K173716 · Quest Medical, Inc. · DTR · Oct 1, 2018 · Cardiovascular
Device Facts
Record ID
K173716
Device Name
MPS2 Myocardial Protection System Console
Applicant
Quest Medical, Inc.
Product Code
DTR · Cardiovascular
Decision Date
Oct 1, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.4240
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Quest Medical MPS®2 Myocardial Protection System, consisting of the MPS2 console and the MPS Delivery Set together, is intended for use by perfusionists and physicians to deliver whole blood (from any arterial source) and/or cardioplegia solutions to the heart during open heart surgery on either an arrested or beart for use up to six hours in duration.
Device Story
MPS2 Myocardial Protection System console manages myocardial perfusion during open heart surgery; used by perfusionists and physicians in clinical settings. Device functions include heat exchange, temperature control, pressure control, flow rate control, automatic priming, and air detection/removal. Operates in three flow modes: normal, cyclic, and low volume. System delivers whole blood and/or cardioplegia solutions to the heart. Provides precise control over delivery parameters to support myocardial protection during procedures. Labeling update includes validated disinfection protocols for the water circulation system to mitigate biofilm formation and microbial contamination risks.
Clinical Evidence
Bench testing only. Validation of disinfection protocols performed per AAMI TIR12, AAMI TIR30, and FDA reprocessing guidance. Testing demonstrated 6-Log reduction for Pseudomonas aeruginosa and 3-Log reduction for Mycobacterium chimaera. Compatibility of disinfectant/inhibitor with water system materials confirmed. Surface cleaning and heat exchanger performance testing also conducted.
Technological Characteristics
Console-based myocardial perfusion system. Features include heat exchanger, temperature/pressure/flow control, and automated priming/air management. Water circulation system materials validated for compatibility with specific disinfectants and microbial growth inhibitors. Intermediate-level disinfection protocol implemented.
Indications for Use
Indicated for use by perfusionists and physicians to deliver whole blood and/or cardioplegia solutions to the heart during open heart surgery on arrested or beating hearts for up to six hours.
Regulatory Classification
Identification
A cardiopulmonary bypass heat exchanger is a device, consisting of a heat exchange system used in extracorporeal circulation to warm or cool the blood or perfusion fluid flowing through the device.
Predicate Devices
MPS2 Myocardial Protection System Console (K041979)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
October 1, 2018
Quest Medical, Inc. % Ronald Warren Sr. Director, Regulatory Affairs Experien Group 224 Airport Parkway San Jose, California 95110
Re: K173716
Trade/Device Name: MPS2 Myocardial Protection System Console Regulation Number: 21 CFR 870.4240 Regulation Name: Cardiopulmonary Bypass Heat Exchanger Regulatory Class: Class II Product Code: DTR, DWC Dated: August 22, 2018 Received: August 23, 2018
Dear Ronald Warren:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Actinclude requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Fernando Aguel Fernando Aguel -S
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
#### K173716
Device Name MPS2 Myocardial Protection System
#### Indications for Use (Describe)
The Quest Medical MPS®2 Myocardial Protection System, consisting of the MPS2 console and the MPS Delivery Set together, is intended for use by perfusionists and physicians to deliver whole blood (from any arterial source) and/or cardioplegia solutions to the heart during open heart surgery on either an arrested or beart for use up to six hours in duration.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### GENERAL INFORMATION [807.92(a)(1)]
### Applicant:
Quest Medical, Inc. One Allentown Parkway Allen, TX 75002-4211 USA Phone: 972-332-6352 Fax: 469-795-2352
# Contact Person:
Ronald S. Warren Vice President, Regulatory Affairs Experien Group 224 Airport Parkway, Suite 250 San Jose, CA 95110 USA Phone: 408-505-3926 Fax: 408-400-0865 Email: ron@experiengroup.com
### Date Prepared: September 25, 2018
### DEVICE INFORMATION [807.92(a)(2)]
Classification: Class II per 21CFR§870.4240
Product Code: DTR
Trade Name: MPS2 Myocardial Protection System Console
Generic/Common Name: Myocardial Perfusion System
# PREDICATE DEVICE(S) [807.92(a)(3)]
MPS2 Myocardial Protection System Console (K041979)
# DEVICE DESCRIPTION [807.92(a)(4)]
The MPS 2 console is a single device for myocardial perfusion that incorporates several different functions: heat exchanger, temperature control, pressure control, flow rate control, automatic priming and air detection / removal, and 3 flow modes - normal, cyclic, low volume.
{4}------------------------------------------------
### INDICATIONS FOR USE [807.92(a)(5)]
The MPS System, MPS console and the MPS delivery set used together, is for use by perfusionists and physicians to deliver whole blood (from any arterial source) and/or cardioplegia solutions to the heart during open heart surgery on either an arrested or beating heart for use up to six hours in duration.
The proposed MPS2 will have the same indication for use as the predicate MPS2 system as cleared under K041979.
### SUBSTANTIAL EQUIVALENCE
This labeling modification for the MPS2 is being submitted in order to provide end-user instructions related to the disinfection of the MPS2 water circulation system. Due to emerging concerns with sources of patient infection during cardiothoracic surgery and increased efforts to reduce biofilm formation in devices that may contribute to microbial contamination in the clinical setting, many perfusionists and other end-users are now requesting validated disinfection protocols. To address this clinical need, Quest Medical has validated a disinfection procedure, the details of which are to be added to the product labeling. This proposed change to product labeling summarized in this Traditional 510(k) premarket notification does not affect the intended use or the indication for use, and does not alter the fundamental scientific technology of the device. The modified MPS2 shares the same indication for use and the same technological characteristics as the predicate device. Therefore, based on the similarities between the two devices, the modified MPS2 is substantially equivalent to the predicate MPS2 device previously cleared under K041979.
# PERFORMANCE DATA [807.92(b)]
All necessary bench testing was conducted on the proposed MPS2 to support a determination of substantial equivalence to the predicate device. The disinfection method added to the product labeling was validated based on relevant guidance from AAMI TIR12. AAMI TIR30. FDA Guidance on Reprocessing Medical Devices, and the FDA Advisory Panel meeting on Circulatory System Devices, which occurred on June 2-3, 2016. The disinfection method meets requirements for an intermediate level of disinfection on Pseudomonas aeruginosa and Mycobacterium chimaera. The disinfection validation demonstrated a 6-Log reduction against the P. aeruginosa and a 3-Log reduction against M. chimaera. The prescribed disinfectant and microbial growth inhibitor were shown to be compatible with the materials of the water circulation system. In addition, this 510(k) notification was supported with results from surface cleaning and disinfection testing and heat exchanger performance testing.
# SUMMARY
The proposed MPS2 device is substantially equivalent to the predicate device.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.