registered trademark symbol next to it. Below that, the word "neurology" is in a smaller, black, italicized sans-serif font.
As noted, the differences do not affect substantial equivalence of safety and effectiveness.
- The Sensor Belt component is modified to support single patient use, disposable variations. The change to a single use disposable Sensor Belt does not raise new questions of safety or effectiveness.
- . Cut-to-Fit and Pre-Sized variations of the XactTrace Single Use Respiratory Effort Belt System are available. The user can cut the Sensor Belt length in the Cut-to-Fit variation. The Pre-Sized XactTrace variations are
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image contains the logo for Natus Neurology. The word "natus" is in a larger, teal-colored font on the top line. Below it, "neurology" is written in a smaller, black font. The logo is simple and modern.
similar lengths as the primary predicate Universal XactTrace lengths. The selection and use of Sensor Belt length are at the discretion of the medical professional.
- . The XactTrace Single Use Respiratory Effort Belt System connectors to the Sensor Belt do not contain any battery or signal conditioning electronics. The battery and signal conditioning is performed in an external, commercially available Inductive Interface Cable or a RIP technology compatible amplifier.
Modifications were performed in compliance with the Embla Systems design control process.
### Performance Testing
To establish the technical equivalency of the XactTrace Single Use Respiratory Effort Belt System, non-clinical, bench evaluations were conducted to confirm compliance with performance requirements, summarized in the following table.
| Requirement | Method | Result |
|-----------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| Dimensional requirements | Measurement of components. | Pass |
| Lock and Snap Sensor compatibility | Components connect properly and perform as<br>intended. | Pass |
| Compatibility with Inductive Interface Cable and<br>RIP technology compatible Embla PSG recording<br>system input connections, key connector inputs | Interface with Inductive Interface Cables and Embla<br>PSG recording system input connections, key<br>connectors, recoding. | Pass |
| Resistance of XactTrace Single Use Cut-to-Fit<br>Effort Belt, variation of XactTrace Single Use<br>Respiratory Effort Belt System | Cut belts to size, confirm resistance measurement. | Pass |
| Resistance of XactTrace Single Use Pre-Sized<br>Effort Belt, variation of XactTrace Single Use<br>Respiratory Effort Belt System | Confirm resistance measurement. | Pass |
| Cleaning of reusable Snap Sensor and Locks | Recommended cleaning method does not cause<br>damage. | Pass |
| Operational temperature and humidity range | No damage with operational temperature and<br>humidity. | Pass |
| Equivalence of breathing effort signal output | Compare breathing effort signal output for the<br>XactTrace Single Use Respiratory Effort Belt System<br>technology under review to primary predicate<br>Universal XactTrace for both Embla amplifier and<br>Inductive Interface Cable. | Pass |
| Equivalence of breathing effort signal output for<br>same displacement of both XactTrace Single Use<br>Sized Effort Belt, variations. | For the same displacement of Sensor Belt, the output<br>from the XactTrace Single Use Cut-to-Fit Effort Belt<br>is equivalent to the output from the XactTrace Single<br>Use Pre-Sized Effort Belt. | Pass |
| EMC compliance with Inductive Interface Cables. | Confirm compliance with IEC 60601-1-2, 4th edition. | Pass |
#### Conclusion
The performance of the XactTrace Single Use Respiratory Effort Belt System is confirmed. All design verification and validation results meet requirements. No new questions of safety or effectiveness are raised by the XactTrace Single Use Respiratory Belt Systems XactTrace Single Use Respiratory Belt System is equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.