← Product Code [HGM](/productcode/HGM) · K173715

# AlertWatch:OB (K173715)

_Alertwatch, Inc. · HGM · Apr 23, 2018 · Obstetrics/Gynecology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K173715

## Device Facts

- **Applicant:** Alertwatch, Inc.
- **Product Code:** [HGM](/productcode/HGM.md)
- **Decision Date:** Apr 23, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 884.2740
- **Device Class:** Class 2
- **Review Panel:** Obstetrics/Gynecology
- **Attributes:** Software as a Medical Device

## Indications for Use

AlertWatch:OB is intended for use by clinicians for secondary monitoring of maternal patients in the labor and delivery unit. AlertWatch:OB is a maternal surveillance system that combines data from validated electronic medical record systems, and displays them in one place. Once alerted by AlertWatch:OB, the clinician must refer to the primary monitor, device, or data source before making a clinical decision.

## Device Story

AlertWatch:OB is a software-only maternal surveillance system for labor and delivery units. It aggregates patient data from validated EMR systems; synthesizes information to identify potential clinical issues; and displays integrated data in a single interface. The system generates clinical advisories and alerts when values exceed preset limits. Clinicians (OB nurses, physicians, anesthesiologists) use the output on PCs, iPads, or iPhones to support clinical decision-making. The device does not replace primary monitors; clinicians must verify alerts against primary monitoring devices or EMR data before acting. Benefits include improved information access and clinical clarity for maternal monitoring.

## Clinical Evidence

No clinical data. Bench testing included a summative human factors study with 17 users demonstrating successful performance of critical tasks. Clinical limits and thresholds were validated via literature review and expert committee consensus (University of Michigan Health System and external anesthesiology experts).

## Technological Characteristics

Software-only device; no patient-contacting components. Operates via EMR data integration. Features include clinical advisory generation, vital sign re-display, and paging integration. Wireless communication evaluated per FDA guidance for coexistence. Thresholds based on clinical literature and expert review. No electrical energy or emissions.

## Regulatory Identification

A perinatal monitoring system is a device used to show graphically the relationship between maternal labor and the fetal heart rate by means of combining and coordinating uterine contraction and fetal heart monitors with appropriate displays of the well-being of the fetus during pregnancy, labor, and delivery. This generic type of device may include any of the devices subject to §§ 884.2600, 884.2640, 884.2660, 884.2675, 884.2700, and 884.2720. This generic type of device may include the following accessories: Central monitoring system and remote repeaters, signal analysis and display equipment, patient and equipment supports, and component parts.

## Predicate Devices

- AlertWatch:OR ([K153335](/device/K153335.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

April 23, 2018

AlertWatch, Inc. % Donna-Bea Tillman Senior Consultant Biologics Consulting Group, Inc. 1555 King Street Suite 300 Alexandria, Virginia 22314

Re: K173715

Trade/Device Name: AlertWatch:OB Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor (Including Cardiotachometer And Rate Alarm) Regulatory Class: Class II Product Code: MWI Dated: March 21, 2018 Received: March 22, 2018

Dear Donna-Bea Tillman:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);

{1}------------------------------------------------

and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

MA Hillemann

for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K173715

Device Name AlertWatch:OB

### Indications for Use (Describe)

AlertWatch:OB is intended for use by clinicians for secondary monitoring of maternal patients in the labor and delivery unit. AlertWatch:OB is a maternal surveillance system that combines data from validated electronic medical record systems, and displays them in one place. Once alerted by AlertWatch:OB, the clinician must refer to the primary monitor, device, or data source before making a clinical decision.

Type of Use (Select one or both, as applicable)

| <span style="font-family: sans-serif;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|----------------------------------------------------------------------------------------------|
|----------------------------------------------------------------------------------------------|

| <span style="font-family: sans-serif;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
|---------------------------------------------------------------------------------------------|
|---------------------------------------------------------------------------------------------|

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

In accordance with 21 CFR 807.87(h) and (21 CFR 807.92) the 510(k) Summary for AlertWatch:OB is provided below.

#### 1. SUBMITTER

| Applicant:                | AlertWatch, Inc<br>330 East Liberty St.<br>Fourth Floor<br>Ann Arbor, MI 48104<br>Tel:<br>Fax:                                                       |
|---------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact:                  | Justin Adams<br>CEO<br>Phone: 734-998-8344<br>E-mail: Justin.adams@alertwatch.com                                                                    |
| Submission Correspondent: | Donna-Bea Tillman, Ph.D.<br>Senior Consultant<br>Biologics Consultant Group, Inc.<br>Phone: 410-531-6542<br>E-mail: dtillman@biologicsconsulting.com |
| Date Prepared:            | December 1, 2017                                                                                                                                     |

#### 2. DEVICE

| Device Trade Name:  | AlertWatch:OB                                                                    |
|---------------------|----------------------------------------------------------------------------------|
| Device Common Name: | Physiological Patient Monitor without arrhythmia<br>detection or alarms          |
| Classification Name | 21 CFR 870.2300 - Cardiac monitor (including<br>cardiotachometer and rate alarm) |
| Regulatory Class:   | II                                                                               |
| Product Code:       | MWI                                                                              |

### PREDICATE DEVICE 3.

Predicate Device: AlertWatch:OR (K153335)

{4}------------------------------------------------

#### 4. DEVICE DESCRIPTION

AlertWatch:OB is a secondary monitoring system used by OB nurses, obstetricians, and OB anesthesiologists to monitor women in the Labor and Delivery (L&D) unit. The purpose of the program is to synthesize a wide range of maternal patient data and inform clinicians of potential problems. Once alerted, the clinician is instructed to refer to the primary monitoring device or EMR before making a clinical decision.

AlertWatch:OB should only be connected to EMR systems that have been validated for use with AlertWatch:OB. AlertWatch, LLC performs the validation for each installation site.

#### INTENDED USE/INDICATIONS FOR USE ട്.

AlertWatch:OB is intended for use by clinicians for secondary monitoring of maternal patients in the labor and delivery unit. AlertWatch:OB is a maternal surveillance system that combines data from validated electronic medical record systems, and displays them in one place. Once alerted by AlertWatch:OB, the clinician must refer to the primary monitor, device, or data source before making a clinical decision.

### SUBSTANTIAL EQUIVALENCE 6.

## Comparison of Indications

Both the subject AlertWatch:OB and the predicate AlertWatch:OR are intended for secondary monitoring of patients in the hospital. With both devices, the user is directed to refer to the primary monitor or device before making a clinical decision. The data collected, the analysis and display technology, and the alerting and paging process for both products is essentially the same. Therefore, this difference in indications for use does not constitute a new intended use, and AlertWatch:OR can be used as a primary predicate for the subject AlertWatch:OB.

# Technological Comparisons

The table below compares the key technological feature of the subject devices to the predicate device (AlertWatch:OR, K153335).

|                                                        | Predicate Device                                                                                                                                                                      | Proposed Device                                                                                                                         |
|--------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number                                          | K153335                                                                                                                                                                               | -                                                                                                                                       |
| Applicant                                              | AlertWatch, Inc.                                                                                                                                                                      | AlertWatch, Inc.                                                                                                                        |
| Device Name                                            | AlertWatch:OR                                                                                                                                                                         | AlertWatch:OB                                                                                                                           |
| Classification Regulation                              | 21 CFR 870.2300 - Cardiac<br>monitor (including<br>cardiotachometer and rate alarm)                                                                                                   | 21 CFR 870.2300 - Cardiac<br>monitor (including<br>cardiotachometer and rate alarm)                                                     |
| Product Code                                           | MWI                                                                                                                                                                                   | MWI                                                                                                                                     |
|                                                        | Predicate Device                                                                                                                                                                      | Proposed Device                                                                                                                         |
| Intended Use environment                               | OR                                                                                                                                                                                    | Labor and Delivery Unit                                                                                                                 |
| Intended Users                                         | Anesthesiologist, Resident,<br>CRNA                                                                                                                                                   | OB Anesthesiologist, OB<br>Physician, OB Nurse, Resident                                                                                |
| Intended for primary<br>monitoring?                    | No - only for secondary<br>monitoring                                                                                                                                                 | Identical                                                                                                                               |
| Data server                                            | AlertWatch:OR accesses data<br>from the AIMS database server.                                                                                                                         | AlertWatch:OB accesses data<br>from the EMR system.                                                                                     |
| Supported AIMS / EMR<br>systems                        | Centricity, version 7.6.3 <span style="font-size: 8px;">SEP</span><br>Picis / Optum, version 421 <span style="font-size: 8px;">SEP</span><br>iMDsoft / Metavision, version<br>5.46.44 | Epic 2015 UI 2                                                                                                                          |
| Supported physiologic<br>monitors                      | GE Solar 9500<br>GE Carespan B850                                                                                                                                                     | None directly.                                                                                                                          |
| Organs Monitored                                       | Brain, lungs, heart, liver,<br>kidneys, skin temperature                                                                                                                              | Identical                                                                                                                               |
| Hardware platforms supported                           | PC, iPad, iPhone                                                                                                                                                                      | Identical                                                                                                                               |
| Generates clinical advisories                          | AlertWatch:OR Analyzes data<br>from patient monitors and other<br>sources and alerts when values<br>exceed preset limits.                                                             | Identical                                                                                                                               |
| Re-display of vital sign data<br>from primary monitors | Yes                                                                                                                                                                                   | Identical                                                                                                                               |
| Intended to replace primary<br>monitors                | No                                                                                                                                                                                    | Identical                                                                                                                               |
| Paging                                                 | Yes. For clinicians to page each<br>other and for AlertWatch:OR to<br>transmit alerts using the hospital-<br>established paging system.                                               | Yes. For clinicians to page each<br>other and for AlertWatch:OB to<br>transmit alerts using the hospital-<br>established paging system. |
| Uses color to display clinical<br>information          | Image: Predicate Device Clinical Information                                                                                                                                          | Image: Proposed Device Clinical Information                                                                                             |

Table 1: Technological Comparison

{5}------------------------------------------------

There are slight differences in technological characteristics that reflect the different clinical needs of the user populations, but the general approach of retrieving data from the EMR system, integrating this data, and performing a series of calculations to assess potential patient clinical

{6}------------------------------------------------

issues is the same. The differences in technological characteristics do not raise different questions of safety and effectiveness.

#### 7. PERFORMANCE DATA

### Biocompatibility Testing

There are no direct or indirect patient-contacting components of the subject device. Therefore, patient contact information is not needed for this device.

## Electrical safety and electromagnetic compatibility (EMC)

Not applicable. The subject device is a software-only device. It contains no electric components, generates no electrical emissions, and uses no electrical energy of any type.

However, because AlertWatch:OB includes wireless communication, the system was evaluated in accordance with the FDA guidance document: Radio Frequency Wireless Technology in Medical Devices: Guidance for Industry and Food and Drug Administration Staff (August 14, 2013). Wireless Co-existence testing was performed to establish that the wireless components work effectively in the hospital environment.

## Software Verification and Validation Testing

Software verification and validation testing was conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." Verification of AlertWatch:OR was conducted to ensure that the product works as designed, and was tested with both constructed data and data from the EMR. Validation was conducted to check the design and performance of the product.

### Bench Testing

- . Human Factors Study: To better understand any potential usability issues that could interfere with the intended use of the product. AlertWatch performed a comprehensive human factors study. AlertWatch first conducted a formative to identify and fix any problems with the usability study plan and AlertWatch:OB product. AlertWatch then recruited 17 users and conducted a summative usability study. The results of the study showed that users with minimal training were able to successfully perform critical tasks and use the device for its intended purpose – to clarify clinical information and support information access.
- . Default Limits and Thresholds: AlertWatch, Inc. used a three-phased approach to ensure that the default limits were clinically valid:
	- 1. Review of References. AlertWatch sought out definitive published studies that highlighted appropriate limits for certain patient conditions. This includes limits and thresholds for the green/yellow/red organ schema, as well as the alerts and formulae.

{7}------------------------------------------------

- 2. Expert Committee. When references were not available, AlertWatch sought out the opinion and confirmation of obstetricians and OB anesthesia physicians at the University of Michigan Health System. Each of these clinicians reviewed the limits, provided feedback, and reviewed the final results.
- 3. External Experts. AlertWatch obtained final review of the default limits from an external group of four anesthesiology and OB anesthesia experts. All approved the clinical limits.
- IEC 60601-1-8: A summary of how AlertWatch:OR addresses the primary issues ● identified in IEC 60601-1-8 is provided in the submission.

## Animal Testing

Not applicable. Animal studies are not necessary to establish the substantial equivalence of this device.

# Clinical Data

Not applicable. Clinical studies are not necessary to establish the substantial equivalence of this device.

#### CONCLUSION 8.

The subject AlertWatch:OB and the predicate AlertWatch:OR have the same intended use, namely to provide secondary monitoring of patients in the hospital environment. There are slight differences in technological characteristics that reflect the different clinical needs of the user populations, but the general approach of retrieving data from the EMR system, integrating this data, and performing a series of calculations to assess potential patient clinical issues is the same. The differences in technological characteristics do not raise different questions of safety and effectiveness, and the results of performance testing demonstrate that the subject device performs in accordance with specifications and meets user needs and intended uses. Therefore, AlertWatch:OB has been demonstrated to be substantially equivalent to the predicate AlertWatch:OR.

---

**Source:** [https://fda.innolitics.com/device/K173715](https://fda.innolitics.com/device/K173715)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
