MicroPort total hip systems are intended for use in total hip arthroplasty for reduction or relief of pain and/or improved hip function in skeletally mature patients with the following conditions: 1. non-inflammatory degenerative joint disease such as osteoarthritis, avascular necrosis, ankylosis, protrusio acetabuli. and painful hip dysplasia; 2. inflammatory degenerative joint disease such as rheumatoid arthritis; 3. correction of functional deformity; and, 4. revision procedures where other treatments or devices have failed.
Device Story
BIOLOX delta Option and Extra-long Heads are orthopedic implants consisting of femoral heads and neck sleeves used in total hip arthroplasty. Femoral heads are manufactured from alumina matrix composite; neck sleeves are manufactured from titanium alloy. The device functions as a modular component of a hip prosthesis system, connecting the femoral stem to the acetabular liner. It is used in clinical settings by orthopedic surgeons during hip replacement or revision surgery. The device provides a bearing surface and structural connection to restore hip function and relieve pain. The system is compatible with various previously cleared MicroPort femoral stems and acetabular shells. No software or AI algorithms are involved.
Clinical Evidence
No clinical data provided. Safety and effectiveness supported by bench testing, including static and post-fatigue burst strength, axial pull-off testing, rotational stability testing, and fretting corrosion evaluation.
Technological Characteristics
Femoral heads: alumina matrix composite; 28-44mm sizes; 16/18 or 12/14 conical bores. Neck sleeves: titanium alloy; short to extra-long sizes; 16/18 external taper, 12/14 internal taper. Modular implant system for total hip arthroplasty.
Indications for Use
Indicated for skeletally mature patients requiring total hip arthroplasty for non-inflammatory degenerative joint disease (osteoarthritis, avascular necrosis, ankylosis, protrusio acetabuli, painful hip dysplasia), inflammatory degenerative joint disease (rheumatoid arthritis), functional deformity correction, or revision procedures.
Regulatory Classification
Identification
A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
July 6, 2018
MicroPort Orthopedics Inc. Allen Mamaril Regulatory Affairs Specialist I 5677 Airline Road Arlington, Tennessee 38002
Re: K173776
Trade/Device Name: BIOLOX delta Option and Extra-long Heads Regulation Number: 21 CFR 888.3353 Regulation Name: Hip Joint Metal/Ceramic/Polymer Semi-Constrained Cemented Or Nonporous Uncemented Prosthesis Regulatory Class: Class II Product Code: LZO Dated: June 1, 2018 Received: June 4, 2018
Dear Allen Mamaril:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Vincent J. Devlin -S
for
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known)
K173776
Device Name BIOLOX® Delta Option and Extra-long Heads
Indications for Use (Describe)
MicroPort total hip systems are intended for use in total hip arthroplasty for reduction or relief of pain and/or improved hip function in skeletally mature patients with the following conditions:
1. non-inflammatory degenerative joint disease such as osteoarthritis, avascular necrosis, ankylosis, protrusio acetabuli. and painful hip dysplasia;
2. inflammatory degenerative joint disease such as rheumatoid arthritis;
3. correction of functional deformity; and,
4. revision procedures where other treatments or devices have failed.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for MicroPort Orthopedics. The logo consists of a blue square with a red star inside, followed by the text "MicroPort" in blue and "Orthopedics" in red. Below the logo is the text "MicroPort Orthopedics Inc." in black.
# 510(K) Summary
In accordance with the Food and Drug Administration Rule to implement provisions of Safe Medical Devices Act of 1990 and in conformance with 21 CFR 807.92, this information serves as a Summary of Safety and Effectiveness for the use of the BIOLOX® delta Option and Extra-long Heads.
| Submitted by: | MicroPort Orthopedics Inc.<br>5677 Airline Rd, Arlington, TN, 38002<br>Phone 866-872-0211<br>Fax: 855-466-2247 |
|--------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date: | December 11, 2017 |
| Contact Person: | Allen Mamaril<br>Regulatory Affairs Specialist I |
| Proprietary Name: | BIOLOX® delta Option and Extra-long Heads |
| Common Name: | Femoral Head, Neck Sleeve |
| Classification Name and Reference: | 21 CFR 888.3353 – Hip joint metal/ceramic/polymer semi-<br>constrained<br>cemented or nonporous uncemented<br>prosthesis—Class II |
| Subject Product Code and Panel Code: | Orthopedics/87/LZO |
| Predicate Device: | Femoral Head:<br>K130376 DYNASTY® Acetabular System with Ceramic<br>Neck Sleeve:<br>K072656 DYNASTY® Acetabular System |
| Reference Device: | Femoral Head:<br>K140043 DYNASTY® Acetabular System with Ceramic<br>K953025 SLT 28mm XXL Femoral Head<br>K141653 Biolox® delta Option Ceramic Heads<br>K131684 Biolox® delta Ceramic heads<br>K131518 Mectacer Biolox Option Heads<br>K112115 Mectacer Biolox Delta Heads<br>Neck Sleeve:<br>K141653 Biolox® delta Option Ceramic Heads<br>K131518 Mectacer Biolox Option Heads |
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for MicroPort Orthopedics. The logo consists of a blue square with a red star inside, followed by the text "MicroPort" in blue and "Orthopedics" in red. Below the logo, the text "MicroPort Orthopedics Inc." is written in a smaller font.
#### DEVICE INFORMATION
# A. Intended Use
MicroPort total hip systems are intended for use in total hip arthroplasty for reduction or relief of pain and/or improved hip function in skeletally mature patients with the following conditions:
- non-inflammatory degenerative joint disease such as osteoarthritis, avascular necrosis, 1. ankylosis, protrusio acetabuli, and painful hip dysplasia;
- 2. inflammatory degenerative joint disease such as rheumatoid arthritis;
- 3. correction of functional deformity; and,
- revision procedures where other treatments or devices have failed. 4.
# B. Device Description
The BIOLOX® delta Option and Extra-long Heads include femoral heads and neck sleeves. The design features are summarized below:
- Femoral Heads ●
- BIOLOX® delta Option: o
- Manufactured from alumina matrix composite
- Head sizes: 28mm, 32mm, 36mm, 40mm, 44mm ■
- Conical Bore: 16/18 Taper
- BIOLOX® delta Extra-long: o
- Manufactured from alumina matrix composite ■
- Head Sizes: 32mm, 36mm, 40mm
- 트 Conical Bore: 12/14 Taper
- Neck Sleeves
- o BIOLOX® delta Option:
- Manufactured from titanium alloy ■
- 트 Neck Sleeve sizes: Short, Medium, Long, and Extra-long
- . Conical Bore: 16/18 External Taper, 12/14 Internal Taper
#### C. Substantial Equivalence Information
The design features and materials of the BIOLOX® delta Option and Extra-long Heads are substantially equivalent to those of the predicate devices cleared under K130376 and K072656. The indications of the subject devices are identical to the predicates. The fundamental scientific technology of the subject devices has not changed relative to the predicate devices. The safety and effectiveness of subject devices are adequately supported by the substantial equivalence information, materials information, and analysis data provided within this Premarket Notification.
# D. Nonclinical Testing
The subject BIOLOX® delta Option femoral heads were evaluated mechanically for static (or pre-fatigue) burst strength and post-fatigue burst strength.
The subject devices were evaluated for the stability of the taper system subjected to traction forces between the BIOLOX® delta Option femoral head and BIOLOX® delta Option neck sleeve by conducting an axial pull off test to measure distraction forces.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image contains the logo for MicroPort Orthopedics. The logo features a blue square with rounded corners, inside of which is a red diamond shape. To the right of the square is the text "MicroPort" in blue, with "Orthopedics" underneath in red. Below the logo is the text "MicroPort Orthopedics Inc."
Rotational stability tests were also performed to measure the torque required to twist the BIOLOX® delta Option femoral head on the tapers.
Lastly, the subject devices were tested for fretting corrosion of the BIOLOX® delta Option sleeve and tapers of the stem/neck mating components.
# E. Clinical Testing
Clinical data was not provided for the subject devices.
# F. Biocompatibility
Biocompatibility information for the subject devices is included in the cited Master File. The intended patient contact, materials, colorants, manufacturing steps and processing additives used in the subject implant devices are identical to those of the predicate and reference devices. Therefore, additional biocompatibility testing was not completed on the subject devices.
The subject devices are implanted using existing specialized trial and impactor instruments, for which colorant and biocompatibility information were reviewed by FDA in K170444.
# G. Component and Accessory Compatibility
Tables 1 and 2 show the compatibility of the subject devices with previously cleared MicroPort Orthopedics products.
#### Table 1. Compatible Shells and Liners, Including 510(k) information
| 510(k) | Device Name |
|---------|--------------------------------------------|
| K061547 | DYNASTY® Beaded Shells and Liners |
| K070785 | DYNASTY® Beaded Shells and Liners |
| K082924 | DYNASTY® BIOFOAM® Shells and Liners |
| K122382 | DYNASTY® BIOFOAM® Shells and Liners |
| K142119 | PROCOTYL® L Beaded Shells and Liners |
| K170444 | PROCOTYL® PRIME Shell and Liners |
| K171181 | PROCOTYL® PRIME E-CLASS™ Shells and Liners |
#### Table 2. Compatible Femoral Components, Including 510(k) Information
| 510(k) | Device Name |
|---------|-------------------------------|
| K003016 | PRO-FEMUR R |
| K012091 | PRO-FEMUR |
| K021346 | STEM HIP REPLACEMENT SYSTEM |
| K041114 | PROFEMUR TAPERED HIP STEM |
| K041586 | PROFEMUR S HIP STEM |
| K051995 | PROFEMUR RENAISSANCE HIP STEM |
| K052915 | PROFEMUR XTR HIP STEM |
| K053588 | PROFEMUR LX HIP STEM |
| K060358 | PROFEMUR TL HIP STEM |
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image contains the logo for MicroPort Orthopedics Inc. The logo features a blue square with rounded corners and a red star inside. To the right of the square is the text "MicroPort" in blue, with "Orthopedics" in red below it. The text "MicroPort Orthopedics Inc." is below the logo.
| K080663 | PROFEMUR LX REVISION 5/8 COATED HIP STEM |
|---------|--------------------------------------------------|
| K081090 | PROFEMUR LX 5/8 COATED HIP STEM |
| K091423 | PROFEMUR HIP SYSTEM MODULAR NECKS |
| K100866 | PROFEMUR HIP SYSTEM MODULAR NECKS |
| K110399 | GLADIATOR PLASMA CLASSIC HIP STEM |
| K111698 | PROFEMUR(R) E CEMENTLESS HIP STEM |
| K111699 | PROFEMUR(R) Z TITANIUM PLASMA SPRAYED HIP STEM |
| K111910 | GLADIATOR HIP STEM |
| K112080 | PRESERVE HIP STEM |
| K112150 | PROFEMUR GLADIATOR HA HIP STEM |
| K121221 | PROFEMUR Z REVISION HIP STEM |
| K123434 | PROFEMUR Z CLASSIC STEM |
| K123688 | PROFEMUR TL CLASSIC STEM |
| K130984 | PROFEMUR RENAISSANCE CLASSIC STEM |
| K140676 | PROFEMUR TL CLASSIC LONG NECK HIP STEMS |
| K141235 | PROFEMUR RENAISSANCE CLASSIC LONG NECK HIP STEMS |
| K150133 | PROFEMUR PRESERVE SIZE 1-3 HIP STEMS |
| K150302 | PROFEMUR PRESERVE CLASSIC STEM |
# H. Conclusions
The safety and effectiveness of the BIOLOX® delta Option and Extra-long Heads are adequately supported by the substantial equivalence information, materials information, and nonclinical testing data provided within this premarket notification.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.